Pilot Study on the Feasibility of High-energy Collimation With Optimized Geometry on a VERITON™ CZT Camera (HELIOS)
Functional imaging using scintigraphy plays a major role in the diagnosis and monitoring of many diseases, particularly thanks to its ability to perform whole-body examinations with high contrast. The indications and use of scintigraphy have increased in recent years, particularly in connection with the development of internal vectorized radiotherapy. This therapeutic approach is based on the administration of radiotracers that enable targeted irradiation of tumor cells, whose biodistribution throughout the body can be analyzed and quantified using scintigraphy, particularly with iodine-131 and lutetium-177.
The recent boom in scintigraphy is also linked to the development of new 360° geometry CZT-SPECT cameras, which enable rapid tomographic acquisitions of the entire body with significantly improved image quality compared to conventional cameras. These systems use mobile CZT semiconductor detectors that dynamically scan the anatomical regions of interest.
CZT detectors are combined with a collimation system consisting of tungsten septa, which are essential for the directional filtering of gamma photons. Unlike conventional scintigraphic cameras, where collimators can be changed or adjusted according to the energy of the detected photons, CZT-SPECT 360° cameras generally rely on fixed collimation, which cannot be changed.
연구 개요
상태
정황
상세 설명
Functional imaging using scintigraphy plays a major role in the diagnosis and monitoring of many diseases, particularly thanks to its ability to perform whole-body examinations with high contrast. The indications and use of scintigraphy have increased in recent years, particularly in connection with the development of internal vectorized radiotherapy. This therapeutic approach is based on the administration of radiotracers that enable targeted irradiation of tumor cells, whose biodistribution throughout the body can be analyzed and quantified using scintigraphy, particularly with iodine-131 and lutetium-177.
The recent boom in scintigraphy is also linked to the development of new 360° geometry CZT-SPECT cameras, which enable rapid tomographic acquisitions of the entire body with significantly improved image quality compared to conventional cameras. These systems use mobile CZT semiconductor detectors that dynamically scan the anatomical regions of interest.
CZT detectors are combined with a collimation system consisting of tungsten septa, which are essential for the directional filtering of gamma photons. Unlike conventional scintigraphic cameras, where collimators can be changed or adjusted according to the energy of the detected photons, CZT-SPECT 360° cameras generally rely on fixed collimation, which cannot be changed.
연구 유형
등록 (추정된)
단계
- 해당 없음
연락처 및 위치
연구 연락처
- 이름: Anne-Sophie Hue Project Manager, MSc
- 전화번호: +33.3.83.15.34.75
- 이메일: a.hue@chru-nancy.fr
참여기준
자격 기준
공부할 수 있는 나이
- 성인
- 고령자
건강한 자원 봉사자를 받아들입니다
설명
Inclusion Criteria:
- patients with multiple known lesions (≥ 2 metastases or lesions of other types) who, as part of their clinical management, require either (i) a 99mTc-HDP bone scan, or (ii) a post-therapeutic 177Lu-PSMA scan, or (iii) a post-therapeutic 131I scintigraphy.
- Person who has undergone a preliminary clinical examination appropriate for clinical research.
- Adult subject who has received complete information about the research and has signed the informed consent form.
- General condition, WHO ≤ 2, and ability to remain lying down without moving for nearly an hour.
Exclusion Criteria:
- Subjects with a contraindication, according to the target examination group, to injection of 99mTc-HDP (Technescan HDP, Curium Pharma), 177Lu-PSMA (Pluvicto®, Novartis), or 131I (131I sodium iodide for therapy, Curium Pharma) .
- Subjects covered by Articles L. 1121-5, L. 1121-7, and L1121-8 of the Public Health Code.
- Pregnant women, women in labor, or breastfeeding mothers.
- Subjects subject to legal protection measures (guardianship, curatorship, judicial protection).
- Subjects unable to give their consent.
- Subjects deprived of their liberty by a judicial or administrative decision, persons receiving psychiatric care pursuant to Articles L. 3212-1 and L.3213-1.
공부 계획
연구는 어떻게 설계됩니까?
디자인 세부사항
- 주 목적: 특수 증상
- 할당: 무작위
- 중재 모델: 병렬 할당
- 마스킹: 없음(오픈 라벨)
무기와 개입
참가자 그룹 / 팔 |
개입 / 치료 |
|---|---|
|
실험적: VERITON-CT New
Optimized collimators for the CZT detectors of the VERITON™ 400 wide-field camera (Spectrum Dynamics Medical) enabling better selection of incident photons
|
- Whole-body 3D recording on the VERITON™ 400 camera equipped with conventional collimators for technetium-99m and lutetium-177.
|
|
활성 비교기: Référence
|
Full-body 3D recording on the VERITON™ 400 camera
Whole-body 2D and cervicothoracic 3D recordings using an Anger camera (SYMBIA camera, Siemens Healthineers) for iodine-131 images.
|
연구는 무엇을 측정합니까?
주요 결과 측정
결과 측정 |
측정값 설명 |
기간 |
|---|---|---|
|
Feasability of whole-body SPECT/CT imaging with iodine-131 using the optimized collimator
기간: 1 month
|
Data analysis obtained with optimized collimator and with conventional collimator
|
1 month
|
|
Perform quantitative analysis of technetium-99m and lutetium-177 with feasability results
기간: 1 month
|
Image sharpness improvment
|
1 month
|
2차 결과 측정
결과 측정 |
측정값 설명 |
기간 |
|---|---|---|
|
Overall quality of whole-body CZT acquisitions with iodine-131
기간: 1 month
|
For iodine-131, the overall quality of images acquired using the CZT-SPECT camera with the new collimator will be described based on an ordinal score for stellar artifacts (absent, present but not interfering, present and interfering), assessed independently by two experienced readers.
The frequency and severity of artifacts will be reported, as well as their association with the interpretable or non-interpretable nature of the examinations defined in the primary endpoint.
Inter-reader agreement will also be described.
|
1 month
|
|
Quantify the detectability and contrast of lesions visualized on whole-body CZT imaging with iodine-131.
기간: 1 month
|
Sharpness, defined as the maximum slope of the intensity profile at the edge of the lesions and the contrast-to-noise ratio
|
1 month
|
공동 작업자 및 조사자
연구 기록 날짜
연구 주요 날짜
연구 시작 (추정된)
기본 완료 (추정된)
연구 완료 (추정된)
연구 등록 날짜
최초 제출
QC 기준을 충족하는 최초 제출
처음 게시됨 (실제)
연구 기록 업데이트
마지막 업데이트 게시됨 (실제)
QC 기준을 충족하는 마지막 업데이트 제출
마지막으로 확인됨
추가 정보
이 연구와 관련된 용어
개별 참가자 데이터(IPD) 계획
개별 참가자 데이터(IPD)를 공유할 계획입니까?
약물 및 장치 정보, 연구 문서
미국 FDA 규제 의약품 연구
미국 FDA 규제 기기 제품 연구
이 정보는 변경 없이 clinicaltrials.gov 웹사이트에서 직접 가져온 것입니다. 귀하의 연구 세부 정보를 변경, 제거 또는 업데이트하도록 요청하는 경우 register@clinicaltrials.gov. 문의하십시오. 변경 사항이 clinicaltrials.gov에 구현되는 즉시 저희 웹사이트에도 자동으로 업데이트됩니다. .