Evaluate the Performance of the Dexcom G7 Continuous Glucose Monitoring (CGM) System in Critically Ill Patients Undergoing Major Abdominal Surgery and Solid Organ Transplantation Which Require Blood Glucose Monitoring for Intensive Insulin Therapy. (DREAM-ICU)

20 de mayo de 2026 actualizado por: Marek Protuš, Institute for Clinical and Experimental Medicine

Determining the Reliability and Accuracy of the Dexcom G7 Continuous Glucose Monitoring System in the ICU

Main objective:

The goal is to to determine the accuracy of Dexcom G7 Continuous glucose monitor system in critically ill patients, undergoing major abdominal surgery and solid organ transplantation, in the intensive care unit.

On admission to the intensive care unit or at the end of surgery, two Dexcom G7 sensors are placed in the infraclavicular position.

Sensor number 1 will be blinded and will only be active when paired with the receiver, without the ability to monitor glucose levels.

Sensor number 2 is used for continuous glucose monitoring and can be used to adjust the insulin dose.

Patients will wear the continuous glucose monitoring sensor for a maximum of 10 days, including the time spent in the intensive care unit and the immediate care unit. The study will not interfere with any usual care clinical workflows.

Calibration is only performed on an unblinded sensor. The reference method for calibration is the glucose from the blood glucose meter.

Calibrate on the first day of monitoring: 4 times a day after approximately 6 hours, or if the blood glucose reading from the blood glucose meter is not within the estimated error range of the continuous glucose monitor (difference greater than 1,5mmol/L).

Calibration on days 2 and 3 of monitoring: once in the morning, or when the meter's glucose reading is not within the estimated error range of the continuous glucose monitoring (difference greater than 1,5mmol/L) On subsequent days when the blood glucose meter reading is not within the estimated error range of continuous glucose monitoring (difference greater than 1,5mmol/L).

Hypothesis:

  • Ho: No difference in mean absolute relative difference (MARD) between calibrated and non-calibrated sensors.
  • Ha: Calibrated sensors have lower mean absolute relative difference (δ = 4%, SD = 5%).

Further objectives:

To evaluate further aspects regarding the Dexcom G7 continuous glucose monitor system's performance in the intensive care unit (ICU), particularly in terms of sensitivity and specificity to detect hypo- and hyperglycaemic events, data availability, device deficiencies and safety.

To identify predictive factors of the accuracy of the Dexcom G7 continuous glucose monitor system in critically ill patients in the ICU.

Descripción general del estudio

Descripción detallada

Design:

  • Each intensive care unit patient wears two sensors (calibrated + non-calibrated).
  • 4 glucose measurements/day × 6 days = 24 measurements/subject.
  • Sequential Organ Failure Assessment (SOFA) score recorded daily.

Tipo de estudio

Intervencionista

Inscripción (Estimado)

62

Fase

  • No aplica

Contactos y Ubicaciones

Esta sección proporciona los datos de contacto de quienes realizan el estudio e información sobre dónde se lleva a cabo este estudio.

Estudio Contacto

  • Nombre: Marek Protus, MD
  • Número de teléfono: +420602278062
  • Correo electrónico: marek.protus@ikem.cz

Copia de seguridad de contactos de estudio

Ubicaciones de estudio

      • Prague, Chequia
        • Reclutamiento
        • Institute of Clinical and Experimental Medice
        • Contacto:
        • Contacto:

Criterios de participación

Los investigadores buscan personas que se ajusten a una determinada descripción, denominada criterio de elegibilidad. Algunos ejemplos de estos criterios son el estado de salud general de una persona o tratamientos previos.

Criterio de elegibilidad

Edades elegibles para estudiar

  • Adulto
  • Adulto Mayor

Acepta Voluntarios Saludables

No

Descripción

Inclusion Criteria:

  • Age 18 years or over
  • Intensive care unit admission
  • Requiring blood glucose monitoring for insulin therapy

Exclusion Criteria:

  • High-dose ascorbic acid or acetaminophen > 4g/day
  • Physical or mental health condition precluding continuous wear of the continuous glucose monitor, as judged by the physician

Plan de estudios

Esta sección proporciona detalles del plan de estudio, incluido cómo está diseñado el estudio y qué mide el estudio.

¿Cómo está diseñado el estudio?

Detalles de diseño

  • Propósito principal: Viabilidad del dispositivo
  • Asignación: No aleatorizado
  • Modelo Intervencionista: Asignación de un solo grupo
  • Enmascaramiento: Triple

Armas e Intervenciones

Grupo de participantes/brazo
Intervención / Tratamiento
Comparador activo: Open labeled CGM
Additional calibrations performed on CGM device during monitoring period
Continuous measurement of glucose from subcutaneous tissue
Comparador falso: Control labeled CGM
No additional calibrations performed on CGM device during monitoring period
Continuous measurement of glucose from subcutaneous tissue

¿Qué mide el estudio?

Medidas de resultado primarias

Medida de resultado
Medida Descripción
Periodo de tiempo
Calculate overall mean absolute relative difference between arms
Periodo de tiempo: From enrolment to discharge from ICU care or up-to day 10
Analyse the difference in mean absolute relative difference (MARD) of Dexcom G7 continuous glucose monitor between calibrated and uncalibrated sensor, in patients undergoing major abdominal surgery and solid organ transplantation.
From enrolment to discharge from ICU care or up-to day 10

Medidas de resultado secundarias

Medida de resultado
Medida Descripción
Periodo de tiempo
Accuracy metrics
Periodo de tiempo: From enrolment to discharge from ICU care or up-to day 10

To evaluate further aspects regarding the Dexcom G7 continuous glucose monitor system's performance in the intensive care unit (ICU), particularly in terms of sensitivity and specificity to detect hypo- and hyperglycaemic events, data availability, device deficiencies and safety.

To identify predictive factors of the accuracy of the Dexcom G7 continuous glucose monitor system in critically ill patients in the ICU.

From enrolment to discharge from ICU care or up-to day 10

Colaboradores e Investigadores

Aquí es donde encontrará personas y organizaciones involucradas en este estudio.

Fechas de registro del estudio

Estas fechas rastrean el progreso del registro del estudio y los envíos de resultados resumidos a ClinicalTrials.gov. Los registros del estudio y los resultados informados son revisados ​​por la Biblioteca Nacional de Medicina (NLM) para asegurarse de que cumplan con los estándares de control de calidad específicos antes de publicarlos en el sitio web público.

Fechas importantes del estudio

Inicio del estudio (Actual)

14 de enero de 2026

Finalización primaria (Estimado)

1 de septiembre de 2026

Finalización del estudio (Estimado)

1 de abril de 2027

Fechas de registro del estudio

Enviado por primera vez

14 de mayo de 2026

Primero enviado que cumplió con los criterios de control de calidad

20 de mayo de 2026

Publicado por primera vez (Actual)

28 de mayo de 2026

Actualizaciones de registros de estudio

Última actualización publicada (Actual)

28 de mayo de 2026

Última actualización enviada que cumplió con los criterios de control de calidad

20 de mayo de 2026

Última verificación

1 de abril de 2026

Más información

Términos relacionados con este estudio

Otros números de identificación del estudio

  • G-25-71
  • ID Project No. LX22NPO5104 (Otro número de subvención/financiamiento: MH CZ-DRO (Institute for Clinical and Experimental Medicine-IKEM, IN 00023001)

Información sobre medicamentos y dispositivos, documentos del estudio

Estudia un producto farmacéutico regulado por la FDA de EE. UU.

No

Estudia un producto de dispositivo regulado por la FDA de EE. UU.

No

Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .

Suscribir