Effect of Different Plasma Insulin Levels on the Accuracy of Continuous Subcutaneous Glucose Monitoring
New Strategies for Automated Glycaemic Control: the Issue of Continuous Glucose Monitoring Accuracy Under Hypoglycaemic Conditions
Achieving near-normoglycaemia has been established as the main objective for most patients with diabetes. However, it is well known that intensification of treatment is associated with an increase in the frequency of hypoglycemia, especially in the context of insulin therapy. The burden of hypoglycemia in terms of psychological implications, morbidity and even mortality, explains why it has been defined as the main limiting factor to achievement of good metabolic control.
Continuous subcutaneous glucose monitoring (CGM) devices have been claimed to be useful in hypoglycemia detection/prevention, allowing theoretically for safer intensification of therapy in diabetic patients. However, accuracy of CGM devices, especially in the hypoglycemic range, raises some concerns. In fact, commercially available CGM devices estimate plasma glucose from measurements in the interstitial fluid and not in plasma. However, the relationship between plasma and interstitial glucose is not fully understood, especially under dynamic conditions, and this may explain the poor CGM performance during rapid changes in blood glucose and hypoglycemia.
In this project, the relationship between plasma and interstitial glucose will be evaluated under conditions of normal glucose concentrations and hypoglycemia. Experiments will be performed to assess the role, if any, of different plasma insulin concentrations on the accuracy of CGM.
All the information obtained may be relevant to the improvement of the ability of CGM devices to detect hypoglycemia and hypoglycemic risk.
연구 개요
상태
연구 유형
등록 (실제)
단계
- 해당 없음
연락처 및 위치
연구 장소
-
-
-
Valencia, 스페인, 46010
- Servicio de Endocrinología y Nutrición, Hospital Clínico Universitario de Valencia
-
-
참여기준
자격 기준
공부할 수 있는 나이
건강한 자원 봉사자를 받아들입니다
연구 대상 성별
설명
Inclusion Criteria:
Subjects with type 1 diabetes mellitus meeting all of the following criteria will be considered for admission to the study:
- Aged between 18 and 60 years
- Under CSII or MDI treatment for at least six months before Visit 1
- Body mass index of between 18 and 30 kg/m2
- HbA1c 6.0-8.5% at Visit 1
- Normal laboratory values, ECG, and vital signs unless the investigator considered an abnormality to be clinically irrelevant
- Women postmenopausal or using contraception judged by the investigator to be adequate (e.g., oral contraceptives, intra-uterine device or surgical treatment)
Exclusion Criteria:
Subjects meeting any of the following criteria will not be included in the study:
- Pregnancy and lactation
- History of hypersensitivity to the study medications or to drugs with similar chemical structures
- Confirmed hypoglycaemia unawareness
- Progressive fatal diseases
- History of drug or alcohol abuse
- History of positive HIV or hepatitis B or C test
- Impaired hepatic function, as shown by, but not limited to, SGPT or SGOT of more than twice the upper limit of the normal range at visit 1
- Impaired renal function, as shown by, but not limited to, serum creatinine > 1.5 mg/dL at visit 1
- Clinically relevant microvascular, cardiovascular, hepatic, neurologic, endocrine or other major systemic diseases other than T1DM which could hinder implementation of the clinical study protocol or interpretation of the study results
- Pre-planned surgery during the study
- Blood donation of more than 500 ml during the past three months for men, or during the past six months for women
- Mental condition rendering the subject unable to understand the nature, scope and possible consequences of the study
- Subject unlikely to comply with clinical study protocol, e.g., uncooperative attitude, inability to return for follow-up visits, or poor likelihood of completing the study
- Receipt of an experimental drug or use of an experimental device during the past 30 days.
공부 계획
연구는 어떻게 설계됩니까?
디자인 세부사항
- 주 목적: 기초 과학
- 할당: 무작위
- 중재 모델: 크로스오버 할당
- 마스킹: 더블
무기와 개입
참가자 그룹 / 팔 |
개입 / 치료 |
|---|---|
|
다른: High plasma insulin-Low plasma insulin
Seven out of the 14 subjects recruited in the study have been randomized to receive on the first study day a High insulin glucose clamp and on the second day a Low insulin glucose clamp (sequence 'AB')
|
Primed intravenous insulin infusion at a rate of 1mU/kg/min is given during the whole study duration.
Glucose is infused at a variable rate to maintain plasma glucose concentrations 'clamped' at pre-specified levels.
Plasma glucose levels will be the same in both study arms.
Primed intravenous insulin infusion at a rate of 0.3mU/kg/min is given during the whole study duration.
Glucose is infused at a variable rate to maintain plasma glucose concentrations 'clamped' at pre-specified levels.
Plasma glucose levels will be the same in both study arms.
|
|
다른: Low plasma insulin-High plasma insulin
Seven out of the 14 subjects recruited in the study have been randomized to receive on the first study day a Low insulin glucose clamp and on the second day a High insulin glucose clamp (sequence 'BA')
|
Primed intravenous insulin infusion at a rate of 1mU/kg/min is given during the whole study duration.
Glucose is infused at a variable rate to maintain plasma glucose concentrations 'clamped' at pre-specified levels.
Plasma glucose levels will be the same in both study arms.
Primed intravenous insulin infusion at a rate of 0.3mU/kg/min is given during the whole study duration.
Glucose is infused at a variable rate to maintain plasma glucose concentrations 'clamped' at pre-specified levels.
Plasma glucose levels will be the same in both study arms.
|
연구는 무엇을 측정합니까?
주요 결과 측정
결과 측정 |
측정값 설명 |
기간 |
|---|---|---|
|
Normalized current signal from the CGM devices
기간: 330 minutes
|
CGM devices are electrochemical sensors. Currently available sensors are placed in the subcutaneous tissue where they react with the glucose of the interstitial fluid giving a current signal (in nano Amperes), which is proportional to glucose concentrations. The current signal is then filtered and transformed into plasma glucose estimations through a calibration algorithm built into the CGM device. In this study we will analyze the current signal from CGM obtained under two different experimental conditions:
This will allow for establishing the role, if any, of different plasma insulin concentrations on the accuracy of CGM to identify plasma glucose falls resulting in hypoglycaemia. Since CGM sensors can exhibit different sensitivities to glucose concentrations, the current signal will be normalized before statistical analysis |
330 minutes
|
공동 작업자 및 조사자
수사관
- 수석 연구원: Francisco Javier Ampudia-Blasco, MD, PhD, Hospital Clínico Universitario de Valencia - Fundación INCLIVA
간행물 및 유용한 링크
연구 기록 날짜
연구 주요 날짜
연구 시작
기본 완료 (실제)
연구 완료 (실제)
연구 등록 날짜
최초 제출
QC 기준을 충족하는 최초 제출
처음 게시됨 (추정)
연구 기록 업데이트
마지막 업데이트 게시됨 (추정)
QC 기준을 충족하는 마지막 업데이트 제출
마지막으로 확인됨
추가 정보
이 연구와 관련된 용어
키워드
추가 관련 MeSH 약관
기타 연구 ID 번호
- FP7-PEOPLE-2009-IEF #252085-2
- Spanish Ministry of Science (기타 보조금/기금 번호: spanish Ministry of Science, DPI2010-20764-C02-01)
이 정보는 변경 없이 clinicaltrials.gov 웹사이트에서 직접 가져온 것입니다. 귀하의 연구 세부 정보를 변경, 제거 또는 업데이트하도록 요청하는 경우 register@clinicaltrials.gov. 문의하십시오. 변경 사항이 clinicaltrials.gov에 구현되는 즉시 저희 웹사이트에도 자동으로 업데이트됩니다. .