- ICH GCP
- Registro de ensayos clínicos de EE. UU.
- Ensayo clínico NCT07706140
Observational Data Collection for Blood Hemoglobin Algorithm Validation in Pediatrics
Prospective Observational Data Collection Study for Validation of a Near-Infrared Spectroscopy-Based Blood Hemoglobin Algorithm in Pediatric Patients Undergoing Cardiac Surgery or Cardiac Catheterization
This study will enroll children and young adults who are scheduled to undergo planned heart surgery or a heart catheterization procedure while under general anesthesia. Researchers will collect information from a forehead sensor that measures oxygen levels in the body's tissues, along with blood hemoglobin measurements obtained from blood samples that are already being collected as part of routine medical care.
The goal of the study is to determine whether a blood hemoglobin monitoring method that has already been cleared by the U.S. Food and Drug Administration (FDA) for use in adults may also work well in pediatric patients.
Descripción general del estudio
Estado
Condiciones
Intervención / Tratamiento
Descripción detallada
Background:
Continuous monitoring of blood hemoglobin levels may help clinicians identify changes in a patient's blood status during surgery and support decisions regarding blood transfusion. A noninvasive blood hemoglobin monitoring method has been developed using near-infrared spectroscopy (NIRS) technology incorporated into the ForeSight Tissue Oximetry System. This technology uses near-infrared light to measure tissue oxygenation and derive information related to blood hemoglobin levels. An algorithm based on these measurements has been validated and cleared by the U.S. Food and Drug Administration (FDA) for use in adult patients.
Study Purpose:
The purpose of this study is to collect synchronized physiologic and laboratory data from pediatric patients to evaluate the feasibility and performance of applying an FDA-cleared adult blood hemoglobin algorithm to pediatric populations. The study will compare hemoglobin values estimated by the algorithm with reference hemoglobin values measured from arterial blood samples using standard blood-gas/co-oximetry analysis.
Study Design:
This is a prospective observational study conducted at Boston Children's Hospital. Pediatric patients undergoing elective cardiac surgery or elective cardiac catheterization procedures under general anesthesia will be eligible for participation. No investigational devices will be used, and participation in the study will not alter routine clinical care.
Study Procedures:
Following enrollment and informed consent, ForeSight tissue oximetry sensors will be placed on the patient's forehead as part of routine monitoring. One or two sensors may be used depending on patient size and available forehead surface area. The sensors will remain in place during the procedure and will continuously record tissue oxygenation and related physiologic signals.
Arterial blood samples obtained as part of routine clinical care will be used to measure blood hemoglobin levels using a blood-gas/co-oximetry analyzer. In some patients, up to two additional blood samples may be collected to strengthen the evaluation dataset. The amount of blood collected for these additional samples will be minimal.
Additional data collected will include demographic information, procedure type, standard physiologic monitoring parameters (such as heart rate, blood pressure, oxygen saturation, respiratory measurements, and temperature), laboratory values, administered medications and blood products, and clinically significant hemodynamic events documented during the procedure.
Data Handling:
Data collected from the monitoring system and clinical records will be linked and subsequently de-identified by authorized study personnel. Personal identifiers will be removed before data are shared with the study sponsor. The sponsor will have access only to de-identified study data.
Outcome Evaluation:
The primary objective is to assess the agreement between hemoglobin values estimated by the NIRS-based algorithm and reference hemoglobin values measured from arterial blood samples. Study analyses will evaluate the accuracy, bias, precision, and agreement of algorithm-derived hemoglobin measurements in pediatric patients. The results will help determine whether the adult FDA-cleared algorithm performs adequately in pediatric populations and will provide information to guide future clinical evaluation and regulatory activities related to noninvasive hemoglobin monitoring in children and adolescents.
Tipo de estudio
Inscripción (Estimado)
Contactos y Ubicaciones
Estudio Contacto
- Nombre: Dima Daaboul, MD
- Número de teléfono: (617) 355-1370
- Correo electrónico: dima.daaboul@childrens.harvard.edu
Copia de seguridad de contactos de estudio
- Nombre: Barry Kussman, MBBCh, FFA(SA)
- Número de teléfono: (617) 355-1370
- Correo electrónico: Barry.Kussman@childrens.harvard.edu
Ubicaciones de estudio
-
-
Massachusetts
-
Boston, Massachusetts, Estados Unidos, 02115
- Boston Children's Hospital
-
Contacto:
- Rachel Bernier, MPH
- Número de teléfono: 857-218-5348
- Correo electrónico: rachel.bernier@childrens.harvard.edu
-
Contacto:
- Jocelyn Booth, BSN
- Número de teléfono: 857-218-4585
- Correo electrónico: jocelyn.booth@childrens.harvard.edu
-
Investigador principal:
- Dima Daaboul, MD
-
Sub-Investigador:
- Barry Kussman, MBBCh, FFA(SA)
-
-
Criterios de participación
Criterio de elegibilidad
Edades elegibles para estudiar
- Niño
- Adulto
Acepta Voluntarios Saludables
Método de muestreo
Población de estudio
Descripción
Inclusion Criteria:
- Age less than 21 years at the time of enrollment.
- Scheduled for elective cardiac surgery or elective cardiac catheterization under general anesthesia.
- Placement of an arterial catheter as part of routine clinical care for the procedure.
Exclusion Criteria:
- Presence of skin, scalp, or craniofacial abnormalities that prevent proper placement of the near-infrared spectroscopy (NIRS) sensor or may significantly affect measurement quality.
- Inability to obtain arterial blood hemoglobin measurements using standard clinical methods.
Plan de estudios
¿Cómo está diseñado el estudio?
Detalles de diseño
Cohortes e Intervenciones
Grupo / Cohorte |
Intervención / Tratamiento |
|---|---|
|
Pediatric Cardiac Surgery and Pediatric Catheterization Patients
This cohort includes pediatric patients undergoing elective cardiac surgery or elective cardiac catheterization under general anesthesia. All participants receive standard clinical care, including routine physiologic monitoring and arterial blood sampling when clinically indicated. As part of routine monitoring, a FDA-cleared near-infrared spectroscopy (NIRS)-based tissue oximetry system (ForeSight Tissue Oximetry System) will be placed on the forehead to continuously record cerebral tissue oxygenation during the procedure. Arterial blood samples obtained for clinical care will be analyzed using standard blood-gas/co-oximetry methods to measure blood hemoglobin levels. In select cases, a small number of additional blood samples may be collected to support data analysis. This is an observational study only. No investigational devices are used, no interventions are assigned, and there is no change to clinical management. All collected data are de-identified prior to analysis. |
This study evaluates noninvasive hemoglobin monitoring using a near-infrared spectroscopy (NIRS)-based tissue oximetry system in pediatric patients undergoing cardiac surgery or cardiac catheterization under general anesthesia. A FDA-cleared ForeSight Tissue Oximetry sensor is placed on the forehead to continuously measure tissue oxygenation, and an algorithm is derived to estimate hemoglobin. Reference hemoglobin values are obtained from arterial blood samples analyzed using standard blood-gas/co-oximetry methods as part of routine clinical care. A small number of additional samples may be collected if needed. The study compares noninvasive hemoglobin estimates with laboratory hemoglobin measurements. No investigational devices are used, and there is no change to clinical care. |
¿Qué mide el estudio?
Medidas de resultado primarias
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
Agreement Between Noninvasive and Invasive Hemoglobin Measurements
Periodo de tiempo: Intraoperative period (from induction of general anesthesia until end of procedure)
|
The primary outcome measure is the agreement between NIRS-estimated hemoglobin and hemoglobin measured from arterial blood samples.
Both assessments measure hemoglobin concentration and will be reported in grams per deciliter (g/dL).
Agreement will be assessed by comparing paired hemoglobin measurements from the two methods.
The reported outcome will include the mean difference between the two measurements (g/dL) and the range of differences observed (g/dL).
|
Intraoperative period (from induction of general anesthesia until end of procedure)
|
Colaboradores e Investigadores
Patrocinador
Colaboradores
Investigadores
- Investigador principal: Dima Daaboul, Boston Children's Hospital
Publicaciones y enlaces útiles
Publicaciones Generales
- Kussman BD, Wypij D, DiNardo JA, Newburger J, Jonas RA, Bartlett J, McGrath E, Laussen PC. An evaluation of bilateral monitoring of cerebral oxygen saturation during pediatric cardiac surgery. Anesth Analg. 2005 Nov;101(5):1294-1300. doi: 10.1213/01.ANE.0000180205.85490.85.
- van Wonderen SF, Aguirre A, Benni P, van den Dool REC, Macias Pingarron JP, Solares GJ, Hatib F, Veelo DP, Vlaar APJ, Jian Z. Continuous and Noninvasive Total Blood Hemoglobin Measurement Using Near-Infrared Reflectance Spectrometry. Anesth Analg. 2025 Nov 1;141(5):1107-1115. doi: 10.1213/ANE.0000000000007618. Epub 2025 Jul 2.
Enlaces Útiles
- Manufacturer information for the HemoSphere advanced patient monitoring platform used to acquire tissue oximetry and hemodynamic data in this study.
- Manufacturer information for the ForeSight tissue oximetry sensor, which provides continuous noninvasive cerebral tissue oxygenation monitoring and tissue hemoglobin measurements used in this study
Fechas de registro del estudio
Fechas importantes del estudio
Inicio del estudio (Estimado)
Finalización primaria (Estimado)
Finalización del estudio (Estimado)
Fechas de registro del estudio
Enviado por primera vez
Primero enviado que cumplió con los criterios de control de calidad
Publicado por primera vez (Actual)
Actualizaciones de registros de estudio
Última actualización publicada (Actual)
Última actualización enviada que cumplió con los criterios de control de calidad
Última verificación
Más información
Términos relacionados con este estudio
Palabras clave
Términos MeSH relevantes adicionales
Otros números de identificación del estudio
- IRB-P00054090
Plan de datos de participantes individuales (IPD)
¿Planea compartir datos de participantes individuales (IPD)?
Descripción del plan IPD
Información sobre medicamentos y dispositivos, documentos del estudio
Estudia un producto farmacéutico regulado por la FDA de EE. UU.
Estudia un producto de dispositivo regulado por la FDA de EE. UU.
producto fabricado y exportado desde los EE. UU.
Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .