- ICH GCP
- Registro de ensayos clínicos de EE. UU.
- Ensayo clínico NCT03795831
Validation Study in Patients of a Non-invasive Blood Pressure Monitor for Perioperative Use
Collecting all available data (waveforms, beat to beat data, status data) generated by a non invasive blood pressure monitor on each hand and compare this to the actual data obtained by intra arterial (radial) monitoring.
This to see if the non invasive bloodpressure monitor can be validated for intraoperative use.
Descripción general del estudio
Estado
Condiciones
Intervención / Tratamiento
Descripción detallada
Although a blood pressure monitoring device could be evaluated by the comparing systolic, diastolic and mean arterial pressure to a reference pressure, it is the specific aim to collect all available data (waveforms, beat to beat data, status data) generated by the blood pressure monitor.
After data analysis and comparison to patient's actual intra-arterial waveform, the outcome will be evaluated by Association for the Advancement of Medical Instrumentation (AAMI) standards to see if the non invasive bloodpressure monitor can be validated for intraoperative use.
The study is targeted at patients already monitored with a system that accurately measures and stores the intra-arterial waveform, such as the FloTrac system in the EV1000 monitor.
This study is a substudy of a feasability protocol evaluating a new algorithm for the Clearsight system.
Tipo de estudio
Inscripción (Anticipado)
Contactos y Ubicaciones
Ubicaciones de estudio
-
-
-
Amsterdam, Países Bajos, 1105AZ
- Reclutamiento
- Academic Medical Center
-
Contacto:
- D.P. Veelo, Dr.
- Número de teléfono: 0031205662533
- Correo electrónico: d.p.veelo@amc.uva.nl
-
-
Criterios de participación
Criterio de elegibilidad
Edades elegibles para estudiar
Acepta Voluntarios Saludables
Géneros elegibles para el estudio
Método de muestreo
Población de estudio
Descripción
Inclusion Criteria:
- Patients whose arterial blood pressure is measured invasively with technology that allows for the storage of the intra-arterial waveform with sufficient quality and resolution (Flotrac).
- Patients older than 18, for elective general surgery.
Exclusion Criteria:
- Patient conditions that are expected to potentially and significantly affect the transfer of pressure between radial artery and the finger, e.g. due to recent finger fractures
Plan de estudios
¿Cómo está diseñado el estudio?
Detalles de diseño
Cohortes e Intervenciones
Grupo / Cohorte |
Intervención / Tratamiento |
---|---|
Cohort
Adult patients, undergoing surgery requiring general anesthesia, planned to be monitored with a system that accurately measures and stores the intra-arterial waveform.
|
A cuff is placed on a finger of the patient.
The ClearSight system uses a volume clamp technology allowing measuring of continuous blood pressure waveform.
Note: There will only be observational recordings, no interventions
|
¿Qué mide el estudio?
Medidas de resultado primarias
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
---|---|---|
validation systole, diastole, mean ABP and pulse pressure
Periodo de tiempo: within 1 hour
|
accuracy and precision of invasive blood pressure curve vs non-invasive blood pressure curves(ipsilateral and contralateral)
|
within 1 hour
|
implementation of software alterations to improve systole, diastole, mean ABP accuracy and pulse pressure
Periodo de tiempo: time during surgery
|
Comparison of the non-invasive blood pressure curve after software alterations to the reference and invasive arterial blood pressure curve
|
time during surgery
|
Colaboradores e Investigadores
Colaboradores
Investigadores
- Investigador principal: D.P. Veelo, Dr, inverstigator
Publicaciones y enlaces útiles
Fechas de registro del estudio
Fechas importantes del estudio
Inicio del estudio (Actual)
Finalización primaria (Anticipado)
Finalización del estudio (Anticipado)
Fechas de registro del estudio
Enviado por primera vez
Primero enviado que cumplió con los criterios de control de calidad
Publicado por primera vez (Actual)
Actualizaciones de registros de estudio
Última actualización publicada (Actual)
Última actualización enviada que cumplió con los criterios de control de calidad
Última verificación
Más información
Términos relacionados con este estudio
Términos MeSH relevantes adicionales
Otros números de identificación del estudio
- 2017-14
Plan de datos de participantes individuales (IPD)
¿Planea compartir datos de participantes individuales (IPD)?
Información sobre medicamentos y dispositivos, documentos del estudio
Estudia un producto farmacéutico regulado por la FDA de EE. UU.
Estudia un producto de dispositivo regulado por la FDA de EE. UU.
producto fabricado y exportado desde los EE. UU.
Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .
Ensayos clínicos sobre ClearSight feasibility testing
-
University of SaskatchewanReclutamiento
-
University of BaselReclutamientoEnfermedad de la válvula aórticaSuiza
-
Clear Cut Medical Ltd.TerminadoCáncer de mamaEstados Unidos, Israel
-
Seoul National University HospitalTerminadoComplicaciones de la cesárea | HipotensiónCorea, república de
-
University Hospital, LilleTerminadoPresión arterial | Anestesia | Hemodinámico, MonitoreoFrancia
-
Nemours Children's ClinicTerminado
-
National University Hospital, SingaporeReclutamientoTrastornos neurocognitivos | Fibrilación auricular | LRASingapur
-
Edwards LifesciencesDesconocido
-
Ludwig-Maximilians - University of MunichTerminadoEstabilidad de la presión arterial en anestesia generalAlemania
-
Fondazione Policlinico Universitario Agostino Gemelli...ReclutamientoTrastorno ortopédico | Inestabilidad hemodinámica | Presión Arterial, BajaItalia