Evaluate the Performance of the Dexcom G7 Continuous Glucose Monitoring (CGM) System in Critically Ill Patients Undergoing Major Abdominal Surgery and Solid Organ Transplantation Which Require Blood Glucose Monitoring for Intensive Insulin Therapy. (DREAM-ICU)
Determining the Reliability and Accuracy of the Dexcom G7 Continuous Glucose Monitoring System in the ICU
Main objective:
The goal is to to determine the accuracy of Dexcom G7 Continuous glucose monitor system in critically ill patients, undergoing major abdominal surgery and solid organ transplantation, in the intensive care unit.
On admission to the intensive care unit or at the end of surgery, two Dexcom G7 sensors are placed in the infraclavicular position.
Sensor number 1 will be blinded and will only be active when paired with the receiver, without the ability to monitor glucose levels.
Sensor number 2 is used for continuous glucose monitoring and can be used to adjust the insulin dose.
Patients will wear the continuous glucose monitoring sensor for a maximum of 10 days, including the time spent in the intensive care unit and the immediate care unit. The study will not interfere with any usual care clinical workflows.
Calibration is only performed on an unblinded sensor. The reference method for calibration is the glucose from the blood glucose meter.
Calibrate on the first day of monitoring: 4 times a day after approximately 6 hours, or if the blood glucose reading from the blood glucose meter is not within the estimated error range of the continuous glucose monitor (difference greater than 1,5mmol/L).
Calibration on days 2 and 3 of monitoring: once in the morning, or when the meter's glucose reading is not within the estimated error range of the continuous glucose monitoring (difference greater than 1,5mmol/L) On subsequent days when the blood glucose meter reading is not within the estimated error range of continuous glucose monitoring (difference greater than 1,5mmol/L).
Hypothesis:
- Ho: No difference in mean absolute relative difference (MARD) between calibrated and non-calibrated sensors.
- Ha: Calibrated sensors have lower mean absolute relative difference (δ = 4%, SD = 5%).
Further objectives:
To evaluate further aspects regarding the Dexcom G7 continuous glucose monitor system's performance in the intensive care unit (ICU), particularly in terms of sensitivity and specificity to detect hypo- and hyperglycaemic events, data availability, device deficiencies and safety.
To identify predictive factors of the accuracy of the Dexcom G7 continuous glucose monitor system in critically ill patients in the ICU.
Przegląd badań
Status
Warunki
Interwencja / Leczenie
Szczegółowy opis
Design:
- Each intensive care unit patient wears two sensors (calibrated + non-calibrated).
- 4 glucose measurements/day × 6 days = 24 measurements/subject.
- Sequential Organ Failure Assessment (SOFA) score recorded daily.
Typ studiów
Zapisy (Szacowany)
Faza
- Nie dotyczy
Kontakty i lokalizacje
Kontakt w sprawie studiów
- Nazwa: Marek Protus, MD
- Numer telefonu: +420602278062
- E-mail: marek.protus@ikem.cz
Kopia zapasowa kontaktu do badania
- Nazwa: Alexander Macek, MD
- Numer telefonu: +421910631876
- E-mail: alexander.macek@ikem.cz
Lokalizacje studiów
-
-
-
Prague, Czechy
- Rekrutacyjny
- Institute of Clinical and Experimental Medice
-
Kontakt:
- Marek Protus, MD
- Numer telefonu: +420602278062
- E-mail: marek.protus@ikem.cz
-
Kontakt:
- Alexander Macek, MD
- Numer telefonu: +421910631876
- E-mail: alexander.macek@ikem.cz
-
-
Kryteria uczestnictwa
Kryteria kwalifikacji
Wiek uprawniający do nauki
- Dorosły
- Starszy dorosły
Akceptuje zdrowych ochotników
Opis
Inclusion Criteria:
- Age 18 years or over
- Intensive care unit admission
- Requiring blood glucose monitoring for insulin therapy
Exclusion Criteria:
- High-dose ascorbic acid or acetaminophen > 4g/day
- Physical or mental health condition precluding continuous wear of the continuous glucose monitor, as judged by the physician
Plan studiów
Jak projektuje się badanie?
Szczegóły projektu
- Główny cel: Wykonalność urządzenia
- Przydział: Nielosowe
- Model interwencyjny: Zadanie dla jednej grupy
- Maskowanie: Potroić
Broń i interwencje
Grupa uczestników / Arm |
Interwencja / Leczenie |
|---|---|
|
Aktywny komparator: Open labeled CGM
Additional calibrations performed on CGM device during monitoring period
|
Continuous measurement of glucose from subcutaneous tissue
|
|
Pozorny komparator: Control labeled CGM
No additional calibrations performed on CGM device during monitoring period
|
Continuous measurement of glucose from subcutaneous tissue
|
Co mierzy badanie?
Podstawowe miary wyniku
Miara wyniku |
Opis środka |
Ramy czasowe |
|---|---|---|
|
Calculate overall mean absolute relative difference between arms
Ramy czasowe: From enrolment to discharge from ICU care or up-to day 10
|
Analyse the difference in mean absolute relative difference (MARD) of Dexcom G7 continuous glucose monitor between calibrated and uncalibrated sensor, in patients undergoing major abdominal surgery and solid organ transplantation.
|
From enrolment to discharge from ICU care or up-to day 10
|
Miary wyników drugorzędnych
Miara wyniku |
Opis środka |
Ramy czasowe |
|---|---|---|
|
Accuracy metrics
Ramy czasowe: From enrolment to discharge from ICU care or up-to day 10
|
To evaluate further aspects regarding the Dexcom G7 continuous glucose monitor system's performance in the intensive care unit (ICU), particularly in terms of sensitivity and specificity to detect hypo- and hyperglycaemic events, data availability, device deficiencies and safety. To identify predictive factors of the accuracy of the Dexcom G7 continuous glucose monitor system in critically ill patients in the ICU. |
From enrolment to discharge from ICU care or up-to day 10
|
Współpracownicy i badacze
Daty zapisu na studia
Główne daty studiów
Rozpoczęcie studiów (Rzeczywisty)
Zakończenie podstawowe (Szacowany)
Ukończenie studiów (Szacowany)
Daty rejestracji na studia
Pierwszy przesłany
Pierwszy przesłany, który spełnia kryteria kontroli jakości
Pierwszy wysłany (Rzeczywisty)
Aktualizacje rekordów badań
Ostatnia wysłana aktualizacja (Rzeczywisty)
Ostatnia przesłana aktualizacja, która spełniała kryteria kontroli jakości
Ostatnia weryfikacja
Więcej informacji
Terminy związane z tym badaniem
Dodatkowe istotne warunki MeSH
Inne numery identyfikacyjne badania
- G-25-71
- ID Project No. LX22NPO5104 (Inny numer grantu/finansowania: MH CZ-DRO (Institute for Clinical and Experimental Medicine-IKEM, IN 00023001)
Informacje o lekach i urządzeniach, dokumenty badawcze
Bada produkt leczniczy regulowany przez amerykańską FDA
Bada produkt urządzenia regulowany przez amerykańską FDA
Te informacje zostały pobrane bezpośrednio ze strony internetowej clinicaltrials.gov bez żadnych zmian. Jeśli chcesz zmienić, usunąć lub zaktualizować dane swojego badania, skontaktuj się z register@clinicaltrials.gov. Gdy tylko zmiana zostanie wprowadzona na stronie clinicaltrials.gov, zostanie ona automatycznie zaktualizowana również na naszej stronie internetowej .