Validation of a Wireless Wearable Vital Signs Monitor in Pediatric Patient

Wireless Skin-Adherent Physiologic Sensor Validation in Pediatric Patients - Phase 1 Clinical Study

This prospective observational study aims to evaluate the accuracy, safety, and feasibility of the ANNE® Chest wireless wearable sensor for continuous monitoring of heart rate, respiratory rate, oxygen saturation, and skin temperature in hospitalized neonates, infants, and children (0 days to 12 years of age) in the neonatal and pediatric intensive care units (NICU and PICU). Measurements obtained from the wireless device will be compared with standard-of-care bedside monitoring and capnography. The study will also assess skin tolerability, signal reliability, and caregiver and healthcare provider perceptions of the device.

調査の概要

研究の種類

観察的

入学 (推定)

42

連絡先と場所

このセクションには、調査を実施する担当者の連絡先の詳細と、この調査が実施されている場所に関する情報が記載されています。

研究連絡先

参加基準

研究者は、適格基準と呼ばれる特定の説明に適合する人を探します。これらの基準のいくつかの例は、人の一般的な健康状態または以前の治療です。

適格基準

就学可能な年齢

  • 子

健康ボランティアの受け入れ

いいえ

サンプリング方法

非確率サンプル

調査対象母集団

Hospitalized neonates, infants, and children (0 days to <12 years of age) admitted to the NICU or PICU at the Montreal Children's Hospital who require continuous physiologic monitoring and whose parent or legal guardian provides informed consent.

説明

Inclusion Criteria:

  • Neonates (0-28 days of life)
  • Infants (29 days of life - 2 years, i.e., not turned 2 years old)
  • Children (2 years old - 12 years old, i.e., not turned 12 years old)

Exclusion Criteria:

  • Contraindication to application of skin sensors (e.g., rash, surgical site, skin breakdown, open wounds, burns, active dermatitis, etc.) at the application site.
  • Severe hemodynamic instability requiring escalating interventions.
  • Congenital airway abnormalities interfere with accurate capnography.
  • Diagnosed neuromuscular disorders affecting respiratory mechanics.
  • Known adhesive allergies.
  • Inability to tolerate device placement due to behavioral issues.
  • Participation in conflicting clinical trials.
  • High-frequency oscillatory or mechanical ventilation.
  • Active bronchiolitis with mask intolerance.
  • Recent facial surgery preventing EtCO₂ nasal line placement.
  • Behavioral conditions preventing adherence to device wearing.

研究計画

このセクションでは、研究がどのように設計され、研究が何を測定しているかなど、研究計画の詳細を提供します。

研究はどのように設計されていますか?

デザインの詳細

コホートと介入

グループ/コホート
介入・治療
Neonates
Hospitalized neonates aged 0-28 days admitted to the neonatal intensive care unit (NICU) or pediatric intensive care unit (PICU).
Participants will wear the ANNE® Chest wireless skin-adherent physiologic sensor for continuous monitoring of heart rate, respiratory rate, electrocardiography (ECG), skin temperature, and body position during the study period. Measurements obtained from the investigational device will be compared with simultaneous standard-of-care bedside monitoring and capnography for validation purposes.
Infants
Hospitalized infants aged 29 days to <2 years admitted to the NICU or PICU
Participants will wear the ANNE® Chest wireless skin-adherent physiologic sensor for continuous monitoring of heart rate, respiratory rate, electrocardiography (ECG), skin temperature, and body position during the study period. Measurements obtained from the investigational device will be compared with simultaneous standard-of-care bedside monitoring and capnography for validation purposes.
Children
Hospitalized children aged 2 to <12 years admitted to the PICU
Participants will wear the ANNE® Chest wireless skin-adherent physiologic sensor for continuous monitoring of heart rate, respiratory rate, electrocardiography (ECG), skin temperature, and body position during the study period. Measurements obtained from the investigational device will be compared with simultaneous standard-of-care bedside monitoring and capnography for validation purposes.

この研究は何を測定していますか?

主要な結果の測定

結果測定
メジャーの説明
時間枠
Respiratory Rate Accuracy
時間枠:Study monitoring period (4 to 8 hours per day over 2 consecutive days)
Agreement between respiratory rate measured by the ANNE® Chest wireless sensor and the reference standard capnography. Accuracy will be evaluated using agreement analyses, with success defined as wearable-derived respiratory rate values within ±3 breaths per minute of the reference standard.
Study monitoring period (4 to 8 hours per day over 2 consecutive days)
Heart Rate Accuracy
時間枠:Study monitoring period (4 to 8 hours per day over 2 consecutive days)
Agreement between heart rate measured by the ANNE® Chest wireless sensor and the reference standard bedside monitor. Accuracy will be evaluated using agreement analyses, with success defined as wearable-derived heart rate values within ±5 beats per minute of the reference monitor.
Study monitoring period (4 to 8 hours per day over 2 consecutive days)

二次結果の測定

結果測定
メジャーの説明
時間枠
Skin Tolerability
時間枠:Baseline (prior to device placement), immediately after device removal, and approximately 30 minutes after device removal.
Skin tolerability of the ANNE® Chest sensor, including the presence of erythema, irritation, or pressure marks following device use. Skin integrity will be assess by comparing before and after photos of sites of placement of wearable sensors using the Berger & Bowman Scoring Scale.
Baseline (prior to device placement), immediately after device removal, and approximately 30 minutes after device removal.
Caregiver and Healthcare Provider Device Usability Perspective
時間枠:Immediately following completion of study participation.
Caregiver and healthcare provider feedback regarding the usability and acceptability of the ANNE®. Parent and clinician perceptions will be assessed using a custom study questionnaire containing Likert-scale, multiple-choice, and open-ended questions addressing usability, comfort, perceived safety, perceived accuracy, and overall acceptability of the wireless monitoring system.
Immediately following completion of study participation.
Respiratory Rate Average Root Mean Squared Error (ARMS)
時間枠:Study monitoring period (4 to 8 hours per day over 2 consecutive days)
Average Root Mean Squared Error (ARMS) between respiratory rate measurements obtained from the ANNE® Chest sensor and the reference monitor within each age cohort.
Study monitoring period (4 to 8 hours per day over 2 consecutive days)

協力者と研究者

ここでは、この調査に関係する人々や組織を見つけることができます。

捜査官

  • 主任研究者:Guilherme Sant'Anna, MD, PhD、RI-MUHC

研究記録日

これらの日付は、ClinicalTrials.gov への研究記録と要約結果の提出の進捗状況を追跡します。研究記録と報告された結果は、国立医学図書館 (NLM) によって審査され、公開 Web サイトに掲載される前に、特定の品質管理基準を満たしていることが確認されます。

主要日程の研究

研究開始 (推定)

2026年9月1日

一次修了 (推定)

2027年9月1日

研究の完了 (推定)

2027年9月1日

試験登録日

最初に提出

2026年7月28日

QC基準を満たした最初の提出物

2026年7月28日

最初の投稿 (実際)

2026年7月31日

学習記録の更新

投稿された最後の更新 (実際)

2026年7月31日

QC基準を満たした最後の更新が送信されました

2026年7月28日

最終確認日

2026年7月1日

詳しくは

本研究に関する用語

個々の参加者データ (IPD) の計画

個々の参加者データ (IPD) を共有する予定はありますか?

いいえ

医薬品およびデバイス情報、研究文書

米国FDA規制医薬品の研究

いいえ

米国FDA規制機器製品の研究

はい

この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。