User Performance Evaluation of Contour® Next One, Accu-Chek® Aviva Connect, FreeStyle Freedom Lite, OneTouch® Verio and GlucoMen® Areo Blood Glucose Monitoring Systems Following ISO 15197:2013
The user performance evaluation shows whether people with diabetes are able to obtain accurate measurement results with a blood glucose monitoring system. In this study, user performance evaluation will be performed for Contour® Next One (Ascensia Diabetes Care GmbH), Accu-Chek® Aviva Connect (Roche Diabetes Care GmbH), FreeStyle Freedom Lite (Abbott Diabetes Care Inc.), OneTouch® Verio (LifeScan Europe) and GlucoMen® areo A. Menarini Diagnostics S.r.l. (based on ISO 15197:2013; EN ISO 15197:2015, clause 8.
For each BGMS, measurement procedures for user performance evaluation will be performed with 1 test meter and 1 reagent system lot by the study subjects.
The same meter and an additional test meter will be used for double measurements performed by study personnel (with the same reagent system lot used by subjects).
調査の概要
状態
条件
介入・治療
- デバイス:Blood glucose monitoring system for self-testing Contour® Next One
- デバイス:Blood glucose monitoring system for self-testing Accu-Chek® Aviva Connect
- デバイス:Blood glucose monitoring system for self-testing FreeStyle Freedom Lite
- デバイス:Blood glucose monitoring system for self-testing OneTouch® Verio
- デバイス:Blood glucose monitoring system for self-testing GlucoMen® areo
研究の種類
入学 (実際)
段階
- 適用できない
連絡先と場所
研究場所
-
-
Baden-Württemberg
-
Ulm、Baden-Württemberg、ドイツ、89081
- Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm
-
-
参加基準
適格基準
就学可能な年齢
健康ボランティアの受け入れ
受講資格のある性別
説明
Inclusion Criteria:
- Male or female, with type 1 diabetes, type 2 diabetes or subjects without diabetes
- Signed informed consent form
- Minimum age of 18 years
- Subjects are legally competent and capable to understand character, meaning and consequences of the study.
- If blood glucose values < 80 mg/dl or > 300 mg/dl shall be measured after short term alteration in insulin therapy:
- Male or female with type 1 diabetes and intensified insulin therapy or insulin pump therapy.
- Signature of subjects to document consent with these procedures on informed consent form.
Exclusion Criteria:
- Pregnancy or lactation period
- Severe acute disease (at the study physician's discretion)
- Severe chronic disease with potential risk during the test procedures (at the study physician's discretion)
- Current constitution that compromises the subject's capability to participate in the study (at the study physician's discretion)
- Being unable to give informed consent
- < 18 years
- Legally incompetent
- Being committed to an institution (e.g. psychiatric clinic)
- Language barriers potentially compromising an adequate compliance with study procedures
- Dependent on investigator or sponsor
- If blood glucose values < 80 mg/dl shall be measured after short term alteration in insulin therapy, subjects with type 1 diabetes, suffering from:
- Coronary heart disease
- Condition after myocardial infarction
- Condition after cerebral events
- Peripheral arterial occlusive disease
- Hypoglycemia unawareness
研究計画
研究はどのように設計されていますか?
武器と介入
参加者グループ / アーム |
介入・治療 |
|---|---|
|
実験的:Blood glucose measurement
Every study subject shall test three out of the five devices.
The testing order of the BGMS will be changed on each subject to minimize any order effects on measurement results.
|
This is a user performance study, that is intended to show whether people with diabetes are able to obtain accurate measurement results with a blood glucose monitoring system (BGMS). The subjects' measurement technique (e.g. applying blood onto the system reagent, reading the result) will be observed and documented by study personnel. Immediately after the measurement by the subject, the study personnel will perform a measurement with the same test meter. Measurement results obtained by the subjects with the BGMS will be compared to measurement results obtained with the comparison method. Sample collection for the comparison measurement will be performed by study personnel within 5 min after the subject's measurement with the test meter. The samples will be collected from the subjects' skin puncture (if possible), an additional skin puncture (if required) will be documented. An additional skin puncture will be performed by study personnel with single-use lancing devices.
他の名前:
|
この研究は何を測定していますか?
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Acceptance criteria defined by ISO 15197:2013 (E) will be applied
時間枠:For each subject, the experimental phase has an expected duration of up to 3 hours
|
95 % of the individual glucose measured values shall fall within ± 15 mg/dl (0.83 mmol/l) of the measured values of the comparison measurement procedure at glucose concentrations < 100 mg/dl (5.55 mmol/l) and within ± 15 % at glucose concentrations ≥ 100 mg/dl (5.55 mmol/l).
|
For each subject, the experimental phase has an expected duration of up to 3 hours
|
協力者と研究者
出版物と役立つリンク
研究記録日
主要日程の研究
研究開始
一次修了 (実際)
研究の完了 (実際)
試験登録日
最初に提出
QC基準を満たした最初の提出物
最初の投稿 (見積もり)
学習記録の更新
投稿された最後の更新 (実際)
QC基準を満たした最後の更新が送信されました
最終確認日
詳しくは
本研究に関する用語
その他の研究ID番号
- IDT-1608-AL
この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。