Clinical Evaluation of Hearing Aid OTC Self-Assessment Methods for Adults With Perceived Mild-to-Moderate Hearing Loss

30 de julho de 2026 atualizado por: Rehear Audiology Company LTD

Clinical Evaluation of Two Rehear RH210 Self-Assessment Methods for Generating NAL-R-Based Prescribed Gain Compared With Conventional Audiometry in Adults With Perceived Mild-to-Moderate Hearing Loss

The goal of this clinical trial is to compare two methods of self assessment of hearing with conventional audiometry in an OTC Hearing Aid for adults with perceived mild to moderate hearing loss. The main questions it aims to answer are:

  1. To evaluate agreement between prescribed gain outputs generated from conventional audiometry and those generated from the self-assessment response method.
  2. Determine whether either self-assessment response method provides acceptable estimated audibility.
  3. Evaluate participant preference between the detection response method and the audibility-release response method.

Researchers will compare the two novel methods with conventional audiometry to determine levels of agreement.

Participants will self assess their hearing methods using two novel methods of hearing assessment as well as receive a conventional audiometric evaluation of hearing levels.

Visão geral do estudo

Descrição detalhada

The purpose of this clinical investigation is to evaluate two RH210 hearing aid software self-assessment response methods used to generate NAL-R-based prescribed gain for adults with perceived mild-to-moderate hearing loss. The study will compare prescribed gain outputs derived from each RH210 self-assessment response method with prescribed gain outputs derived from conventional audiometry using the same NAL-R prescription approach. The study will also assess whether the resulting prescribed gain provides estimated audibility using Speech Intelligibility Index (SII) analysis, and will evaluate participant preference between the two self-assessment response methods. The RH210A is an air-conduction hearing aid designed for adults with perceived mild-to-moderate sensorineural hearing loss.

The two RH210 self-assessment response methods evaluated in this study are:

  1. Detection response method (Option 1): the participant presses a response control when they hear the tone or beep.
  2. Audibility-release response method (Option 2): the participant maintains a press while the tone or beep is audible and releases the response control when they no longer hear it.

The study is designed to:

  1. Evaluate agreement between NAL-R prescribed gain outputs generated from conventional audiometry and NAL-R prescribed gain outputs generated from each RH210 self-assessment response method.
  2. Compare the detection response method and audibility-release response method with respect to prescribed gain agreement by ear and frequency.
  3. Assess estimated audibility associated with the self-assessment-derived NAL-R prescriptions using Speech Intelligibility Index (SII) analysis.
  4. Determine whether either self-assessment response method provides acceptable estimated audibility, including in cases where prescribed gain does not exactly match the conventional-audiometry-derived NAL-R prescription.
  5. Evaluate participant preference between the detection response method and the audibility-release response method.
  6. Assess general agreement, completeness, invalid-test rate, usability and response consistency for each self-assessment response method.
  7. Support validation of the Rehear assessment-to-fitting pathway without positioning the RH210 self-assessment function as a diagnostic audiometer.

This study is not intended to establish Human Factors performance. However, observations arising during execution of the assessment workflow may identify opportunities to improve the user interface, workflow or user guidance before the final Human Factors validation study. These observations are exploratory development outputs and are not formal study endpoints.

Tipo de estudo

Intervencional

Inscrição (Estimado)

24

Estágio

  • Não aplicável

Contactos e Locais

Esta seção fornece os detalhes de contato para aqueles que conduzem o estudo e informações sobre onde este estudo está sendo realizado.

Contato de estudo

Estude backup de contato

  • Nome: Sanna Hou, Master of Audiology
  • Número de telefone: +61 2 9412 6872
  • E-mail: sann.hou@nal.gov.au

Locais de estudo

    • New South Wales
      • Macquarie Park, New South Wales, Austrália, 2113
        • National Acoustic Laboratories
        • Contato:

Critérios de participação

Os pesquisadores procuram pessoas que se encaixem em uma determinada descrição, chamada de critérios de elegibilidade. Alguns exemplos desses critérios são a condição geral de saúde de uma pessoa ou tratamentos anteriores.

Critérios de elegibilidade

Idades elegíveis para estudo

  • Adulto
  • Adulto mais velho

Aceita Voluntários Saudáveis

Não

Descrição

Inclusion Criteria:

  • Age 18 years
  • Self-perceived mild-to-moderate hearing difficulties/ hearing loss
  • Competent smartphone use (able to complete self-fitting via the app)
  • Able and willing to provide informed consent
  • HHIA screening consistent with perceived hearing difficulty
  • Willingness and ability to attend two in-person research appointments
  • Ability to operate a SmartPhone

Exclusion Criteria:

  • Conductive or mixed hearing loss
  • Occluding cerumen
  • Otologic disease/ active ear pathology
  • Severe or profound hearing loss (PTA >S0dBHL) (WHO, 2021)
  • Presence of (red-flag) conditions relevant to OTC hearing aid use (per 21 CFR 801.421)
  • Visible deformity of the ear (congenital or traumatic).
  • Fluid, pus, or blood coming from the ear (e.g., active drainage).
  • Sudden or rapidly progressive hearing loss within the recent past (often defined as within ~90 days).
  • Any acute or chronic dizziness or vertigo associated with hearing issues.
  • Pain or discomfort in or around the ear.
  • Visible evidence of significant cerumen (ear wax) accumulation or foreign body in the ear canal.
  • Unilateral hearing loss of sudden onset or large asymmetry between ears.
  • Audiometric air-bone gap suggestive of possible conductive pathology (e.g., gap equal or more than 15 dB at key frequencies)
  • Visible evidence of significant cerumen (ear wax) accumulation or foreign body in the ear canal.
  • Unilateral hearing loss of sudden onset or large asymmetry between ears

Exclusion Criteria:

-

Plano de estudo

Esta seção fornece detalhes do plano de estudo, incluindo como o estudo é projetado e o que o estudo está medindo.

Como o estudo é projetado?

Detalhes do projeto

  • Finalidade Principal: Tratamento
  • Alocação: Randomizado
  • Modelo Intervencional: Atribuição cruzada
  • Mascaramento: Nenhum (rótulo aberto)

Armas e Intervenções

Grupo de Participantes / Braço
Intervenção / Tratamento
Comparador Ativo: Conventional manual pure-tone audiometry
The comparator for this clinical investigation is conventional manual pure-tone audiometry performed in accordance with ANSI/ASA S3.21-2004 (R2023), Methods for Manual Pure-Tone Threshold Audiometry, followed by generation of a NAL-R-based prescribed gain output.
Experimental: Mode A
RH210 self-test method 1: detection response method, where the participant presses when they hear the tone or beep

The Investigational Medical Device (IMD) consists of the RH210 hearing aid software self-test function accessed through the Rehear smartphone application and used in combination with the RH210A compatible in-ear hardware platform.

For this clinical investigation, the RH210A hardware platform is used to present calibrated test sounds (pure tones) to the participant during the app-based self-test. The RH210A is not being evaluated as a diagnostic audiometer, and the study is not designed to evaluate the hearing aid's amplification performance.

The two RH210 self-test response methods evaluated in this study are:

  • Detection response method: the participant presses the response control when they hear the tone or beep (Mode A).
  • Audibility-release response method: the participant maintains the response control while the tone or beep is audible and releases the response control when they no longer hear the tone or beep (Mode B).
Experimental: Mode B
RH210 self-test method 2: audibility-release response method, where the participant releases the response control when they no longer hear the tone or beep

The Investigational Medical Device (IMD) consists of the RH210 hearing aid software self-test function accessed through the Rehear smartphone application and used in combination with the RH210A compatible in-ear hardware platform.

For this clinical investigation, the RH210A hardware platform is used to present calibrated test sounds (pure tones) to the participant during the app-based self-test. The RH210A is not being evaluated as a diagnostic audiometer, and the study is not designed to evaluate the hearing aid's amplification performance.

The two RH210 self-test response methods evaluated in this study are:

  • Detection response method: the participant presses the response control when they hear the tone or beep (Mode A).
  • Audibility-release response method: the participant maintains the response control while the tone or beep is audible and releases the response control when they no longer hear the tone or beep (Mode B).

O que o estudo está medindo?

Medidas de resultados primários

Medida de resultado
Descrição da medida
Prazo
Agreement of NAL-R prescribed gain outputs (dB) between the comparator conditions
Prazo: Day 1

The primary endpoint is the agreement of NAL-R prescribed gain outputs (dB), by ear (R/L) and frequency (Hz), between:

  • Conventional-audiometry-derived NAL-R prescribed gain (dB), and
  • RH210 self-test-derived NAL-R prescribed gain (dB), generated using each of the two RH210 response methods.

The frequency-specific analysis will include: 500 Hz; 1 kHz; 2 kHz; and 4 kHz.

The two RH210 response methods are:

  • Detection response method: participant presses when they hear the tone or beep (Mode A).
  • Audibility-release response method: participant releases the response control when they no longer hear the tone or beep Mode B).
Day 1

Medidas de resultados secundários

Medida de resultado
Descrição da medida
Prazo
Number of Participants with Treatment-Related Adverse Events as classified by ISO14155:2026
Prazo: Day 1

The secondary safety objective is to assess the safety of participant exposure to the RH210 app-based self-test procedure when the RH210A in-ear hardware platform is used to present test sounds.

Safety will be assessed by the incidence, severity, seriousness and relationship to the investigational device or study procedure of:

  • adverse events;
  • adverse device effects;
  • anticipated adverse device effects;
  • discomfort associated with in-ear device use;
  • excessive loudness or sound intolerance;

No formal statistical hypothesis is proposed for safety. Safety outcomes will be summarized descriptively.

Day 1
Difference in Speech Intelligibility Index (SII) and SRT in quiet (CNC words) between the three conditions
Prazo: Day 1

The secondary objective is to evaluate whether RH210 self-test-derived NAL-R prescribed gain outputs (dB) provide estimated audibility comparable to the conventional-audiometry-derived NAL-R reference pathway.

The secondary endpoint is the difference in Speech Intelligibility Index / Audibility Index and SRT in quiet between:

(1) The conventional-audiometry-derived NAL-R prescribed gain condition (dB); and (2) Each RH210 self-test-derived NAL-R prescribed gain (dB) condition.

The analysis will be conducted under a pre-specified standard aided speech condition (SRT50% dB), for example adaptive speech in quiet SRT50, with the final method defined in the Statistical Analysis Plan

Day 1

Outras medidas de resultado

Medida de resultado
Descrição da medida
Prazo
Participant preference (A vs B questionnaire) will be assessed after completion of both self fit response methods.
Prazo: Day 1

The first exploratory objective is to evaluate participant preference between the two RH210 self-test response methods.

Participant preference will be assessed after completion of both RH210 response methods. The endpoint will include:

  • proportion of participants preferring the detection response method;
  • proportion preferring the audibility-release response method;
  • proportion reporting no preference;
  • reasons for preference, where collected. No formal hypothesis is proposed. Results will be summarized descriptively.
Day 1
Compare the practical performance of the two RH210 self-test response methods using NAL checklist.
Prazo: Day 1

Exploratory endpoints will include, for each response method:

  • completion rate;
  • invalid or unusable test rate;
  • missing frequency-specific data rate;
  • test duration;
  • participant-reported ease of use;
  • observed difficulty or need for clarification;
  • repeatability or response consistency, where repeat data are available.

No formal hypothesis is proposed. Results will be summarized descriptively.

Day 1

Colaboradores e Investigadores

É aqui que você encontrará pessoas e organizações envolvidas com este estudo.

Investigadores

  • Investigador principal: Bettina Turnbull, Doctor of Audiology, National Acoustic Laboratories

Datas de registro do estudo

Essas datas acompanham o progresso do registro do estudo e os envios de resumo dos resultados para ClinicalTrials.gov. Os registros do estudo e os resultados relatados são revisados ​​pela National Library of Medicine (NLM) para garantir que atendam aos padrões específicos de controle de qualidade antes de serem publicados no site público.

Datas Principais do Estudo

Início do estudo (Estimado)

15 de agosto de 2026

Conclusão Primária (Estimado)

1 de outubro de 2026

Conclusão do estudo (Estimado)

1 de outubro de 2026

Datas de inscrição no estudo

Enviado pela primeira vez

26 de julho de 2026

Enviado pela primeira vez que atendeu aos critérios de CQ

26 de julho de 2026

Primeira postagem (Real)

31 de julho de 2026

Atualizações de registro de estudo

Última Atualização Postada (Real)

3 de agosto de 2026

Última atualização enviada que atendeu aos critérios de controle de qualidade

30 de julho de 2026

Última verificação

1 de julho de 2026

Mais Informações

Termos relacionados a este estudo

Plano para dados de participantes individuais (IPD)

Planeja compartilhar dados de participantes individuais (IPD)?

NÃO

Descrição do plano IPD

The study is a commercial sponsor investigation of a product that is not yet cleared by regulatory agencies

Informações sobre medicamentos e dispositivos, documentos de estudo

Estuda um medicamento regulamentado pela FDA dos EUA

Não

Estuda um produto de dispositivo regulamentado pela FDA dos EUA

Não

Essas informações foram obtidas diretamente do site clinicaltrials.gov sem nenhuma alteração. Se você tiver alguma solicitação para alterar, remover ou atualizar os detalhes do seu estudo, entre em contato com register@clinicaltrials.gov. Assim que uma alteração for implementada em clinicaltrials.gov, ela também será atualizada automaticamente em nosso site .

Se inscrever