Methods for the Measurement of Respiratory Rate

23 mars 2016 mis à jour par: Niels Egholm Pedersen, Herlev Hospital

This study aims to investigate these research questions:

  1. Is there a difference in the results obtained, when respiratory rate is measured automatically by an electronic device (SensiumVitals® system, Sensium Healthcare) compared with a research assistant using a criterion standard approach?
  2. Is there a difference in the results obtained, when respiratory rate is measured automatically by an electronic device (SensiumVitals® system, Sensium Healthcare) compared with hospital staff's current clinical practice?

Aperçu de l'étude

Statut

Complété

Intervention / Traitement

Description détaillée

Respiratory rate is among the first vital signs to change when a deterioration in patient conditions occurs, and that it is highly useful in identifying high-risk patients, patients in pain, risk of death in patients treated for myocardial infarctions and medical patients at risk of cardiac arrest.

Respiratory rate is one of the seven physiological parameters included in the Early Warning Score (EWS), an aggregated track- and trigger-system for risk-stratification of patients in hospital wards at risk of imminent clinical deterioration. EWS includes an escalation protocol, defining when it is necessary for the ward staff to call for assistance and when vital parameters should be measured again.

Manual measurement of respiratory rates is the standard in hospital wards in the Capital Region of Denmark. During the measurement of respiratory rate, patients ideally have to lie still and refrain from talking. This poses a difficulty in daily, clinical practice. Several studies in hospital wards both using and not using an EWS-system report that respiratory rate is often omitted when vital signs are measured.

Even if the respiratory rate is measured regularly, poor inter-observer agreement and reproducibility of measured respiratory rates have been found. Respiratory rates recorded by nurses have been reported to be generally higher than those measured by observers using a standardised approach.

The EWS currently in use is validated using measurements obtained by normal nursing practise. If there is an element of human influence on the measurement of respiratory rate, automation could affect the prognostic properties of EWS in an unpredictable way, even if the automatically measured respiratory rates more reliably reflect the patients' physiology than current clinical practise.

If a systematic difference between the recordings by staff and the electronic devices currently on the market exists, this could have an implication on the EWS scores and subsequently, escalation protocols for patients with elevated EWS could be affected. If this is the case, EWS might even have to be re-validated using machine-measured respiratory rates.

Thus, even as there could be a number of advantages of automating the measurements of respiratory rates in hospital wards, due considerations are required before ward staff's manual measurement of respiratory rate can be replaced by measurements obtained from electronic devices. The purpose of this study is to evaluate if there is a systematic difference in the respiratory rate measurements performed by an electronic device (SensiumVitals® system, Sensium Healthcare) compared with the measurements performed by a research assistant using a criterion standard approach and the ward staff's usual measurements.

Type d'étude

Interventionnel

Inscription (Réel)

50

Phase

  • N'est pas applicable

Contacts et emplacements

Cette section fournit les coordonnées de ceux qui mènent l'étude et des informations sur le lieu où cette étude est menée.

Lieux d'étude

      • Herlev, Danemark, 2730
        • Copenhagen Academy for Medical Education and Simulation and Medical-Short Time Ward, Herlev Hospital

Critères de participation

Les chercheurs recherchent des personnes qui correspondent à une certaine description, appelée critères d'éligibilité. Certains exemples de ces critères sont l'état de santé général d'une personne ou des traitements antérieurs.

Critère d'éligibilité

Âges éligibles pour étudier

18 ans et plus (Adulte, Adulte plus âgé)

Accepte les volontaires sains

Non

Sexes éligibles pour l'étude

Tout

La description

Inclusion Criteria:

  • Age 18 years or older.
  • Able and willing to give informed consent to participation in the study and necessary data collection.
  • Admitted to the general medical short-time ward and planned stay there at the time of inclusion greater than two hours.

Exclusion Criteria:

  • Wounds, burns or other conditions that prevent proper attachment of the measurement device, or where wearing the device might cause or increase discomfort from wearing the device.
  • Allergies to medical graded tape or ECG electrodes.
  • Patients with cardiac pacemaker or implanted defibrillator.
  • Any condition or situation preventing proper information, consent and enrolment.
  • Respiratory rate measurements by another device using impedance pneumography.

Plan d'étude

Cette section fournit des détails sur le plan d'étude, y compris la façon dont l'étude est conçue et ce que l'étude mesure.

Comment l'étude est-elle conçue ?

Détails de conception

  • Objectif principal: Diagnostique
  • Répartition: N / A
  • Modèle interventionnel: Affectation à un seul groupe

Armes et Interventions

Groupe de participants / Bras
Intervention / Traitement
Autre: Included patients
All included patients will have respiratory rates measured by a research assistant using a criterion standard approach, by the SensiumVitals device and by the ward staff.
The patients will have their respiratory rates, heart rates and temperature (axilla) measured continuously by the SensiumVitals patch.

Que mesure l'étude ?

Principaux critères de jugement

Mesure des résultats
Description de la mesure
Délai
Difference in respiratory rate recorded by the research assistant using a criterion standard approach and the SensiumVitals device
Délai: 24 Hours
After enrolment at the general, medical short-time ward a research assistant will measure the patient's respiratory rate using a criterion standard approach. The time of this measurement will be registered, and the corresponding measurement made by the SensiumVitals device will be extracted and used for the comparison. Reported with Bland-Altman plots.
24 Hours

Mesures de résultats secondaires

Mesure des résultats
Description de la mesure
Délai
Difference in respiratory rate recorded by the ward staff using standard clinical practice and the SensiumVitals device
Délai: 24 Hours
The ward staff's first measurement of respiratory rate after the patient is enrolled in the study at the general, medical short-time ward as registered in the electronic patient journal. The time of this measurement will be registered, and the corresponding measurement made by the SensiumVitals device will be extracted and used for the comparison. Reported with Bland-Altman plots.
24 Hours

Autres mesures de résultats

Mesure des résultats
Description de la mesure
Délai
Early Warning Score recorded by the research assistant
Délai: 24 Hours
Early Warning Score (point score) will be recorded using standard practice by the research assistant. Parameters included in the Early Warning Score: Peripheal arterial oxygen satuation (%), pulse (beats per minute), use of supplementary oxygen (liters per minute), systolic and diastolic blood pressure (mmHg), level of consciousness (AVPU-scale), temperature (degrees celcius) and pulse regularity (yes/no).
24 Hours
Early Warning Score recorded by the ward staff
Délai: 24 Hours
Early Warning Score (point score) recorded using standard practice by the ward staff. Early Warning Score consists of: Peripheal arterial oxygen satuation (%), pulse (beats per minute), use of supplementary oxygen (liters per minute), systolic and diastolic blood pressure (mmHg), level of consciousness (AVPU-scale) and temperature (degrees celcius).
24 Hours

Collaborateurs et enquêteurs

C'est ici que vous trouverez les personnes et les organisations impliquées dans cette étude.

Parrainer

Les enquêteurs

  • Chercheur principal: Niels E Pedersen, MD, Copenhagen Academy for Medical Education and Simulation, Herlev Hospital, Center for HR, Capital Region of Denmark

Dates d'enregistrement des études

Ces dates suivent la progression des dossiers d'étude et des soumissions de résultats sommaires à ClinicalTrials.gov. Les dossiers d'étude et les résultats rapportés sont examinés par la Bibliothèque nationale de médecine (NLM) pour s'assurer qu'ils répondent à des normes de contrôle de qualité spécifiques avant d'être publiés sur le site Web public.

Dates principales de l'étude

Début de l'étude

1 janvier 2016

Achèvement primaire (Réel)

1 février 2016

Achèvement de l'étude (Réel)

1 février 2016

Dates d'inscription aux études

Première soumission

11 janvier 2016

Première soumission répondant aux critères de contrôle qualité

29 janvier 2016

Première publication (Estimation)

3 février 2016

Mises à jour des dossiers d'étude

Dernière mise à jour publiée (Estimation)

24 mars 2016

Dernière mise à jour soumise répondant aux critères de contrôle qualité

23 mars 2016

Dernière vérification

1 mars 2016

Plus d'information

Termes liés à cette étude

Autres numéros d'identification d'étude

  • CAMES-16-001

Ces informations ont été extraites directement du site Web clinicaltrials.gov sans aucune modification. Si vous avez des demandes de modification, de suppression ou de mise à jour des détails de votre étude, veuillez contacter register@clinicaltrials.gov. Dès qu'un changement est mis en œuvre sur clinicaltrials.gov, il sera également mis à jour automatiquement sur notre site Web .

S'abonner