此页面是自动翻译的,不保证翻译的准确性。请参阅 英文版 对于源文本。

Effect of Different Plasma Insulin Levels on the Accuracy of Continuous Subcutaneous Glucose Monitoring

New Strategies for Automated Glycaemic Control: the Issue of Continuous Glucose Monitoring Accuracy Under Hypoglycaemic Conditions

Achieving near-normoglycaemia has been established as the main objective for most patients with diabetes. However, it is well known that intensification of treatment is associated with an increase in the frequency of hypoglycemia, especially in the context of insulin therapy. The burden of hypoglycemia in terms of psychological implications, morbidity and even mortality, explains why it has been defined as the main limiting factor to achievement of good metabolic control.

Continuous subcutaneous glucose monitoring (CGM) devices have been claimed to be useful in hypoglycemia detection/prevention, allowing theoretically for safer intensification of therapy in diabetic patients. However, accuracy of CGM devices, especially in the hypoglycemic range, raises some concerns. In fact, commercially available CGM devices estimate plasma glucose from measurements in the interstitial fluid and not in plasma. However, the relationship between plasma and interstitial glucose is not fully understood, especially under dynamic conditions, and this may explain the poor CGM performance during rapid changes in blood glucose and hypoglycemia.

In this project, the relationship between plasma and interstitial glucose will be evaluated under conditions of normal glucose concentrations and hypoglycemia. Experiments will be performed to assess the role, if any, of different plasma insulin concentrations on the accuracy of CGM.

All the information obtained may be relevant to the improvement of the ability of CGM devices to detect hypoglycemia and hypoglycemic risk.

研究概览

研究类型

介入性

注册 (实际的)

14

阶段

  • 不适用

联系人和位置

本节提供了进行研究的人员的详细联系信息,以及有关进行该研究的地点的信息。

学习地点

      • Valencia、西班牙、46010
        • Servicio de Endocrinología y Nutrición, Hospital Clínico Universitario de Valencia

参与标准

研究人员寻找符合特定描述的人,称为资格标准。这些标准的一些例子是一个人的一般健康状况或先前的治疗。

资格标准

适合学习的年龄

18年 至 60年 (成人)

接受健康志愿者

不

有资格学习的性别

全部

描述

Inclusion Criteria:

  • Subjects with type 1 diabetes mellitus meeting all of the following criteria will be considered for admission to the study:

    • Aged between 18 and 60 years
    • Under CSII or MDI treatment for at least six months before Visit 1
    • Body mass index of between 18 and 30 kg/m2
    • HbA1c 6.0-8.5% at Visit 1
    • Normal laboratory values, ECG, and vital signs unless the investigator considered an abnormality to be clinically irrelevant
    • Women postmenopausal or using contraception judged by the investigator to be adequate (e.g., oral contraceptives, intra-uterine device or surgical treatment)

Exclusion Criteria:

  • Subjects meeting any of the following criteria will not be included in the study:

    • Pregnancy and lactation
    • History of hypersensitivity to the study medications or to drugs with similar chemical structures
    • Confirmed hypoglycaemia unawareness
    • Progressive fatal diseases
    • History of drug or alcohol abuse
    • History of positive HIV or hepatitis B or C test
    • Impaired hepatic function, as shown by, but not limited to, SGPT or SGOT of more than twice the upper limit of the normal range at visit 1
    • Impaired renal function, as shown by, but not limited to, serum creatinine > 1.5 mg/dL at visit 1
    • Clinically relevant microvascular, cardiovascular, hepatic, neurologic, endocrine or other major systemic diseases other than T1DM which could hinder implementation of the clinical study protocol or interpretation of the study results
    • Pre-planned surgery during the study
    • Blood donation of more than 500 ml during the past three months for men, or during the past six months for women
    • Mental condition rendering the subject unable to understand the nature, scope and possible consequences of the study
    • Subject unlikely to comply with clinical study protocol, e.g., uncooperative attitude, inability to return for follow-up visits, or poor likelihood of completing the study
    • Receipt of an experimental drug or use of an experimental device during the past 30 days.

学习计划

本节提供研究计划的详细信息,包括研究的设计方式和研究的衡量标准。

研究是如何设计的?

设计细节

  • 主要用途:基础科学
  • 分配:随机化
  • 介入模型:交叉作业
  • 屏蔽:双倍的

武器和干预

参与者组/臂
干预/治疗
其他:High plasma insulin-Low plasma insulin
Seven out of the 14 subjects recruited in the study have been randomized to receive on the first study day a High insulin glucose clamp and on the second day a Low insulin glucose clamp (sequence 'AB')
Primed intravenous insulin infusion at a rate of 1mU/kg/min is given during the whole study duration. Glucose is infused at a variable rate to maintain plasma glucose concentrations 'clamped' at pre-specified levels. Plasma glucose levels will be the same in both study arms.
Primed intravenous insulin infusion at a rate of 0.3mU/kg/min is given during the whole study duration. Glucose is infused at a variable rate to maintain plasma glucose concentrations 'clamped' at pre-specified levels. Plasma glucose levels will be the same in both study arms.
其他:Low plasma insulin-High plasma insulin
Seven out of the 14 subjects recruited in the study have been randomized to receive on the first study day a Low insulin glucose clamp and on the second day a High insulin glucose clamp (sequence 'BA')
Primed intravenous insulin infusion at a rate of 1mU/kg/min is given during the whole study duration. Glucose is infused at a variable rate to maintain plasma glucose concentrations 'clamped' at pre-specified levels. Plasma glucose levels will be the same in both study arms.
Primed intravenous insulin infusion at a rate of 0.3mU/kg/min is given during the whole study duration. Glucose is infused at a variable rate to maintain plasma glucose concentrations 'clamped' at pre-specified levels. Plasma glucose levels will be the same in both study arms.

研究衡量的是什么?

主要结果指标

结果测量
措施说明
大体时间
Normalized current signal from the CGM devices
大体时间:330 minutes

CGM devices are electrochemical sensors. Currently available sensors are placed in the subcutaneous tissue where they react with the glucose of the interstitial fluid giving a current signal (in nano Amperes), which is proportional to glucose concentrations. The current signal is then filtered and transformed into plasma glucose estimations through a calibration algorithm built into the CGM device.

In this study we will analyze the current signal from CGM obtained under two different experimental conditions:

  1. Eu-hypoglycemic clamp with High plasma insulin concentrations
  2. Eu-hypoglycemic clamp with relatively Low plasma insulin concentrations

This will allow for establishing the role, if any, of different plasma insulin concentrations on the accuracy of CGM to identify plasma glucose falls resulting in hypoglycaemia.

Since CGM sensors can exhibit different sensitivities to glucose concentrations, the current signal will be normalized before statistical analysis

330 minutes

合作者和调查者

在这里您可以找到参与这项研究的人员和组织。

调查人员

  • 首席研究员:Francisco Javier Ampudia-Blasco, MD, PhD、Hospital Clínico Universitario de Valencia - Fundación INCLIVA

出版物和有用的链接

负责输入研究信息的人员自愿提供这些出版物。这些可能与研究有关。

研究记录日期

这些日期跟踪向 ClinicalTrials.gov 提交研究记录和摘要结果的进度。研究记录和报告的结果由国家医学图书馆 (NLM) 审查,以确保它们在发布到公共网站之前符合特定的质量控制标准。

研究主要日期

学习开始

2011年10月1日

初级完成 (实际的)

2012年9月1日

研究完成 (实际的)

2012年10月1日

研究注册日期

首次提交

2012年10月24日

首先提交符合 QC 标准的

2012年10月24日

首次发布 (估计)

2012年10月26日

研究记录更新

最后更新发布 (估计)

2012年10月26日

上次提交的符合 QC 标准的更新

2012年10月24日

最后验证

2012年10月1日

更多信息

与本研究相关的术语

其他研究编号

  • FP7-PEOPLE-2009-IEF #252085-2
  • Spanish Ministry of Science (其他赠款/资助编号:spanish Ministry of Science, DPI2010-20764-C02-01)

此信息直接从 clinicaltrials.gov 网站检索,没有任何更改。如果您有任何更改、删除或更新研究详细信息的请求,请联系 register@clinicaltrials.gov. clinicaltrials.gov 上实施更改,我们的网站上也会自动更新.

订阅