R Wave Sensing After VT Ablation (RASA-VT)

18 dicembre 2018 aggiornato da: Ivan Zeljkovic, University Hospital "Sestre Milosrdnice"

R Wave Sensing After Radiofrequency Ablation of Ventricular Tachycardia Substrate in Patients With Structural Heart Disease

Sustained, monomorphic ventricular tachycardia (VT) is most commonly encountered in patients with structural heart disease, usually with ischemic aetiology. It has been proven that repeated episodes of sustained VT contribute to the mortality of patients with structural heart disease. These patients are usually implanted with implantable cardioverter defibrillator without (ICD) or with cardiac resynchronisation therapy (CRT-D). According to the current guidelines 3D mapping with radiofrequency (RF) ablation of the tachycardia substrate is an established therapeutic option. As part of the clinical follow-up of patients implanted with cardiac implantable electronic devices (CIED), CIED function parameters are monitored every 6 months, which is always done before and after any procedure involving RF ablation. Furthermore, a transient drop of R wave sensing has been demonstrated after the atrioventricular (AV) node ablation in patients with single-chamber pacemakers and fast atrial fibrillation. The aim of this study is to assess the change in R wave sensing after the RF ablation of VT substrate.

This study will be performed as a clinical, prospective, multi-centre, observational cohort study with a structured follow-up period of 12 months. All consecutive patients with sustained VT implanted with ICD or CRT-D undergoing RF ablation procedure of myocardial substrate, who are able to understand and sign informed consent, will be enrolled. Primary objective is a R wave sensing drop > 30% after VT substrate ablation procedure. Recruiting should not exceed 12 months with the minimal follow-up period of 12 months (24 months in total). Standardized statistical methods and test will be done using SPSS Software Version 22.0 or newer.

This unique study offers the possibility to show the impact of RF ablation on short-term and long-term R wave sensing change assessed by ICD or CRT-D's ventricle electrode in patients with sustained VT and structural heart disease undergoing ablation procedure. This observational data is needed to further refine the treatment of these patients and to prevent possible ICD/CRT-D dysfunction which could endanger this patient population.

Panoramica dello studio

Descrizione dettagliata

Sustained, monomorphic ventricular tachycardia (VT) is most commonly encountered in patients with structural heart disease, usually with ischemic aetiology. It has been proven that repeated episodes of sustained VT contribute to the mortality of patients with structural heart disease. These patients are usually implanted with implantable cardioverter defibrillator without (ICD) or with cardiac resynchronisation therapy (CRT-D) as part of sudden cardiac death (SCD) prevention and have had activation of the device due to the sustained VT episodes. According to the current guidelines 3D mapping with radiofrequency (RF) ablation of the tachycardia substrate is an established therapeutic option. As part of the clinical follow-up of patients implanted with cardiac implantable electronic devices (CIED), including ICDs and CRT-Ds, function parameters are monitored every 6 months. The same check-up is done before and after any procedure involving RF ablation. Furthermore, a transient drop of R wave sensing has been demonstrated after the RF ablation of the atrioventricular (AV) node in patients with uncontrollable fast atrial fibrillation implanted with single-chamber pacemakers. The above mentioned have not been studied yet in patients ablated due to sustained VT. Hence, the aim of this study is to assess the change in R wave sensing after the RF ablation of VT substrate in patients with structural heart disease.

This study will be performed as a clinical, prospective, multi-centre, observational (analytic, longitudinal) cohort study with patient enrolment in 4 electrophysiology centres in Croatia and Switzerland and a structured follow-up period of 12 months.

All consecutive patients with sustained VT implanted with ICD or CRT-D undergoing RF ablation procedure of myocardial substrate, who are able to understand and sign informed consent, will be enrolled.

Exclusion criteria:

  • Patients with combined mechanical aortic and mitral valve prostheses
  • Patients with RF ablation of VT done in the last 12 months
  • Patients with known damaged ICD/CRT-D electrode or battery
  • Patients undergoing VT ablation who have not been implanted with ICD or CRT-D
  • Patients not willing or able to undergo clinical follow-up of ICD/CRT-D device
  • Patients with life expectancy < 1 year
  • Not able to understand or willing to sign informed consent
  • Age < 18 years Primary objective is a R wave sensing drop > 30% after VT substrate ablation procedure.

Specific (secondary) end-points:

  • incidence of VT recurrence after the ablation procedure;
  • risk factors for R wave sensing drop (> 30%) after the RF ablation of VT substrate;
  • time needed for R wave sensing recovery after the RF ablation procedure. Recruiting should not exceed 12 months with the minimal follow-up period of 12 months (24 months in total).

The risk categorisation of the study was done and the study was awarded a risk category A.

Sample size was chosen according to annual procedure volumes in Croatia and Switzerland. In addition, it is confirmed by power test (number needed to recruit: 52). Recruiting should not exceed 12 months. Categorical variables will be presented by contingency tables in absolute values and percentages. Categorical variables will be compared by the chi-square with Yates corrections or Fisher's exact test. Continuous data will be expressed as means and standard deviations or median with corresponding interquartile range. The differences in quantitative variables will be tested by Mann-Whitney U-test (non-parametric test for independent samples). Logistic regression will be used to analyse the association between all variables of interest and the R wave sensing drop. Multivariate logistic regression will be used to adjust for confounding factors. Predictors of the R wave sensing drop will be assessed in a univariate and multivariate Cox proportional hazards model. The statistical analysis will be done using SPSS Software Version 22.0 or newer (IBM SPSS Statistics, New York, USA).

This unique study offers the possibility to show the impact of RF ablation on short-term and long-term R wave sensing change assessed by ICD or CRT-D's ventricle electrode in patients with sustained VT and structural heart disease undergoing ablation procedure. This observational data is needed to further refine the treatment of these patients and to prevent possible ICD/CRT-D dysfunction which could endanger this patient population. Despite a high number of cohort registries for cardiovascular diseases (e.g. Danish heart registry, ORBIT-AF, GARFIELD AF registry, Swiss AF registry), there are scarce data on patients with structural VT and no data regarding R wave sensing changes influenced by RF energy application in ventricles. In addition, this could influence obligatory rhythm-monitoring duration and consequently hospitalization duration of patients with VT undergoing RF ablation procedures. Apart from the above mentioned, in those patients with non-optimal preprocedural R wave sensing, this could mean contraindication for procedure or having the need to do postprocedural defibrillation testing. It is likely, that data from our study will lead to future changes in periprocedural planning and modalities.

This study will be conducted in compliance with the current version of the Declaration of Helsinki, the ICH-GCP or ISO EN 14155 (as far as applicable) as well as all national legal and regulatory requirements. According to ethical regulations, the study plan with the proposed participant information and consent form as well as other project-specific documents will be submitted to Ethic committee of the participating centres.

Tipo di studio

Osservativo

Iscrizione (Anticipato)

100

Contatti e Sedi

Questa sezione fornisce i recapiti di coloro che conducono lo studio e informazioni su dove viene condotto lo studio.

Luoghi di studio

      • Rijeka, Croazia, 51000
        • University Hospital Rijeka
      • Zagreb, Croazia, 10000
        • University Hospital Sestre Milosrdnice
      • Basel, Svizzera, 4031
        • University Hospital Basel
      • Bern, Svizzera, 3012
        • University Hospital Bern

Criteri di partecipazione

I ricercatori cercano persone che corrispondano a una certa descrizione, chiamata criteri di ammissibilità. Alcuni esempi di questi criteri sono le condizioni generali di salute di una persona o trattamenti precedenti.

Criteri di ammissibilità

Età idonea allo studio

18 anni e precedenti (Adulto, Adulto più anziano)

Accetta volontari sani

No

Sessi ammissibili allo studio

Tutto

Metodo di campionamento

Campione non probabilistico

Popolazione di studio

Patients with structural heart disease (ischemic and non-ischemic aetiology) implanted with implantable cardioverter defibrillator (ICD) or cardiac resynchronization therapy cardioverter defibrillator (CRT-D) and recurrent, sustained, monomorphic ventricular tachycardia, who are, according to their cardiologist, planned for 3D mapping and radiofrequency ablation of ventricular tachycardia myocardial substrate.

Descrizione

Inclusion Criteria:

  • patients with recurrent, sustained ventricular tachycardia
  • patients implanted with implantable cardioverter defibrillator (ICD) or cardiac resynchronization therapy cardioverter defibrillator (CRT-D)
  • patients planned for 3D mapping and radiofrequency ablation of ventricular tachycardia myocardial substrate
  • patients who are able to understand and sign informed consent

Exclusion Criteria:

  • Patients with combined mechanical aortic and mitral valve prostheses
  • Patients with RF ablation of VT done in the last 12 months
  • Patients with known damaged ICD/CRT-D electrode or battery
  • Patients undergoing VT ablation who have not been implanted with ICD or CRT-D
  • Patients not willing or able to undergo clinical follow-up of ICD/CRT-D device
  • Patients with life expectancy < 1 year
  • Not able to understand or willing to sign informed consent
  • Age < 18 years

Piano di studio

Questa sezione fornisce i dettagli del piano di studio, compreso il modo in cui lo studio è progettato e ciò che lo studio sta misurando.

Come è strutturato lo studio?

Dettagli di progettazione

Cosa sta misurando lo studio?

Misure di risultato primarie

Misura del risultato
Misura Descrizione
Lasso di tempo
R wave sensing change
Lasso di tempo: immediately after the ablation procedure
R wave sensing drop > 30% after RF ablation of VT substrate in left and/or right ventricle assessed by ventricular electrode of the CIED
immediately after the ablation procedure

Misure di risultato secondarie

Misura del risultato
Misura Descrizione
Lasso di tempo
R wave sensing change 24h follow-up
Lasso di tempo: 24 hours after ablation procedure
R wave sensing drop > 30% after RF ablation of VT substrate in left and/or right ventricle assessed by ventricular electrode of the CIED
24 hours after ablation procedure
R wave sensing change 3-6 month follow-up
Lasso di tempo: 3-6 months after ablation procedure
R wave sensing drop > 30% after RF ablation of VT substrate in left and/or right ventricle assessed by ventricular electrode of the CIED
3-6 months after ablation procedure
R wave sensing change 9-12 month follow-up
Lasso di tempo: 9-12 months after the ablation procedure
R wave sensing drop > 30% after RF ablation of VT substrate in left and/or right ventricle assessed by ventricular electrode of the CIED
9-12 months after the ablation procedure
Incidence of VT recurrence after the ablation procedure
Lasso di tempo: 12 months after the ablation procedure
Assessing the VT recurrence after the ablation procedure by ICD or CRT-D interrogation adn function testing
12 months after the ablation procedure
Risk factors for R wave sensing change after the RF ablation of VT substrate
Lasso di tempo: 12 months
assessing risk factors for R wave sensing drop after the RF ablation of VT substrate
12 months

Collaboratori e investigatori

Qui è dove troverai le persone e le organizzazioni coinvolte in questo studio.

Investigatori

  • Investigatore principale: Ivan Zeljkovic, MD, University Hospital "Sestre Milosrdnice"

Pubblicazioni e link utili

La persona responsabile dell'inserimento delle informazioni sullo studio fornisce volontariamente queste pubblicazioni. Questi possono riguardare qualsiasi cosa relativa allo studio.

Studiare le date dei record

Queste date tengono traccia dell'avanzamento della registrazione dello studio e dell'invio dei risultati di sintesi a ClinicalTrials.gov. I record degli studi e i risultati riportati vengono esaminati dalla National Library of Medicine (NLM) per assicurarsi che soddisfino specifici standard di controllo della qualità prima di essere pubblicati sul sito Web pubblico.

Studia le date principali

Inizio studio (Anticipato)

1 dicembre 2018

Completamento primario (Anticipato)

1 dicembre 2019

Completamento dello studio (Anticipato)

1 dicembre 2020

Date di iscrizione allo studio

Primo inviato

17 dicembre 2018

Primo inviato che soddisfa i criteri di controllo qualità

17 dicembre 2018

Primo Inserito (Effettivo)

19 dicembre 2018

Aggiornamenti dei record di studio

Ultimo aggiornamento pubblicato (Effettivo)

20 dicembre 2018

Ultimo aggiornamento inviato che soddisfa i criteri QC

18 dicembre 2018

Ultimo verificato

1 dicembre 2018

Maggiori informazioni

Termini relativi a questo studio

Altri numeri di identificazione dello studio

  • ZAPSA01
  • UTN U1111-1223-2331 (Altro identificatore: Universal Trial Number - World Health Organization)

Piano per i dati dei singoli partecipanti (IPD)

Hai intenzione di condividere i dati dei singoli partecipanti (IPD)?

NO

Informazioni su farmaci e dispositivi, documenti di studio

Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti

No

Studia un dispositivo regolamentato dalla FDA degli Stati Uniti

No

Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .

Sottoscrivi