Using the CompuFlo® CathCheck™ System to Check Epidural Catheter Function Before Tubal Ligation: A Pilot Study

perjantai 17. heinäkuuta 2026 päivittänyt: Ling-Qun Hu, Ohio State University

Assessing Labor Epidural Catheter Dysfunction for Tubal Ligation in Parturients Using CompuFlo® CathCheck™ System: A Pilot Study

This pilot study evaluates the use of the CompuFlo® CathCheck™ System to assess labor epidural catheter function before reactivation for postpartum tubal ligation. The study aims to determine whether this technology can accurately identify functional versus dysfunctional epidural catheters compared with standard clinical assessment methods. Improving catheter assessment may reduce failed epidural reactivation, the need for additional anesthesia techniques, and improve patient safety.

Tutkimuksen yleiskatsaus

Yksityiskohtainen kuvaus

Postpartum tubal ligation is a common surgical procedure performed after delivery, and labor epidural catheters are frequently used for anesthesia during this procedure. However, reactivation of labor epidural catheters after vaginal delivery can fail, requiring conversion to spinal or general anesthesia. Reported failure rates are significant, and factors such as prolonged time since delivery, prior need for additional epidural medication during labor, and poor analgesic satisfaction may increase the risk of failure.

Currently, there is no standardized method to confirm the continued function and location of an epidural catheter after childbirth and before reactivation for postpartum procedures. Standard clinical practice typically relies on catheter aspiration, epidural test dosing, and assessment of sensory blockade, but these methods may not reliably identify catheter dysfunction or migration.

This prospective pilot cohort study will evaluate the use of the CompuFlo® CathCheck™ System as an additional objective tool for assessing labor epidural catheter function before postpartum tubal ligation. The system uses computerized pressure measurements and pulsatile waveform detection to evaluate epidural catheter placement and patency. Study assessments will be compared with standard clinical evaluation methods to determine the ability of the device to identify functional versus dysfunctional epidural catheters.

The primary objective of this study is to assess the diagnostic accuracy of the CompuFlo® CathCheck™ System, including its sensitivity and specificity, in predicting successful epidural catheter reactivation. This study aims to determine whether incorporating objective pressure and waveform measurements can improve identification of catheter dysfunction, reduce failed epidural reactivation, minimize the need for additional anesthesia procedures, and enhance patient safety during postpartum tubal ligation.

Opintotyyppi

Interventio

Ilmoittautuminen (Arvioitu)

120

Vaihe

  • Ei sovellettavissa

Yhteystiedot ja paikat

Tässä osiossa on tutkimuksen suorittajien yhteystiedot ja tiedot siitä, missä tämä tutkimus suoritetaan.

Opiskeluyhteys

Tutki yhteystietojen varmuuskopiointi

Opiskelupaikat

    • Ohio
      • Columbus, Ohio, Yhdysvallat, 43210
        • Rekrytointi
        • The Ohio State University

Osallistumiskriteerit

Tutkijat etsivät ihmisiä, jotka sopivat tiettyyn kuvaukseen, jota kutsutaan kelpoisuuskriteereiksi. Joitakin esimerkkejä näistä kriteereistä ovat henkilön yleinen terveydentila tai aiemmat hoidot.

Kelpoisuusvaatimukset

Opintokelpoiset iät

  • Aikuinen
  • Vanhempi Aikuinen

Hyväksyy terveitä vapaaehtoisia

Ei

Kuvaus

Inclusion Criteria

  • Age ≥ 18 years
  • Postpartum women electing to undergo tubal ligation with an indwelling labor epidural catheter
  • Ability to provide informed consent in English
  • Vaginal delivery

Exclusion Criteria

  • Parturients not receiving labor epidural analgesia or receiving an alternative anesthetic technique during labor and delivery
  • Epidural catheter located >2 cm from the documented insertion depth at initial placement
  • Planned elective cesarean delivery
  • Chronic pain conditions (e.g., chronic low back pain, sickle cell disease)
  • Chronic opioid use disorder
  • Chronic gabapentin use
  • Uncontrolled mood disorders that, in the investigator's judgment, may interfere with study participation or interpretation of study outcomes
  • Prisoners or incarcerated individuals

Opintosuunnitelma

Tässä osiossa on tietoja tutkimussuunnitelmasta, mukaan lukien kuinka tutkimus on suunniteltu ja mitä tutkimuksella mitataan.

Miten tutkimus on suunniteltu?

Suunnittelun yksityiskohdat

  • Ensisijainen käyttötarkoitus: Diagnostiikka
  • Jako: Ei käytössä
  • Inventiomalli: Yksittäinen ryhmätehtävä
  • Naamiointi: Ei mitään (avoin tarra)

Aseet ja interventiot

Osallistujaryhmä / Arm
Interventio / Hoito
Kokeellinen: Labor Epidural Reactivation Cohort
Postpartum women undergoing elective tubal ligation after vaginal delivery with an indwelling labor epidural catheter. Participants will undergo standard epidural catheter assessment and reactivation, including evaluation with the CompuFlo® CathCheck™ System to determine catheter functionality and predict successful epidural anesthesia for the tubal ligation procedure.
Evaluation of labor epidural catheter functionality using the CompuFlo® CathCheck™ System following standard epidural test-dose administration and prior to epidural reactivation for postpartum tubal ligation.

Mitä tutkimuksessa mitataan?

Ensisijaiset tulostoimenpiteet

Tulosmittaus
Toimenpiteen kuvaus
Aikaikkuna
Rate of Successful Labor Epidural Catheter Reactivation
Aikaikkuna: Periprocedural (from the start of epidural reactivation assessment until the completion of the postpartum tubal ligation procedure)
The percentage of participants with successful use of the existing labor epidural catheter for postpartum tubal ligation without evidence of catheter failure or the need for alternative anesthesia techniques (e.g., conversion to general anesthesia or spinal anesthesia).
Periprocedural (from the start of epidural reactivation assessment until the completion of the postpartum tubal ligation procedure)

Toissijaiset tulostoimenpiteet

Tulosmittaus
Toimenpiteen kuvaus
Aikaikkuna
Sensitivity of the CompuFlo® CathCheck™ System
Aikaikkuna: Pre-procedural (prior to epidural reactivation)
The sensitivity percentage of the CompuFlo® CathCheck™ System for correctly identifying functional labor epidural catheters that successfully provide anesthesia for postpartum tubal ligation.
Pre-procedural (prior to epidural reactivation)
Specificity of the CompuFlo® CathCheck™ System
Aikaikkuna: Pre-procedural (prior to epidural reactivation)
The specificity percentage of the CompuFlo® CathCheck™ System for correctly identifying non-functional labor epidural catheters.
Pre-procedural (prior to epidural reactivation)
Area Under the Receiver Operating Characteristic (ROC) Curve (AUC) of the CompuFlo® CathCheck™ System
Aikaikkuna: Pre-procedural (prior to epidural reactivation)
The AUC value of the CompuFlo® CathCheck™ System to evaluate its overall diagnostic performance in discriminating between functional and non-functional labor epidural catheters.
Pre-procedural (prior to epidural reactivation)
Number of Dermatomal Sensory Block Levels After Epidural Test Dose
Aikaikkuna: Periprocedural (at the time of labor epidural placement and at the time of epidural reactivation)
The number of dermatomal levels demonstrating loss of cold sensation following a standard epidural local anesthetic test dose.
Periprocedural (at the time of labor epidural placement and at the time of epidural reactivation)
Duration of Labor Epidural Catheter Unused Time
Aikaikkuna: Periprocedural (from delivery until epidural reactivation)
The total duration of time (measured in hours) between the completion of labor analgesia and the attempted epidural reactivation for the tubal ligation.
Periprocedural (from delivery until epidural reactivation)
Frequency of Labor Epidural Top-Up Doses
Aikaikkuna: Intrapartum (from labor epidural placement until delivery)
The total number of supplemental epidural local anesthetic bolus doses administered during labor prior to delivery.
Intrapartum (from labor epidural placement until delivery)
Postoperative Pain Intensity via Numeric Rating Scale (NRS-11)
Aikaikkuna: Periprocedural (upon admission to the post-anesthesia care unit [PACU] and at discharge from the PACU)
Pain intensity measured using the 11-point Numeric Rating Scale (NRS-11), where 0 represents "no pain" and 10 represents "the worst pain imaginable."
Periprocedural (upon admission to the post-anesthesia care unit [PACU] and at discharge from the PACU)
Patient Satisfaction Score With Labor Analgesia
Aikaikkuna: Periprocedural (prior to discharge from the post-anesthesia care unit [PACU])
Patient-reported satisfaction with labor epidural analgesia, assessed using a standard Likert scale (e.g., 1-5 scale, where 1 is "highly dissatisfied" and 5 is "highly satisfied").
Periprocedural (prior to discharge from the post-anesthesia care unit [PACU])
Patient Satisfaction Score With Tubal Ligation Anesthesia
Aikaikkuna: Periprocedural (prior to discharge from the post-anesthesia care unit [PACU])
Patient-reported satisfaction with anesthetic management during the postpartum tubal ligation procedure, assessed using a standard Likert scale (e.g., 1-5 scale, where 1 is "highly dissatisfied" and 5 is "highly satisfied").
Periprocedural (prior to discharge from the post-anesthesia care unit [PACU])

Yhteistyökumppanit ja tutkijat

Täältä löydät tähän tutkimukseen osallistuvat ihmiset ja organisaatiot.

Opintojen ennätyspäivät

Nämä päivämäärät seuraavat ClinicalTrials.gov-sivustolle lähetettyjen tutkimustietueiden ja yhteenvetojen edistymistä. National Library of Medicine (NLM) tarkistaa tutkimustiedot ja raportoidut tulokset varmistaakseen, että ne täyttävät tietyt laadunvalvontastandardit, ennen kuin ne julkaistaan ​​julkisella verkkosivustolla.

Opi tärkeimmät päivämäärät

Opiskelun aloitus (Todellinen)

Torstai 19. joulukuuta 2024

Ensisijainen valmistuminen (Arvioitu)

Tiistai 31. lokakuuta 2028

Opintojen valmistuminen (Arvioitu)

Sunnuntai 31. joulukuuta 2028

Opintoihin ilmoittautumispäivät

Ensimmäinen lähetetty

Maanantai 13. heinäkuuta 2026

Ensimmäinen toimitettu, joka täytti QC-kriteerit

Perjantai 17. heinäkuuta 2026

Ensimmäinen Lähetetty (Todellinen)

Keskiviikko 22. heinäkuuta 2026

Tutkimustietojen päivitykset

Viimeisin päivitys julkaistu (Todellinen)

Keskiviikko 22. heinäkuuta 2026

Viimeisin lähetetty päivitys, joka täytti QC-kriteerit

Perjantai 17. heinäkuuta 2026

Viimeksi vahvistettu

Keskiviikko 1. heinäkuuta 2026

Lisää tietoa

Tähän tutkimukseen liittyvät termit

Lääke- ja laitetiedot, tutkimusasiakirjat

Tutkii yhdysvaltalaista FDA sääntelemää lääkevalmistetta

Ei

Tutkii yhdysvaltalaista FDA sääntelemää laitetuotetta

Joo

Yhdysvalloissa valmistettu ja sieltä viety tuote

Ei

Nämä tiedot haettiin suoraan verkkosivustolta clinicaltrials.gov ilman muutoksia. Jos sinulla on pyyntöjä muuttaa, poistaa tai päivittää tutkimustietojasi, ota yhteyttä register@clinicaltrials.gov. Heti kun muutos on otettu käyttöön osoitteessa clinicaltrials.gov, se päivitetään automaattisesti myös verkkosivustollemme .