Ta strona została przetłumaczona automatycznie i dokładność tłumaczenia nie jest gwarantowana. Proszę odnieść się do angielska wersja za tekst źródłowy.

Development of a Mobile Terminal-Based Intelligent Detection System for Multiple Anterior Segment Diseases of the Eye (LENS)

8 czerwca 2026 zaktualizowane przez: Zhongshan Ophthalmic Center, Sun Yat-sen University

This is a multi-center, cross-sectional study evaluating a smartphone-based artificial intelligence (AI) system for anterior segment eye disease screening. The system is designed to identify 16 clinically important anterior segment conditions from images captured using a standard Android smartphone. A core design feature of the system is that all image analysis is performed entirely on the smartphone itself, without requiring internet connectivity or cloud-based server infrastructure.

The study is motivated by a structural challenge in the deployment of medical AI: systems that depend on cloud infrastructure for inference are non-functional in settings without reliable internet access, which disproportionately excludes populations in low-resource regions where the burden of preventable eye disease is highest. This study evaluates whether an on-device AI system, designed with operational constraints as a primary engineering objective, can deliver clinically acceptable diagnostic performance while remaining operable under real-world connectivity limitations.

The study comprises five evaluation components. First, the diagnostic performance of the AI system is benchmarked against board-certified ophthalmologists of varying seniority on a standardized set of smartphone-captured anterior segment images. Second, the usability of the system is evaluated among non-medical users who perform self-administered screening with minimal instruction, with per-screening time recorded across consecutive attempts to characterize the learning curve. Third, a head-to-head field trial directly compares the on-device AI system against a functionally equivalent cloud-based deployment of the same model architecture across key operational dimensions including screening duration, diagnostic performance, and user acceptability. Fourth, population-level screening is conducted among consecutively enrolled community residents at two low-resource sites, with per-disease sensitivity and specificity calculated against reference-standard slit-lamp examinations. Fifth, pre-specified health-economic and environmental analyses compare the two deployment modalities in terms of per-person screening cost, cost-effectiveness, per-inference electricity consumption, and projected carbon emissions at scale.

The reference standard for all diagnostic comparisons is slit-lamp biomicroscopic examination performed by board-certified ophthalmologists. The study is designed and reported in accordance with the DECIDE-AI reporting guideline for early-stage clinical evaluation of AI-driven decision-support systems.

Przegląd badań

Typ studiów

Obserwacyjny

Zapisy (Szacowany)

3000

Kontakty i lokalizacje

Ta sekcja zawiera dane kontaktowe osób prowadzących badanie oraz informacje o tym, gdzie badanie jest przeprowadzane.

Kontakt w sprawie studiów

Kopia zapasowa kontaktu do badania

  • Nazwa: Longhui Li

Lokalizacje studiów

    • Guangdong
      • Guangzhou, Guangdong, Chiny, 510060
        • Rekrutacyjny
        • Zhongshan Ophthalmic Center, Sun Yat-sen University
        • Kontakt:

Kryteria uczestnictwa

Badacze szukają osób, które pasują do określonego opisu, zwanego kryteriami kwalifikacyjnymi. Niektóre przykłady tych kryteriów to ogólny stan zdrowia danej osoby lub wcześniejsze leczenie.

Kryteria kwalifikacji

Wiek uprawniający do nauki

  • Dorosły
  • Starszy dorosły

Akceptuje zdrowych ochotników

Tak

Metoda próbkowania

Próbka prawdopodobieństwa

Badana populacja

Participants are adults aged 18 years or older enrolled from four source populations: consecutive outpatients at a tertiary ophthalmic center and community hospitals; non-medical adults with no ophthalmic training recruited from a tertiary clinic for usability evaluation; community residents aged 60 years or older from two low-resource sites in western and southwestern China for field deployment evaluation; and board-certified ophthalmologists of varying seniority serving as diagnostic raters for AI benchmarking.

Opis

Inclusion Criteria:

  • Adults aged 18 years or older;
  • Willing to participate and able to provide written informed consent prior to enrollment.

Exclusion Criteria:

  • Unable to cooperate with anterior segment image capture (including smartphone-based photography or slit-lamp biomicroscopy).

Plan studiów

Ta sekcja zawiera szczegółowe informacje na temat planu badania, w tym sposób zaprojektowania badania i jego pomiary.

Jak projektuje się badanie?

Szczegóły projektu

Kohorty i interwencje

Grupa / Kohorta
Interwencja / Leczenie
On-Device Deployment Group
Participants screened using a smartphone-based AI system that performs all image inference locally on the device without requiring internet connectivity. The AI system analyzes smartphone-captured anterior segment images and generates diagnostic outputs entirely on the smartphone hardware, independent of cloud infrastructure or network access.
A structured-pruned one-stage object-detection model deployed as a standalone Android application, performing all image inference on-device without internet connectivity, designed to detect 16 anterior segment eye diseases from smartphone-captured images.
Cloud-Based Deployment Group
Participants screened using a functionally equivalent deployment of the same AI model architecture, in which smartphone-captured anterior segment images are transmitted to a remote server for inference, with diagnostic outputs returned via internet connection. This group serves as the comparator to evaluate the operational differences attributable solely to deployment modality, as the underlying model architecture is identical between the two groups.
A structured-pruned one-stage object-detection model deployed as a standalone Android application, performing all image inference on-device without internet connectivity, designed to detect 16 anterior segment eye diseases from smartphone-captured images.

Co mierzy badanie?

Podstawowe miary wyniku

Miara wyniku
Opis środka
Ramy czasowe
Case-level diagnostic accuracy of the AI system compared with board-certified ophthalmologists
Ramy czasowe: Day 1
Case-level accuracy is defined as the proportion of images with fully correct diagnostic labels concordant with the reference standard. The AI system and board-certified ophthalmologists stratified by clinical seniority (junior: fewer than 5 years of independent practice; intermediate: 5 to 15 years; senior: more than 15 years) independently evaluate the same standardized set of smartphone-captured anterior segment images, sampled to ensure balanced representation of all 16 disease categories and normal eyes. The reference standard for each image is established by a senior ophthalmologist with more than 30 years of clinical experience who does not participate in the benchmarking exercise. Clinicians are masked to the AI system output and to each other's assessments throughout.
Day 1
Diagnostic accuracy of the AI system when operated by non-medical users
Ramy czasowe: Day 1
Diagnostic accuracy is defined as the proportion of images correctly classified by non-medical users operating the AI system independently. Non-medical users, including patients and their family members attending the outpatient clinic of Zhongshan Ophthalmic Center, are instructed to install the AI application on their own smartphones, follow the in-app guidelines, and capture anterior segment images of an accompanying person to receive a screening result. All inference is performed on-device without internet connectivity. AI-generated diagnostic outputs are compared with reference-standard diagnoses obtained from subsequent slit-lamp examinations performed by ophthalmologists at the same clinic.
Day 1
Sensitivity of the on-device AI system in population-level community screening
Ramy czasowe: Day 1
Sensitivity is defined as the proportion of participants with a given anterior segment disease who are correctly identified as positive by the on-device AI system (true positives divided by the sum of true positives and false negatives), calculated separately for each target disease category. Village staff without medical training use the on-device AI system to screen consecutively enrolled local residents. The reference standard is established through slit-lamp examinations performed by ophthalmologists.
Day 1
Incremental cost-effectiveness ratio of on-device versus cloud-based screening
Ramy czasowe: Day 1
The incremental cost-effectiveness ratio (ICER) is defined as the difference in lifetime costs between on-device and cloud-based screening divided by the difference in quality-adjusted life-years (QALYs) between the two strategies, estimated from a pre-specified decision-analytic model comprising a decision tree with a downstream Markov state-transition structure applied to a simulated cohort of 100,000 individuals.
Day 1
Per-inference electricity consumption of on-device versus cloud-based deployment
Ramy czasowe: Day 1
Electricity consumption in joules per inference cycle is measured over 10,000 inference cycles on both the on-device smartphone and a standardized cloud-server configuration (Intel Xeon Gold 6248 CPU, NVIDIA Tesla V100 GPU) using the Experiment Impact Tracker toolkit. Cloud-server measurements are averaged across low-traffic and high-traffic server conditions. Results are reported separately for the on-device and cloud-based deployment modalities and compared as a ratio.
Day 1

Miary wyników drugorzędnych

Miara wyniku
Opis środka
Ramy czasowe
Total image evaluation time of the AI system compared with board-certified ophthalmologists
Ramy czasowe: Day 1
Total time in seconds required to complete independent evaluation of all images in the standardized benchmarking set, recorded separately for the AI system and for the board-certified ophthalmologists. Reported as mean with standard deviation for both the AI system and the clinician group.
Day 1
Per-screening time learning curve among non-medical operators
Ramy czasowe: Day 1
The duration of each screening attempt in seconds is recorded for analysis. Participants repeat the screening process and self-evaluate their operating proficiency after each attempt. The procedure is concluded once a participant deems himself or herself proficient for three consecutive attempts. The learning curve is characterized by the change in mean per-screening duration across consecutive attempts.
Day 1
Screening duration comparing on-device and cloud-based deployment in resource-limited field settings
Ramy czasowe: Day 1
Mean screening duration in seconds per participant, recorded for each of the two deployment modalities: on-device inference and cloud-based inference using a functionally equivalent deployment of the identical model architecture with inference performed on a remote server. Local residents are divided into two equal groups and screened by village staff without medical training using either the on-device or the cloud-based system.
Day 1
User acceptability comparing on-device and cloud-based deployment in resource-limited field settings
Ramy czasowe: Day 1
User acceptability is assessed using a questionnaire investigating patient satisfaction, privacy concerns, and willingness to recommend the AI system, administered to participants in both the on-device and cloud-based deployment groups. Local residents are divided into two equal groups and screened by village staff without medical training using either the on-device or the cloud-based system. Questionnaire responses are compared between the two deployment groups.
Day 1
Diagnostic accuracy of the on-device AI system in population-level community screening
Ramy czasowe: Day 1
Diagnostic accuracy is defined as the proportion of participants for whom the AI system assigns a correct diagnostic label concordant with the reference standard. Village staff without medical training use the on-device AI system to screen consecutively enrolled local residents. The reference standard is established through slit-lamp examinations performed by ophthalmologists.
Day 1
Specificity of the on-device AI system in population-level community screening
Ramy czasowe: Day 1
Specificity is defined as the proportion of participants without a given anterior segment disease who are correctly identified as negative by the on-device AI system (true negatives divided by the sum of true negatives and false positives), calculated separately for each target disease category. Village staff without medical training use the on-device AI system to screen consecutively enrolled local residents. The reference standard is established through slit-lamp examinations performed by ophthalmologists.
Day 1

Współpracownicy i badacze

Tutaj znajdziesz osoby i organizacje zaangażowane w to badanie.

Śledczy

  • Główny śledczy: Longhui Li, Zhongshan Ophthalmic Center, Sun Yat-sen University

Daty zapisu na studia

Daty te śledzą postęp w przesyłaniu rekordów badań i podsumowań wyników do ClinicalTrials.gov. Zapisy badań i zgłoszone wyniki są przeglądane przez National Library of Medicine (NLM), aby upewnić się, że spełniają określone standardy kontroli jakości, zanim zostaną opublikowane na publicznej stronie internetowej.

Główne daty studiów

Rozpoczęcie studiów (Rzeczywisty)

12 grudnia 2023

Zakończenie podstawowe (Szacowany)

1 maja 2028

Ukończenie studiów (Szacowany)

1 grudnia 2028

Daty rejestracji na studia

Pierwszy przesłany

28 maja 2026

Pierwszy przesłany, który spełnia kryteria kontroli jakości

8 czerwca 2026

Pierwszy wysłany (Rzeczywisty)

9 czerwca 2026

Aktualizacje rekordów badań

Ostatnia wysłana aktualizacja (Rzeczywisty)

9 czerwca 2026

Ostatnia przesłana aktualizacja, która spełniała kryteria kontroli jakości

8 czerwca 2026

Ostatnia weryfikacja

1 maja 2026

Więcej informacji

Terminy związane z tym badaniem

Plan dla danych uczestnika indywidualnego (IPD)

Planujesz udostępniać dane poszczególnych uczestników (IPD)?

NIE

Informacje o lekach i urządzeniach, dokumenty badawcze

Bada produkt leczniczy regulowany przez amerykańską FDA

Nie

Bada produkt urządzenia regulowany przez amerykańską FDA

Nie

Te informacje zostały pobrane bezpośrednio ze strony internetowej clinicaltrials.gov bez żadnych zmian. Jeśli chcesz zmienić, usunąć lub zaktualizować dane swojego badania, skontaktuj się z register@clinicaltrials.gov. Gdy tylko zmiana zostanie wprowadzona na stronie clinicaltrials.gov, zostanie ona automatycznie zaktualizowana również na naszej stronie internetowej .

Subskrybuj