Evaluating Air Leak Detection in Intubated Patients
A Study to Evaluate Air Leak Detection Above the Endotracheal Cuff in Intubated Patients Using the AnapnoGuard 100 System
This trial is designed to assess the effectiveness of the AnapnoGaurd 100 system relative to standard of care (SOC) Control, in the course of mechanical ventilation and intubation. The main effectiveness aspect of this trial is the AnapnoGaurd 100's ability to reduce leaks between the endotracheal tube and the trachea walls as a result of the cuff pressure management. Since the primary outcome can't be measured by any method other than the investigational device, the study groups will be treated as follows:
- The Treatment group in this study will be treated with the fully functional AnapnoGaurd 100 device (per its intended use),
- The Control group will be treated with the AnapnoGuard device, where the cuff control will be disabled (off) and the cuff management will be done according to SOC.
Effectiveness will be measured by the duration and level of carbon dioxide (CO2) leakage around the endotracheal tube (ETT) cuff, from the lungs to the subglottic space and safety will be measured by adverse events (AEs).
Přehled studie
Postavení
Podmínky
Intervence / Léčba
Typ studie
Zápis (Aktuální)
Fáze
- Nelze použít
Kontakty a umístění
Studijní místa
-
-
-
Bnei Brak, Izrael
- Mayanei Hayeshua Medical Center, General ICU
-
Haifa, Izrael
- Rambam Medical Center, Cardiac Surgery Department
-
Haifa, Izrael
- Rambam Medical Center, Neurosurgery Department
-
Holon, Izrael
- Wolfson Medical Center, General ICU
-
-
Kritéria účasti
Kritéria způsobilosti
Věk způsobilý ke studiu
Přijímá zdravé dobrovolníky
Pohlaví způsobilá ke studiu
Popis
Inclusion Criteria:
- Age above 18 (men and women);
- Subject is expected to receive mechanical ventilation for more than 12 hours;
- Connection of the ETT to the AnapnoGuard system within 12 hours from intubation initiation;
- Subject or subject's legally acceptable representative signed the Informed Consent Form
Exclusion Criteria:
1. Subjects with facial, oropharyngeal or neck trauma 2. BMI>40 3. Pregnant women 4. Patient ventilated in prone position 5. Difficult intubation (defined as more than 3 intubation attempts)
-
Studijní plán
Jak je studie koncipována?
Detaily designu
- Primární účel: Prevence
- Přidělení: Randomizované
- Intervenční model: Paralelní přiřazení
- Maskování: Singl
Zbraně a zásahy
Skupina účastníků / Arm |
Intervence / Léčba |
|---|---|
|
Experimentální: Study group
Subjects in the study group will be connected to the AnapnoGuard 100 control unit operating in the normal clinical mode (automatic CO2 leak measurement above the cuff, cuff pressure control, evacuation of secretions and tracheal rinsing)
|
|
|
Aktivní komparátor: Control group
Subjects in the control group will be connected to the AnapnoGuard 100 control unit.
In the control group, the cuff pressure control of the AnapnoGuard 100 control unit will be disabled (OFF).
CO2 level above the cuff will be recorded.
Suction and rinsing function will operate after the system detects no CO2.
|
Co je měření studie?
Primární výstupní opatření
Měření výsledku |
Popis opatření |
Časové okno |
|---|---|---|
|
CO2 Leakage Above the ETT Cuff, Measured Over Time.
Časové okno: Subjects will be followed for the duration of endotracheal intubation, an expected average of 4 days
|
Trial outcome measure represents patients' exposure to CO2 leakage standardized per hour. Specifically, each patients' area under the curve (AUC) of CO2 leakage (mmHg) is computed over the whole trial by multiplying leakage duration (X-axis; continuous time) by CO2 level (Y-axis; mmHg), divided by the patient's number of hours in the trial. Outcome measure is thus [CO2 mmHg*hour]/hour. CO2 leakage is the concentration of CO2 above the cuff as measured by the AG100 system every few minutes (in clinical mode). Time points which created the curve were every two adjacent time intervals having a valid CO2 reading. |
Subjects will be followed for the duration of endotracheal intubation, an expected average of 4 days
|
Sekundární výstupní opatření
Měření výsledku |
Popis opatření |
Časové okno |
|---|---|---|
|
Number of the Cuff Pressure Measurements Within the Safety Accepted Range (24 and 40cmH2O)
Časové okno: Subjects will be followed for the duration of endotracheal intubation, an expected average of 4 days
|
mean number of cuff pressure measurements within the safety accepted range of 24 and 40 cmH2O, normalized using the total number of valid cuff pressure measurements.
|
Subjects will be followed for the duration of endotracheal intubation, an expected average of 4 days
|
Spolupracovníci a vyšetřovatelé
Sponzor
Vyšetřovatelé
- Ředitel studie: Shai Efrati, MD, Hospitech Respiration
- Vrchní vyšetřovatel: Gil Bolotin, MD
Termíny studijních záznamů
Hlavní termíny studia
Začátek studia
Primární dokončení (Aktuální)
Dokončení studie (Aktuální)
Termíny zápisu do studia
První předloženo
První předloženo, které splnilo kritéria kontroly kvality
První zveřejněno (Odhad)
Aktualizace studijních záznamů
Poslední zveřejněná aktualizace (Aktuální)
Odeslaná poslední aktualizace, která splnila kritéria kontroly kvality
Naposledy ověřeno
Více informací
Termíny související s touto studií
Klíčová slova
Další identifikační čísla studie
- HST-AG-07
Tyto informace byly beze změn načteny přímo z webu clinicaltrials.gov. Máte-li jakékoli požadavky na změnu, odstranění nebo aktualizaci podrobností studie, kontaktujte prosím register@clinicaltrials.gov. Jakmile bude změna implementována na clinicaltrials.gov, bude automaticky aktualizována i na našem webu .