Using the CompuFlo® CathCheck™ System to Check Epidural Catheter Function Before Tubal Ligation: A Pilot Study

17. července 2026 aktualizováno: Ling-Qun Hu, Ohio State University

Assessing Labor Epidural Catheter Dysfunction for Tubal Ligation in Parturients Using CompuFlo® CathCheck™ System: A Pilot Study

This pilot study evaluates the use of the CompuFlo® CathCheck™ System to assess labor epidural catheter function before reactivation for postpartum tubal ligation. The study aims to determine whether this technology can accurately identify functional versus dysfunctional epidural catheters compared with standard clinical assessment methods. Improving catheter assessment may reduce failed epidural reactivation, the need for additional anesthesia techniques, and improve patient safety.

Přehled studie

Detailní popis

Postpartum tubal ligation is a common surgical procedure performed after delivery, and labor epidural catheters are frequently used for anesthesia during this procedure. However, reactivation of labor epidural catheters after vaginal delivery can fail, requiring conversion to spinal or general anesthesia. Reported failure rates are significant, and factors such as prolonged time since delivery, prior need for additional epidural medication during labor, and poor analgesic satisfaction may increase the risk of failure.

Currently, there is no standardized method to confirm the continued function and location of an epidural catheter after childbirth and before reactivation for postpartum procedures. Standard clinical practice typically relies on catheter aspiration, epidural test dosing, and assessment of sensory blockade, but these methods may not reliably identify catheter dysfunction or migration.

This prospective pilot cohort study will evaluate the use of the CompuFlo® CathCheck™ System as an additional objective tool for assessing labor epidural catheter function before postpartum tubal ligation. The system uses computerized pressure measurements and pulsatile waveform detection to evaluate epidural catheter placement and patency. Study assessments will be compared with standard clinical evaluation methods to determine the ability of the device to identify functional versus dysfunctional epidural catheters.

The primary objective of this study is to assess the diagnostic accuracy of the CompuFlo® CathCheck™ System, including its sensitivity and specificity, in predicting successful epidural catheter reactivation. This study aims to determine whether incorporating objective pressure and waveform measurements can improve identification of catheter dysfunction, reduce failed epidural reactivation, minimize the need for additional anesthesia procedures, and enhance patient safety during postpartum tubal ligation.

Typ studie

Intervenční

Zápis (Odhadovaný)

120

Fáze

  • Nelze použít

Kontakty a umístění

Tato část poskytuje kontaktní údaje pro ty, kteří studii provádějí, a informace o tom, kde se tato studie provádí.

Studijní kontakt

Studijní záloha kontaktů

Studijní místa

    • Ohio
      • Columbus, Ohio, Spojené státy, 43210
        • Nábor
        • The Ohio State University

Kritéria účasti

Výzkumníci hledají lidi, kteří odpovídají určitému popisu, kterému se říká kritéria způsobilosti. Některé příklady těchto kritérií jsou celkový zdravotní stav osoby nebo předchozí léčba.

Kritéria způsobilosti

Věk způsobilý ke studiu

  • Dospělý
  • Starší dospělý

Přijímá zdravé dobrovolníky

Ne

Popis

Inclusion Criteria

  • Age ≥ 18 years
  • Postpartum women electing to undergo tubal ligation with an indwelling labor epidural catheter
  • Ability to provide informed consent in English
  • Vaginal delivery

Exclusion Criteria

  • Parturients not receiving labor epidural analgesia or receiving an alternative anesthetic technique during labor and delivery
  • Epidural catheter located >2 cm from the documented insertion depth at initial placement
  • Planned elective cesarean delivery
  • Chronic pain conditions (e.g., chronic low back pain, sickle cell disease)
  • Chronic opioid use disorder
  • Chronic gabapentin use
  • Uncontrolled mood disorders that, in the investigator's judgment, may interfere with study participation or interpretation of study outcomes
  • Prisoners or incarcerated individuals

Studijní plán

Tato část poskytuje podrobnosti o studijním plánu, včetně toho, jak je studie navržena a co studie měří.

Jak je studie koncipována?

Detaily designu

  • Primární účel: Diagnostický
  • Přidělení: N/A
  • Intervenční model: Přiřazení jedné skupiny
  • Maskování: Žádné (otevřený štítek)

Zbraně a zásahy

Skupina účastníků / Arm
Intervence / Léčba
Experimentální: Labor Epidural Reactivation Cohort
Postpartum women undergoing elective tubal ligation after vaginal delivery with an indwelling labor epidural catheter. Participants will undergo standard epidural catheter assessment and reactivation, including evaluation with the CompuFlo® CathCheck™ System to determine catheter functionality and predict successful epidural anesthesia for the tubal ligation procedure.
Evaluation of labor epidural catheter functionality using the CompuFlo® CathCheck™ System following standard epidural test-dose administration and prior to epidural reactivation for postpartum tubal ligation.

Co je měření studie?

Primární výstupní opatření

Měření výsledku
Popis opatření
Časové okno
Rate of Successful Labor Epidural Catheter Reactivation
Časové okno: Periprocedural (from the start of epidural reactivation assessment until the completion of the postpartum tubal ligation procedure)
The percentage of participants with successful use of the existing labor epidural catheter for postpartum tubal ligation without evidence of catheter failure or the need for alternative anesthesia techniques (e.g., conversion to general anesthesia or spinal anesthesia).
Periprocedural (from the start of epidural reactivation assessment until the completion of the postpartum tubal ligation procedure)

Sekundární výstupní opatření

Měření výsledku
Popis opatření
Časové okno
Sensitivity of the CompuFlo® CathCheck™ System
Časové okno: Pre-procedural (prior to epidural reactivation)
The sensitivity percentage of the CompuFlo® CathCheck™ System for correctly identifying functional labor epidural catheters that successfully provide anesthesia for postpartum tubal ligation.
Pre-procedural (prior to epidural reactivation)
Specificity of the CompuFlo® CathCheck™ System
Časové okno: Pre-procedural (prior to epidural reactivation)
The specificity percentage of the CompuFlo® CathCheck™ System for correctly identifying non-functional labor epidural catheters.
Pre-procedural (prior to epidural reactivation)
Area Under the Receiver Operating Characteristic (ROC) Curve (AUC) of the CompuFlo® CathCheck™ System
Časové okno: Pre-procedural (prior to epidural reactivation)
The AUC value of the CompuFlo® CathCheck™ System to evaluate its overall diagnostic performance in discriminating between functional and non-functional labor epidural catheters.
Pre-procedural (prior to epidural reactivation)
Number of Dermatomal Sensory Block Levels After Epidural Test Dose
Časové okno: Periprocedural (at the time of labor epidural placement and at the time of epidural reactivation)
The number of dermatomal levels demonstrating loss of cold sensation following a standard epidural local anesthetic test dose.
Periprocedural (at the time of labor epidural placement and at the time of epidural reactivation)
Duration of Labor Epidural Catheter Unused Time
Časové okno: Periprocedural (from delivery until epidural reactivation)
The total duration of time (measured in hours) between the completion of labor analgesia and the attempted epidural reactivation for the tubal ligation.
Periprocedural (from delivery until epidural reactivation)
Frequency of Labor Epidural Top-Up Doses
Časové okno: Intrapartum (from labor epidural placement until delivery)
The total number of supplemental epidural local anesthetic bolus doses administered during labor prior to delivery.
Intrapartum (from labor epidural placement until delivery)
Postoperative Pain Intensity via Numeric Rating Scale (NRS-11)
Časové okno: Periprocedural (upon admission to the post-anesthesia care unit [PACU] and at discharge from the PACU)
Pain intensity measured using the 11-point Numeric Rating Scale (NRS-11), where 0 represents "no pain" and 10 represents "the worst pain imaginable."
Periprocedural (upon admission to the post-anesthesia care unit [PACU] and at discharge from the PACU)
Patient Satisfaction Score With Labor Analgesia
Časové okno: Periprocedural (prior to discharge from the post-anesthesia care unit [PACU])
Patient-reported satisfaction with labor epidural analgesia, assessed using a standard Likert scale (e.g., 1-5 scale, where 1 is "highly dissatisfied" and 5 is "highly satisfied").
Periprocedural (prior to discharge from the post-anesthesia care unit [PACU])
Patient Satisfaction Score With Tubal Ligation Anesthesia
Časové okno: Periprocedural (prior to discharge from the post-anesthesia care unit [PACU])
Patient-reported satisfaction with anesthetic management during the postpartum tubal ligation procedure, assessed using a standard Likert scale (e.g., 1-5 scale, where 1 is "highly dissatisfied" and 5 is "highly satisfied").
Periprocedural (prior to discharge from the post-anesthesia care unit [PACU])

Spolupracovníci a vyšetřovatelé

Zde najdete lidi a organizace zapojené do této studie.

Termíny studijních záznamů

Tato data sledují průběh záznamů studie a předkládání souhrnných výsledků na ClinicalTrials.gov. Záznamy ze studií a hlášené výsledky jsou před zveřejněním na veřejné webové stránce přezkoumány Národní lékařskou knihovnou (NLM), aby se ujistily, že splňují specifické standardy kontroly kvality.

Hlavní termíny studia

Začátek studia (Aktuální)

19. prosince 2024

Primární dokončení (Odhadovaný)

31. října 2028

Dokončení studie (Odhadovaný)

31. prosince 2028

Termíny zápisu do studia

První předloženo

13. července 2026

První předloženo, které splnilo kritéria kontroly kvality

17. července 2026

První zveřejněno (Aktuální)

22. července 2026

Aktualizace studijních záznamů

Poslední zveřejněná aktualizace (Aktuální)

22. července 2026

Odeslaná poslední aktualizace, která splnila kritéria kontroly kvality

17. července 2026

Naposledy ověřeno

1. července 2026

Více informací

Termíny související s touto studií

Další identifikační čísla studie

  • 2023H0188

Informace o lécích a zařízeních, studijní dokumenty

Studuje lékový produkt regulovaný americkým FDA

Ne

Studuje produkt zařízení regulovaný americkým úřadem FDA

Ano

produkt vyrobený a vyvážený z USA

Ne

Tyto informace byly beze změn načteny přímo z webu clinicaltrials.gov. Máte-li jakékoli požadavky na změnu, odstranění nebo aktualizaci podrobností studie, kontaktujte prosím register@clinicaltrials.gov. Jakmile bude změna implementována na clinicaltrials.gov, bude automaticky aktualizována i na našem webu .

3
Předplatit