Effect of Different Plasma Insulin Levels on the Accuracy of Continuous Subcutaneous Glucose Monitoring

New Strategies for Automated Glycaemic Control: the Issue of Continuous Glucose Monitoring Accuracy Under Hypoglycaemic Conditions

Achieving near-normoglycaemia has been established as the main objective for most patients with diabetes. However, it is well known that intensification of treatment is associated with an increase in the frequency of hypoglycemia, especially in the context of insulin therapy. The burden of hypoglycemia in terms of psychological implications, morbidity and even mortality, explains why it has been defined as the main limiting factor to achievement of good metabolic control.

Continuous subcutaneous glucose monitoring (CGM) devices have been claimed to be useful in hypoglycemia detection/prevention, allowing theoretically for safer intensification of therapy in diabetic patients. However, accuracy of CGM devices, especially in the hypoglycemic range, raises some concerns. In fact, commercially available CGM devices estimate plasma glucose from measurements in the interstitial fluid and not in plasma. However, the relationship between plasma and interstitial glucose is not fully understood, especially under dynamic conditions, and this may explain the poor CGM performance during rapid changes in blood glucose and hypoglycemia.

In this project, the relationship between plasma and interstitial glucose will be evaluated under conditions of normal glucose concentrations and hypoglycemia. Experiments will be performed to assess the role, if any, of different plasma insulin concentrations on the accuracy of CGM.

All the information obtained may be relevant to the improvement of the ability of CGM devices to detect hypoglycemia and hypoglycemic risk.

Panoramica dello studio

Tipo di studio

Interventistico

Iscrizione (Effettivo)

14

Fase

  • Non applicabile

Contatti e Sedi

Questa sezione fornisce i recapiti di coloro che conducono lo studio e informazioni su dove viene condotto lo studio.

Luoghi di studio

      • Valencia, Spagna, 46010
        • Servicio de Endocrinología y Nutrición, Hospital Clínico Universitario de Valencia

Criteri di partecipazione

I ricercatori cercano persone che corrispondano a una certa descrizione, chiamata criteri di ammissibilità. Alcuni esempi di questi criteri sono le condizioni generali di salute di una persona o trattamenti precedenti.

Criteri di ammissibilità

Età idonea allo studio

Da 18 anni a 60 anni (Adulto)

Accetta volontari sani

No

Sessi ammissibili allo studio

Tutto

Descrizione

Inclusion Criteria:

  • Subjects with type 1 diabetes mellitus meeting all of the following criteria will be considered for admission to the study:

    • Aged between 18 and 60 years
    • Under CSII or MDI treatment for at least six months before Visit 1
    • Body mass index of between 18 and 30 kg/m2
    • HbA1c 6.0-8.5% at Visit 1
    • Normal laboratory values, ECG, and vital signs unless the investigator considered an abnormality to be clinically irrelevant
    • Women postmenopausal or using contraception judged by the investigator to be adequate (e.g., oral contraceptives, intra-uterine device or surgical treatment)

Exclusion Criteria:

  • Subjects meeting any of the following criteria will not be included in the study:

    • Pregnancy and lactation
    • History of hypersensitivity to the study medications or to drugs with similar chemical structures
    • Confirmed hypoglycaemia unawareness
    • Progressive fatal diseases
    • History of drug or alcohol abuse
    • History of positive HIV or hepatitis B or C test
    • Impaired hepatic function, as shown by, but not limited to, SGPT or SGOT of more than twice the upper limit of the normal range at visit 1
    • Impaired renal function, as shown by, but not limited to, serum creatinine > 1.5 mg/dL at visit 1
    • Clinically relevant microvascular, cardiovascular, hepatic, neurologic, endocrine or other major systemic diseases other than T1DM which could hinder implementation of the clinical study protocol or interpretation of the study results
    • Pre-planned surgery during the study
    • Blood donation of more than 500 ml during the past three months for men, or during the past six months for women
    • Mental condition rendering the subject unable to understand the nature, scope and possible consequences of the study
    • Subject unlikely to comply with clinical study protocol, e.g., uncooperative attitude, inability to return for follow-up visits, or poor likelihood of completing the study
    • Receipt of an experimental drug or use of an experimental device during the past 30 days.

Piano di studio

Questa sezione fornisce i dettagli del piano di studio, compreso il modo in cui lo studio è progettato e ciò che lo studio sta misurando.

Come è strutturato lo studio?

Dettagli di progettazione

  • Scopo principale: Scienza basilare
  • Assegnazione: Randomizzato
  • Modello interventistico: Assegnazione incrociata
  • Mascheramento: Doppio

Armi e interventi

Gruppo di partecipanti / Arm
Intervento / Trattamento
Altro: High plasma insulin-Low plasma insulin
Seven out of the 14 subjects recruited in the study have been randomized to receive on the first study day a High insulin glucose clamp and on the second day a Low insulin glucose clamp (sequence 'AB')
Primed intravenous insulin infusion at a rate of 1mU/kg/min is given during the whole study duration. Glucose is infused at a variable rate to maintain plasma glucose concentrations 'clamped' at pre-specified levels. Plasma glucose levels will be the same in both study arms.
Primed intravenous insulin infusion at a rate of 0.3mU/kg/min is given during the whole study duration. Glucose is infused at a variable rate to maintain plasma glucose concentrations 'clamped' at pre-specified levels. Plasma glucose levels will be the same in both study arms.
Altro: Low plasma insulin-High plasma insulin
Seven out of the 14 subjects recruited in the study have been randomized to receive on the first study day a Low insulin glucose clamp and on the second day a High insulin glucose clamp (sequence 'BA')
Primed intravenous insulin infusion at a rate of 1mU/kg/min is given during the whole study duration. Glucose is infused at a variable rate to maintain plasma glucose concentrations 'clamped' at pre-specified levels. Plasma glucose levels will be the same in both study arms.
Primed intravenous insulin infusion at a rate of 0.3mU/kg/min is given during the whole study duration. Glucose is infused at a variable rate to maintain plasma glucose concentrations 'clamped' at pre-specified levels. Plasma glucose levels will be the same in both study arms.

Cosa sta misurando lo studio?

Misure di risultato primarie

Misura del risultato
Misura Descrizione
Lasso di tempo
Normalized current signal from the CGM devices
Lasso di tempo: 330 minutes

CGM devices are electrochemical sensors. Currently available sensors are placed in the subcutaneous tissue where they react with the glucose of the interstitial fluid giving a current signal (in nano Amperes), which is proportional to glucose concentrations. The current signal is then filtered and transformed into plasma glucose estimations through a calibration algorithm built into the CGM device.

In this study we will analyze the current signal from CGM obtained under two different experimental conditions:

  1. Eu-hypoglycemic clamp with High plasma insulin concentrations
  2. Eu-hypoglycemic clamp with relatively Low plasma insulin concentrations

This will allow for establishing the role, if any, of different plasma insulin concentrations on the accuracy of CGM to identify plasma glucose falls resulting in hypoglycaemia.

Since CGM sensors can exhibit different sensitivities to glucose concentrations, the current signal will be normalized before statistical analysis

330 minutes

Collaboratori e investigatori

Qui è dove troverai le persone e le organizzazioni coinvolte in questo studio.

Investigatori

  • Investigatore principale: Francisco Javier Ampudia-Blasco, MD, PhD, Hospital Clínico Universitario de Valencia - Fundación INCLIVA

Pubblicazioni e link utili

La persona responsabile dell'inserimento delle informazioni sullo studio fornisce volontariamente queste pubblicazioni. Questi possono riguardare qualsiasi cosa relativa allo studio.

Studiare le date dei record

Queste date tengono traccia dell'avanzamento della registrazione dello studio e dell'invio dei risultati di sintesi a ClinicalTrials.gov. I record degli studi e i risultati riportati vengono esaminati dalla National Library of Medicine (NLM) per assicurarsi che soddisfino specifici standard di controllo della qualità prima di essere pubblicati sul sito Web pubblico.

Studia le date principali

Inizio studio

1 ottobre 2011

Completamento primario (Effettivo)

1 settembre 2012

Completamento dello studio (Effettivo)

1 ottobre 2012

Date di iscrizione allo studio

Primo inviato

24 ottobre 2012

Primo inviato che soddisfa i criteri di controllo qualità

24 ottobre 2012

Primo Inserito (Stima)

26 ottobre 2012

Aggiornamenti dei record di studio

Ultimo aggiornamento pubblicato (Stima)

26 ottobre 2012

Ultimo aggiornamento inviato che soddisfa i criteri QC

24 ottobre 2012

Ultimo verificato

1 ottobre 2012

Maggiori informazioni

Termini relativi a questo studio

Altri numeri di identificazione dello studio

  • FP7-PEOPLE-2009-IEF #252085-2
  • Spanish Ministry of Science (Altro numero di sovvenzione/finanziamento: spanish Ministry of Science, DPI2010-20764-C02-01)

Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .