Evaluating Air Leak Detection in Intubated Patients
A Study to Evaluate Air Leak Detection Above the Endotracheal Cuff in Intubated Patients Using the AnapnoGuard 100 System
This trial is designed to assess the effectiveness of the AnapnoGaurd 100 system relative to standard of care (SOC) Control, in the course of mechanical ventilation and intubation. The main effectiveness aspect of this trial is the AnapnoGaurd 100's ability to reduce leaks between the endotracheal tube and the trachea walls as a result of the cuff pressure management. Since the primary outcome can't be measured by any method other than the investigational device, the study groups will be treated as follows:
- The Treatment group in this study will be treated with the fully functional AnapnoGaurd 100 device (per its intended use),
- The Control group will be treated with the AnapnoGuard device, where the cuff control will be disabled (off) and the cuff management will be done according to SOC.
Effectiveness will be measured by the duration and level of carbon dioxide (CO2) leakage around the endotracheal tube (ETT) cuff, from the lungs to the subglottic space and safety will be measured by adverse events (AEs).
調査の概要
研究の種類
入学 (実際)
段階
- 適用できない
連絡先と場所
研究場所
-
-
-
Bnei Brak、イスラエル
- Mayanei Hayeshua Medical Center, General ICU
-
Haifa、イスラエル
- Rambam Medical Center, Cardiac Surgery Department
-
Haifa、イスラエル
- Rambam Medical Center, Neurosurgery Department
-
Holon、イスラエル
- Wolfson Medical Center, General ICU
-
-
参加基準
適格基準
就学可能な年齢
健康ボランティアの受け入れ
受講資格のある性別
説明
Inclusion Criteria:
- Age above 18 (men and women);
- Subject is expected to receive mechanical ventilation for more than 12 hours;
- Connection of the ETT to the AnapnoGuard system within 12 hours from intubation initiation;
- Subject or subject's legally acceptable representative signed the Informed Consent Form
Exclusion Criteria:
1. Subjects with facial, oropharyngeal or neck trauma 2. BMI>40 3. Pregnant women 4. Patient ventilated in prone position 5. Difficult intubation (defined as more than 3 intubation attempts)
-
研究計画
研究はどのように設計されていますか?
デザインの詳細
- 主な目的:防止
- 割り当て:ランダム化
- 介入モデル:並列代入
- マスキング:独身
武器と介入
参加者グループ / アーム |
介入・治療 |
|---|---|
|
実験的:Study group
Subjects in the study group will be connected to the AnapnoGuard 100 control unit operating in the normal clinical mode (automatic CO2 leak measurement above the cuff, cuff pressure control, evacuation of secretions and tracheal rinsing)
|
|
|
アクティブコンパレータ:Control group
Subjects in the control group will be connected to the AnapnoGuard 100 control unit.
In the control group, the cuff pressure control of the AnapnoGuard 100 control unit will be disabled (OFF).
CO2 level above the cuff will be recorded.
Suction and rinsing function will operate after the system detects no CO2.
|
この研究は何を測定していますか?
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
CO2 Leakage Above the ETT Cuff, Measured Over Time.
時間枠:Subjects will be followed for the duration of endotracheal intubation, an expected average of 4 days
|
Trial outcome measure represents patients' exposure to CO2 leakage standardized per hour. Specifically, each patients' area under the curve (AUC) of CO2 leakage (mmHg) is computed over the whole trial by multiplying leakage duration (X-axis; continuous time) by CO2 level (Y-axis; mmHg), divided by the patient's number of hours in the trial. Outcome measure is thus [CO2 mmHg*hour]/hour. CO2 leakage is the concentration of CO2 above the cuff as measured by the AG100 system every few minutes (in clinical mode). Time points which created the curve were every two adjacent time intervals having a valid CO2 reading. |
Subjects will be followed for the duration of endotracheal intubation, an expected average of 4 days
|
二次結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Number of the Cuff Pressure Measurements Within the Safety Accepted Range (24 and 40cmH2O)
時間枠:Subjects will be followed for the duration of endotracheal intubation, an expected average of 4 days
|
mean number of cuff pressure measurements within the safety accepted range of 24 and 40 cmH2O, normalized using the total number of valid cuff pressure measurements.
|
Subjects will be followed for the duration of endotracheal intubation, an expected average of 4 days
|
協力者と研究者
スポンサー
捜査官
- スタディディレクター:Shai Efrati, MD、Hospitech Respiration
- 主任研究者:Gil Bolotin, MD
研究記録日
主要日程の研究
研究開始
一次修了 (実際)
研究の完了 (実際)
試験登録日
最初に提出
QC基準を満たした最初の提出物
最初の投稿 (見積もり)
学習記録の更新
投稿された最後の更新 (実際)
QC基準を満たした最後の更新が送信されました
最終確認日
詳しくは
この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。