Prospective, Randomized, Parallel-group Comparison of Two Refractive EDOF Intraocular Lenses: Johnson & Johnson Tecnis PureSee Versus 1stQ Comfort Z-EDOF (1stQvsPureSee)

30 de junio de 2026 actualizado por: Prim. Univ.-Prof. Dr. Pia Veronika Vécsei-Marlovits, MSc, MB, Vienna Hospital Association

This study compares two types of artificial lens (intraocular lens, or IOL) used during cataract surgery. Both are extended depth-of-focus (EDOF) lenses, designed to give clear vision across a range of distances - far and intermediate - so that people depend less on glasses after surgery. The two lenses, the Tecnis PureSee and the 1stQ Comfort Z-EDOF, use different optical designs, and it is not yet known whether one provides a wider range of clear vision than the other.

Adults having cataract surgery in both eyes may take part. Each participant is randomly assigned to receive one of the two lens types, and the same lens type is implanted in both eyes. Neither the participant nor the staff measuring vision knows which lens was used until the main 6-month assessment is complete.

After surgery, participants attend check-ups at 1, 3, and 6 months. The study measures how wide a range of distances each person can see clearly, along with vision at different distances, quality of vision, side effects such as glare or halos, and overall satisfaction. The results will help doctors and patients choose between these lens types.

Descripción general del estudio

Descripción detallada

Background and rationale: Extended depth-of-focus (EDOF) intraocular lenses (IOLs) extend the range of functional vision beyond that of conventional monofocal IOLs by modifying the eye's wavefront to create an elongated focal range, while aiming to limit the photic phenomena (halos, glare, starbursts) often associated with multifocal IOLs. Different EDOF platforms use fundamentally different optical strategies, yet direct comparative evidence between EDOF designs from prospective randomized trials is limited; most published studies compare an EDOF IOL with monofocal or multifocal lenses rather than with another EDOF design. This trial provides a head-to-head comparison of two purely refractive EDOF platforms that employ different optical principles: the Tecnis PureSee (Johnson & Johnson Vision), a refractive EDOF design with a continuous power profile and negative spherical aberration, and the 1stQ Comfort Z-EDOF (1stQ GmbH), a refractive aspheric EDOF design with a modified spherical aberration profile.

Design: This is a single-center, prospective, randomized, parallel-group comparative study conducted at the Department of Ophthalmology, Klinik Hietzing, Vienna, Austria. Patients undergoing bilateral cataract surgery are randomized 1:1 to receive one of the two EDOF IOL platforms, implanted bilaterally (the same platform in both eyes). Randomization uses permuted blocks and is stratified by toric versus non-toric IOL variant; patients with preoperative corneal astigmatism greater than 0.75 D receive the toric version of the assigned platform. Target refraction is emmetropia in the dominant eye and approximately -0.50 D in the non-dominant eye. Participants and outcome assessors are masked to lens allocation until the 6-month primary endpoint assessment is complete; the operating surgeon cannot be masked.

The study compares monocular (eye-level) and binocular (patient-level) outcomes between the two platforms. The primary endpoint is the monocular distance-corrected defocus range at the 0.2 logMAR threshold, measured 6 months after surgery from a standardized defocus curve (4-meter ETDRS chart, defocus +1.0 D to -3.5 D in 0.5 D steps). Secondary assessments include visual acuity at multiple distances, the complete defocus curve, manifest refraction and astigmatic vector components, higher-order wavefront aberrations, rotational stability of the toric variants, patient-reported visual function and quality of vision (including halos, glare, and starbursts), spectacle independence, satisfaction, and surgical safety.

Assessments are performed preoperatively and at 1, 3, and 6 months postoperatively. The primary analysis compares the mean monocular defocus range between groups using one eye per patient (the first-operated eye), with sensitivity analyses using both eyes and appropriate adjustment for within-patient correlation. The study is regulated as a post-market clinical follow-up (PMCF) investigation under Article 74(1) of Regulation (EU) 2017/745; both lenses are CE-marked and used within their approved intended purpose.

Tipo de estudio

Intervencionista

Inscripción (Estimado)

110

Fase

  • No aplica

Contactos y Ubicaciones

Esta sección proporciona los datos de contacto de quienes realizan el estudio e información sobre dónde se lleva a cabo este estudio.

Estudio Contacto

  • Nombre: Michal Klimek, MD
  • Número de teléfono: +43 650 460 7496
  • Correo electrónico: klimek.michael@gmx.at

Ubicaciones de estudio

    • State of Vienna
      • Vienna, State of Vienna, Austria, 1130
        • Augenabteilung, Klinik Hietzing, Wiener Gesundheitsverbund
        • Contacto:

Criterios de participación

Los investigadores buscan personas que se ajusten a una determinada descripción, denominada criterio de elegibilidad. Algunos ejemplos de estos criterios son el estado de salud general de una persona o tratamientos previos.

Criterio de elegibilidad

Edades elegibles para estudiar

  • Adulto
  • Adulto Mayor

Acepta Voluntarios Saludables

No

Descripción

Inclusion Criteria:

  • Age ≥40 years, able to understand the study and give written informed consent
  • Clinically significant bilateral age-related cataract suitable for phacoemulsification with posterior chamber IOL implantation
  • Expected postoperative best-corrected visual acuity ≤0.2 logMAR in both eyes
  • Regular corneal astigmatism ≤4.5 D in both eyes (IOLMaster 700)
  • Axial length 21.5-26.5 mm in both eyes
  • Clear cornea without clinically significant opacities, scars, or stromal dystrophies
  • Normal pupil function; no significant anisocoria (≥1 mm) and no afferent pupillary defect
  • Acceptance of the target refraction strategy: emmetropia in the dominant eye and -0.50 D in the non-dominant eye
  • Willing and able to attend all follow-up visits (Month 1, 3, and 6)
  • Written informed consent obtained before any study-related procedure

Exclusion Criteria:

  • Prior corneal refractive surgery, corneal transplantation, glaucoma filtering surgery, or vitreoretinal surgery in either eye (uncomplicated intravitreal injections allowed)
  • Irregular astigmatism, keratoconus, forme fruste keratoconus, corneal ectasia, significant corneal opacities or scars, or endothelial dystrophy
  • Keratometric astigmatism >4.5 D in either eye
  • Axial length <21.5 mm or >26.5 mm in either eye
  • Calculated IOL power outside the available range of either study lens
  • Macular pathology expected to limit visual acuity (e.g., age-related macular degeneration, diabetic maculopathy, epiretinal membrane, macular hole)
  • Advanced glaucoma with central visual field involvement, or IOP not controlled on ≤2 topical medications
  • Optic neuropathy expected to limit visual potential
  • Pre-existing amblyopia in either eye
  • Significant anisocoria, afferent pupillary defect, non-dilating pupil, or iris abnormalities affecting IOL function or measurement
  • Endothelial cell density <1500 cells/mm²
  • Uncontrolled diabetes mellitus (HbA1c >9%) or autoimmune disease with active ocular involvement
  • Systemic medications affecting pupil size that cannot be discontinued (e.g., tamsulosin)
  • Pregnancy or breastfeeding
  • Intraoperative complication precluding safe in-the-bag IOL implantation (affected eyes withdrawn)
  • Cognitive impairment, language barrier, or other factors preventing reliable participation
  • Current enrollment in another interventional clinical study affecting visual outcomes

Plan de estudios

Esta sección proporciona detalles del plan de estudio, incluido cómo está diseñado el estudio y qué mide el estudio.

¿Cómo está diseñado el estudio?

Detalles de diseño

  • Propósito principal: Tratamiento
  • Asignación: Aleatorizado
  • Modelo Intervencionista: Asignación paralela
  • Enmascaramiento: Doble

Armas e Intervenciones

Grupo de participantes/brazo
Intervención / Tratamiento
Experimental: Tecnis PureSee
Bilateral implantation of the Tecnis PureSee refractive extended depth-of-focus IOL (Johnson & Johnson Vision; non-toric model ZEN00V or toric DET series). The same lens platform is implanted in both eyes. Target refraction: emmetropia in the dominant eye and approximately -0.50 D in the non-dominant eye.
Phacoemulsification with bilateral implantation of the Tecnis PureSee refractive EDOF IOL in the capsular bag. Toric variant used for preoperative corneal astigmatism greater than 0.75 D.
Comparador activo: 1stQ Comfort Z-EDOF
Bilateral implantation of the 1stQ Comfort Z-EDOF refractive extended depth-of-focus IOL (1stQ GmbH; non-toric models B1XBY0/S1XBY0 or toric B1ZBY0). The same lens platform is implanted in both eyes. Target refraction: emmetropia in the dominant eye and approximately -0.50 D in the non-dominant eye.
Phacoemulsification with bilateral implantation of the 1stQ Comfort Z-EDOF refractive EDOF IOL in the capsular bag. Toric variant used for preoperative corneal astigmatism greater than 0.75 D.

¿Qué mide el estudio?

Medidas de resultado primarias

Medida de resultado
Medida Descripción
Periodo de tiempo
Monocular distance-corrected defocus range (DCDR) at 0.2 logMAR
Periodo de tiempo: 6 months after surgery
Maximum contiguous defocus span (in diopters) over which monocular distance-corrected visual acuity remains ≤0.2 logMAR, derived from a standardized defocus curve (4 m ETDRS chart; +1.0 D to -3.5 D in 0.5 D steps).
6 months after surgery

Medidas de resultado secundarias

Medida de resultado
Medida Descripción
Periodo de tiempo
Monocular visual acuity at multiple distances
Periodo de tiempo: 6 months after surgery
logMAR acuity: UDVA and CDVA (4 m); UIVA at 66 cm and 80 cm and DCIVA (intermediate); UNVA at 40 cm and DCNVA (near).
6 months after surgery

Colaboradores e Investigadores

Aquí es donde encontrará personas y organizaciones involucradas en este estudio.

Fechas de registro del estudio

Estas fechas rastrean el progreso del registro del estudio y los envíos de resultados resumidos a ClinicalTrials.gov. Los registros del estudio y los resultados informados son revisados ​​por la Biblioteca Nacional de Medicina (NLM) para asegurarse de que cumplan con los estándares de control de calidad específicos antes de publicarlos en el sitio web público.

Fechas importantes del estudio

Inicio del estudio (Estimado)

1 de agosto de 2026

Finalización primaria (Estimado)

1 de abril de 2029

Finalización del estudio (Estimado)

1 de abril de 2029

Fechas de registro del estudio

Enviado por primera vez

30 de junio de 2026

Primero enviado que cumplió con los criterios de control de calidad

30 de junio de 2026

Publicado por primera vez (Actual)

7 de julio de 2026

Actualizaciones de registros de estudio

Última actualización publicada (Actual)

7 de julio de 2026

Última actualización enviada que cumplió con los criterios de control de calidad

30 de junio de 2026

Última verificación

1 de junio de 2026

Más información

Términos relacionados con este estudio

Palabras clave

Términos MeSH relevantes adicionales

Otros números de identificación del estudio

  • MA_15_EK_26_094_VK

Plan de datos de participantes individuales (IPD)

¿Planea compartir datos de participantes individuales (IPD)?

NO

Información sobre medicamentos y dispositivos, documentos del estudio

Estudia un producto farmacéutico regulado por la FDA de EE. UU.

No

Estudia un producto de dispositivo regulado por la FDA de EE. UU.

No

Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .