Clinical Validation of SMD-AFECG for Predicting New-Onset Atrial Fibrillation Within 2 Hours Using Single-Lead ECG Data

20 de julio de 2026 actualizado por: HUINNO Co., Ltd

A Retrospective, Single-Center, Single-Arm, Superiority Confirmatory Study to Evaluate the Clinical Validity of SMD-AFECG, Artificial Intelligence Software to Predict the Risk of Atrial Fibrillation, Including First Episode Within 2 Hours, Using Single-Lead-Based Electrocardiogram Analysis

The purpose of this retrospective study is to evaluate the clinical performance of SMD-AFECG, an artificial intelligence-based medical device software that predicts the risk of atrial fibrillation occurring within 2 hours using single-lead electrocardiogram data. A total of 797 eligible electrocardiogram datasets collected through VitalDB at Seoul National University Hospital will be included. The performance of SMD-AFECG will be evaluated separately for new-onset atrial fibrillation in patients without a previous history of atrial fibrillation (NOAF) and atrial fibrillation episodes in patients with a previous history of atrial fibrillation (RAF). Two physicians blinded to the software results will review the electrocardiogram data and relevant medical records to establish the reference-standard classification. The blinded electrocardiogram datasets will then be analyzed using SMD-AFECG, and the software-generated predictions will be compared with the reference standard to evaluate the area under the receiver operating characteristic curve for NOAF and RAF.

Descripción general del estudio

Descripción detallada

This clinical study is designed as a retrospective, single-center, single-arm, superiority confirmatory clinical performance study.

Pre-screening Electronic medical records and the VitalDB database at Seoul National University Hospital will be reviewed to identify potentially eligible data. The study will use continuous single-lead electrocardiogram data collected through VitalDB between September 1, 2022, and before September 30, 2025, from patients who were 19 years of age or older at the time of data collection.

Screening and Data Collection A screening number will be assigned sequentially to each potentially eligible case. Eligibility will be determined by reviewing the predefined inclusion and exclusion criteria. For eligible cases, continuous single-lead electrocardiogram data, signal loss, heart rate, the occurrence and timing of atrial fibrillation episodes, the electrocardiogram collection period, demographic information, previous history of atrial fibrillation, and pregnancy status will be collected. Directly identifying personal information will not be collected.

Test Dataset Creation Eligible data will be classified according to the presence or absence of a previous history of atrial fibrillation and the occurrence of an atrial fibrillation episode. The datasets will be categorized as new-onset atrial fibrillation positive or negative and repeated atrial fibrillation positive or negative. A subject number will then be assigned to each eligible dataset.

The planned sample size is 797 cases, including 462 cases for the evaluation of new-onset atrial fibrillation and 335 cases for the evaluation of repeated atrial fibrillation.

Reference Standard Establishment Two physicians with at least 5 years of relevant clinical experience will independently review the electrocardiogram data and relevant medical records. They will assess the suitability of the electrocardiogram data, confirm the presence or absence and timing of atrial fibrillation episodes, and verify the previous history of atrial fibrillation. Any disagreement between the physicians will be resolved according to the procedures specified in the study protocol. The final agreed classification will be used as the reference standard.

Application of the Investigational Device The medical device operator will receive the test electrocardiogram datasets labeled only with screening numbers and will not be provided with the reference-standard results or positive or negative group classifications. The blinded datasets will be analyzed using SMD-AFECG to generate the predicted risk of atrial fibrillation occurring within 2 hours. The software-generated results will be recorded in the electronic data capture system.

Statistical Analysis The software-generated predictions will be compared with the reference-standard classifications. The primary performance measures are the areas under the receiver operating characteristic curves for the prediction of new-onset atrial fibrillation and repeated atrial fibrillation. Two-sided 95% confidence intervals will be calculated using DeLong's method with a Wald-type confidence interval.

The performance criterion for new-onset atrial fibrillation will be met if the lower bound of the 95% confidence interval for the area under the curve is greater than 0.7753. The performance criterion for repeated atrial fibrillation will be met if the corresponding lower bound is greater than 0.8299. The study will be considered successful only if both primary performance criteria are met.

Secondary performance measures include the area under the precision-recall curve, precision, recall, F1 score, sensitivity, specificity, positive predictive value, negative predictive value, confusion-matrix results at different thresholds, and prediction time horizon.

Because this is a retrospective study using previously collected data, there will be no direct participant contact, additional examination, treatment, or change in clinical care.

Tipo de estudio

De observación

Inscripción (Estimado)

797

Contactos y Ubicaciones

Esta sección proporciona los datos de contacto de quienes realizan el estudio e información sobre dónde se lleva a cabo este estudio.

Estudio Contacto

Criterios de participación

Los investigadores buscan personas que se ajusten a una determinada descripción, denominada criterio de elegibilidad. Algunos ejemplos de estos criterios son el estado de salud general de una persona o tratamientos previos.

Criterio de elegibilidad

Edades elegibles para estudiar

  • Adulto
  • Adulto Mayor

Acepta Voluntarios Saludables

No

Método de muestreo

Muestra no probabilística

Población de estudio

Patients aged 19 years or older whose continuous single-lead electrocardiogram data were collected through VitalDB at Seoul National University Hospital from September 1, 2022, to before September 30, 2025. Eligible datasets will be identified retrospectively from the VitalDB registry and corresponding clinical records. The study population includes patients with and without a previous history of atrial fibrillation and with or without an atrial fibrillation episode during the predefined assessment window. Eligible datasets will be classified into the NOAF-positive, NOAF-negative, RAF-positive, and RAF-negative groups.

Descripción

Inclusion Criteria:

  • Existing continuous single-lead electrocardiogram data collected through VitalDB at Seoul National University Hospital from September 1, 2022, to before September 30, 2025
  • Participant aged 19 years or older at the time of electrocardiogram data collection
  • Availability of an electrocardiogram dataset meeting one of the following criteria:

    • No previous history of atrial fibrillation and a first atrial fibrillation episode, with continuous atrial fibrillation-free electrocardiogram data for the 2 hours before onset and electrocardiogram data for 1 hour after onset (NOAF-positive group)
    • No previous history of atrial fibrillation and no atrial fibrillation episode, with continuous atrial fibrillation-free electrocardiogram data for the 2 hours before a selected index time and electrocardiogram data for 1 hour after the index time (NOAF-negative group)
    • Previous history of atrial fibrillation and an atrial fibrillation episode, with continuous atrial fibrillation-free electrocardiogram data for the 2 hours before onset and electrocardiogram data for 1 hour after onset (RAF-positive group)
    • Previous history of atrial fibrillation and no atrial fibrillation episode, with continuous atrial fibrillation-free electrocardiogram data for the 2 hours before a selected index time and electrocardiogram data for 1 hour after the index time (RAF-negative group)
  • If atrial fibrillation occurred within 2 hours after the start of electrocardiogram recording, availability of electrocardiogram data from the start of recording through up to 1 hour after atrial fibrillation onset, with a minimum recorded electrocardiogram duration of 10 minutes

Exclusion Criteria:

  • Electrocardiogram data considered to be of inadequate quality because of noise, poor equipment handling, unclear signals, artifacts, or similar problems
  • Electrocardiogram data with signal loss of 5 percent or greater
  • Electrocardiogram data with a heart rate below 30 beats per minute or above 300 beats per minute
  • Data collected from a participant who was pregnant or breastfeeding at the time of data collection
  • Data previously used for training or validation of the artificial intelligence model
  • Data considered unsuitable for this clinical study by the principal investigator

Plan de estudios

Esta sección proporciona detalles del plan de estudio, incluido cómo está diseñado el estudio y qué mide el estudio.

¿Cómo está diseñado el estudio?

Detalles de diseño

Cohortes e Intervenciones

Grupo / Cohorte
NOAF Positive Group
Single-lead electrocardiogram datasets from patients without a previous history of atrial fibrillation that include the first atrial fibrillation episode. The datasets include an atrial fibrillation-free electrocardiogram period before the onset of atrial fibrillation and electrocardiogram data following its onset. SMD-AFECG will be retrospectively applied to estimate the risk of atrial fibrillation occurring within 2 hours.
NOAF Negative Group
Single-lead electrocardiogram datasets from patients without a previous history of atrial fibrillation and without an atrial fibrillation episode during the predefined assessment period. The datasets include a continuous atrial fibrillation-free electrocardiogram period before a selected index time and electrocardiogram data following the index time. SMD-AFECG will be retrospectively applied to estimate the risk of atrial fibrillation occurring within 2 hours.
RAF Positive Group
Single-lead electrocardiogram datasets from patients with a previous history of atrial fibrillation that include an atrial fibrillation episode. The datasets include an atrial fibrillation-free electrocardiogram period before the onset of atrial fibrillation and electrocardiogram data following its onset. SMD-AFECG will be retrospectively applied to estimate the risk of atrial fibrillation occurring within 2 hours.
RAF Negative Group
Single-lead electrocardiogram datasets from patients with a previous history of atrial fibrillation and without an atrial fibrillation episode during the predefined assessment period. The datasets include a continuous atrial fibrillation-free electrocardiogram period before a selected index time and electrocardiogram data following the index time. SMD-AFECG will be retrospectively applied to estimate the risk of atrial fibrillation occurring within 2 hours.

¿Qué mide el estudio?

Medidas de resultado primarias

Medida de resultado
Medida Descripción
Periodo de tiempo
AUROC of SMD-AFECG for Predicting New-Onset Atrial Fibrillation Within 2 Hours
Periodo de tiempo: During the AF-free ECG window immediately preceding AF onset for positive cases or the selected index time for negative cases (up to 2 hours)
The area under the receiver operating characteristic curve (AUROC) will be calculated by comparing the maximum risk score generated by SMD-AFECG with the reference-standard classification of the NOAF-positive and NOAF-negative ECG datasets. A two-sided 95% confidence interval will be estimated using DeLong's method with a Wald-type confidence interval. The performance criterion will be met if the lower bound of the 95% confidence interval is greater than 0.7753.
During the AF-free ECG window immediately preceding AF onset for positive cases or the selected index time for negative cases (up to 2 hours)
AUROC of SMD-AFECG for Predicting Repeated Atrial Fibrillation Within 2 Hours
Periodo de tiempo: During the AF-free ECG window immediately preceding AF onset for positive cases or the selected index time for negative cases (up to 2 hours)
The area under the receiver operating characteristic curve (AUROC) will be calculated by comparing the maximum risk score generated by SMD-AFECG with the reference-standard classification of the RAF-positive and RAF-negative ECG datasets. A two-sided 95% confidence interval will be estimated using DeLong's method with a Wald-type confidence interval. The performance criterion will be met if the lower bound of the 95% confidence interval is greater than 0.8299.
During the AF-free ECG window immediately preceding AF onset for positive cases or the selected index time for negative cases (up to 2 hours)

Medidas de resultado secundarias

Medida de resultado
Medida Descripción
Periodo de tiempo
AUPRC of SMD-AFECG for Predicting New-Onset Atrial Fibrillation Within 2 Hours
Periodo de tiempo: During the AF-free ECG window immediately preceding AF onset for positive cases or the selected index time for negative cases, up to 2 hours
The area under the precision-recall curve (AUPRC) will be calculated from the relationship between precision and recall using the reference-standard NOAF-positive and NOAF-negative ECG datasets. AUPRC ranges from 0 to 1, with higher values indicating better identification of positive cases.
During the AF-free ECG window immediately preceding AF onset for positive cases or the selected index time for negative cases, up to 2 hours
AUPRC of SMD-AFECG for Predicting Repeated Atrial Fibrillation Within 2 Hours
Periodo de tiempo: During the AF-free ECG window immediately preceding AF onset for positive cases or the selected index time for negative cases, up to 2 hours
The area under the precision-recall curve (AUPRC) will be calculated from the relationship between precision and recall using the reference-standard RAF-positive and RAF-negative ECG datasets. AUPRC ranges from 0 to 1, with higher values indicating better identification of positive cases.
During the AF-free ECG window immediately preceding AF onset for positive cases or the selected index time for negative cases, up to 2 hours
Threshold-Specific Sensitivity of SMD-AFECG for Predicting New-Onset Atrial Fibrillation Within 2 Hours
Periodo de tiempo: During the AF-free ECG window immediately preceding AF onset for positive cases or the selected index time for negative cases, up to 2 hours
At each evaluated threshold, sensitivity, also referred to as recall, will be calculated as the proportion of reference-standard NOAF-positive ECG datasets classified as positive by SMD-AFECG.
During the AF-free ECG window immediately preceding AF onset for positive cases or the selected index time for negative cases, up to 2 hours
Threshold-Specific Sensitivity of SMD-AFECG for Predicting Repeated Atrial Fibrillation Within 2 Hours
Periodo de tiempo: During the AF-free ECG window immediately preceding AF onset for positive cases or the selected index time for negative cases, up to 2 hours
At each evaluated threshold, sensitivity, also referred to as recall, will be calculated as the proportion of reference-standard RAF-positive ECG datasets classified as positive by SMD-AFECG.
During the AF-free ECG window immediately preceding AF onset for positive cases or the selected index time for negative cases, up to 2 hours
Threshold-Specific Specificity of SMD-AFECG for Predicting New-Onset Atrial Fibrillation Within 2 Hours
Periodo de tiempo: During the AF-free ECG window immediately preceding AF onset for positive cases or the selected index time for negative cases, up to 2 hours
At each evaluated threshold, specificity will be calculated as the proportion of reference-standard NOAF-negative ECG datasets classified as negative by SMD-AFECG.
During the AF-free ECG window immediately preceding AF onset for positive cases or the selected index time for negative cases, up to 2 hours
Threshold-Specific Specificity of SMD-AFECG for Predicting Repeated Atrial Fibrillation Within 2 Hours
Periodo de tiempo: During the AF-free ECG window immediately preceding AF onset for positive cases or the selected index time for negative cases, up to 2 hours
At each evaluated threshold, specificity will be calculated as the proportion of reference-standard RAF-negative ECG datasets classified as negative by SMD-AFECG.
During the AF-free ECG window immediately preceding AF onset for positive cases or the selected index time for negative cases, up to 2 hours
Threshold-Specific Positive Predictive Value of SMD-AFECG for New-Onset Atrial Fibrillation
Periodo de tiempo: During the AF-free ECG window immediately preceding AF onset for positive cases or the selected index time for negative cases, up to 2 hours
At each evaluated threshold, positive predictive value, also referred to as precision, will be calculated as the proportion of ECG datasets classified as positive by SMD-AFECG that are NOAF-positive according to the reference standard.
During the AF-free ECG window immediately preceding AF onset for positive cases or the selected index time for negative cases, up to 2 hours
Threshold-Specific Positive Predictive Value of SMD-AFECG for Repeated Atrial Fibrillation
Periodo de tiempo: During the AF-free ECG window immediately preceding AF onset for positive cases or the selected index time for negative cases, up to 2 hours
At each evaluated threshold, positive predictive value, also referred to as precision, will be calculated as the proportion of ECG datasets classified as positive by SMD-AFECG that are RAF-positive according to the reference standard.
During the AF-free ECG window immediately preceding AF onset for positive cases or the selected index time for negative cases, up to 2 hours
Threshold-Specific Negative Predictive Value of SMD-AFECG for New-Onset Atrial Fibrillation
Periodo de tiempo: During the AF-free ECG window immediately preceding AF onset for positive cases or the selected index time for negative cases, up to 2 hours
At each evaluated threshold, negative predictive value will be calculated as the proportion of ECG datasets classified as negative by SMD-AFECG that are NOAF-negative according to the reference standard.
During the AF-free ECG window immediately preceding AF onset for positive cases or the selected index time for negative cases, up to 2 hours
Threshold-Specific Negative Predictive Value of SMD-AFECG for Repeated Atrial Fibrillation
Periodo de tiempo: At each evaluated threshold, negative predictive value will be calculated as the proportion of ECG datasets classified as negative by SMD-AFECG that are RAF-negative according to the reference standard.
At each evaluated threshold, negative predictive value will be calculated as the proportion of ECG datasets classified as negative by SMD-AFECG that are RAF-negative according to the reference standard.
At each evaluated threshold, negative predictive value will be calculated as the proportion of ECG datasets classified as negative by SMD-AFECG that are RAF-negative according to the reference standard.
Threshold-Specific F1 Score of SMD-AFECG for Predicting New-Onset Atrial Fibrillation
Periodo de tiempo: During the AF-free ECG window immediately preceding AF onset for positive cases or the selected index time for negative cases, up to 2 hours
At each evaluated threshold, the F1 score will be calculated as the harmonic mean of precision and recall for the classification of NOAF-positive and NOAF-negative ECG datasets. Values range from 0 to 1, with higher values indicating better performance.
During the AF-free ECG window immediately preceding AF onset for positive cases or the selected index time for negative cases, up to 2 hours
Threshold-Specific F1 Score of SMD-AFECG for Predicting Repeated Atrial Fibrillation
Periodo de tiempo: During the AF-free ECG window immediately preceding AF onset for positive cases or the selected index time for negative cases, up to 2 hours
At each evaluated threshold, the F1 score will be calculated as the harmonic mean of precision and recall for the classification of RAF-positive and RAF-negative ECG datasets. Values range from 0 to 1, with higher values indicating better performance.
During the AF-free ECG window immediately preceding AF onset for positive cases or the selected index time for negative cases, up to 2 hours
Prediction Time Horizon of SMD-AFECG for New-Onset Atrial Fibrillation
Periodo de tiempo: Within the 2-hour period before the onset of the first atrial fibrillation episode
The prediction time horizon will be calculated as the time interval, in minutes, between the SMD-AFECG prediction or alarm and the onset of the reference-standard first atrial fibrillation episode among NOAF-positive ECG datasets.
Within the 2-hour period before the onset of the first atrial fibrillation episode
Prediction Time Horizon of SMD-AFECG for Repeated Atrial Fibrillation
Periodo de tiempo: Within the 2-hour period before the onset of the atrial fibrillation episode
The prediction time horizon will be calculated as the time interval, in minutes, between the SMD-AFECG prediction or alarm and the onset of the reference-standard atrial fibrillation episode among RAF-positive ECG datasets.
Within the 2-hour period before the onset of the atrial fibrillation episode

Colaboradores e Investigadores

Aquí es donde encontrará personas y organizaciones involucradas en este estudio.

Fechas de registro del estudio

Estas fechas rastrean el progreso del registro del estudio y los envíos de resultados resumidos a ClinicalTrials.gov. Los registros del estudio y los resultados informados son revisados ​​por la Biblioteca Nacional de Medicina (NLM) para asegurarse de que cumplan con los estándares de control de calidad específicos antes de publicarlos en el sitio web público.

Fechas importantes del estudio

Inicio del estudio (Estimado)

1 de octubre de 2026

Finalización primaria (Estimado)

30 de noviembre de 2026

Finalización del estudio (Estimado)

31 de diciembre de 2026

Fechas de registro del estudio

Enviado por primera vez

20 de julio de 2026

Primero enviado que cumplió con los criterios de control de calidad

20 de julio de 2026

Publicado por primera vez (Actual)

23 de julio de 2026

Actualizaciones de registros de estudio

Última actualización publicada (Actual)

23 de julio de 2026

Última actualización enviada que cumplió con los criterios de control de calidad

20 de julio de 2026

Última verificación

1 de julio de 2026

Más información

Términos relacionados con este estudio

Plan de datos de participantes individuales (IPD)

¿Planea compartir datos de participantes individuales (IPD)?

NO

Descripción del plan IPD

Individual participant data collected in this study are not planned to be shared with external researchers. Any future sharing would require separate sponsor approval and compliance with applicable privacy, ethical, and institutional requirements.

Información sobre medicamentos y dispositivos, documentos del estudio

Estudia un producto farmacéutico regulado por la FDA de EE. UU.

No

Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .

Suscribir