Prospective Evaluation of iCam-OCT

September 18, 2026 updated by: Siloam Vision

This is a prospective, non-significant-risk clinical study sponsored by Siloam Vision, Inc. to evaluate whether ophthalmologists can assess retinopathy of prematurity (ROP) using images from the iCam-OCT device as effectively as with the current standard, binocular indirect ophthalmoscopy (BIO). Results will support an FDA 510(k) submission. A separate follow-up study (undilated OCT imaging, with dilation only before the BIO exam) is planned once this study completes.

Device. The iCam-OCT is a handheld, swept-source OCT (SS-OCT) system operating at 1040nm and 400kHz, with 7-micron depth resolution and a 12mm scan depth. It captures an ultra-widefield (197°) field of view - the widest reported for OCT - in a contact-based, non-mydriatic mode, with scans under one second and a full bilateral exam averaging 1-2 minutes. It's designed for supine imaging of infants/small children who can't use conventional tabletop OCT. It's a Class 1 laser product conforming to IEC 60601-1, IEC 60601-1-2, IEC 60825-1, and ANSI Z80.36-2021. Over a dozen prior publications using earlier iCam-OCT prototypes (500+ patients, 1700+ imaging sessions) reported no adverse events and found the images clinically useful for ROP staging, zone assessment, and detection of related pathologies.

Sites and population. Two sites: Oregon Health & Science University (Doernbecher NICU) and Children's Hospital Colorado Anschutz. Both sites are new to the device and unaffiliated with the sponsor. Enrollment target is a minimum of 40 premature infants meeting AAP ROP-screening criteria (birth weight <1500g or gestational age ≤30 weeks; or birth weight 1500-2000g or gestational age >30 weeks if deemed at-risk). At least 30 infants must have ROP (stage 1 or worse in any zone) and at least 10 without ROP, with at least 20 subjects per site. Main exclusions are inability to obtain parental consent or clinical instability precluding research imaging.

Study design. After informed consent, each infant undergoes both a BIO exam (by an independent, ROP-experienced ophthalmologist unaffiliated with the sponsor) and iCam-OCT imaging, in randomized order, following pupil dilation per routine care. Imaging is performed by a trained/certified operator (ophthalmologist, nurse, technician, or coordinator) using a standardized protocol with up to six suggested scan views (posterior, temporal, nasal, temporal/nasal ora serrata, posterior HD). Anonymized OCT images are then graded independently by three masked, ROP-experienced ophthalmologists for image quality (0-3 scale) and clinical utility (usefulness for confirming normal structures or detecting pathology), and separately for ROP zone/stage/plus disease, without knowledge of the BIO findings.

Endpoints and performance goals.

Primary: percent of OCT image sets graded acceptable (score ≥2) for both quality and clinical utility - goal ≥85%.

Secondary: sensitivity/specificity of OCT-based ROP detection vs. BIO (goals >90% sensitivity, >70% specificity); documentation/visualization of full retinal vascularization to the ora serrata (same goals); and documentation of the vascular/avascular border in eyes with incomplete vascularization (same goals).

Safety: tracking of systemic AEs (nurse-initiated pauses), ocular AEs (corneal abrasion, conjunctival irritation, infection, need for topical medication), and any other AEs.

Discordant cases - infants classified "no ROP" by BIO but flagged as stage 1+ by OCT (potential false positives relative to the BIO gold standard) - get extra scrutiny: a coordinator reviews subsequent hospital records for later-diagnosed ROP, and all three graders reconvene to reach consensus on whether the OCT findings represent true early-stage ROP that BIO missed, versus immature retina with a defined vascular border. Clinical management throughout the study is driven by BIO only, not by OCT findings.

Statistics. The eye is the unit of analysis; each eye is graded independently by all three masked graders, and sensitivity/specificity are computed per grader and averaged. Sample size (n=30 for the ROP-positive group) was powered via a non-inferiority test assuming 85% BIO sensitivity (from a cited 2018 Biten et al. study), a 5% non-inferiority margin, and one-sided alpha of 0.05, yielding ~80% power. Supportive analyses include per-grader McNemar's tests for OCT/BIO discordance and non-inferiority testing of OCT sensitivity against the fixed 85% literature benchmark (not a within-study BIO comparison).

Safety monitoring. Risks are characterized as low (corneal abrasion, subconjunctival hemorrhage, infection - similar to standard contact ROP exams) plus general risks of apnea/bradycardia inherent to examining fragile premature infants. Dr. Benjamin Young serves as the independent safety monitor overseeing adverse event review; AEs/SAEs must be reported to the IRB within 5 days.

Study Overview

Status

Completed

Intervention / Treatment

Study Type

Observational

Enrollment (Actual)

68

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Oregon
      • Portland, Oregon, United States, 97239
        • OHSU

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

4 weeks and older (Child, Adult, Older Adult)

Accepts Healthy Volunteers

Yes

Genders Eligible for Study

All

Sampling Method

Non-Probability Sample

Study Population

The study population is premature infants indicated for ROP screening - approximately 40 such infants will be enrolled in total, matching AAP screening guidelines (birth weight <1500g or gestational age ≤30 weeks; or birth weight 1500-2000g or gestational age >30 weeks if judged at-risk by the attending pediatrician or neonatologist).

Description

Inclusion Criteria:The patient must be an infant matching American Academy of Pediatrics (AAP) screening criteria for ROP evaluation, defined as one of the following:

  • A birth weight of <1500g, or a gestational age of ≤30 weeks; or
  • A birth weight between 1500-2000g, or
  • A gestational age of >30 weeks, in infants who are believed to be at risk by their attending pediatrician or neonatologist.

Exclusion Criteria:

  • Inability to obtain informed consent from the parent or guardian or
  • The infant is deemed clinically unstable for research imaging by the neonatology team.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Observational Models: COHORT
  • Time Perspectives: PROSPECTIVE

Cohorts and Interventions

Group / Cohort
Intervention / Treatment
Patient cohort

The study participants are premature infants who meet the American Academy of Pediatrics (AAP) screening criteria for ROP evaluation. Specifically, eligible infants must have either a birth weight under 1500g or a gestational age of 30 weeks or less, or, alternatively, a birth weight between 1500-2000g or a gestational age over 30 weeks if their attending pediatrician or neonatologist believes they're at risk for ROP.

enrolled infant undergoes both a BIO exam and iCam-OCT imaging (order randomized) during the same dilated eye exam

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Clinical Quality
Time Frame: study visit - day 1

The primary outcome measure is the clinical quality and utility of iCam-OCT images, as judged by three independent, masked, ROP-experienced ophthalmologist graders.

study visit - day 1

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Presence or absence of ROP
Time Frame: study visit - day 1

Agreement between OCT image grading and BIO exam findings on whether ROP is present, defined as stage 1 or worse in any zone, with or without pre-plus or plus disease. Measured as sensitivity (target >90%) and specificity (target >70%), with BIO as the reference standard.

study visit - day 1

Other Outcome Measures

Outcome Measure
Measure Description
Time Frame
Full vascularization documentation
Time Frame: study visit - day 1

In eyes that BIO determines to be fully vascularized, whether OCT can similarly document full retinal vascularization - specifically, visualization out to zone III and within two disc diameters of the ora serrata. Measured as sensitivity (target >90%) and specificity (target >70%); this assessment requires visualization of the ora serrata on the OCT images.

study visit - day 1
Vascular/avascular border documentation
Time Frame: study visit - day 1

In eyes with incomplete retinal vascularization (areas of avascular retina) per BIO, whether OCT can document the vascular/avascular border. Again measured as sensitivity (target >90%) and specificity (target >70%). Visibility of the ora serrata is not required for this particular assessment.

study visit - day 1

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Sponsor

Investigators

  • Study Chair: John P Campbell, MD, Siloam Vision

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

January 28, 2026

Primary Completion (Actual)

July 15, 2026

Study Completion (Actual)

August 15, 2026

Study Registration Dates

First Submitted

September 14, 2026

First Submitted That Met QC Criteria

September 18, 2026

First Posted (Actual)

September 24, 2026

Study Record Updates

Last Update Posted (Actual)

September 24, 2026

Last Update Submitted That Met QC Criteria

September 18, 2026

Last Verified

December 1, 2025

More Information

Terms related to this study

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

YES

IPD Plan Description

There is a plan to make IPD and related data dictionaries available.

IPD Sharing Supporting Information Type

  • STUDY_PROTOCOL
  • SAP
  • ICF
  • CSR

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

Yes

product manufactured in and exported from the U.S.

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.