Calibre vs. COSMED

2026년 6월 16일 업데이트: Duke University

Correlation of Cardiopulmonary Exercise Testing (CPET) VO2 Max Parameters Between the Calibre Device and COSMED Metabolic Cart

This study evaluates agreement between the Calibre device and the COSMED Quark CPET system for measurement of cardiorespiratory fitness parameters during maximal exercise testing in adults aged 18-55 years.

연구 개요

연구 유형

중재적

등록 (추정된)

40

단계

  • 해당 없음

참여기준

연구원은 적격성 기준이라는 특정 설명에 맞는 사람을 찾습니다. 이러한 기준의 몇 가지 예는 개인의 일반적인 건강 상태 또는 이전 치료입니다.

자격 기준

공부할 수 있는 나이

  • 성인

건강한 자원 봉사자를 받아들입니다

예

설명

Inclusion Criteria:

  • Age 18-55
  • Able to speak English
  • Able to provide informed consent

Exclusion Criteria:

  • Acute myocardial infarction
  • Unstable angina
  • Syncope
  • Uncontrolled asthma
  • Arterial desaturation at rest on room air <85%
  • Severe untreated arterial hypertension at rest (>200 mm Hg systolic, >120 mm Hg diastolic)
  • Significant pulmonary hypertension
  • Active thrombosis of the lower extremity, unless treated for at least 2 weeks prior to participation.
  • Within 2 weeks of acute symptomatic pulmonary embolus
  • Considered inappropriate to participate by Principal Investigator
  • Previously experienced a heart attack
  • Pregnancy

공부 계획

이 섹션에서는 연구 설계 방법과 연구가 측정하는 내용을 포함하여 연구 계획에 대한 세부 정보를 제공합니다.

연구는 어떻게 설계됩니까?

디자인 세부사항

  • 주 목적: 다른
  • 할당: 무작위
  • 중재 모델: 크로스오버 할당
  • 마스킹: 없음(오픈 라벨)

무기와 개입

참가자 그룹 / 팔
개입 / 치료
실험적: Calibre First Sequence
Participants undergo cardiopulmonary exercise testing using the investigational Calibre device during Visit 1, followed by testing with the COSMED Quark CPET reference system during Visit 2 (at least 2 days later).
Investigational wearable facemask device with integrated sensors for breath-by-breath measurement of respiratory gas exchange and ventilatory parameters during cardiopulmonary exercise testing, evaluated against the COSMED Quark CPET reference system.
실험적: COSMED First Sequence
Participants undergo cardiopulmonary exercise testing using the COSMED Quark CPET reference system during Visit 1, followed by testing with the investigational Calibre device during Visit 2 (at least 2 days later).
FDA-cleared cardiopulmonary exercise testing (CPET) metabolic cart used as the reference comparator for measurement of respiratory gas exchange and ventilatory parameters during ramp cycle ergometry. Participants wear a facemask connected to the system for breath-by-breath measurement of oxygen consumption (VO2), carbon dioxide production (VCO2), minute ventilation (VE), anaerobic threshold, and ventilatory efficiency during maximal exercise testing.

연구는 무엇을 측정합니까?

주요 결과 측정

결과 측정
측정값 설명
기간
VO₂ peak correlation, as measured by Pearson or Spearman correlation coefficient between Calibre and COSMED
기간: During CPET testing across Visit 1 and Visit 2, up to 7 days.
VO₂ peak measured during CPET will be compared between devices using Pearson or Spearman correlation coefficient, depending on data distribution.
During CPET testing across Visit 1 and Visit 2, up to 7 days.
VO₂ peak agreement, as measured by Bland-Altman bias and limits of agreement between Calibre and COSMED
기간: During CPET testing across Visit 1 and Visit 2, up to 7 days.
VO₂ peak measured during CPET will be compared between devices using Bland-Altman analysis to estimate average bias and limits of agreement.
During CPET testing across Visit 1 and Visit 2, up to 7 days.
Anaerobic threshold correlation, as measured by Pearson or Spearman correlation coefficient between Calibre and COSMED
기간: During CPET testing across Visit 1 and Visit 2, up to 7 days
Anaerobic threshold measured during CPET will be compared between devices using Pearson or Spearman correlation coefficient, depending on data distribution.
During CPET testing across Visit 1 and Visit 2, up to 7 days
Anaerobic threshold agreement, as measured by Bland-Altman bias and limits of agreement between Calibre and COSMED
기간: During CPET testing across Visit 1 and Visit 2, up to 7 days
Anaerobic threshold measured during CPET will be compared between devices using Bland-Altman analysis to estimate average bias and limits of agreement.
During CPET testing across Visit 1 and Visit 2, up to 7 days
VE/VCO₂ correlation, as measured by Pearson or Spearman correlation coefficient between Calibre and COSMED
기간: During CPET testing across Visit 1 and Visit 2, up to 7 days
Ventilatory equivalent for carbon dioxide (VE/VCO₂) measured during CPET will be compared between devices using Pearson or Spearman correlation coefficient, depending on data distribution.
During CPET testing across Visit 1 and Visit 2, up to 7 days
VE/VCO₂ agreement, as measured by Bland-Altman bias and limits of agreement between Calibre and COSMED
기간: During CPET testing across Visit 1 and Visit 2, up to 7 days
Ventilatory equivalent for carbon dioxide (VE/VCO₂) measured during CPET will be compared between devices using Bland-Altman analysis to estimate average bias and limits of agreement.
During CPET testing across Visit 1 and Visit 2, up to 7 days

공동 작업자 및 조사자

여기에서 이 연구와 관련된 사람과 조직을 찾을 수 있습니다.

스폰서

수사관

  • 수석 연구원: David MacLeod, M.B.B.S., Duke University

연구 기록 날짜

이 날짜는 ClinicalTrials.gov에 대한 연구 기록 및 요약 결과 제출의 진행 상황을 추적합니다. 연구 기록 및 보고된 결과는 공개 웹사이트에 게시되기 전에 특정 품질 관리 기준을 충족하는지 확인하기 위해 국립 의학 도서관(NLM)에서 검토합니다.

연구 주요 날짜

연구 시작 (추정된)

2026년 9월 1일

기본 완료 (추정된)

2027년 7월 1일

연구 완료 (추정된)

2027년 7월 1일

연구 등록 날짜

최초 제출

2026년 5월 5일

QC 기준을 충족하는 최초 제출

2026년 5월 12일

처음 게시됨 (실제)

2026년 5월 19일

연구 기록 업데이트

마지막 업데이트 게시됨 (실제)

2026년 6월 17일

QC 기준을 충족하는 마지막 업데이트 제출

2026년 6월 16일

마지막으로 확인됨

2026년 5월 1일

추가 정보

이 연구와 관련된 용어

기타 연구 ID 번호

  • Pro00118791

개별 참가자 데이터(IPD) 계획

개별 참가자 데이터(IPD)를 공유할 계획입니까?

아니요

약물 및 장치 정보, 연구 문서

미국 FDA 규제 의약품 연구

아니

미국 FDA 규제 기기 제품 연구

예

미국에서 제조되어 미국에서 수출되는 제품

예

이 정보는 변경 없이 clinicaltrials.gov 웹사이트에서 직접 가져온 것입니다. 귀하의 연구 세부 정보를 변경, 제거 또는 업데이트하도록 요청하는 경우 register@clinicaltrials.gov. 문의하십시오. 변경 사항이 clinicaltrials.gov에 구현되는 즉시 저희 웹사이트에도 자동으로 업데이트됩니다. .

구독하다