- ICH GCP
- Yhdysvaltain kliinisten tutkimusten rekisteri
- Kliininen tutkimus NCT07611721
Evaluate the Performance of the Dexcom G7 Continuous Glucose Monitoring (CGM) System in Critically Ill Patients Undergoing Major Abdominal Surgery and Solid Organ Transplantation Which Require Blood Glucose Monitoring for Intensive Insulin Therapy. (DREAM-ICU)
Determining the Reliability and Accuracy of the Dexcom G7 Continuous Glucose Monitoring System in the ICU
Main objective:
The goal is to to determine the accuracy of Dexcom G7 Continuous glucose monitor system in critically ill patients, undergoing major abdominal surgery and solid organ transplantation, in the intensive care unit.
On admission to the intensive care unit or at the end of surgery, two Dexcom G7 sensors are placed in the infraclavicular position.
Sensor number 1 will be blinded and will only be active when paired with the receiver, without the ability to monitor glucose levels.
Sensor number 2 is used for continuous glucose monitoring and can be used to adjust the insulin dose.
Patients will wear the continuous glucose monitoring sensor for a maximum of 10 days, including the time spent in the intensive care unit and the immediate care unit. The study will not interfere with any usual care clinical workflows.
Calibration is only performed on an unblinded sensor. The reference method for calibration is the glucose from the blood glucose meter.
Calibrate on the first day of monitoring: 4 times a day after approximately 6 hours, or if the blood glucose reading from the blood glucose meter is not within the estimated error range of the continuous glucose monitor (difference greater than 1,5mmol/L).
Calibration on days 2 and 3 of monitoring: once in the morning, or when the meter's glucose reading is not within the estimated error range of the continuous glucose monitoring (difference greater than 1,5mmol/L) On subsequent days when the blood glucose meter reading is not within the estimated error range of continuous glucose monitoring (difference greater than 1,5mmol/L).
Hypothesis:
- Ho: No difference in mean absolute relative difference (MARD) between calibrated and non-calibrated sensors.
- Ha: Calibrated sensors have lower mean absolute relative difference (δ = 4%, SD = 5%).
Further objectives:
To evaluate further aspects regarding the Dexcom G7 continuous glucose monitor system's performance in the intensive care unit (ICU), particularly in terms of sensitivity and specificity to detect hypo- and hyperglycaemic events, data availability, device deficiencies and safety.
To identify predictive factors of the accuracy of the Dexcom G7 continuous glucose monitor system in critically ill patients in the ICU.
Tutkimuksen yleiskatsaus
Tila
Ehdot
Interventio / Hoito
Yksityiskohtainen kuvaus
Design:
- Each intensive care unit patient wears two sensors (calibrated + non-calibrated).
- 4 glucose measurements/day × 6 days = 24 measurements/subject.
- Sequential Organ Failure Assessment (SOFA) score recorded daily.
Opintotyyppi
Ilmoittautuminen (Arvioitu)
Vaihe
- Ei sovellettavissa
Yhteystiedot ja paikat
Opiskeluyhteys
- Nimi: Marek Protus, MD
- Puhelinnumero: +420602278062
- Sähköposti: marek.protus@ikem.cz
Tutki yhteystietojen varmuuskopiointi
- Nimi: Alexander Macek, MD
- Puhelinnumero: +421910631876
- Sähköposti: alexander.macek@ikem.cz
Opiskelupaikat
-
-
-
Prague, Tšekki
- Rekrytointi
- Institute of Clinical and Experimental Medice
-
Ottaa yhteyttä:
- Marek Protus, MD
- Puhelinnumero: +420602278062
- Sähköposti: marek.protus@ikem.cz
-
Ottaa yhteyttä:
- Alexander Macek, MD
- Puhelinnumero: +421910631876
- Sähköposti: alexander.macek@ikem.cz
-
-
Osallistumiskriteerit
Kelpoisuusvaatimukset
Opintokelpoiset iät
- Aikuinen
- Vanhempi Aikuinen
Hyväksyy terveitä vapaaehtoisia
Kuvaus
Inclusion Criteria:
- Age 18 years or over
- Intensive care unit admission
- Requiring blood glucose monitoring for insulin therapy
Exclusion Criteria:
- High-dose ascorbic acid or acetaminophen > 4g/day
- Physical or mental health condition precluding continuous wear of the continuous glucose monitor, as judged by the physician
Opintosuunnitelma
Miten tutkimus on suunniteltu?
Suunnittelun yksityiskohdat
- Ensisijainen käyttötarkoitus: Laitteen toteutettavuus
- Jako: Ei satunnaistettu
- Inventiomalli: Yksittäinen ryhmätehtävä
- Naamiointi: Kolminkertaistaa
Aseet ja interventiot
Osallistujaryhmä / Arm |
Interventio / Hoito |
|---|---|
|
Active Comparator: Open labeled CGM
Additional calibrations performed on CGM device during monitoring period
|
Continuous measurement of glucose from subcutaneous tissue
|
|
Huijausvertailija: Control labeled CGM
No additional calibrations performed on CGM device during monitoring period
|
Continuous measurement of glucose from subcutaneous tissue
|
Mitä tutkimuksessa mitataan?
Ensisijaiset tulostoimenpiteet
Tulosmittaus |
Toimenpiteen kuvaus |
Aikaikkuna |
|---|---|---|
|
Calculate overall mean absolute relative difference between arms
Aikaikkuna: From enrolment to discharge from ICU care or up-to day 10
|
Analyse the difference in mean absolute relative difference (MARD) of Dexcom G7 continuous glucose monitor between calibrated and uncalibrated sensor, in patients undergoing major abdominal surgery and solid organ transplantation.
|
From enrolment to discharge from ICU care or up-to day 10
|
Toissijaiset tulostoimenpiteet
Tulosmittaus |
Toimenpiteen kuvaus |
Aikaikkuna |
|---|---|---|
|
Accuracy metrics
Aikaikkuna: From enrolment to discharge from ICU care or up-to day 10
|
To evaluate further aspects regarding the Dexcom G7 continuous glucose monitor system's performance in the intensive care unit (ICU), particularly in terms of sensitivity and specificity to detect hypo- and hyperglycaemic events, data availability, device deficiencies and safety. To identify predictive factors of the accuracy of the Dexcom G7 continuous glucose monitor system in critically ill patients in the ICU. |
From enrolment to discharge from ICU care or up-to day 10
|
Yhteistyökumppanit ja tutkijat
Opintojen ennätyspäivät
Opi tärkeimmät päivämäärät
Opiskelun aloitus (Todellinen)
Ensisijainen valmistuminen (Arvioitu)
Opintojen valmistuminen (Arvioitu)
Opintoihin ilmoittautumispäivät
Ensimmäinen lähetetty
Ensimmäinen toimitettu, joka täytti QC-kriteerit
Ensimmäinen Lähetetty (Todellinen)
Tutkimustietojen päivitykset
Viimeisin päivitys julkaistu (Todellinen)
Viimeisin lähetetty päivitys, joka täytti QC-kriteerit
Viimeksi vahvistettu
Lisää tietoa
Tähän tutkimukseen liittyvät termit
Muita asiaankuuluvia MeSH-ehtoja
Muut tutkimustunnusnumerot
- G-25-71
- ID Project No. LX22NPO5104 (Muu apuraha/rahoitusnumero: MH CZ-DRO (Institute for Clinical and Experimental Medicine-IKEM, IN 00023001)
Lääke- ja laitetiedot, tutkimusasiakirjat
Tutkii yhdysvaltalaista FDA sääntelemää lääkevalmistetta
Tutkii yhdysvaltalaista FDA sääntelemää laitetuotetta
Nämä tiedot haettiin suoraan verkkosivustolta clinicaltrials.gov ilman muutoksia. Jos sinulla on pyyntöjä muuttaa, poistaa tai päivittää tutkimustietojasi, ota yhteyttä register@clinicaltrials.gov. Heti kun muutos on otettu käyttöön osoitteessa clinicaltrials.gov, se päivitetään automaattisesti myös verkkosivustollemme .