- ICH GCP
- Amerikanska kliniska prövningsregistret
- Klinisk prövning NCT07710365
Individualized Supramaximal Stimulation Current During Continuous Neuromonitoring in Thyroid Surgery (CIONM-SSC)
Patient-Specific Determination of Supramaximal Stimulation Current During Continuous Intraoperative Neuromonitoring of the Recurrent Laryngeal Nerve in Thyroid Surgery
Continuous intraoperative neuromonitoring (C-IONM) is increasingly used during thyroid surgery to detect early changes in recurrent laryngeal nerve function and reduce the risk of nerve injury. Current clinical practice recommends the use of fixed supramaximal stimulation currents; however, these recommendations are based on limited experimental evidence and may not account for individual variability between patients.
This retrospective observational cohort study aims to determine patient-specific supramaximal stimulation currents (SSC) during C-IONM and to evaluate whether individualized stimulation provides monitoring quality comparable to conventional fixed-current stimulation while potentially reducing unnecessary stimulation intensity. The study also investigates factors influencing electromyographic response amplitudes during surgery and assesses the safety of individualized stimulation parameters.
Studieöversikt
Status
Betingelser
Detaljerad beskrivning
Continuous intraoperative neuromonitoring (C-IONM) of the vagus nerve is increasingly used during thyroid surgery to facilitate early recognition of progressive recurrent laryngeal nerve (RLN) injury. Current international recommendations advocate the use of a fixed supramaximal stimulation current, typically between 1 and 2 mA. However, these recommendations are based on limited experimental and clinical evidence and do not account for potential interindividual differences in nerve excitability.
The aim of this retrospective observational cohort study was to characterize the distribution of patient-specific supramaximal stimulation current (SSC) during C-IONM and to determine whether individualized SSC selection maintains monitoring performance compared with conventional fixed-current stimulation. Secondary objectives included evaluation of factors affecting electromyographic response amplitude and assessment of the safety of individualized stimulation.
The study included consecutive adult patients undergoing thyroid surgery with C-IONM at a tertiary endocrine surgery center. Patients operated before implementation of an updated stimulation protocol constituted the control cohort and underwent monitoring using a fixed stimulation current of 1.5 mA. After protocol implementation, SSC was individually determined for each vagus nerve at the beginning of the procedure by gradually increasing stimulation current until no further increase in electromyographic response amplitude was observed. Continuous monitoring was subsequently performed using the individualized SSC.
Neuromonitoring recordings were exported from the monitoring system, anonymized, and analyzed using dedicated Python software developed for this study. The software identified individual stimulation events, excluded incomplete or technically inadequate recordings, and extracted stimulation current and electromyographic amplitude for quantitative analysis. More than one million individual stimulation responses were evaluated.
The primary objective was to determine the distribution of individualized SSC values and compare monitoring performance between individualized and fixed-current stimulation. Secondary analyses evaluated changes in response amplitude during surgery, variability between right and left vagus nerves, factors influencing amplitude, and the occurrence of monitoring-related adverse events.
Statistical analyses included descriptive statistics, non-parametric comparisons, regression analysis, and multivariable modeling to identify independent predictors of electromyographic response amplitude.
Studietyp
Inskrivning (Faktisk)
Kontakter och platser
Studieorter
-
-
-
Gdansk, Polen, 80-135
- The University Clinical Centre in Gdansk
-
-
Deltagandekriterier
Urvalskriterier
Åldrar som är berättigade till studier
- Vuxen
- Äldre vuxen
Tar emot friska volontärer
Testmetod
Studera befolkning
Beskrivning
Inclusion Criteria:
- Adults aged 18 years or older.
- Patients undergoing thyroid surgery with continuous intraoperative neuromonitoring (C-IONM) of the vagus nerve.
- Availability of complete intraoperative neuromonitoring recordings suitable for analysis.
- Availability of complete clinical and operative data.
Exclusion Criteria:
- Age younger than 18 years.
- Incomplete or technically inadequate neuromonitoring recordings.
- Loss of signal or recurrent laryngeal nerve palsy preventing reliable amplitude analysis.
- Missing or incomplete clinical data required for the study.
Studieplan
Hur är studien utformad?
Designdetaljer
Kohorter och interventioner
Grupp / Kohort |
Intervention / Behandling |
|---|---|
|
Control cohort
Patients who underwent thyroid surgery with continuous intraoperative neuromonitoring (C-IONM) using a fixed vagus nerve stimulation current of 1.5 mA throughout the procedure.
These patients were operated before implementation of the individualized stimulation protocol.
|
Continuous vagus nerve stimulation during thyroid surgery using a fixed supramaximal stimulation current of 1.5 mA.
|
|
Individualized SSC cohort
Patients who underwent thyroid surgery with continuous intraoperative neuromonitoring (C-IONM) using an individually determined supramaximal stimulation current (SSC).
SSC was established at the beginning of the procedure for each vagus nerve and subsequently used throughout continuous monitoring.
|
Continuous vagus nerve stimulation during thyroid surgery using an individually determined supramaximal stimulation current established at the beginning of the procedure.
|
Vad mäter studien?
Primära resultatmått
Resultatmått |
Åtgärdsbeskrivning |
Tidsram |
|---|---|---|
|
Patient-specific supramaximal stimulation current
Tidsram: During surgery
|
Distribution of the individualized supramaximal stimulation current required to achieve maximal electromyographic response during continuous intraoperative neuromonitoring of the vagus nerve.
|
During surgery
|
|
Patient-specific supramaximal stimulation current (SSC)
Tidsram: During surgery (at the beginning of continuous intraoperative neuromonitoring)
|
Distribution of the individualized supramaximal stimulation current (SSC) required to achieve maximal electromyographic response during continuous intraoperative neuromonitoring of the vagus nerve.
|
During surgery (at the beginning of continuous intraoperative neuromonitoring)
|
Sekundära resultatmått
Resultatmått |
Åtgärdsbeskrivning |
Tidsram |
|---|---|---|
|
Electromyographic response amplitude
Tidsram: During surgery
|
Comparison of electromyographic response amplitudes between individualized and fixed stimulation current protocols during continuous intraoperative neuromonitoring.
|
During surgery
|
Samarbetspartners och utredare
Sponsor
Studieavstämningsdatum
Studera stora datum
Studiestart (Faktisk)
Primärt slutförande (Faktisk)
Avslutad studie (Faktisk)
Studieregistreringsdatum
Först inskickad
Först inskickad som uppfyllde QC-kriterierna
Första postat (Faktisk)
Uppdateringar av studier
Senaste uppdatering publicerad (Faktisk)
Senaste inskickade uppdateringen som uppfyllde QC-kriterierna
Senast verifierad
Mer information
Termer relaterade till denna studie
Nyckelord
Ytterligare relevanta MeSH-villkor
Andra studie-ID-nummer
- KB/202/2025
Plan för individuella deltagardata (IPD)
Planerar du att dela individuella deltagardata (IPD)?
IPD-planbeskrivning
Läkemedels- och apparatinformation, studiedokument
Studerar en amerikansk FDA-reglerad läkemedelsprodukt
Studerar en amerikansk FDA-reglerad produktprodukt
Denna information hämtades direkt från webbplatsen clinicaltrials.gov utan några ändringar. Om du har några önskemål om att ändra, ta bort eller uppdatera dina studieuppgifter, vänligen kontakta register@clinicaltrials.gov. Så snart en ändring har implementerats på clinicaltrials.gov, kommer denna att uppdateras automatiskt även på vår webbplats .