Evaluate the Performance of the Dexcom G7 Continuous Glucose Monitoring (CGM) System in Critically Ill Patients Undergoing Major Abdominal Surgery and Solid Organ Transplantation Which Require Blood Glucose Monitoring for Intensive Insulin Therapy. (DREAM-ICU)
Determining the Reliability and Accuracy of the Dexcom G7 Continuous Glucose Monitoring System in the ICU
Main objective:
The goal is to to determine the accuracy of Dexcom G7 Continuous glucose monitor system in critically ill patients, undergoing major abdominal surgery and solid organ transplantation, in the intensive care unit.
On admission to the intensive care unit or at the end of surgery, two Dexcom G7 sensors are placed in the infraclavicular position.
Sensor number 1 will be blinded and will only be active when paired with the receiver, without the ability to monitor glucose levels.
Sensor number 2 is used for continuous glucose monitoring and can be used to adjust the insulin dose.
Patients will wear the continuous glucose monitoring sensor for a maximum of 10 days, including the time spent in the intensive care unit and the immediate care unit. The study will not interfere with any usual care clinical workflows.
Calibration is only performed on an unblinded sensor. The reference method for calibration is the glucose from the blood glucose meter.
Calibrate on the first day of monitoring: 4 times a day after approximately 6 hours, or if the blood glucose reading from the blood glucose meter is not within the estimated error range of the continuous glucose monitor (difference greater than 1,5mmol/L).
Calibration on days 2 and 3 of monitoring: once in the morning, or when the meter's glucose reading is not within the estimated error range of the continuous glucose monitoring (difference greater than 1,5mmol/L) On subsequent days when the blood glucose meter reading is not within the estimated error range of continuous glucose monitoring (difference greater than 1,5mmol/L).
Hypothesis:
- Ho: No difference in mean absolute relative difference (MARD) between calibrated and non-calibrated sensors.
- Ha: Calibrated sensors have lower mean absolute relative difference (δ = 4%, SD = 5%).
Further objectives:
To evaluate further aspects regarding the Dexcom G7 continuous glucose monitor system's performance in the intensive care unit (ICU), particularly in terms of sensitivity and specificity to detect hypo- and hyperglycaemic events, data availability, device deficiencies and safety.
To identify predictive factors of the accuracy of the Dexcom G7 continuous glucose monitor system in critically ill patients in the ICU.
調査の概要
状態
詳細な説明
Design:
- Each intensive care unit patient wears two sensors (calibrated + non-calibrated).
- 4 glucose measurements/day × 6 days = 24 measurements/subject.
- Sequential Organ Failure Assessment (SOFA) score recorded daily.
研究の種類
入学 (推定)
段階
- 適用できない
連絡先と場所
研究連絡先
- 名前:Marek Protus, MD
- 電話番号:+420602278062
- メール:marek.protus@ikem.cz
研究連絡先のバックアップ
- 名前:Alexander Macek, MD
- 電話番号:+421910631876
- メール:alexander.macek@ikem.cz
研究場所
-
-
-
Prague、チェコ
- 募集
- Institute of Clinical and Experimental Medice
-
コンタクト:
- Marek Protus, MD
- 電話番号:+420602278062
- メール:marek.protus@ikem.cz
-
コンタクト:
- Alexander Macek, MD
- 電話番号:+421910631876
- メール:alexander.macek@ikem.cz
-
-
参加基準
適格基準
就学可能な年齢
- 大人
- 高齢者
健康ボランティアの受け入れ
説明
Inclusion Criteria:
- Age 18 years or over
- Intensive care unit admission
- Requiring blood glucose monitoring for insulin therapy
Exclusion Criteria:
- High-dose ascorbic acid or acetaminophen > 4g/day
- Physical or mental health condition precluding continuous wear of the continuous glucose monitor, as judged by the physician
研究計画
研究はどのように設計されていますか?
デザインの詳細
- 主な目的:デバイスの実現可能性
- 割り当て:非ランダム化
- 介入モデル:単一グループの割り当て
- マスキング:トリプル
武器と介入
参加者グループ / アーム |
介入・治療 |
|---|---|
|
アクティブコンパレータ:Open labeled CGM
Additional calibrations performed on CGM device during monitoring period
|
Continuous measurement of glucose from subcutaneous tissue
|
|
偽コンパレータ:Control labeled CGM
No additional calibrations performed on CGM device during monitoring period
|
Continuous measurement of glucose from subcutaneous tissue
|
この研究は何を測定していますか?
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Calculate overall mean absolute relative difference between arms
時間枠:From enrolment to discharge from ICU care or up-to day 10
|
Analyse the difference in mean absolute relative difference (MARD) of Dexcom G7 continuous glucose monitor between calibrated and uncalibrated sensor, in patients undergoing major abdominal surgery and solid organ transplantation.
|
From enrolment to discharge from ICU care or up-to day 10
|
二次結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Accuracy metrics
時間枠:From enrolment to discharge from ICU care or up-to day 10
|
To evaluate further aspects regarding the Dexcom G7 continuous glucose monitor system's performance in the intensive care unit (ICU), particularly in terms of sensitivity and specificity to detect hypo- and hyperglycaemic events, data availability, device deficiencies and safety. To identify predictive factors of the accuracy of the Dexcom G7 continuous glucose monitor system in critically ill patients in the ICU. |
From enrolment to discharge from ICU care or up-to day 10
|
協力者と研究者
研究記録日
主要日程の研究
研究開始 (実際)
一次修了 (推定)
研究の完了 (推定)
試験登録日
最初に提出
QC基準を満たした最初の提出物
最初の投稿 (実際)
学習記録の更新
投稿された最後の更新 (実際)
QC基準を満たした最後の更新が送信されました
最終確認日
詳しくは
この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。