Validation Study in Patients of a Non-invasive Blood Pressure Monitor for Perioperative Use
Collecting all available data (waveforms, beat to beat data, status data) generated by a non invasive blood pressure monitor on each hand and compare this to the actual data obtained by intra arterial (radial) monitoring.
This to see if the non invasive bloodpressure monitor can be validated for intraoperative use.
調査の概要
詳細な説明
Although a blood pressure monitoring device could be evaluated by the comparing systolic, diastolic and mean arterial pressure to a reference pressure, it is the specific aim to collect all available data (waveforms, beat to beat data, status data) generated by the blood pressure monitor.
After data analysis and comparison to patient's actual intra-arterial waveform, the outcome will be evaluated by Association for the Advancement of Medical Instrumentation (AAMI) standards to see if the non invasive bloodpressure monitor can be validated for intraoperative use.
The study is targeted at patients already monitored with a system that accurately measures and stores the intra-arterial waveform, such as the FloTrac system in the EV1000 monitor.
This study is a substudy of a feasability protocol evaluating a new algorithm for the Clearsight system.
研究の種類
入学 (予想される)
連絡先と場所
研究場所
-
-
-
Amsterdam、オランダ、1105AZ
- 募集
- Academic Medical Center
-
コンタクト:
- D.P. Veelo, Dr.
- 電話番号:0031205662533
- メール:d.p.veelo@amc.uva.nl
-
-
参加基準
適格基準
就学可能な年齢
健康ボランティアの受け入れ
受講資格のある性別
サンプリング方法
調査対象母集団
説明
Inclusion Criteria:
- Patients whose arterial blood pressure is measured invasively with technology that allows for the storage of the intra-arterial waveform with sufficient quality and resolution (Flotrac).
- Patients older than 18, for elective general surgery.
Exclusion Criteria:
- Patient conditions that are expected to potentially and significantly affect the transfer of pressure between radial artery and the finger, e.g. due to recent finger fractures
研究計画
研究はどのように設計されていますか?
デザインの詳細
コホートと介入
グループ/コホート |
介入・治療 |
|---|---|
|
Cohort
Adult patients, undergoing surgery requiring general anesthesia, planned to be monitored with a system that accurately measures and stores the intra-arterial waveform.
|
A cuff is placed on a finger of the patient.
The ClearSight system uses a volume clamp technology allowing measuring of continuous blood pressure waveform.
Note: There will only be observational recordings, no interventions
|
この研究は何を測定していますか?
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
validation systole, diastole, mean ABP and pulse pressure
時間枠:within 1 hour
|
accuracy and precision of invasive blood pressure curve vs non-invasive blood pressure curves(ipsilateral and contralateral)
|
within 1 hour
|
|
implementation of software alterations to improve systole, diastole, mean ABP accuracy and pulse pressure
時間枠:time during surgery
|
Comparison of the non-invasive blood pressure curve after software alterations to the reference and invasive arterial blood pressure curve
|
time during surgery
|
協力者と研究者
捜査官
- 主任研究者:D.P. Veelo, Dr、inverstigator
出版物と役立つリンク
研究記録日
主要日程の研究
研究開始 (実際)
一次修了 (予想される)
研究の完了 (予想される)
試験登録日
最初に提出
QC基準を満たした最初の提出物
最初の投稿 (実際)
学習記録の更新
投稿された最後の更新 (実際)
QC基準を満たした最後の更新が送信されました
最終確認日
詳しくは
この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。