Safety, Tolerability, and Accuracy of a Hydrogel-Based Wearable Dynamic Intraocular Pressure Sensor in Healthy Adults

Safety, Tolerability, and Accuracy of a Novel Hydrogel-Based Wearable Dynamic Intraocular Pressure Sensor in Healthy Adults: A Single-Center, Single-Arm Exploratory Clinical Trial

The occurrence and progression of eye diseases like glaucoma are closely related to intraocular pressure fluctuations. Traditional single-point intraocular pressure measurement methods cannot reflect daily dynamic changes and are prone to missing hidden intraocular pressure peaks, limiting precise diagnosis and individualized treatment. This project aims to evaluate the wearing safety and measurement accuracy of a new low-modulus, high-strength hydrogel wearable dynamic intraocular pressure sensor. This product adopts the form of smart contact lenses, fully embedding a micro sensor into a hydrogel lens, combined with an external antenna mount and a portable recorder to continuously detect and wirelessly record even slight changes in corneal curvature, thereby obtaining real-time intraocular pressure fluctuation trend data. Based on this, the objective of this study is to evaluate the safety, measurement accuracy and tolerabilityof a wearable dynamic intraocular pressure sensor based on a novel hydrogel in healthy volunteers.

Přehled studie

Postavení

Zatím nenabíráme

Podmínky

Typ studie

Intervenční

Zápis (Odhadovaný)

20

Fáze

  • Nelze použít

Kontakty a umístění

Tato část poskytuje kontaktní údaje pro ty, kteří studii provádějí, a informace o tom, kde se tato studie provádí.

Studijní kontakt

Studijní místa

    • Shanghai Municipality
      • Shanghai, Shanghai Municipality, Čína, 200080
        • Shanghai General Hospital

Kritéria účasti

Výzkumníci hledají lidi, kteří odpovídají určitému popisu, kterému se říká kritéria způsobilosti. Některé příklady těchto kritérií jsou celkový zdravotní stav osoby nebo předchozí léčba.

Kritéria způsobilosti

Věk způsobilý ke studiu

  • Dospělý

Přijímá zdravé dobrovolníky

Ano

Popis

Inclusion Criteria:

  • Voluntarily agrees to participate in the study and provides written informed consent.
  • Has previously worn contact lenses.
  • Has no ocular disease other than refractive error.
  • Central corneal thickness between 540 and 560 μm, inclusive.
  • Corneal curvature between 42 and 44 D, inclusive.

Exclusion Criteria:

  • Unable to tolerate the hydrogel-based wearable dynamic intraocular pressure sensor.
  • Pregnant or breastfeeding.
  • History of allergy to relevant medical materials or known hypersensitivity to materials used in the investigational device.

Studijní plán

Tato část poskytuje podrobnosti o studijním plánu, včetně toho, jak je studie navržena a co studie měří.

Jak je studie koncipována?

Detaily designu

  • Primární účel: Jiný
  • Přidělení: N/A
  • Intervenční model: Přiřazení jedné skupiny
  • Maskování: Žádné (otevřený štítek)

Zbraně a zásahy

Skupina účastníků / Arm
Intervence / Léčba
Experimentální: Hydrogel-Based Wearable Dynamic Intraocular Pressure Sensor

Participants will undergo a single-arm evaluation of a hydrogel-based wearable dynamic intraocular pressure sensor (CLS). After 2 hours of fasting from food and water, baseline ocular safety examinations will be performed. Intraocular pressure will first be measured using non-contact tonometry (NCT) and Goldmann applanation tonometry (GAT). Participants will then wear the CLS and undergo a 10-minute adaptation period before the first sensor signal is recorded.

Participants will drink 1 L of water within 5 minutes. At 10-15 minutes after completion of water intake, a second sensor signal will be recorded, followed by CLS removal and repeat GAT and NCT measurements. After an interval of at least 45 minutes, NCT and GAT will be repeated, the CLS will be worn again for a 10-minute adaptation period, and a third sensor signal will be recorded. After completion of the measurement procedure, the CLS will be removed and ocular safety examinations will be repeated to evaluate changes associate

The investigational device is a wearable hydrogel-based contact lens sensor designed to record electrical signals associated with changes in intraocular pressure. The CLS is worn on the study eye and allowed to stabilize for 10 minutes before signal acquisition. Sensor signals are obtained at three predefined measurement points and compared with Goldmann applanation tonometry measurements to assess measurement accuracy.

Co je měření studie?

Primární výstupní opatření

Měření výsledku
Popis opatření
Časové okno
Absolute Error of CLS-Predicted Intraocular Pressure Compared With Goldmann Applanation Tonometry
Časové okno: Day 1, at the third validation measurement
The first two paired measurements of the contact lens sensor (CLS) signal and Goldmann applanation tonometry (GAT) intraocular pressure will be used to establish an individualized linear calibration equation for each participant. The third CLS signal will be converted to a predicted intraocular pressure using the individualized calibration equation. Absolute error is defined as the absolute difference between the CLS-predicted intraocular pressure and the corresponding GAT measurement, expressed in mmHg.
Day 1, at the third validation measurement

Sekundární výstupní opatření

Měření výsledku
Popis opatření
Časové okno
Agreement Between CLS-Predicted Intraocular Pressure and Non-Contact Tonometry
Časové okno: Day 1, at the third validation measurement
At the third validation assessment, intraocular pressure predicted from the contact lens sensor (CLS) using the individualized calibration equation will be compared with the corresponding non-contact tonometry (NCT) measurement.
Day 1, at the third validation measurement
Increased incidence of corneal fluorescent staining grade
Časové okno: Baseline and at the post-study ocular examination on Day 1
Corneal fluorescence staining grading (National Eye Institute (NEI) standards) The cornea is divided into five regions (central, top, and above).Square, inferior, nasal side, temporal side), each area scores 0-3 points (0 = unstained, 1 = sparse dots, 2 = dense dots, 3 = fused or patchy)
Baseline and at the post-study ocular examination on Day 1
Conjunctival congestion score
Časové okno: Baseline and at the post-study ocular examination on Day 1
The Ora Calibra® scale was used to measure conjunctival redness in each eye. Each eye will be rated on a scale from 0 to 4, with 0 being best and 4 being the worst.
Baseline and at the post-study ocular examination on Day 1
Change From Baseline in Tear Break-Up Time After CLS Wear
Časové okno: Baseline and at the post-study ocular examination on Day 1
Tear break-up time (TBUT) will be measured before and after CLS wear. After blinking several times, the participant will keep the eye open while looking straight ahead, and the time until the first dry spot appears in the tear film will be recorded in seconds.
Baseline and at the post-study ocular examination on Day 1
Change From Baseline in Best-Corrected Visual Acuity After CLS Wear
Časové okno: Baseline and at the post-study ocular examination on Day 1
Best-corrected visual acuity (BCVA) will be assessed before and after CLS wear. Change from baseline will be calculated by comparing the post-study BCVA with the baseline BCVA.
Baseline and at the post-study ocular examination on Day 1
CLDEQ-8-Based Immediate Ocular Tolerability Questionnaire
Časové okno: Immediately after completion of CLS wear on Day 1

Immediate ocular tolerability during wear of the investigational contact lens sensor (CLS) will be assessed using a study-specific questionnaire adapted from the Contact Lens Dry Eye Questionnaire-8 (CLDEQ-8).

The questionnaire contains eight items assessing: frequency and intensity of ocular discomfort, frequency and intensity of ocular dryness, frequency and intensity of fluctuating or blurred vision, frequency of wanting to close the eye because of discomfort, and the desire to remove the CLS because of discomfort.

Frequency items are scored from 0 to 4, intensity items from 0 to 5, and the item assessing the desire to remove the CLS from 1 to 6. Higher scores indicate worse ocular tolerability.

Immediately after completion of CLS wear on Day 1

Spolupracovníci a vyšetřovatelé

Zde najdete lidi a organizace zapojené do této studie.

Termíny studijních záznamů

Tato data sledují průběh záznamů studie a předkládání souhrnných výsledků na ClinicalTrials.gov. Záznamy ze studií a hlášené výsledky jsou před zveřejněním na veřejné webové stránce přezkoumány Národní lékařskou knihovnou (NLM), aby se ujistily, že splňují specifické standardy kontroly kvality.

Hlavní termíny studia

Začátek studia (Odhadovaný)

30. srpna 2026

Primární dokončení (Odhadovaný)

30. listopadu 2026

Dokončení studie (Odhadovaný)

18. února 2027

Termíny zápisu do studia

První předloženo

12. srpna 2026

První předloženo, které splnilo kritéria kontroly kvality

15. srpna 2026

První zveřejněno (Aktuální)

18. srpna 2026

Aktualizace studijních záznamů

Poslední zveřejněná aktualizace (Aktuální)

18. srpna 2026

Odeslaná poslední aktualizace, která splnila kritéria kontroly kvality

15. srpna 2026

Naposledy ověřeno

1. srpna 2026

Více informací

Termíny související s touto studií

Další identifikační čísla studie

  • 20260424074539870

Plán pro data jednotlivých účastníků (IPD)

Plánujete sdílet data jednotlivých účastníků (IPD)?

ANO

Informace o lécích a zařízeních, studijní dokumenty

Studuje lékový produkt regulovaný americkým FDA

Ne

Studuje produkt zařízení regulovaný americkým úřadem FDA

Ne

Tyto informace byly beze změn načteny přímo z webu clinicaltrials.gov. Máte-li jakékoli požadavky na změnu, odstranění nebo aktualizaci podrobností studie, kontaktujte prosím register@clinicaltrials.gov. Jakmile bude změna implementována na clinicaltrials.gov, bude automaticky aktualizována i na našem webu .