B40 Non-Invasive Blood Pressure Equivalency Study
調査の概要
詳細な説明
Automated non-invasive blood pressure devices have a long history. These devices first made their appearance in the mid-1970's. Since that time, there have been many improvements in the performance of these devices. Specifically, when making changes to the algorithms that control these devices, the goals are: to improve accuracy and precision in the blood pressure (BP) estimates, to reduce the time needed to determine the BP estimates, to increase comfort for the patient, and to work through artifacts and arrhythmias that are often present in the clinical environment.
The purpose of this study is to demonstrate equivalency of the GE Healthcare B40 Patient Monitor noninvasive blood pressure (NIBP) parameter compared to the reference device.
研究の種類
入学 (実際)
段階
- 適用できない
連絡先と場所
研究場所
-
-
Colorado
-
Louisville、Colorado、アメリカ、80027
- Clinimark
-
-
参加基準
適格基準
就学可能な年齢
健康ボランティアの受け入れ
受講資格のある性別
説明
Inclusion Criteria:
- Ability to provide written informed consent or have a legally authorized representative including a parent or guardian provide written informed consent
- Within the age range from older than 3 years old to 75 years old (adult and child)
- Ability to maintain proper placement of cuff and/or probe and/or sensor
Exclusion Criteria:
- Neonates (ages 0-29 days) and infants (30 days to 3 years old) shall be excluded
- Any subject deemed too unstable, by clinician's discretion, or for another reason deemed unacceptable for study by the clinician
- Known dysrhythmias (when applicable)
- Known disease state that compromises circulation to the extremity
- Vigorous exercise prior to participating in the study
- Excessive movement or excitability causing false values or no determinations
- Known allergy to latex when latex products will be in contact with subject
- Pregnant females
研究計画
研究はどのように設計されていますか?
デザインの詳細
- 主な目的:診断
- 割り当て:なし
- 介入モデル:単一グループの割り当て
- マスキング:なし(オープンラベル)
武器と介入
参加者グループ / アーム |
介入・治療 |
|---|---|
|
実験的:Blood Pressure Reading
Simultaneous blood pressure readings with DINAMAP PRO1000 and B40 monitor; total of 6 successful readings
|
Simultaneous blood pressure reading on the Investigational B40 patient monitor and DINAMAP PRO1000 patient monitor
|
この研究は何を測定していますか?
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Difference Between the B40 Monitor and Reference Device DINAMAP PRO1000 on NIBP Measurements
時間枠:End of each blood pressure reading
|
The primary endpoints are the difference in systolic, diastolic and mean blood pressure values between the B40 Monitor and PRO1000.
The mean of the difference should be no more than 5mmHg and the standard deviation of the difference should be no more than 8mmHg per the AAMI SP-10 standard.
|
End of each blood pressure reading
|
協力者と研究者
スポンサー
研究記録日
主要日程の研究
研究開始
一次修了 (実際)
研究の完了 (実際)
試験登録日
最初に提出
QC基準を満たした最初の提出物
最初の投稿 (見積もり)
学習記録の更新
投稿された最後の更新 (見積もり)
QC基準を満たした最後の更新が送信されました
最終確認日
詳しくは
本研究に関する用語
この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。
血圧の臨床試験
-
Hangzhou Agile Groups Network Technology Co., Ltd.完了