Diese Seite wurde automatisch übersetzt und die Genauigkeit der Übersetzung wird nicht garantiert. Bitte wende dich an die englische Version für einen Quelltext.

Prospective Evaluation of iCam-OCT

18. September 2026 aktualisiert von: Siloam Vision

This is a prospective, non-significant-risk clinical study sponsored by Siloam Vision, Inc. to evaluate whether ophthalmologists can assess retinopathy of prematurity (ROP) using images from the iCam-OCT device as effectively as with the current standard, binocular indirect ophthalmoscopy (BIO). Results will support an FDA 510(k) submission. A separate follow-up study (undilated OCT imaging, with dilation only before the BIO exam) is planned once this study completes.

Device. The iCam-OCT is a handheld, swept-source OCT (SS-OCT) system operating at 1040nm and 400kHz, with 7-micron depth resolution and a 12mm scan depth. It captures an ultra-widefield (197°) field of view - the widest reported for OCT - in a contact-based, non-mydriatic mode, with scans under one second and a full bilateral exam averaging 1-2 minutes. It's designed for supine imaging of infants/small children who can't use conventional tabletop OCT. It's a Class 1 laser product conforming to IEC 60601-1, IEC 60601-1-2, IEC 60825-1, and ANSI Z80.36-2021. Over a dozen prior publications using earlier iCam-OCT prototypes (500+ patients, 1700+ imaging sessions) reported no adverse events and found the images clinically useful for ROP staging, zone assessment, and detection of related pathologies.

Sites and population. Two sites: Oregon Health & Science University (Doernbecher NICU) and Children's Hospital Colorado Anschutz. Both sites are new to the device and unaffiliated with the sponsor. Enrollment target is a minimum of 40 premature infants meeting AAP ROP-screening criteria (birth weight <1500g or gestational age ≤30 weeks; or birth weight 1500-2000g or gestational age >30 weeks if deemed at-risk). At least 30 infants must have ROP (stage 1 or worse in any zone) and at least 10 without ROP, with at least 20 subjects per site. Main exclusions are inability to obtain parental consent or clinical instability precluding research imaging.

Study design. After informed consent, each infant undergoes both a BIO exam (by an independent, ROP-experienced ophthalmologist unaffiliated with the sponsor) and iCam-OCT imaging, in randomized order, following pupil dilation per routine care. Imaging is performed by a trained/certified operator (ophthalmologist, nurse, technician, or coordinator) using a standardized protocol with up to six suggested scan views (posterior, temporal, nasal, temporal/nasal ora serrata, posterior HD). Anonymized OCT images are then graded independently by three masked, ROP-experienced ophthalmologists for image quality (0-3 scale) and clinical utility (usefulness for confirming normal structures or detecting pathology), and separately for ROP zone/stage/plus disease, without knowledge of the BIO findings.

Endpoints and performance goals.

Primary: percent of OCT image sets graded acceptable (score ≥2) for both quality and clinical utility - goal ≥85%.

Secondary: sensitivity/specificity of OCT-based ROP detection vs. BIO (goals >90% sensitivity, >70% specificity); documentation/visualization of full retinal vascularization to the ora serrata (same goals); and documentation of the vascular/avascular border in eyes with incomplete vascularization (same goals).

Safety: tracking of systemic AEs (nurse-initiated pauses), ocular AEs (corneal abrasion, conjunctival irritation, infection, need for topical medication), and any other AEs.

Discordant cases - infants classified "no ROP" by BIO but flagged as stage 1+ by OCT (potential false positives relative to the BIO gold standard) - get extra scrutiny: a coordinator reviews subsequent hospital records for later-diagnosed ROP, and all three graders reconvene to reach consensus on whether the OCT findings represent true early-stage ROP that BIO missed, versus immature retina with a defined vascular border. Clinical management throughout the study is driven by BIO only, not by OCT findings.

Statistics. The eye is the unit of analysis; each eye is graded independently by all three masked graders, and sensitivity/specificity are computed per grader and averaged. Sample size (n=30 for the ROP-positive group) was powered via a non-inferiority test assuming 85% BIO sensitivity (from a cited 2018 Biten et al. study), a 5% non-inferiority margin, and one-sided alpha of 0.05, yielding ~80% power. Supportive analyses include per-grader McNemar's tests for OCT/BIO discordance and non-inferiority testing of OCT sensitivity against the fixed 85% literature benchmark (not a within-study BIO comparison).

Safety monitoring. Risks are characterized as low (corneal abrasion, subconjunctival hemorrhage, infection - similar to standard contact ROP exams) plus general risks of apnea/bradycardia inherent to examining fragile premature infants. Dr. Benjamin Young serves as the independent safety monitor overseeing adverse event review; AEs/SAEs must be reported to the IRB within 5 days.

Studienübersicht

Status

Abgeschlossen

Intervention / Behandlung

Studientyp

Beobachtungs

Einschreibung (Tatsächlich)

68

Kontakte und Standorte

Dieser Abschnitt enthält die Kontaktdaten derjenigen, die die Studie durchführen, und Informationen darüber, wo diese Studie durchgeführt wird.

Studienorte

Teilnahmekriterien

Forscher suchen nach Personen, die einer bestimmten Beschreibung entsprechen, die als Auswahlkriterien bezeichnet werden. Einige Beispiele für diese Kriterien sind der allgemeine Gesundheitszustand einer Person oder frühere Behandlungen.

Zulassungskriterien

Studienberechtigtes Alter

4 Wochen und älter (Kind, Erwachsene, Älterer Erwachsener)

Akzeptiert gesunde Freiwillige

Ja

Studienberechtigte Geschlechter

Alle

Probenahmeverfahren

Nicht-Wahrscheinlichkeitsprobe

Studienpopulation

The study population is premature infants indicated for ROP screening - approximately 40 such infants will be enrolled in total, matching AAP screening guidelines (birth weight <1500g or gestational age ≤30 weeks; or birth weight 1500-2000g or gestational age >30 weeks if judged at-risk by the attending pediatrician or neonatologist).

Beschreibung

Inclusion Criteria:The patient must be an infant matching American Academy of Pediatrics (AAP) screening criteria for ROP evaluation, defined as one of the following:

  • A birth weight of <1500g, or a gestational age of ≤30 weeks; or
  • A birth weight between 1500-2000g, or
  • A gestational age of >30 weeks, in infants who are believed to be at risk by their attending pediatrician or neonatologist.

Exclusion Criteria:

  • Inability to obtain informed consent from the parent or guardian or
  • The infant is deemed clinically unstable for research imaging by the neonatology team.

Studienplan

Dieser Abschnitt enthält Einzelheiten zum Studienplan, einschließlich des Studiendesigns und der Messung der Studieninhalte.

Wie ist die Studie aufgebaut?

Designdetails

  • Beobachtungsmodelle: COHORT
  • Zeitperspektiven: PROSPECTIVE

Kohorten und Interventionen

Gruppe / Kohorte
Intervention / Behandlung
Patient cohort

The study participants are premature infants who meet the American Academy of Pediatrics (AAP) screening criteria for ROP evaluation. Specifically, eligible infants must have either a birth weight under 1500g or a gestational age of 30 weeks or less, or, alternatively, a birth weight between 1500-2000g or a gestational age over 30 weeks if their attending pediatrician or neonatologist believes they're at risk for ROP.

enrolled infant undergoes both a BIO exam and iCam-OCT imaging (order randomized) during the same dilated eye exam

Was misst die Studie?

Primäre Ergebnismessungen

Ergebnis Maßnahme
Maßnahmenbeschreibung
Zeitfenster
Clinical Quality
Zeitfenster: study visit - day 1

The primary outcome measure is the clinical quality and utility of iCam-OCT images, as judged by three independent, masked, ROP-experienced ophthalmologist graders.

study visit - day 1

Sekundäre Ergebnismessungen

Ergebnis Maßnahme
Maßnahmenbeschreibung
Zeitfenster
Presence or absence of ROP
Zeitfenster: study visit - day 1

Agreement between OCT image grading and BIO exam findings on whether ROP is present, defined as stage 1 or worse in any zone, with or without pre-plus or plus disease. Measured as sensitivity (target >90%) and specificity (target >70%), with BIO as the reference standard.

study visit - day 1

Andere Ergebnismessungen

Ergebnis Maßnahme
Maßnahmenbeschreibung
Zeitfenster
Full vascularization documentation
Zeitfenster: study visit - day 1

In eyes that BIO determines to be fully vascularized, whether OCT can similarly document full retinal vascularization - specifically, visualization out to zone III and within two disc diameters of the ora serrata. Measured as sensitivity (target >90%) and specificity (target >70%); this assessment requires visualization of the ora serrata on the OCT images.

study visit - day 1
Vascular/avascular border documentation
Zeitfenster: study visit - day 1

In eyes with incomplete retinal vascularization (areas of avascular retina) per BIO, whether OCT can document the vascular/avascular border. Again measured as sensitivity (target >90%) and specificity (target >70%). Visibility of the ora serrata is not required for this particular assessment.

study visit - day 1

Mitarbeiter und Ermittler

Hier finden Sie Personen und Organisationen, die an dieser Studie beteiligt sind.

Sponsor

Ermittler

  • Studienstuhl: John P Campbell, MD, Siloam Vision

Studienaufzeichnungsdaten

Diese Daten verfolgen den Fortschritt der Übermittlung von Studienaufzeichnungen und zusammenfassenden Ergebnissen an ClinicalTrials.gov. Studienaufzeichnungen und gemeldete Ergebnisse werden von der National Library of Medicine (NLM) überprüft, um sicherzustellen, dass sie bestimmten Qualitätskontrollstandards entsprechen, bevor sie auf der öffentlichen Website veröffentlicht werden.

Haupttermine studieren

Studienbeginn (Tatsächlich)

28. Januar 2026

Primärer Abschluss (Tatsächlich)

15. Juli 2026

Studienabschluss (Tatsächlich)

15. August 2026

Studienanmeldedaten

Zuerst eingereicht

14. September 2026

Zuerst eingereicht, das die QC-Kriterien erfüllt hat

18. September 2026

Zuerst gepostet (Tatsächlich)

24. September 2026

Studienaufzeichnungsaktualisierungen

Letztes Update gepostet (Tatsächlich)

24. September 2026

Letztes eingereichtes Update, das die QC-Kriterien erfüllt

18. September 2026

Zuletzt verifiziert

1. Dezember 2025

Mehr Informationen

Begriffe im Zusammenhang mit dieser Studie

Plan für individuelle Teilnehmerdaten (IPD)

Planen Sie, individuelle Teilnehmerdaten (IPD) zu teilen?

JA

Beschreibung des IPD-Plans

There is a plan to make IPD and related data dictionaries available.

Art der unterstützenden IPD-Freigabeinformationen

  • STUDIENPROTOKOLL
  • SAFT
  • ICF
  • CSR

Arzneimittel- und Geräteinformationen, Studienunterlagen

Studiert ein von der US-amerikanischen FDA reguliertes Arzneimittelprodukt

Nein

Studiert ein von der US-amerikanischen FDA reguliertes Geräteprodukt

Ja

Produkt, das in den USA hergestellt und aus den USA exportiert wird

Nein

Diese Informationen wurden ohne Änderungen direkt von der Website clinicaltrials.gov abgerufen. Wenn Sie Ihre Studiendaten ändern, entfernen oder aktualisieren möchten, wenden Sie sich bitte an register@clinicaltrials.gov. Sobald eine Änderung auf clinicaltrials.gov implementiert wird, wird diese automatisch auch auf unserer Website aktualisiert .

Abonnieren