Using the CompuFlo® CathCheck™ System to Check Epidural Catheter Function Before Tubal Ligation: A Pilot Study

17 luglio 2026 aggiornato da: Ling-Qun Hu, Ohio State University

Assessing Labor Epidural Catheter Dysfunction for Tubal Ligation in Parturients Using CompuFlo® CathCheck™ System: A Pilot Study

This pilot study evaluates the use of the CompuFlo® CathCheck™ System to assess labor epidural catheter function before reactivation for postpartum tubal ligation. The study aims to determine whether this technology can accurately identify functional versus dysfunctional epidural catheters compared with standard clinical assessment methods. Improving catheter assessment may reduce failed epidural reactivation, the need for additional anesthesia techniques, and improve patient safety.

Panoramica dello studio

Stato

Reclutamento

Descrizione dettagliata

Postpartum tubal ligation is a common surgical procedure performed after delivery, and labor epidural catheters are frequently used for anesthesia during this procedure. However, reactivation of labor epidural catheters after vaginal delivery can fail, requiring conversion to spinal or general anesthesia. Reported failure rates are significant, and factors such as prolonged time since delivery, prior need for additional epidural medication during labor, and poor analgesic satisfaction may increase the risk of failure.

Currently, there is no standardized method to confirm the continued function and location of an epidural catheter after childbirth and before reactivation for postpartum procedures. Standard clinical practice typically relies on catheter aspiration, epidural test dosing, and assessment of sensory blockade, but these methods may not reliably identify catheter dysfunction or migration.

This prospective pilot cohort study will evaluate the use of the CompuFlo® CathCheck™ System as an additional objective tool for assessing labor epidural catheter function before postpartum tubal ligation. The system uses computerized pressure measurements and pulsatile waveform detection to evaluate epidural catheter placement and patency. Study assessments will be compared with standard clinical evaluation methods to determine the ability of the device to identify functional versus dysfunctional epidural catheters.

The primary objective of this study is to assess the diagnostic accuracy of the CompuFlo® CathCheck™ System, including its sensitivity and specificity, in predicting successful epidural catheter reactivation. This study aims to determine whether incorporating objective pressure and waveform measurements can improve identification of catheter dysfunction, reduce failed epidural reactivation, minimize the need for additional anesthesia procedures, and enhance patient safety during postpartum tubal ligation.

Tipo di studio

Interventistico

Iscrizione (Stimato)

120

Fase

  • Non applicabile

Contatti e Sedi

Questa sezione fornisce i recapiti di coloro che conducono lo studio e informazioni su dove viene condotto lo studio.

Contatto studio

Backup dei contatti dello studio

Luoghi di studio

    • Ohio
      • Columbus, Ohio, Stati Uniti, 43210
        • Reclutamento
        • The Ohio State University

Criteri di partecipazione

I ricercatori cercano persone che corrispondano a una certa descrizione, chiamata criteri di ammissibilità. Alcuni esempi di questi criteri sono le condizioni generali di salute di una persona o trattamenti precedenti.

Criteri di ammissibilità

Età idonea allo studio

  • Adulto
  • Adulto più anziano

Accetta volontari sani

No

Descrizione

Inclusion Criteria

  • Age ≥ 18 years
  • Postpartum women electing to undergo tubal ligation with an indwelling labor epidural catheter
  • Ability to provide informed consent in English
  • Vaginal delivery

Exclusion Criteria

  • Parturients not receiving labor epidural analgesia or receiving an alternative anesthetic technique during labor and delivery
  • Epidural catheter located >2 cm from the documented insertion depth at initial placement
  • Planned elective cesarean delivery
  • Chronic pain conditions (e.g., chronic low back pain, sickle cell disease)
  • Chronic opioid use disorder
  • Chronic gabapentin use
  • Uncontrolled mood disorders that, in the investigator's judgment, may interfere with study participation or interpretation of study outcomes
  • Prisoners or incarcerated individuals

Piano di studio

Questa sezione fornisce i dettagli del piano di studio, compreso il modo in cui lo studio è progettato e ciò che lo studio sta misurando.

Come è strutturato lo studio?

Dettagli di progettazione

  • Scopo principale: Diagnostico
  • Assegnazione: N / A
  • Modello interventistico: Assegnazione di gruppo singolo
  • Mascheramento: Nessuno (etichetta aperta)

Armi e interventi

Gruppo di partecipanti / Arm
Intervento / Trattamento
Sperimentale: Labor Epidural Reactivation Cohort
Postpartum women undergoing elective tubal ligation after vaginal delivery with an indwelling labor epidural catheter. Participants will undergo standard epidural catheter assessment and reactivation, including evaluation with the CompuFlo® CathCheck™ System to determine catheter functionality and predict successful epidural anesthesia for the tubal ligation procedure.
Evaluation of labor epidural catheter functionality using the CompuFlo® CathCheck™ System following standard epidural test-dose administration and prior to epidural reactivation for postpartum tubal ligation.

Cosa sta misurando lo studio?

Misure di risultato primarie

Misura del risultato
Misura Descrizione
Lasso di tempo
Rate of Successful Labor Epidural Catheter Reactivation
Lasso di tempo: Periprocedural (from the start of epidural reactivation assessment until the completion of the postpartum tubal ligation procedure)
The percentage of participants with successful use of the existing labor epidural catheter for postpartum tubal ligation without evidence of catheter failure or the need for alternative anesthesia techniques (e.g., conversion to general anesthesia or spinal anesthesia).
Periprocedural (from the start of epidural reactivation assessment until the completion of the postpartum tubal ligation procedure)

Misure di risultato secondarie

Misura del risultato
Misura Descrizione
Lasso di tempo
Sensitivity of the CompuFlo® CathCheck™ System
Lasso di tempo: Pre-procedural (prior to epidural reactivation)
The sensitivity percentage of the CompuFlo® CathCheck™ System for correctly identifying functional labor epidural catheters that successfully provide anesthesia for postpartum tubal ligation.
Pre-procedural (prior to epidural reactivation)
Specificity of the CompuFlo® CathCheck™ System
Lasso di tempo: Pre-procedural (prior to epidural reactivation)
The specificity percentage of the CompuFlo® CathCheck™ System for correctly identifying non-functional labor epidural catheters.
Pre-procedural (prior to epidural reactivation)
Area Under the Receiver Operating Characteristic (ROC) Curve (AUC) of the CompuFlo® CathCheck™ System
Lasso di tempo: Pre-procedural (prior to epidural reactivation)
The AUC value of the CompuFlo® CathCheck™ System to evaluate its overall diagnostic performance in discriminating between functional and non-functional labor epidural catheters.
Pre-procedural (prior to epidural reactivation)
Number of Dermatomal Sensory Block Levels After Epidural Test Dose
Lasso di tempo: Periprocedural (at the time of labor epidural placement and at the time of epidural reactivation)
The number of dermatomal levels demonstrating loss of cold sensation following a standard epidural local anesthetic test dose.
Periprocedural (at the time of labor epidural placement and at the time of epidural reactivation)
Duration of Labor Epidural Catheter Unused Time
Lasso di tempo: Periprocedural (from delivery until epidural reactivation)
The total duration of time (measured in hours) between the completion of labor analgesia and the attempted epidural reactivation for the tubal ligation.
Periprocedural (from delivery until epidural reactivation)
Frequency of Labor Epidural Top-Up Doses
Lasso di tempo: Intrapartum (from labor epidural placement until delivery)
The total number of supplemental epidural local anesthetic bolus doses administered during labor prior to delivery.
Intrapartum (from labor epidural placement until delivery)
Postoperative Pain Intensity via Numeric Rating Scale (NRS-11)
Lasso di tempo: Periprocedural (upon admission to the post-anesthesia care unit [PACU] and at discharge from the PACU)
Pain intensity measured using the 11-point Numeric Rating Scale (NRS-11), where 0 represents "no pain" and 10 represents "the worst pain imaginable."
Periprocedural (upon admission to the post-anesthesia care unit [PACU] and at discharge from the PACU)
Patient Satisfaction Score With Labor Analgesia
Lasso di tempo: Periprocedural (prior to discharge from the post-anesthesia care unit [PACU])
Patient-reported satisfaction with labor epidural analgesia, assessed using a standard Likert scale (e.g., 1-5 scale, where 1 is "highly dissatisfied" and 5 is "highly satisfied").
Periprocedural (prior to discharge from the post-anesthesia care unit [PACU])
Patient Satisfaction Score With Tubal Ligation Anesthesia
Lasso di tempo: Periprocedural (prior to discharge from the post-anesthesia care unit [PACU])
Patient-reported satisfaction with anesthetic management during the postpartum tubal ligation procedure, assessed using a standard Likert scale (e.g., 1-5 scale, where 1 is "highly dissatisfied" and 5 is "highly satisfied").
Periprocedural (prior to discharge from the post-anesthesia care unit [PACU])

Collaboratori e investigatori

Qui è dove troverai le persone e le organizzazioni coinvolte in questo studio.

Studiare le date dei record

Queste date tengono traccia dell'avanzamento della registrazione dello studio e dell'invio dei risultati di sintesi a ClinicalTrials.gov. I record degli studi e i risultati riportati vengono esaminati dalla National Library of Medicine (NLM) per assicurarsi che soddisfino specifici standard di controllo della qualità prima di essere pubblicati sul sito Web pubblico.

Studia le date principali

Inizio studio (Effettivo)

19 dicembre 2024

Completamento primario (Stimato)

31 ottobre 2028

Completamento dello studio (Stimato)

31 dicembre 2028

Date di iscrizione allo studio

Primo inviato

13 luglio 2026

Primo inviato che soddisfa i criteri di controllo qualità

17 luglio 2026

Primo Inserito (Effettivo)

22 luglio 2026

Aggiornamenti dei record di studio

Ultimo aggiornamento pubblicato (Effettivo)

22 luglio 2026

Ultimo aggiornamento inviato che soddisfa i criteri QC

17 luglio 2026

Ultimo verificato

1 luglio 2026

Maggiori informazioni

Termini relativi a questo studio

Altri numeri di identificazione dello studio

  • 2023H0188

Informazioni su farmaci e dispositivi, documenti di studio

Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti

No

Studia un dispositivo regolamentato dalla FDA degli Stati Uniti

prodotto fabbricato ed esportato dagli Stati Uniti

No

Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .

Prove cliniche su CompuFlo® CathCheck™ System Assessment

3
Sottoscrivi