- ICH GCP
- Yhdysvaltain kliinisten tutkimusten rekisteri
- Kliininen tutkimus NCT07706504
Comparison of Two Quantitative EMG Monitors for Deep Neuromuscular Block in Laparoscopic/Robotic and VATS Surgery
torstai 9. heinäkuuta 2026 päivittänyt: Boris Mraovic, Ohio State University
Comparison of Two Quantitative Electromyography Monitors for Deep Neuro Muscular Block in Patients Undergoing Abdominal Laparoscopic/Robotic or Thoracic Video-Assisted (VATS) Surgeries
The main objective of the study is to compare the frequency of intraoperative spontaneous diaphragmatic movements and breath-initiation efforts in participants undergoing robotic and/or thoracic surgery under deep neuromuscular blockade (NMB), using either the TetraGraph or TwitchView applied to the dominant hand.
These events will serve as objective indicators to evaluate each device's ability to monitor and maintain adequate deep NMB.
Tutkimuksen yleiskatsaus
Tila
Rekrytointi
Yksityiskohtainen kuvaus
This study is a prospective intraoperative device comparison trial evaluating neuromuscular blockade monitoring using two quantitative EMG-based devices, the TetraGraph (Senzime) and TwitchView (Blink), during elective laparoscopic, robotic, or video-assisted thoracoscopic (VATS) abdominal surgeries requiring neuromuscular blockade.
Participants will be assigned to receive both monitoring devices simultaneously, placed on opposite upper extremities, allowing each subject to serve as their own control.
Standard anesthesia care will be maintained throughout the procedure, including volatile anesthetic titration, opioid administration as needed, mechanical ventilation targeting normocapnia, temperature regulation, and routine neuromuscular blockade management with rocuronium and reversal with sugammadex per institutional practice.
Neuromuscular function will be continuously assessed using train-of-four (TOF), train-of-four ratio, and post-tetanic count (PTC) measurements, with data recorded at predefined intraoperative time points including baseline, post-paralytic administration, maintenance, reversal, and prior to extubation.
Additional data will include dosing of neuromuscular blocking agents, physiologic parameters, and intraoperative clinical observations relevant to depth of blockade and recovery.
Provider usability and workflow integration will be evaluated postoperatively through brief standardized surveys assessing device usability, signal quality, and overall satisfaction.
All collected data will be derived from device outputs and routine perioperative clinical documentation without altering standard patient care.
Opintotyyppi
Interventio
Ilmoittautuminen (Arvioitu)
35
Vaihe
- Ei sovellettavissa
Yhteystiedot ja paikat
Tässä osiossa on tutkimuksen suorittajien yhteystiedot ja tiedot siitä, missä tämä tutkimus suoritetaan.
Opiskeluyhteys
- Nimi: Alberto Uribe, MD
- Puhelinnumero: 6142933559
- Sähköposti: alberto.uribe@osumc.edu
Tutki yhteystietojen varmuuskopiointi
- Nimi: Elvia Vera, MD
- Puhelinnumero: 6142933559
- Sähköposti: Elvia.VeraMiquilena@osumc.edu
Opiskelupaikat
-
-
Ohio
-
Columbus, Ohio, Yhdysvallat, 43201
- Rekrytointi
- The Ohio State University Wexner Medical Center
-
Ottaa yhteyttä:
- Alberto Uribe, MD
- Puhelinnumero: 614-293-0775
- Sähköposti: alberto.uribe@osumc.edu
-
-
Osallistumiskriteerit
Tutkijat etsivät ihmisiä, jotka sopivat tiettyyn kuvaukseen, jota kutsutaan kelpoisuuskriteereiksi. Joitakin esimerkkejä näistä kriteereistä ovat henkilön yleinen terveydentila tai aiemmat hoidot.
Kelpoisuusvaatimukset
Opintokelpoiset iät
- Aikuinen
- Vanhempi Aikuinen
Hyväksyy terveitä vapaaehtoisia
Joo
Kuvaus
Inclusion Criteria:
- Adult male or female patients aged ≥ 18 years old.
- Patients undergo elective abdominal laparoscopic/robotic or thoracic video-assisted (VAT) surgeries requiring general anesthesia for greater than 3 hours at The Ohio State University Wexner Medical Center.
- Able to provide a signed, written informed consent.
- Able to speak, read, and write in English.
- ASA physical status I-III.
Exclusion Criteria:
- Patients who require emergency surgery or an emergent intervention.
- Any documented cognitive or psychological disorders that, in the investigator's opinion, can interfere with the patient's pain perception.
- Vulnerable populations: pregnant females, prisoners, breastfeeding.
- Presence of any medical condition that, in the opinion of the principal investigator, should exclude the patient from the study (patients with pre-existing neuromuscular diseases.
- Allergy or contraindications to any of the anesthetics, NMB agents, or sugammadex.
- Patients with anatomical abnormalities of the hands or arms that prevent proper placement of the sensors for nerve stimulation.
- Limited access to the monitoring area due to surgical positioning.
- Patients with peripheral vascular disease, since it may affect measurement accuracy.
Opintosuunnitelma
Tässä osiossa on tietoja tutkimussuunnitelmasta, mukaan lukien kuinka tutkimus on suunniteltu ja mitä tutkimuksella mitataan.
Miten tutkimus on suunniteltu?
Suunnittelun yksityiskohdat
- Ensisijainen käyttötarkoitus: Diagnostiikka
- Jako: Satunnaistettu
- Inventiomalli: Rinnakkaistehtävä
- Naamiointi: Kaksinkertainen
Aseet ja interventiot
Osallistujaryhmä / Arm |
Interventio / Hoito |
|---|---|
|
Kokeellinen: TetraGraph (Senzime) Dominant Hand
Participants will have the TetraGraph (Senzime) electromyography (EMG) neuromuscular monitoring device placed on the dominant hand and the TwitchView (Blink) EMG neuromuscular monitoring device placed on the non-dominant hand.
The device placed on the dominant hand will be used to guide intraoperative neuromuscular blockade management, while measurements from both devices will be collected for comparison of quantitative neuromuscular monitoring performance.
|
The TetraGraph (Senzime) quantitative electromyography (EMG) neuromuscular monitoring device will be applied to the participant's dominant hand after induction of general anesthesia and prior to neuromuscular blocking agent administration.
Surface electrodes will be placed over the ulnar nerve per manufacturer instructions, and the device will be calibrated to obtain baseline Train-of-Four (TOF) measurements.
Following rocuronium administration, the device will continuously monitor neuromuscular function, recording TOF approximately every 15 seconds and post-tetanic count (PTC) every 5 minutes when TOF count is zero.
As the dominant-hand monitor, its readings will guide rocuronium redosing to maintain deep neuromuscular blockade (PTC 0-3) and assess recovery.
During emergence, it will guide reversal, with extubation at TOF ratio ≥0.9.
|
|
Kokeellinen: TwitchView (Blink) Dominant Hand
Participants will have the TwitchView (Blink) electromyography (EMG) neuromuscular monitoring device placed on the dominant hand and the TetraGraph (Senzime) EMG neuromuscular monitoring device placed on the non-dominant hand.
The device placed on the dominant hand will be used to guide intraoperative neuromuscular blockade management, while measurements from both devices will be collected for comparison of quantitative neuromuscular monitoring performance.
|
The TwitchView (Blink) quantitative electromyography (EMG) neuromuscular monitoring device will be applied to the participant's dominant hand after induction of general anesthesia and prior to administration of neuromuscular blocking agents.
Surface electrodes will be placed over the ulnar nerve per manufacturer instructions, and the device will be calibrated to obtain baseline Train-of-Four (TOF) measurements.
Following rocuronium administration, neuromuscular function will be continuously monitored, with TOF recorded approximately every 15 seconds and post-tetanic count (PTC) every 5 minutes when TOF count is zero.
As the dominant-hand device, its measurements will guide rocuronium redosing to maintain deep neuromuscular blockade (PTC 0-3) and assess recovery.
During emergence, it will guide reversal, with extubation at TOF ratio ≥0.9.
|
Mitä tutkimuksessa mitataan?
Ensisijaiset tulostoimenpiteet
Tulosmittaus |
Toimenpiteen kuvaus |
Aikaikkuna |
|---|---|---|
|
Incidence of intraoperative diaphragmatic movement during deep neuromuscular blockade
Aikaikkuna: During maintenance of deep neuromuscular blockade throughout the surgical procedure, from achievement of deep neuromuscular blockade after induction of general anesthesia until completion of surgery prior to emergence.
|
The number of spontaneous diaphragmatic movement events observed during maintenance of deep neuromuscular blockade will be recorded and compared between participants monitored with the TetraGraph (Senzime) and TwitchView (Blink) electromyography (EMG) neuromuscular monitoring devices.
|
During maintenance of deep neuromuscular blockade throughout the surgical procedure, from achievement of deep neuromuscular blockade after induction of general anesthesia until completion of surgery prior to emergence.
|
|
Incidence of spontaneous breath-initiation efforts during deep neuromuscular blockade
Aikaikkuna: During maintenance of deep neuromuscular blockade throughout the surgical procedure, from achievement of deep neuromuscular blockade after induction of general anesthesia until completion of surgery prior to emergence.
|
The number of spontaneous breath-initiation efforts observed during maintenance of deep neuromuscular blockade will be recorded and compared between participants monitored with the TetraGraph (Senzime) and TwitchView (Blink) EMG neuromuscular monitoring devices.
|
During maintenance of deep neuromuscular blockade throughout the surgical procedure, from achievement of deep neuromuscular blockade after induction of general anesthesia until completion of surgery prior to emergence.
|
Toissijaiset tulostoimenpiteet
Tulosmittaus |
Toimenpiteen kuvaus |
Aikaikkuna |
|---|---|---|
|
Intubation Difficulty Scale (IDS) score
Aikaikkuna: During tracheal intubation following induction of general anesthesia and administration of neuromuscular blockade, prior to surgical incision.
|
Intubation difficulty will be assessed using the Intubation Difficulty Scale (IDS), a composite score ranging from 0 upward, where 0 represents easy intubation and higher scores indicate greater intubation difficulty.
|
During tracheal intubation following induction of general anesthesia and administration of neuromuscular blockade, prior to surgical incision.
|
|
Incidence of patient movement or bucking during intubation
Aikaikkuna: During tracheal intubation following induction of general anesthesia and administration of neuromuscular blockade, prior to surgical incision.
|
Percentage of participants with patient movement or bucking during tracheal intubation, assessed as a clinical indicator of inadequate neuromuscular blockade.
|
During tracheal intubation following induction of general anesthesia and administration of neuromuscular blockade, prior to surgical incision.
|
|
Cormack-Lehane grade of vocal cord visibility during intubation
Aikaikkuna: During tracheal intubation following induction of general anesthesia and administration of neuromuscular blockade, prior to surgical incision.
|
Vocal cord visibility will be assessed during tracheal intubation using the Cormack-Lehane laryngoscopic grading system (Grades I-IV), where Grade I indicates the best glottic view and Grade IV indicates the poorest view.
|
During tracheal intubation following induction of general anesthesia and administration of neuromuscular blockade, prior to surgical incision.
|
|
Vocal cord movement during intubation
Aikaikkuna: During tracheal intubation following induction of general anesthesia and administration of neuromuscular blockade, prior to surgical incision.
|
Vocal cord movement will be assessed during tracheal intubation using the predefined study classification (e.g., open, moving, closing, or closed) as an indicator of intubating conditions.
|
During tracheal intubation following induction of general anesthesia and administration of neuromuscular blockade, prior to surgical incision.
|
|
Post-tetanic count (PTC)
Aikaikkuna: Throughout the intraoperative period during maintenance of neuromuscular blockade, from induction of general anesthesia until initiation of emergence from anesthesia.
|
Post-tetanic count (PTC), reported as the number of muscle responses (count), will be measured using the TetraGraph and TwitchView neuromuscular monitoring devices and compared for agreement and correlation.
|
Throughout the intraoperative period during maintenance of neuromuscular blockade, from induction of general anesthesia until initiation of emergence from anesthesia.
|
|
Train-of-four (TOF) count
Aikaikkuna: Throughout the intraoperative period during maintenance of neuromuscular blockade, from induction of general anesthesia until initiation of emergence from anesthesia.
|
Train-of-four (TOF) count, reported as the number of twitch responses (0-4), will be measured using the TetraGraph and TwitchView devices and compared for agreement and correlation.
|
Throughout the intraoperative period during maintenance of neuromuscular blockade, from induction of general anesthesia until initiation of emergence from anesthesia.
|
|
Time to recovery of train-of-four ratio >0.9
Aikaikkuna: During emergence from anesthesia, from administration of neuromuscular blockade reversal agent until extubation.
|
Time, in minutes, from administration of neuromuscular blockade reversal agent to achievement of a train-of-four (TOF) ratio >0.9, as measured by the TetraGraph and TwitchView devices.
|
During emergence from anesthesia, from administration of neuromuscular blockade reversal agent until extubation.
|
|
Train-of-four ratio before extubation
Aikaikkuna: Immediately before extubation following completion of surgery and recovery from neuromuscular blockade.
|
Train-of-four (TOF) ratio, reported as a decimal value from 0 to 1.0 (or percentage), measured immediately before extubation using the TetraGraph and TwitchView devices.
Higher values indicate greater recovery of neuromuscular function.
|
Immediately before extubation following completion of surgery and recovery from neuromuscular blockade.
|
|
System Usability Scale (SUS) score
Aikaikkuna: Immediately after completion of the intraoperative procedure and device use.
|
Anesthesia care providers will evaluate device usability using the System Usability Scale (SUS), which ranges from 0 to 100, with higher scores indicating better perceived usability.
|
Immediately after completion of the intraoperative procedure and device use.
|
|
Leiden Surgical Rating Scale (L-SRS) score
Aikaikkuna: At completion of the surgical procedure, prior to emergence from general anesthesia.
|
Surgeon-rated surgical conditions will be assessed using the Leiden Surgical Rating Scale (L-SRS), a 5-point ordinal scale ranging from 1 to 5, where 1 indicates extremely poor surgical conditions and 5 indicates optimal surgical conditions.
Higher scores indicate better surgical conditions.
|
At completion of the surgical procedure, prior to emergence from general anesthesia.
|
Yhteistyökumppanit ja tutkijat
Täältä löydät tähän tutkimukseen osallistuvat ihmiset ja organisaatiot.
Sponsori
Julkaisuja ja hyödyllisiä linkkejä
Tutkimusta koskevien tietojen syöttämisestä vastaava henkilö toimittaa nämä julkaisut vapaaehtoisesti. Nämä voivat koskea mitä tahansa tutkimukseen liittyvää.
Yleiset julkaisut
- Kopman AF, Brull SJ. Etiology of Acceleromyographic Train-of-four Overshoot: A Hypothesis. Anesthesiology. 2025 Jun 1;142(6):1171-1173. doi: 10.1097/ALN.0000000000005377. Epub 2025 Apr 7. No abstract available.
- Blobner M, Frick CG, Stauble RB, Feussner H, Schaller SJ, Unterbuchner C, Lingg C, Geisler M, Fink H. Neuromuscular blockade improves surgical conditions (NISCO). Surg Endosc. 2015 Mar;29(3):627-36. doi: 10.1007/s00464-014-3711-7. Epub 2014 Aug 15.
- Ahluwalia JS, Morley CJ, Mockridge JN. Computerised determination of spontaneous inspiratory and expiratory times in premature neonates during intermittent positive pressure ventilation. II: Results from 20 babies. Arch Dis Child Fetal Neonatal Ed. 1994 Nov;71(3):F161-4. doi: 10.1136/fn.71.3.f161.
- Bussey L, Jelacic S, Togashi K, Hulvershorn J, Bowdle A. Train-of-four monitoring with the twitchview monitor electctromyograph compared to the GE NMT electromyograph and manual palpation. J Clin Monit Comput. 2021 Dec;35(6):1477-1483. doi: 10.1007/s10877-020-00615-7. Epub 2020 Nov 9.
- Matsuda K. [Mechanical properties of dental alloys for clasps]. Kokubyo Gakkai Zasshi. 1976 Jun;43(2):192. No abstract available. Japanese.
Opintojen ennätyspäivät
Nämä päivämäärät seuraavat ClinicalTrials.gov-sivustolle lähetettyjen tutkimustietueiden ja yhteenvetojen edistymistä. National Library of Medicine (NLM) tarkistaa tutkimustiedot ja raportoidut tulokset varmistaakseen, että ne täyttävät tietyt laadunvalvontastandardit, ennen kuin ne julkaistaan julkisella verkkosivustolla.
Opi tärkeimmät päivämäärät
Opiskelun aloitus (Todellinen)
Maanantai 9. helmikuuta 2026
Ensisijainen valmistuminen (Arvioitu)
Keskiviikko 9. helmikuuta 2028
Opintojen valmistuminen (Arvioitu)
Keskiviikko 9. elokuuta 2028
Opintoihin ilmoittautumispäivät
Ensimmäinen lähetetty
Torstai 14. toukokuuta 2026
Ensimmäinen toimitettu, joka täytti QC-kriteerit
Torstai 9. heinäkuuta 2026
Ensimmäinen Lähetetty (Todellinen)
Keskiviikko 15. heinäkuuta 2026
Tutkimustietojen päivitykset
Viimeisin päivitys julkaistu (Todellinen)
Keskiviikko 15. heinäkuuta 2026
Viimeisin lähetetty päivitys, joka täytti QC-kriteerit
Torstai 9. heinäkuuta 2026
Viimeksi vahvistettu
Perjantai 1. toukokuuta 2026
Lisää tietoa
Tähän tutkimukseen liittyvät termit
Muut tutkimustunnusnumerot
- STUDY20252768
Yksittäisten osallistujien tietojen suunnitelma (IPD)
Aiotko jakaa yksittäisten osallistujien tietoja (IPD)?
PÄÄTTÄMÄTÖN
Lääke- ja laitetiedot, tutkimusasiakirjat
Tutkii yhdysvaltalaista FDA sääntelemää lääkevalmistetta
Ei
Tutkii yhdysvaltalaista FDA sääntelemää laitetuotetta
Joo
Yhdysvalloissa valmistettu ja sieltä viety tuote
Ei
Nämä tiedot haettiin suoraan verkkosivustolta clinicaltrials.gov ilman muutoksia. Jos sinulla on pyyntöjä muuttaa, poistaa tai päivittää tutkimustietojasi, ota yhteyttä register@clinicaltrials.gov. Heti kun muutos on otettu käyttöön osoitteessa clinicaltrials.gov, se päivitetään automaattisesti myös verkkosivustollemme .