糖尿病足評価のためのサーモグラフィカメラ検証 (HIK-DM-2026)
糖尿病足評価のためのサーモグラフィカメラ検証
調査の概要
詳細な説明
Each participant is assessed in a single session in a thermally stable room at 20 to 25 degrees Celsius, with relative humidity recorded as a covariate. After 15 minutes of acclimatization in a semi-Fowler position with the backrest at 30 to 45 degrees, both cameras acquire plantar thermograms from a static tripod at a fixed distance of 75 cm, with the tripod head at 90 degrees to the plantar plane and an emissivity setting of 0.98. The two radiometric acquisitions are consecutive and always less than 10 seconds apart. The order of the cameras is randomized 1:1 at participant level (sequence A: M60 then Mini2 V2; sequence B: Mini2 V2 then M60), from a list generated in advance by a team member not involved in image acquisition and kept in a password-protected file until recruitment is complete. No specific background is used.
Thermograms are exported to HIKMICRO Analyzer version 2.0.1. Six anatomical regions of interest, defined a priori and identical for both cameras, are delimited on each foot: hallux, first, third and fifth metatarsal heads, midfoot and heel. There is no post-hoc region selection and no indexing to either device. Contralateral thermal asymmetry (Delta T) is computed per region as the absolute difference in mean temperature between homologous regions of the right and left foot.
Images are renamed with an alphanumeric code that removes any explicit reference to the source device before analysis, and regions are delimited by an independent evaluator blinded to the neurological status of the participant. Technical blinding is acknowledged as partial, because the difference in spatial resolution between the two cameras may allow a trained evaluator to infer the source device.
The sample size follows the method of Walter, Eliasziw and Donner for the intraclass correlation coefficient, with a null hypothesis of ICC0 = 0.65 against an alternative of ICC1 = 0.80, alpha = 0.05, nominal power 0.80 and two observations per participant, which yields a minimum of 77 participants. The study enrolls 130 participants, which raises the actual power for this contrast to 95.7 percent and absorbs expected losses.
研究の種類
入学 (推定)
連絡先と場所
研究連絡先
- 名前:Eduardo Luis Martín Javier
- 電話番号:+34 647 773 675
- メール:eduardo@avanza-salud.com
研究場所
-
-
Santa Cruz de Tenerife
-
San Cristóbal de La Laguna、Santa Cruz de Tenerife、スペイン、38201
- Clínica Avanza Salud
-
コンタクト:
- Eduardo Luis Martín Javier
- 電話番号:+34 647 773 675
- メール:eduardo@avanza-salud.com
-
-
参加基準
適格基準
就学可能な年齢
- 大人
- 高齢者
健康ボランティアの受け入れ
サンプリング方法
調査対象母集団
説明
参加基準:
- 18歳以上。
- 1型または2型糖尿病と診断されていること。
- インフォームドコンセントを理解し、署名できること。
除外基準:
- 活動性の足潰瘍。
- 下肢の最近の手術。
- 研究プロトコルに従うことを妨げる認知障害。
研究計画
研究はどのように設計されていますか?
デザインの詳細
コホートと介入
グループ/コホート |
介入・治療 |
|---|---|
|
Diabetes patients
Adults with diagnosed type 1 or type 2 diabetes mellitus, without active plantar ulceration, recruited from the active caseload of general podiatry and physiotherapy consultations.
|
Low-cost infrared thermographic camera (256 x 192 pixels, NETD below 40 mK) used solely as a measuring instrument.
It is not a medical device and no clinical decision is derived from its readings.
High-resolution infrared thermographic camera (640 x 480 pixels, NETD below 35 mK) used as a technical comparator.
It is not designated as a clinical or diagnostic reference standard.
|
この研究は何を測定していますか?
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Intraclass correlation coefficient for absolute agreement, ICC(A,1), between the two cameras for contralateral thermal asymmetry (Delta T) in each of six pre-specified plantar regions of interest
時間枠:Single measurement session (Day 1)
|
Delta T is computed per region as the absolute difference in mean temperature between homologous regions of the right and left foot.
One ICC(A,1) is reported for each of the six regions defined a priori (hallux, first, third and fifth metatarsal heads, midfoot and heel), with its 95 percent confidence interval, using the McGraw and Wong (1996) formulation, equivalent to ICC(2,1) of Shrout and Fleiss (1979), two-way random effects, single measurement.
Each region contributes 130 paired measurements.
The study hypothesis is ICC greater than or equal to 0.80, tested against a null of ICC0 = 0.65.
Neither device is treated as a reference standard.
|
Single measurement session (Day 1)
|
二次結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Cohen's Kappa for inter-device agreement in the binary classification of thermal asymmetry at the 2.2 degree Celsius cut-off, per region of interest
時間枠:Single measurement session (Day 1)
|
Overall agreement proportion and Kappa with 95 percent confidence interval for the dichotomous classification Delta T greater than or equal to 2.2 degrees Celsius versus below it, in each of the six regions.
The cut-off is used as an operational category without clinical connotation, and neither device is designated as the diagnostic reference.
|
Single measurement session (Day 1)
|
|
Consistency intraclass correlation coefficient, ICC(C,1), between the two cameras for mean plantar temperature (Tmean)
時間枠:Single measurement session (Day 1)
|
McGraw and Wong (1996) consistency formulation, equivalent to ICC(3,1) of Shrout and Fleiss (1979), reported with its 95 percent confidence interval.
Exploratory endpoint with a hypothesis of ICC greater than or equal to 0.75.
It is not powered by an independent sample size calculation.
|
Single measurement session (Day 1)
|
|
Bland-Altman systematic bias and 95 percent limits of agreement between the two cameras for mean plantar temperature (Tmean) and for contralateral thermal asymmetry (Delta T)
時間枠:Single measurement session (Day 1)
|
Systematic bias and 95 percent limits of agreement computed across the six pre-specified plantar regions of interest, for both the mean plantar temperature and the contralateral thermal asymmetry.
Neither device is treated as a reference standard.
|
Single measurement session (Day 1)
|
|
Influence of participant clinical variables on the magnitude of inter-device discordance
時間枠:Single measurement session (Day 1)
|
The absolute inter-device difference is analyzed against the presence of peripheral sensory neuropathy, body mass index and type of diabetes mellitus.
|
Single measurement session (Day 1)
|
|
Image acquisition time
時間枠:Single measurement session (Day 1)
|
Time elapsed from the end of the acclimatization period to the correct export of both thermograms, measured with a stopwatch for each participant and reported in seconds.
|
Single measurement session (Day 1)
|
|
Proportion of technically valid thermograms
時間枠:Single measurement session (Day 1)
|
Percentage of thermographic acquisitions that did not require repetition because of a technical error, namely defocus, participant movement or a saving failure.
|
Single measurement session (Day 1)
|
|
Participant acceptability of the measurement protocol
時間枠:Single measurement session (Day 1)
|
Number of participants answering yes to a single dichotomous question asked verbally at the end of the session, on whether they would consider repeating this procedure at a future appointment.
|
Single measurement session (Day 1)
|
|
Descriptive distribution of the maximum contralateral thermal asymmetry per participant
時間枠:Single measurement session (Day 1)
|
Median and interquartile range reported separately for each camera.
This measure is descriptive only.
It is not the primary criterion and it is not indexed to either device.
|
Single measurement session (Day 1)
|
協力者と研究者
研究記録日
主要日程の研究
研究開始 (推定)
一次修了 (推定)
研究の完了 (推定)
試験登録日
最初に提出
QC基準を満たした最初の提出物
最初の投稿 (実際)
学習記録の更新
投稿された最後の更新 (実際)
QC基準を満たした最後の更新が送信されました
最終確認日
詳しくは
本研究に関する用語
キーワード
その他の研究ID番号
- HIK-DM-2026-01
- CHUC_2026_24 (その他の識別子:CEIm Provincial de Santa Cruz de Tenerife, Complejo Hospitalario Universitario de Canarias)
個々の参加者データ (IPD) の計画
個々の参加者データ (IPD) を共有する予定はありますか?
医薬品およびデバイス情報、研究文書
米国FDA規制医薬品の研究
米国FDA規制機器製品の研究
この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。