Pilot Study on the Feasibility of High-energy Collimation With Optimized Geometry on a VERITON™ CZT Camera (HELIOS)
Functional imaging using scintigraphy plays a major role in the diagnosis and monitoring of many diseases, particularly thanks to its ability to perform whole-body examinations with high contrast. The indications and use of scintigraphy have increased in recent years, particularly in connection with the development of internal vectorized radiotherapy. This therapeutic approach is based on the administration of radiotracers that enable targeted irradiation of tumor cells, whose biodistribution throughout the body can be analyzed and quantified using scintigraphy, particularly with iodine-131 and lutetium-177.
The recent boom in scintigraphy is also linked to the development of new 360° geometry CZT-SPECT cameras, which enable rapid tomographic acquisitions of the entire body with significantly improved image quality compared to conventional cameras. These systems use mobile CZT semiconductor detectors that dynamically scan the anatomical regions of interest.
CZT detectors are combined with a collimation system consisting of tungsten septa, which are essential for the directional filtering of gamma photons. Unlike conventional scintigraphic cameras, where collimators can be changed or adjusted according to the energy of the detected photons, CZT-SPECT 360° cameras generally rely on fixed collimation, which cannot be changed.
Tutkimuksen yleiskatsaus
Tila
Ehdot
Yksityiskohtainen kuvaus
Functional imaging using scintigraphy plays a major role in the diagnosis and monitoring of many diseases, particularly thanks to its ability to perform whole-body examinations with high contrast. The indications and use of scintigraphy have increased in recent years, particularly in connection with the development of internal vectorized radiotherapy. This therapeutic approach is based on the administration of radiotracers that enable targeted irradiation of tumor cells, whose biodistribution throughout the body can be analyzed and quantified using scintigraphy, particularly with iodine-131 and lutetium-177.
The recent boom in scintigraphy is also linked to the development of new 360° geometry CZT-SPECT cameras, which enable rapid tomographic acquisitions of the entire body with significantly improved image quality compared to conventional cameras. These systems use mobile CZT semiconductor detectors that dynamically scan the anatomical regions of interest.
CZT detectors are combined with a collimation system consisting of tungsten septa, which are essential for the directional filtering of gamma photons. Unlike conventional scintigraphic cameras, where collimators can be changed or adjusted according to the energy of the detected photons, CZT-SPECT 360° cameras generally rely on fixed collimation, which cannot be changed.
Opintotyyppi
Ilmoittautuminen (Arvioitu)
Vaihe
- Ei sovellettavissa
Yhteystiedot ja paikat
Opiskeluyhteys
- Nimi: Anne-Sophie Hue Project Manager, MSc
- Puhelinnumero: +33.3.83.15.34.75
- Sähköposti: a.hue@chru-nancy.fr
Osallistumiskriteerit
Kelpoisuusvaatimukset
Opintokelpoiset iät
- Aikuinen
- Vanhempi Aikuinen
Hyväksyy terveitä vapaaehtoisia
Kuvaus
Inclusion Criteria:
- patients with multiple known lesions (≥ 2 metastases or lesions of other types) who, as part of their clinical management, require either (i) a 99mTc-HDP bone scan, or (ii) a post-therapeutic 177Lu-PSMA scan, or (iii) a post-therapeutic 131I scintigraphy.
- Person who has undergone a preliminary clinical examination appropriate for clinical research.
- Adult subject who has received complete information about the research and has signed the informed consent form.
- General condition, WHO ≤ 2, and ability to remain lying down without moving for nearly an hour.
Exclusion Criteria:
- Subjects with a contraindication, according to the target examination group, to injection of 99mTc-HDP (Technescan HDP, Curium Pharma), 177Lu-PSMA (Pluvicto®, Novartis), or 131I (131I sodium iodide for therapy, Curium Pharma) .
- Subjects covered by Articles L. 1121-5, L. 1121-7, and L1121-8 of the Public Health Code.
- Pregnant women, women in labor, or breastfeeding mothers.
- Subjects subject to legal protection measures (guardianship, curatorship, judicial protection).
- Subjects unable to give their consent.
- Subjects deprived of their liberty by a judicial or administrative decision, persons receiving psychiatric care pursuant to Articles L. 3212-1 and L.3213-1.
Opintosuunnitelma
Miten tutkimus on suunniteltu?
Suunnittelun yksityiskohdat
- Ensisijainen käyttötarkoitus: Diagnostiikka
- Jako: Satunnaistettu
- Inventiomalli: Rinnakkaistehtävä
- Naamiointi: Ei mitään (avoin tarra)
Aseet ja interventiot
Osallistujaryhmä / Arm |
Interventio / Hoito |
|---|---|
|
Kokeellinen: VERITON-CT New
Optimized collimators for the CZT detectors of the VERITON™ 400 wide-field camera (Spectrum Dynamics Medical) enabling better selection of incident photons
|
- Whole-body 3D recording on the VERITON™ 400 camera equipped with conventional collimators for technetium-99m and lutetium-177.
|
|
Active Comparator: Référence
|
Full-body 3D recording on the VERITON™ 400 camera
Whole-body 2D and cervicothoracic 3D recordings using an Anger camera (SYMBIA camera, Siemens Healthineers) for iodine-131 images.
|
Mitä tutkimuksessa mitataan?
Ensisijaiset tulostoimenpiteet
Tulosmittaus |
Toimenpiteen kuvaus |
Aikaikkuna |
|---|---|---|
|
Feasability of whole-body SPECT/CT imaging with iodine-131 using the optimized collimator
Aikaikkuna: 1 month
|
Data analysis obtained with optimized collimator and with conventional collimator
|
1 month
|
|
Perform quantitative analysis of technetium-99m and lutetium-177 with feasability results
Aikaikkuna: 1 month
|
Image sharpness improvment
|
1 month
|
Toissijaiset tulostoimenpiteet
Tulosmittaus |
Toimenpiteen kuvaus |
Aikaikkuna |
|---|---|---|
|
Overall quality of whole-body CZT acquisitions with iodine-131
Aikaikkuna: 1 month
|
For iodine-131, the overall quality of images acquired using the CZT-SPECT camera with the new collimator will be described based on an ordinal score for stellar artifacts (absent, present but not interfering, present and interfering), assessed independently by two experienced readers.
The frequency and severity of artifacts will be reported, as well as their association with the interpretable or non-interpretable nature of the examinations defined in the primary endpoint.
Inter-reader agreement will also be described.
|
1 month
|
|
Quantify the detectability and contrast of lesions visualized on whole-body CZT imaging with iodine-131.
Aikaikkuna: 1 month
|
Sharpness, defined as the maximum slope of the intensity profile at the edge of the lesions and the contrast-to-noise ratio
|
1 month
|
Yhteistyökumppanit ja tutkijat
Sponsori
Opintojen ennätyspäivät
Opi tärkeimmät päivämäärät
Opiskelun aloitus (Arvioitu)
Ensisijainen valmistuminen (Arvioitu)
Opintojen valmistuminen (Arvioitu)
Opintoihin ilmoittautumispäivät
Ensimmäinen lähetetty
Ensimmäinen toimitettu, joka täytti QC-kriteerit
Ensimmäinen Lähetetty (Todellinen)
Tutkimustietojen päivitykset
Viimeisin päivitys julkaistu (Todellinen)
Viimeisin lähetetty päivitys, joka täytti QC-kriteerit
Viimeksi vahvistettu
Lisää tietoa
Tähän tutkimukseen liittyvät termit
Avainsanat
Muita asiaankuuluvia MeSH-ehtoja
Muut tutkimustunnusnumerot
- 2026-A00456-45
Yksittäisten osallistujien tietojen suunnitelma (IPD)
Aiotko jakaa yksittäisten osallistujien tietoja (IPD)?
Lääke- ja laitetiedot, tutkimusasiakirjat
Tutkii yhdysvaltalaista FDA sääntelemää lääkevalmistetta
Tutkii yhdysvaltalaista FDA sääntelemää laitetuotetta
Nämä tiedot haettiin suoraan verkkosivustolta clinicaltrials.gov ilman muutoksia. Jos sinulla on pyyntöjä muuttaa, poistaa tai päivittää tutkimustietojasi, ota yhteyttä register@clinicaltrials.gov. Heti kun muutos on otettu käyttöön osoitteessa clinicaltrials.gov, se päivitetään automaattisesti myös verkkosivustollemme .