Head-to-Head Comparison Study Between Different FDA Registered Allergy Skin Test Applicators

February 9, 2026 updated by: QHSLab, Inc.
This clinical study, titled "Head-to-Head Comparison Study between Different FDA Registered Allergy Skin Test Applicators," aims to compare the reliability and clinical performance of three skin prick test (SPT) devices: Allertest™ Multiple Skin Test Applicator, Lincoln Multi-Test II, and Greer Skintestor OMNI Applicator. The primary objective is to evaluate the consistency and accuracy of these devices in allergy testing.

Study Overview

Detailed Description

Background and Rationale:

Percutaneous allergy skin testing remains the gold standard diagnostic method for IgE-mediated allergic disease. Multiple-head skin prick test (SPT) devices enable simultaneous application of multiple allergens, improving testing efficiency while potentially reducing patient discomfort. However, significant inter-device variability exists among commercially available devices, with important implications for diagnostic accuracy and clinical interpretation.

Intra-device variability-inconsistent responses across individual test heads within a single device-represents a particularly concerning phenomenon that may lead to physician misinterpretation and potentially inappropriate treatment decisions. Such variability can be attributed to manufacturing factors including tooling mold precision, molding equipment quality, and process controls during production. Prior research has demonstrated that devices with lower finished product variability exhibit improved clinical performance characterized by reduced false-positive reactions to negative controls.

Study Design:

This prospective, single-visit, head-to-head device comparison study employed a paired design in which each participant served as their own control, receiving testing with two devices simultaneously-one device applied to each forearm. This within-subject design minimizes inter-individual variability and provides robust comparative data.

Participants were allocated into two comparison groups based on device head configuration: a 10-head device comparison (n=13) and an 8-head device comparison (n=17). Device assignment to right versus left forearm was randomized. Testing personnel were blinded to solution contents (histamine versus glycerin) during application, and a separate evaluator blinded to device assignment recorded wheal measurements.

Study Procedures:

All testing was performed by trained personnel in accordance with standard percutaneous skin testing protocols. Histamine dihydrochloride (1 mg/mL; ALK-Abello) served as the positive control and glycerin solution (ALK-Abello) served as the negative control. Devices were applied to the volar forearm surfaces with standardized pressure and technique. Test sites were maintained at least 2 cm apart to prevent cross-contamination between histamine and glycerin reactions.

Wheal measurements were obtained 15-20 minutes post-application using the mean diameter method (average of longest diameter and its perpendicular midpoint), consistent with established international guidelines. Pain assessment was performed immediately following device application.

Statistical Considerations:

Paired comparisons between devices were analyzed using Wilcoxon signed-rank tests for continuous variables and McNemar's test for dichotomous outcomes. Intra-device variability was quantified using the coefficient of variation (CV) calculated across test heads, with lower values indicating greater consistency. Exact binomial (Clopper-Pearson) confidence intervals were calculated for sensitivity and specificity estimates. Statistical significance was set at α=0.05 (two-sided).

Clinical Significance:

This study addresses an important knowledge gap regarding comparative clinical performance of multi-head allergy testing devices. Results provide evidence-based data to inform device selection decisions that may impact diagnostic accuracy, patient comfort, and reliability of allergen identification in clinical allergy practice.

Study Type

Interventional

Enrollment (Actual)

30

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Florida
      • Miami, Florida, United States, 33155
        • QHS Health center

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

  • Adult
  • Older Adult

Accepts Healthy Volunteers

Yes

Description

Inclusion Criteria:

  • Patients will be entered into this study only if they meet all of the following criteria:

    1. Written and signed informed consent obtained before starting any protocol-specific procedures.
    2. Subjects willing to withhold antihistamines, Leukotriene antagonists and H2 antagonists for at least 1 week before testing to avoid masking the histamine reactions on the skin.
    3. Male or female between 18 to 60 65 years, inclusive.
    4. Ability to comply with the study procedures and visit schedule.

Exclusion Criteria:

  • Patients will NOT be entered into this study if they meet any of the following criteria:

    1. Anaphylactic reaction (constitutional) after the previous skin test with the same allergen
    2. Acute fever
    3. Systemic disease with acute manifestation or decompensation
    4. Abnormal skin condition in the area to be tested (such as eczema sunburn condition after sunbathing)
    5. Pregnancy
    6. Presence of dermatographism, severe atopic dermatitis, or use of tricyclic antidepressants.
    7. Current use of beta-blockers
    8. Known hypersensitivity to any component of the test solutions (histamine or glycerin).
    9. Participation in another clinical study within 30 days before enrollment. Participation in another clinical study within 30 days before enrollment.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Diagnostic
  • Allocation: Randomized
  • Interventional Model: Crossover Assignment
  • Masking: Quadruple

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: AllerTest Device
Participants receive percutaneous allergy skin testing using an AllerTest multi-head device (either AllerTest-10 or AllerTest-8, manufactured by MedScience Research Group, Inc., distributed by ALK Abello, Inc.). The device is applied to one randomized forearm with a standardized histamine positive control (10 mg/mL histamine base) and a glycerin negative control (50% glycerin in saline). Outcomes measured include procedural pain (VAS 0-10), histamine wheal size at 15 minutes, glycerin wheal size at 15 minutes, sensitivity (proportion of histamine sites ≥3 mm), specificity (proportion of glycerin sites <3 mm), and intra-device variability (coefficient of variation across test heads).
FDA-cleared 10-head percutaneous skin test applicator manufactured by MedScience Research Group, Inc., distributed by ALK Abello, Inc. The device consists of 10 individual test heads arranged in a linear array, each with a standardized needle length designed to penetrate the epidermis without drawing blood. Applied with standardized pressure for simultaneous delivery of 10 test substances (histamine and glycerin controls in this study). Single-use, disposable device.
Other Names:
  • AllerTest 10-head
FDA-cleared 8-head percutaneous skin test applicator manufactured by MedScience Research Group, Inc., distributed by ALK Abello, Inc. The device consists of 8 individual test heads arranged in a linear array, each with a standardized needle length designed to penetrate the epidermis without drawing blood. Applied with standardized pressure for simultaneous delivery of 8 test substances (histamine and glycerin controls in this study). Single-use, disposable device.
Active Comparator: Comparator Device
Participants receive percutaneous allergy skin testing using a comparator multi-head device (either Skintestor OMNI manufactured by Greer Laboratories, Inc., or Multi-Test II manufactured by Lincoln Diagnostics, Inc.). The device is applied simultaneously to the opposite forearm with identical standardized histamine positive control (10 mg/mL histamine base) and glycerin negative control (50% glycerin in saline). The same outcomes are measured to enable direct paired comparison: procedural pain (VAS 0-10), histamine wheal size at 15 minutes, glycerin wheal size at 15 minutes, sensitivity, specificity, and intra-device variability.
FDA-cleared 10-head percutaneous skin test applicator manufactured by Greer Laboratories, Inc. The device consists of 10 individual test heads arranged in a linear array for simultaneous application of multiple allergens and controls. Applied with standardized pressure for delivery of histamine and glycerin controls in this study. Single-use, disposable device.
FDA-cleared 8-head percutaneous skin test applicator manufactured by Lincoln Diagnostics, Inc. The device consists of 8 individual test heads arranged in a linear array for simultaneous application of multiple allergens and controls. Applied with standardized pressure for delivery of histamine and glycerin controls in this study. Single-use, disposable device.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Pain Assessment Using Visual Analog Scale (VAS)
Time Frame: Immediately after device application (within 1 minute of skin test procedure)
Procedural pain assessed immediately after device application using a 0-10 Visual Analog Scale (VAS), where 0 represents no pain and 10 represents unspeakable pain. Lower scores indicate less pain and better tolerability.
Immediately after device application (within 1 minute of skin test procedure)
Sensitivity (Histamine Positive Control Performance)
Time Frame: 15 minutes after device application
Proportion of histamine control sites producing wheal ≥3 mm diameter at 15 minutes post-application. A wheal of ≥3 mm is the standard threshold for a positive allergic reaction. Higher sensitivity indicates better device performance in reliably eliciting positive control responses. Calculated as: (number of histamine sites with wheal ≥3 mm) / (total number of histamine sites tested) × 100%.
15 minutes after device application
Specificity (Glycerin Negative Control Performance)
Time Frame: 15 minutes after device application
Proportion of glycerin control sites producing wheal <3 mm diameter at 15 minutes post-application. Glycerin should not produce a wheal response; a result <3 mm confirms appropriate negative control performance and absence of non-specific mechanical irritation. Higher specificity indicates better device performance. Calculated as: (number of glycerin sites with wheal <3 mm) / (total number of glycerin sites tested) × 100%.
15 minutes after device application

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Intra-Device Variability
Time Frame: 15 minutes after device application
Coefficient of variation (CV) of histamine wheal sizes across all test heads within each device, calculated as CV = (standard deviation / mean) × 100%. Lower CV values indicate greater consistency and precision across test heads within the same device. This measure reflects the reliability of allergen delivery across multiple heads of a single device application.
15 minutes after device application
Mean Histamine Wheal Size
Time Frame: 15 minutes after device application
Mean diameter (in millimeters) of wheals produced by histamine positive control across all test heads of each device. Wheal diameter measured using the mean diameter method (average of longest diameter and its perpendicular midpoint). Larger wheals indicate stronger positive control response.
15 minutes after device application
Mean Glycerin Wheal Size
Time Frame: 15 minutes after device application
Mean diameter (in millimeters) of wheals produced by glycerin negative control across all test heads of each device. Wheal diameter measured using the mean diameter method (average of longest diameter and its perpendicular midpoint). Smaller measurements (ideally 0 mm) indicate appropriate negative control performance with minimal non-specific irritation.
15 minutes after device application
Proportion of Subjects with Clinically Significant Pain
Time Frame: Immediately after device application (within 1 minute of skin test procedure)
Proportion of participants experiencing more than mild pain, defined as a score greater than 4 on the Visual Analog Scale (VAS) for pain. The VAS ranges from 0 (no pain) to 10 (unspeakable pain), with higher scores indicating worse pain. A threshold of >4 represents approximately 40% of the scale and indicates clinically meaningful discomfort. Lower percentages indicate better device tolerability across the study population. Reported as percentage with 95% exact binomial confidence intervals.
Immediately after device application (within 1 minute of skin test procedure)

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Sponsor

Collaborators

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

General Publications

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

November 12, 2025

Primary Completion (Actual)

November 30, 2025

Study Completion (Actual)

January 30, 2026

Study Registration Dates

First Submitted

February 3, 2026

First Submitted That Met QC Criteria

February 3, 2026

First Posted (Actual)

February 10, 2026

Study Record Updates

Last Update Posted (Actual)

February 12, 2026

Last Update Submitted That Met QC Criteria

February 9, 2026

Last Verified

February 1, 2026

More Information

Terms related to this study

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

NO

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

Yes

product manufactured in and exported from the U.S.

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.