Does Gender Play a Role in Bone-mineral Density Measurement Precision?
2015年10月1日 更新者:University of Wisconsin, Madison
Bone mineral density (BMD) measurement using dual-energy x-ray absorptiometry (DXA) is the current gold standard for osteoporosis diagnosis and therapy monitoring.
Like all quantitative tests, there is some variability in BMD results obtained when scanning a person more than once.
As such, it is current clinical practice, based on the recommendation of the International Society for Clinical Densitometry, that each technologist perform a precision assessment.
This approach consists of scanning 30 people twice; the data from which allow determination of what constitutes a real difference in BMD with 95% confidence.
A precision assessment typically evaluates a specific clinic's population, using the age range and genders seen at that clinic.
However men generally have larger, but often more arthritic, bones than women which may impact the precision results.
Therefore, it is possible that gender-specific precision values should be used in clinical practice, however this issue has never been investigated.
調査の概要
状態
完了
条件
研究の種類
観察的
入学 (実際)
180
連絡先と場所
このセクションには、調査を実施する担当者の連絡先の詳細と、この調査が実施されている場所に関する情報が記載されています。
研究場所
-
-
Wisconsin
-
Madison、Wisconsin、アメリカ、53705
- University of Wisconsin Osteoporosis Clinical Center and Research Program
-
-
参加基準
研究者は、適格基準と呼ばれる特定の説明に適合する人を探します。これらの基準のいくつかの例は、人の一般的な健康状態または以前の治療です。
適格基準
就学可能な年齢
65年歳以上 (高齢者)
健康ボランティアの受け入れ
はい
受講資格のある性別
全て
サンプリング方法
非確率サンプル
調査対象母集団
The study population will consist of 90 men and 90 women volunteers who are aged 65 years and older, able and willing to provide informed consent.
説明
Inclusion Criteria:
- Age ≥ 65 years
- Able and willing to sign informed consent
Exclusion Criteria:
- Inability to have DXA scans performed due to weight ≥ 450 pounds (exceeds densitometer table limit)
- Metallic hardware in, or overlaying, any of the measured skeletal sites
研究計画
このセクションでは、研究がどのように設計され、研究が何を測定しているかなど、研究計画の詳細を提供します。
研究はどのように設計されていますか?
デザインの詳細
コホートと介入
グループ/コホート |
|---|
|
男性
|
|
女性
|
この研究は何を測定していますか?
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
DXA Bone Mineral Density precision
時間枠:1 day
|
The primary analysis will compare DXA BMD precision (mean square error/variance) in males and females, pooled across technicians, using the two-sample F-test for equality of variances.
Separate analyses will also be performed for each technician.
In addition, due to the sensitivity of the F-test to the normality assumption, a secondary analysis will perform the same comparisons using Levene's test.
A nominal two-sided p-value of 0.05 will be regarded as statistically significant.
|
1 day
|
協力者と研究者
ここでは、この調査に関係する人々や組織を見つけることができます。
研究記録日
これらの日付は、ClinicalTrials.gov への研究記録と要約結果の提出の進捗状況を追跡します。研究記録と報告された結果は、国立医学図書館 (NLM) によって審査され、公開 Web サイトに掲載される前に、特定の品質管理基準を満たしていることが確認されます。
主要日程の研究
研究開始
2010年7月1日
一次修了 (実際)
2011年1月1日
研究の完了 (実際)
2011年1月1日
試験登録日
最初に提出
2010年9月30日
QC基準を満たした最初の提出物
2011年3月25日
最初の投稿 (見積もり)
2011年3月29日
学習記録の更新
投稿された最後の更新 (見積もり)
2015年10月5日
QC基準を満たした最後の更新が送信されました
2015年10月1日
最終確認日
2011年3月1日
詳しくは
この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。