Accuracy of Positioning Guide and Customized Plate in Zygomaticomaxillary Complex Fracture
Accuracy of Positioning Guide and Customized Titanium Plate for Reduction and Fixation of Zygomaticomaxillary Complex Fracture: A Case Series
This prospective, single-arm case series aims to evaluate the accuracy of a virtual planning-based positioning guide and customized titanium plate for the reduction and fixation of unilateral zygomaticomaxillary complex (ZMC) fractures.
The primary objective is to assess fracture reduction accuracy by comparing postoperative CT findings with the preoperative virtual surgical plan. Secondary objectives are to evaluate fixation stability and patient satisfaction.
Participants undergo preoperative CT imaging for virtual surgical planning, followed by surgical reduction and fixation using a patient-specific positioning guide and customized titanium plate. Postoperative CT imaging is used to assess reduction accuracy, with subsequent follow-up to evaluate fixation stability and patient satisfaction.
Descripción general del estudio
Estado
Intervención / Tratamiento
Descripción detallada
The study evaluates the accuracy of a patient-specific positioning guide and customized titanium plate for the reduction and fixation of unilateral zygomaticomaxillary complex (ZMC) fractures.
All patients undergo a thorough medical and dental history and comprehensive clinical examination. Clinical measurements and photographic documentation are obtained to assess facial symmetry, ocular function, and mouth opening. A preoperative computed tomography (CT) scan (T0) is obtained within one week before surgery to assess the fracture pattern and the extent of displacement. The Digital Imaging and Communications in Medicine (DICOM) data are imported into specialized medical imaging software (e.g., Mimics, 3-matic, or equivalent) for three-dimensional reconstruction and virtual surgical planning (VSP).
During the virtual planning phase, the fractured zygomaticomaxillary complex is first virtually reduced by mirroring the unaffected contralateral side to establish the planned anatomical position. If mirroring is not feasible, established anatomical landmarks and anatomical norms are used to guide the virtual reduction. The customized titanium plate is then designed on the virtually reduced model to fit the planned bone contours. The screw positions and trajectories are also planned, and the screws are virtually positioned within their corresponding fracture segments to define the intended fixation configuration.
After finalizing the plate and screw design in the reduced position, the zygomatic complex is virtually returned to its original displaced fracture position while maintaining the planned screw positions relative to their respective bone segments. The patient-specific positioning guide is then designed on this unreduced model. The guide incorporates the planned drilling locations, allowing the screw holes to be predrilled in the original fracture position. These predrilled holes subsequently guide the reduction and allow the customized plate to be positioned and fixed according to the original virtual surgical plan.
The customized titanium plate is designed using computer-aided design (CAD) and manufactured using computer-aided manufacturing (CAM). The design is exported in STL format, and machining programs are generated for production. The plate is fabricated from medical-grade Titanium Grade 4 using a three-axis CNC milling machine (MILLSTAR, Taiwan), through a subtractive manufacturing process. The machined plate undergoes deburring, polishing, anodizing, cleaning, and packing before use. Digital planning and fabrication are performed by Arab Engineers for Designs and Medical Instrumentation in collaboration with the research team.
Surgical treatment is performed under general anesthesia using lateral eyebrow and intraoral vestibular incisions to expose the fracture site. The patient-specific positioning guide is placed on the zygomatic complex in its original displaced fracture position. The planned screw holes are predrilled through the guide. After removal of the guide, the fracture segments are reduced using the predrilled holes as references to reproduce the planned anatomical position. The customized titanium plate is then positioned according to the planned screw positions and secured with screws to achieve fixation. The surgical site is irrigated and the incisions are closed in layers.
A postoperative CT scan (T1) is obtained within one week after surgery to assess the accuracy of fracture reduction and plate placement. The preoperative virtual surgical plan and postoperative CT model are imported into imaging software and superimposed. Linear and angular deviations between the planned and achieved positions of anatomical landmarks and plate positions are measured in millimeters and degrees. Measurements are performed by the researcher and verified by the main supervisor.
Clinical follow-up is conducted at 1 week, 1 month, 3 months, 6 months to assess wound healing, facial symmetry, ocular function, mouth opening, and postoperative complications, including infection, nerve paresthesia, malunion, and non-union. A follow-up CT scan (T2) is planned at 6 months to evaluate bone healing and fixation stability, including any signs of loosening or displacement. Patient satisfaction is assessed using a Visual Analog Scale (VAS) ranging from 0 to 10.
Tipo de estudio
Inscripción (Estimado)
Fase
- No aplica
Contactos y Ubicaciones
Ubicaciones de estudio
-
-
-
Cairo, Egipto
- department of oral and maxillofacial surgery, faculty of dentistry, Cairo university
-
-
Criterios de participación
Criterio de elegibilidad
Edades elegibles para estudiar
Acepta Voluntarios Saludables
Géneros elegibles para el estudio
Descripción
Inclusion Criteria:
- Patients diagnosed with Zygomatico-maxillary complex fractures requiring surgical intervention.
- Patients aged 18 years or older.
- Patients willing to provide informed consent and comply with follow-up protocols.
- Patients with adequate bone quality and quantity for stable fixation.
Exclusion Criteria:
- Patients with comminuted ZMC fractures that preclude the use of custom-made guides based on mirroring the contra lateral side.
- Patients with pre-existing facial deformities or asymmetries that would interfere with accurate assessment of fracture reduction.
- Patients with systemic diseases or conditions that contraindicate elective surgery or impair bone healing (e.g., uncontrolled diabetes, severe osteoporosis, active malignancy, and immunosuppression).
- Patients on medications affecting bone metabolism (e.g., long-term corticosteroids, bisphosphonates). Patients with active infections in the maxillofacial region.
- Pregnant or lactating women.
- Patients with psychiatric disorders or unrealistic expectations.
- Patients who are unable to undergo CT scans or follow post-operative instructions.
Plan de estudios
¿Cómo está diseñado el estudio?
Detalles de diseño
- Propósito principal: Tratamiento
- Asignación: N / A
- Modelo Intervencionista: Asignación de un solo grupo
- Enmascaramiento: Ninguno (etiqueta abierta)
Armas e Intervenciones
Grupo de participantes/brazo |
Intervención / Tratamiento |
|---|---|
|
Experimental: Virtual Planning and Customized Plates
Patients with ZMC fractures undergoing surgical treatment using 3D virtual planning and patient-specific titanium plates. |
A 3D-printed positioning guide designed through virtual surgical planning to assist in the accurate reduction of ZMC fractures.
Otros nombres:
A custom-fabricated titanium plate designed to fit the patient's specific anatomy for rigid fixation of ZMC fractures.
Otros nombres:
|
¿Qué mide el estudio?
Medidas de resultado primarias
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
Accuracy of Fracture Reduction
Periodo de tiempo: one week postoperative
|
Evaluation of the accuracy of fracture reduction achieved using the positioning guide and customized titanium plate. This will be assessed by superimposing the post-operative computed tomography (CT) scans onto the pre-operative virtual surgical plan. The difference will be reported as linear deviation, measured in millimeters (mm). |
one week postoperative
|
Medidas de resultado secundarias
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
Fixation stability
Periodo de tiempo: 6 months postoperatively
|
Measured in millimeters by comparing the 6-month follow-up CT to the immediate postoperative CT to check for hardware or bone displacement. |
6 months postoperatively
|
|
Patient satisfaction
Periodo de tiempo: 6 months postoperatively
|
Evaluation of the patient's overall satisfaction with the functional and aesthetic outcomes of the procedure. This will be assessed using a Visual Analog Scale (VAS) ranging from 0 to 10, where 0 indicates "completely dissatisfied" and 10 indicates "completely satisfied". The reported value will be the numerical score on this scale |
6 months postoperatively
|
Colaboradores e Investigadores
Patrocinador
Investigadores
- Director de estudio: Prof. Ramy Ragab Elbialy, prof. of maxillofacial surgery, Department of Oral and maxillofacial surgery,Faculty of dentistry, Cairo university
- Director de estudio: Dr. Mohamed Ali Zain Elabden, lecturer of maxillofacial surg, Department of Oral and maxillofacial surgery, Faculty of dentistry, Cairo university
Fechas de registro del estudio
Fechas importantes del estudio
Inicio del estudio (Actual)
Finalización primaria (Estimado)
Finalización del estudio (Estimado)
Fechas de registro del estudio
Enviado por primera vez
Primero enviado que cumplió con los criterios de control de calidad
Publicado por primera vez (Actual)
Actualizaciones de registros de estudio
Última actualización publicada (Actual)
Última actualización enviada que cumplió con los criterios de control de calidad
Última verificación
Más información
Términos relacionados con este estudio
Palabras clave
Términos MeSH relevantes adicionales
Otros números de identificación del estudio
- OMFS3319
Información sobre medicamentos y dispositivos, documentos del estudio
Estudia un producto farmacéutico regulado por la FDA de EE. UU.
Estudia un producto de dispositivo regulado por la FDA de EE. UU.
Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .