Safety, Tolerability, and Accuracy of a Hydrogel-Based Wearable Dynamic Intraocular Pressure Sensor in Healthy Adults

Safety, Tolerability, and Accuracy of a Novel Hydrogel-Based Wearable Dynamic Intraocular Pressure Sensor in Healthy Adults: A Single-Center, Single-Arm Exploratory Clinical Trial

The occurrence and progression of eye diseases like glaucoma are closely related to intraocular pressure fluctuations. Traditional single-point intraocular pressure measurement methods cannot reflect daily dynamic changes and are prone to missing hidden intraocular pressure peaks, limiting precise diagnosis and individualized treatment. This project aims to evaluate the wearing safety and measurement accuracy of a new low-modulus, high-strength hydrogel wearable dynamic intraocular pressure sensor. This product adopts the form of smart contact lenses, fully embedding a micro sensor into a hydrogel lens, combined with an external antenna mount and a portable recorder to continuously detect and wirelessly record even slight changes in corneal curvature, thereby obtaining real-time intraocular pressure fluctuation trend data. Based on this, the objective of this study is to evaluate the safety, measurement accuracy and tolerabilityof a wearable dynamic intraocular pressure sensor based on a novel hydrogel in healthy volunteers.

Studieöversikt

Status

Har inte rekryterat ännu

Betingelser

Studietyp

Interventionell

Inskrivning (Beräknad)

20

Fas

  • Inte tillämpbar

Kontakter och platser

Det här avsnittet innehåller kontaktuppgifter för dem som genomför studien och information om var denna studie genomförs.

Studiekontakt

Studieorter

    • Shanghai Municipality
      • Shanghai, Shanghai Municipality, Kina, 200080
        • Shanghai General Hospital

Deltagandekriterier

Forskare letar efter personer som passar en viss beskrivning, så kallade behörighetskriterier. Några exempel på dessa kriterier är en persons allmänna hälsotillstånd eller tidigare behandlingar.

Urvalskriterier

Åldrar som är berättigade till studier

  • Vuxen

Tar emot friska volontärer

Ja

Beskrivning

Inclusion Criteria:

  • Voluntarily agrees to participate in the study and provides written informed consent.
  • Has previously worn contact lenses.
  • Has no ocular disease other than refractive error.
  • Central corneal thickness between 540 and 560 μm, inclusive.
  • Corneal curvature between 42 and 44 D, inclusive.

Exclusion Criteria:

  • Unable to tolerate the hydrogel-based wearable dynamic intraocular pressure sensor.
  • Pregnant or breastfeeding.
  • History of allergy to relevant medical materials or known hypersensitivity to materials used in the investigational device.

Studieplan

Det här avsnittet ger detaljer om studieplanen, inklusive hur studien är utformad och vad studien mäter.

Hur är studien utformad?

Designdetaljer

  • Primärt syfte: Övrig
  • Tilldelning: N/A
  • Interventionsmodell: Enskild gruppuppgift
  • Maskning: Ingen (Open Label)

Vapen och interventioner

Deltagargrupp / Arm
Intervention / Behandling
Experimentell: Hydrogel-Based Wearable Dynamic Intraocular Pressure Sensor

Participants will undergo a single-arm evaluation of a hydrogel-based wearable dynamic intraocular pressure sensor (CLS). After 2 hours of fasting from food and water, baseline ocular safety examinations will be performed. Intraocular pressure will first be measured using non-contact tonometry (NCT) and Goldmann applanation tonometry (GAT). Participants will then wear the CLS and undergo a 10-minute adaptation period before the first sensor signal is recorded.

Participants will drink 1 L of water within 5 minutes. At 10-15 minutes after completion of water intake, a second sensor signal will be recorded, followed by CLS removal and repeat GAT and NCT measurements. After an interval of at least 45 minutes, NCT and GAT will be repeated, the CLS will be worn again for a 10-minute adaptation period, and a third sensor signal will be recorded. After completion of the measurement procedure, the CLS will be removed and ocular safety examinations will be repeated to evaluate changes associate

The investigational device is a wearable hydrogel-based contact lens sensor designed to record electrical signals associated with changes in intraocular pressure. The CLS is worn on the study eye and allowed to stabilize for 10 minutes before signal acquisition. Sensor signals are obtained at three predefined measurement points and compared with Goldmann applanation tonometry measurements to assess measurement accuracy.

Vad mäter studien?

Primära resultatmått

Resultatmått
Åtgärdsbeskrivning
Tidsram
Absolute Error of CLS-Predicted Intraocular Pressure Compared With Goldmann Applanation Tonometry
Tidsram: Day 1, at the third validation measurement
The first two paired measurements of the contact lens sensor (CLS) signal and Goldmann applanation tonometry (GAT) intraocular pressure will be used to establish an individualized linear calibration equation for each participant. The third CLS signal will be converted to a predicted intraocular pressure using the individualized calibration equation. Absolute error is defined as the absolute difference between the CLS-predicted intraocular pressure and the corresponding GAT measurement, expressed in mmHg.
Day 1, at the third validation measurement

Sekundära resultatmått

Resultatmått
Åtgärdsbeskrivning
Tidsram
Agreement Between CLS-Predicted Intraocular Pressure and Non-Contact Tonometry
Tidsram: Day 1, at the third validation measurement
At the third validation assessment, intraocular pressure predicted from the contact lens sensor (CLS) using the individualized calibration equation will be compared with the corresponding non-contact tonometry (NCT) measurement.
Day 1, at the third validation measurement
Increased incidence of corneal fluorescent staining grade
Tidsram: Baseline and at the post-study ocular examination on Day 1
Corneal fluorescence staining grading (National Eye Institute (NEI) standards) The cornea is divided into five regions (central, top, and above).Square, inferior, nasal side, temporal side), each area scores 0-3 points (0 = unstained, 1 = sparse dots, 2 = dense dots, 3 = fused or patchy)
Baseline and at the post-study ocular examination on Day 1
Conjunctival congestion score
Tidsram: Baseline and at the post-study ocular examination on Day 1
The Ora Calibra® scale was used to measure conjunctival redness in each eye. Each eye will be rated on a scale from 0 to 4, with 0 being best and 4 being the worst.
Baseline and at the post-study ocular examination on Day 1
Change From Baseline in Tear Break-Up Time After CLS Wear
Tidsram: Baseline and at the post-study ocular examination on Day 1
Tear break-up time (TBUT) will be measured before and after CLS wear. After blinking several times, the participant will keep the eye open while looking straight ahead, and the time until the first dry spot appears in the tear film will be recorded in seconds.
Baseline and at the post-study ocular examination on Day 1
Change From Baseline in Best-Corrected Visual Acuity After CLS Wear
Tidsram: Baseline and at the post-study ocular examination on Day 1
Best-corrected visual acuity (BCVA) will be assessed before and after CLS wear. Change from baseline will be calculated by comparing the post-study BCVA with the baseline BCVA.
Baseline and at the post-study ocular examination on Day 1
CLDEQ-8-Based Immediate Ocular Tolerability Questionnaire
Tidsram: Immediately after completion of CLS wear on Day 1

Immediate ocular tolerability during wear of the investigational contact lens sensor (CLS) will be assessed using a study-specific questionnaire adapted from the Contact Lens Dry Eye Questionnaire-8 (CLDEQ-8).

The questionnaire contains eight items assessing: frequency and intensity of ocular discomfort, frequency and intensity of ocular dryness, frequency and intensity of fluctuating or blurred vision, frequency of wanting to close the eye because of discomfort, and the desire to remove the CLS because of discomfort.

Frequency items are scored from 0 to 4, intensity items from 0 to 5, and the item assessing the desire to remove the CLS from 1 to 6. Higher scores indicate worse ocular tolerability.

Immediately after completion of CLS wear on Day 1

Samarbetspartners och utredare

Det är här du hittar personer och organisationer som är involverade i denna studie.

Studieavstämningsdatum

Dessa datum spårar framstegen för inlämningar av studieposter och sammanfattande resultat till ClinicalTrials.gov. Studieposter och rapporterade resultat granskas av National Library of Medicine (NLM) för att säkerställa att de uppfyller specifika kvalitetskontrollstandarder innan de publiceras på den offentliga webbplatsen.

Studera stora datum

Studiestart (Beräknad)

30 augusti 2026

Primärt slutförande (Beräknad)

30 november 2026

Avslutad studie (Beräknad)

18 februari 2027

Studieregistreringsdatum

Först inskickad

12 augusti 2026

Först inskickad som uppfyllde QC-kriterierna

15 augusti 2026

Första postat (Faktisk)

18 augusti 2026

Uppdateringar av studier

Senaste uppdatering publicerad (Faktisk)

18 augusti 2026

Senaste inskickade uppdateringen som uppfyllde QC-kriterierna

15 augusti 2026

Senast verifierad

1 augusti 2026

Mer information

Termer relaterade till denna studie

Andra studie-ID-nummer

  • 20260424074539870

Plan för individuella deltagardata (IPD)

Planerar du att dela individuella deltagardata (IPD)?

JA

Läkemedels- och apparatinformation, studiedokument

Studerar en amerikansk FDA-reglerad läkemedelsprodukt

Nej

Studerar en amerikansk FDA-reglerad produktprodukt

Nej

Denna information hämtades direkt från webbplatsen clinicaltrials.gov utan några ändringar. Om du har några önskemål om att ändra, ta bort eller uppdatera dina studieuppgifter, vänligen kontakta register@clinicaltrials.gov. Så snart en ändring har implementerats på clinicaltrials.gov, kommer denna att uppdateras automatiskt även på vår webbplats .