Questa pagina è stata tradotta automaticamente e l'accuratezza della traduzione non è garantita. Si prega di fare riferimento al Versione inglese per un testo di partenza.

Prospective Evaluation of iCam-OCT

18 settembre 2026 aggiornato da: Siloam Vision

This is a prospective, non-significant-risk clinical study sponsored by Siloam Vision, Inc. to evaluate whether ophthalmologists can assess retinopathy of prematurity (ROP) using images from the iCam-OCT device as effectively as with the current standard, binocular indirect ophthalmoscopy (BIO). Results will support an FDA 510(k) submission. A separate follow-up study (undilated OCT imaging, with dilation only before the BIO exam) is planned once this study completes.

Device. The iCam-OCT is a handheld, swept-source OCT (SS-OCT) system operating at 1040nm and 400kHz, with 7-micron depth resolution and a 12mm scan depth. It captures an ultra-widefield (197°) field of view - the widest reported for OCT - in a contact-based, non-mydriatic mode, with scans under one second and a full bilateral exam averaging 1-2 minutes. It's designed for supine imaging of infants/small children who can't use conventional tabletop OCT. It's a Class 1 laser product conforming to IEC 60601-1, IEC 60601-1-2, IEC 60825-1, and ANSI Z80.36-2021. Over a dozen prior publications using earlier iCam-OCT prototypes (500+ patients, 1700+ imaging sessions) reported no adverse events and found the images clinically useful for ROP staging, zone assessment, and detection of related pathologies.

Sites and population. Two sites: Oregon Health & Science University (Doernbecher NICU) and Children's Hospital Colorado Anschutz. Both sites are new to the device and unaffiliated with the sponsor. Enrollment target is a minimum of 40 premature infants meeting AAP ROP-screening criteria (birth weight <1500g or gestational age ≤30 weeks; or birth weight 1500-2000g or gestational age >30 weeks if deemed at-risk). At least 30 infants must have ROP (stage 1 or worse in any zone) and at least 10 without ROP, with at least 20 subjects per site. Main exclusions are inability to obtain parental consent or clinical instability precluding research imaging.

Study design. After informed consent, each infant undergoes both a BIO exam (by an independent, ROP-experienced ophthalmologist unaffiliated with the sponsor) and iCam-OCT imaging, in randomized order, following pupil dilation per routine care. Imaging is performed by a trained/certified operator (ophthalmologist, nurse, technician, or coordinator) using a standardized protocol with up to six suggested scan views (posterior, temporal, nasal, temporal/nasal ora serrata, posterior HD). Anonymized OCT images are then graded independently by three masked, ROP-experienced ophthalmologists for image quality (0-3 scale) and clinical utility (usefulness for confirming normal structures or detecting pathology), and separately for ROP zone/stage/plus disease, without knowledge of the BIO findings.

Endpoints and performance goals.

Primary: percent of OCT image sets graded acceptable (score ≥2) for both quality and clinical utility - goal ≥85%.

Secondary: sensitivity/specificity of OCT-based ROP detection vs. BIO (goals >90% sensitivity, >70% specificity); documentation/visualization of full retinal vascularization to the ora serrata (same goals); and documentation of the vascular/avascular border in eyes with incomplete vascularization (same goals).

Safety: tracking of systemic AEs (nurse-initiated pauses), ocular AEs (corneal abrasion, conjunctival irritation, infection, need for topical medication), and any other AEs.

Discordant cases - infants classified "no ROP" by BIO but flagged as stage 1+ by OCT (potential false positives relative to the BIO gold standard) - get extra scrutiny: a coordinator reviews subsequent hospital records for later-diagnosed ROP, and all three graders reconvene to reach consensus on whether the OCT findings represent true early-stage ROP that BIO missed, versus immature retina with a defined vascular border. Clinical management throughout the study is driven by BIO only, not by OCT findings.

Statistics. The eye is the unit of analysis; each eye is graded independently by all three masked graders, and sensitivity/specificity are computed per grader and averaged. Sample size (n=30 for the ROP-positive group) was powered via a non-inferiority test assuming 85% BIO sensitivity (from a cited 2018 Biten et al. study), a 5% non-inferiority margin, and one-sided alpha of 0.05, yielding ~80% power. Supportive analyses include per-grader McNemar's tests for OCT/BIO discordance and non-inferiority testing of OCT sensitivity against the fixed 85% literature benchmark (not a within-study BIO comparison).

Safety monitoring. Risks are characterized as low (corneal abrasion, subconjunctival hemorrhage, infection - similar to standard contact ROP exams) plus general risks of apnea/bradycardia inherent to examining fragile premature infants. Dr. Benjamin Young serves as the independent safety monitor overseeing adverse event review; AEs/SAEs must be reported to the IRB within 5 days.

Panoramica dello studio

Stato

Completato

Intervento / Trattamento

Tipo di studio

Osservativo

Iscrizione (Effettivo)

68

Contatti e Sedi

Questa sezione fornisce i recapiti di coloro che conducono lo studio e informazioni su dove viene condotto lo studio.

Luoghi di studio

    • Oregon
      • Portland, Oregon, Stati Uniti, 97239
        • OHSU

Criteri di partecipazione

I ricercatori cercano persone che corrispondano a una certa descrizione, chiamata criteri di ammissibilità. Alcuni esempi di questi criteri sono le condizioni generali di salute di una persona o trattamenti precedenti.

Criteri di ammissibilità

Età idonea allo studio

4 settimane e precedenti (Bambino, Adulto, Adulto più anziano)

Accetta volontari sani

Sì

Sessi ammissibili allo studio

Tutto

Metodo di campionamento

Campione non probabilistico

Popolazione di studio

The study population is premature infants indicated for ROP screening - approximately 40 such infants will be enrolled in total, matching AAP screening guidelines (birth weight <1500g or gestational age ≤30 weeks; or birth weight 1500-2000g or gestational age >30 weeks if judged at-risk by the attending pediatrician or neonatologist).

Descrizione

Inclusion Criteria:The patient must be an infant matching American Academy of Pediatrics (AAP) screening criteria for ROP evaluation, defined as one of the following:

  • A birth weight of <1500g, or a gestational age of ≤30 weeks; or
  • A birth weight between 1500-2000g, or
  • A gestational age of >30 weeks, in infants who are believed to be at risk by their attending pediatrician or neonatologist.

Exclusion Criteria:

  • Inability to obtain informed consent from the parent or guardian or
  • The infant is deemed clinically unstable for research imaging by the neonatology team.

Piano di studio

Questa sezione fornisce i dettagli del piano di studio, compreso il modo in cui lo studio è progettato e ciò che lo studio sta misurando.

Come è strutturato lo studio?

Dettagli di progettazione

  • Modelli osservazionali: COHORT
  • Prospettive temporali: PROSPECTIVE

Coorti e interventi

Gruppo / Coorte
Intervento / Trattamento
Patient cohort

The study participants are premature infants who meet the American Academy of Pediatrics (AAP) screening criteria for ROP evaluation. Specifically, eligible infants must have either a birth weight under 1500g or a gestational age of 30 weeks or less, or, alternatively, a birth weight between 1500-2000g or a gestational age over 30 weeks if their attending pediatrician or neonatologist believes they're at risk for ROP.

enrolled infant undergoes both a BIO exam and iCam-OCT imaging (order randomized) during the same dilated eye exam

Cosa sta misurando lo studio?

Misure di risultato primarie

Misura del risultato
Misura Descrizione
Lasso di tempo
Clinical Quality
Lasso di tempo: study visit - day 1

The primary outcome measure is the clinical quality and utility of iCam-OCT images, as judged by three independent, masked, ROP-experienced ophthalmologist graders.

study visit - day 1

Misure di risultato secondarie

Misura del risultato
Misura Descrizione
Lasso di tempo
Presence or absence of ROP
Lasso di tempo: study visit - day 1

Agreement between OCT image grading and BIO exam findings on whether ROP is present, defined as stage 1 or worse in any zone, with or without pre-plus or plus disease. Measured as sensitivity (target >90%) and specificity (target >70%), with BIO as the reference standard.

study visit - day 1

Altre misure di risultato

Misura del risultato
Misura Descrizione
Lasso di tempo
Full vascularization documentation
Lasso di tempo: study visit - day 1

In eyes that BIO determines to be fully vascularized, whether OCT can similarly document full retinal vascularization - specifically, visualization out to zone III and within two disc diameters of the ora serrata. Measured as sensitivity (target >90%) and specificity (target >70%); this assessment requires visualization of the ora serrata on the OCT images.

study visit - day 1
Vascular/avascular border documentation
Lasso di tempo: study visit - day 1

In eyes with incomplete retinal vascularization (areas of avascular retina) per BIO, whether OCT can document the vascular/avascular border. Again measured as sensitivity (target >90%) and specificity (target >70%). Visibility of the ora serrata is not required for this particular assessment.

study visit - day 1

Collaboratori e investigatori

Qui è dove troverai le persone e le organizzazioni coinvolte in questo studio.

Sponsor

Investigatori

  • Cattedra di studio: John P Campbell, MD, Siloam Vision

Studiare le date dei record

Queste date tengono traccia dell'avanzamento della registrazione dello studio e dell'invio dei risultati di sintesi a ClinicalTrials.gov. I record degli studi e i risultati riportati vengono esaminati dalla National Library of Medicine (NLM) per assicurarsi che soddisfino specifici standard di controllo della qualità prima di essere pubblicati sul sito Web pubblico.

Studia le date principali

Inizio studio (Effettivo)

28 gennaio 2026

Completamento primario (Effettivo)

15 luglio 2026

Completamento dello studio (Effettivo)

15 agosto 2026

Date di iscrizione allo studio

Primo inviato

14 settembre 2026

Primo inviato che soddisfa i criteri di controllo qualità

18 settembre 2026

Primo Inserito (Effettivo)

24 settembre 2026

Aggiornamenti dei record di studio

Ultimo aggiornamento pubblicato (Effettivo)

24 settembre 2026

Ultimo aggiornamento inviato che soddisfa i criteri QC

18 settembre 2026

Ultimo verificato

1 dicembre 2025

Maggiori informazioni

Termini relativi a questo studio

Piano per i dati dei singoli partecipanti (IPD)

Hai intenzione di condividere i dati dei singoli partecipanti (IPD)?

SÌ

Descrizione del piano IPD

There is a plan to make IPD and related data dictionaries available.

Tipo di informazioni di supporto alla condivisione IPD

  • STUDIO_PROTOCOLLO
  • LINFA
  • ICF
  • RSI

Informazioni su farmaci e dispositivi, documenti di studio

Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti

No

Studia un dispositivo regolamentato dalla FDA degli Stati Uniti

Sì

prodotto fabbricato ed esportato dagli Stati Uniti

No

Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .