이 페이지는 자동 번역되었으며 번역의 정확성을 보장하지 않습니다. 참조하십시오 영문판 원본 텍스트의 경우.

Investigating the Relationship Between Measures of the Hearing Nerve and Speech Perception Outcomes in Adult Nexa Cochlear Implant Recipients (NODE-SDN)

2026년 11월 16일 업데이트: Cochlear

Prospective, Interventional, Multicentre Feasibility Investigation Evaluating Objective Measures and Clinical Outcomes in Adult Nexa® Cochlear Implant Recipients

This study aims to learn more about the information that can be obtained directly from measurements through a Nexa cochlear implant during and after surgery. Researchers will investigate whether these measurements can help understand how the implant is interacting with the inner ear and hearing nerve, where the electrodes are positioned inside the cochlea, and potentially how the implant should be programmed and how well a person detects sounds.

The study will also evaluate how well two methods agree in determining the position of the electrode contacts relative to the anatomy of the inner ear and the hearing nerve. One method uses measurements after surgery made through the implant itself. The other uses medical Computed Tomography(CT) imaging made after the surgery. The information collected will also contribute to a clinical research database that may be used to support future research and development. Participants will be told about this potential future use when they provide consent.

연구 개요

상태

아직 모집하지 않음

연구 유형

중재적

등록 (추정된)

200

단계

  • 해당 없음

연락처 및 위치

이 섹션에서는 연구를 수행하는 사람들의 연락처 정보와 이 연구가 수행되는 장소에 대한 정보를 제공합니다.

연구 연락처

연구 연락처 백업

연구 장소

    • Lower Saxony
      • Hanover, Lower Saxony, 독일, 30625
        • Medizinische Hochschule Hannover
        • 수석 연구원:
          • Andreas Büchner, PhD
    • Missouri
      • St Louis, Missouri, 미국, 63141
        • Washington University School of Medicine Ear, Nose & Throat Centre
        • 수석 연구원:
          • Amit Walia, MD
    • Utah
      • Salt Lake City, Utah, 미국, 84132
        • University of Utah Hospitals and Clinics
        • 수석 연구원:
          • Craig Buchman, MD
    • Antwerp
      • Wilrijk, Antwerp, 벨기에, 2610
        • Sint Augustinus ENT European Institute for ORL-HNS
        • 수석 연구원:
          • Andrzej Zarowski, MD
    • Victoria
      • East Melbourne, Victoria, 호주, 3002
        • The University of Melbourne, Royal Victorian Eye and Ear Hospital
        • 수석 연구원:
          • Stephen O'Leary, MBBS, PhD, FRACS

참여기준

연구원은 적격성 기준이라는 특정 설명에 맞는 사람을 찾습니다. 이러한 기준의 몇 가지 예는 개인의 일반적인 건강 상태 또는 이전 치료입니다.

자격 기준

공부할 수 있는 나이

18년 이상 (성인, 고령자)

건강한 자원 봉사자를 받아들입니다

아니

연구 대상 성별

모두

설명

Inclusion Criteria:

  1. Aged 18 years or older (no upper age limit).
  2. Candidate for CI1032 cochlear implantation as per the local sensorineural hearing loss criteria.
  3. Sufficient proficiency in the language used to conduct the study procedures, as determined by the investigator.
  4. Availability of pre-op CT scan of the cochlear region meeting quality requirements as determined by the sponsor.
  5. Willing and able to provide written informed consent and to comply with all requirements of the protocol.

Exclusion Criteria:

  1. Previous cochlear implantation in the same ear.
  2. Cochlear implant in the contralateral ear.
  3. Have an unaided air conduction pure tone average (average of unaided thresholds at 0.5, 1, 2 and 4 kHz) ≤40 dB HL in the contralateral ear measured within the last 12 months
  4. Severe or greater sensorineural hearing loss prior to 2 years of age, as documented in medical history.
  5. Deafness due to lesions of the auditory nerve or central auditory pathway.
  6. Cochlear ossification or anatomical anomaly precluding full electrode array insertion.
  7. Diagnosis of auditory neuropathy.
  8. Additional conditions preventing or restricting participation in the audiological evaluations, as determined by the investigator.
  9. Pregnant or breastfeeding women.
  10. Unrealistic expectations on the part of the subject regarding the potential benefits, risks, and limitations of the surgical procedure and device.
  11. Unable or unwilling to comply with the requirements of the clinical investigation as determined by the Investigator.
  12. Investigator site personnel directly affiliated with this study and/or their immediate families; immediate family is defined as a spouse, parent, child, or sibling.
  13. Cochlear employees or employees of Contract Research Organisations or contractors engaged by Cochlear for the purposes of this investigation.
  14. Current participation, or participation in another interventional clinical study/trial in the past 30 days, involving an investigational drug or device (unless determined by the Sponsor to not impact this investigation).

공부 계획

이 섹션에서는 연구 설계 방법과 연구가 측정하는 내용을 포함하여 연구 계획에 대한 세부 정보를 제공합니다.

연구는 어떻게 설계됩니까?

디자인 세부사항

  • 주 목적: 특수 증상
  • 할당: 해당 없음
  • 중재 모델: 단일 그룹 할당
  • 마스킹: 없음(오픈 라벨)

무기와 개입

참가자 그룹 / 팔
개입 / 치료
실험적: Cochlear implant recipients

Adult participants receiving a commercial Cochlear Nucleus Nexa cochlear implant, who undergo additional investigational objective measure procedures performed intra- and post-operatively.

Participants will be tested with intra- and post-operative objective measures, including Transimpedance Matrix (TIM) and Neural Response telemetry (NRT) recordings, using monopolar and focused multipolar stimulation.

연구는 무엇을 측정합니까?

주요 결과 측정

결과 측정
측정값 설명
기간
Correlation Between CT-Derived Electrode-to-Modiolus Distance and MAP T-Levels
기간: Postoperative CT assessment and cochlear implant programming assessment at 6 months post-activation

The correlation between electrode-to-modiolus distance (in millimeters, mm) derived from postoperative computed tomography (CT) imaging and cochlear implant MAP threshold (T) levels (in current level, CL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between electrode position and MAP T-levels.

Postoperative CT assessment and cochlear implant programming assessment at 6 months post-activation
Correlation Between CT-Derived Electrode-to-Modiolus Distance and MAP C-Levels
기간: Postoperative CT assessment and cochlear implant programming assessment at 6 months post-activation

The correlation between electrode-to-modiolus distance (in millimeters, mm) derived from postoperative CT imaging and cochlear implant MAP maximum comfortable (C) levels (in current level, CL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between electrode position and MAP C-levels.

Postoperative CT assessment and cochlear implant programming assessment at 6 months post-activation
Correlation Between CT-Derived Electrode-to-Modiolus Distance and Aided Pure-Tone Thresholds
기간: Postoperative CT assessment and aided hearing assessment at 6 months post-activation.

The correlation between electrode-to-modiolus distance (in millimeters, mm) derived from postoperative CT imaging and aided pure-tone thresholds (in dBHL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between electrode position and aided hearing thresholds.

Postoperative CT assessment and aided hearing assessment at 6 months post-activation.
Correlation Between Monopolar T-NRT Thresholds and MAP T-Levels
기간: At 6 months post-activation.

The correlation between monopolar telemetry neural response threshold (T-NRT) measurements and cochlear implant MAP threshold (T) levels (in current level, CL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between monopolar T-NRT thresholds and MAP T-levels.

At 6 months post-activation.
Correlation Between Monopolar T-NRT Thresholds and MAP C-Levels
기간: At 6 months post-activation.

The correlation between monopolar telemetry neural response threshold (T-NRT) measurements and cochlear implant MAP maximum comfortable (C) levels (in current level, CL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between monopolar T-NRT thresholds and MAP C-levels.

At 6 months post-activation.
Correlation Between Monopolar T-NRT Thresholds and Aided Pure-Tone Thresholds
기간: At 6 months post-activation.

The correlation between monopolar telemetry neural response threshold (T-NRT) measurements (in current level, CL) and aided pure-tone thresholds (in dBHL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between monopolar T-NRT thresholds and aided hearing thresholds.

At 6 months post-activation.
Correlation Between Focused T-NRT Thresholds and MAP T-Levels
기간: At 6 months post-activation.

The correlation between focused telemetry neural response threshold (T-NRT) measurements (in current level, CL) and cochlear implant MAP threshold (T) levels (in current level, CL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between focused T-NRT thresholds and MAP T-levels.

At 6 months post-activation.
Correlation Between Focused T-NRT Thresholds and MAP C-Levels
기간: At 6 months post-activation.

The correlation between focused telemetry neural response threshold (T-NRT) measurements (in current level, CL) and cochlear implant MAP maximum comfortable (C) levels (in current level, CL) will be assessed. The correlation coefficient with 95% confidence intervals. will be reported to characterize the strength and direction of the association between focused T-NRT thresholds and MAP C-levels.

At 6 months post-activation.
Correlation Between Focused T-NRT Thresholds and Aided Pure-Tone Thresholds
기간: At 6 months post-activation.

The correlation between focused telemetry neural response threshold (T-NRT) measurements (in current level, CL) and aided pure-tone thresholds (in dBHL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between focused T-NRT thresholds and aided hearing thresholds.

At 6 months post-activation.

2차 결과 측정

결과 측정
측정값 설명
기간
Mean Absolute Error Between TIM-Derived and CT-Derived Electrode-to-Modiolus Distance
기간: At 6 months post-activation

Postoperative electrode-to-modiolus distance for individual electrode contacts will be determined using postoperative computed tomography (CT) imaging and using Transimpedance Matrix (TIM) measurements analysed with an algorithm. The mean absolute error between paired TIM-derived and CT-derived electrode-to-modiolus distance measurements will be reported in millimetres (mm).

At 6 months post-activation

공동 작업자 및 조사자

여기에서 이 연구와 관련된 사람과 조직을 찾을 수 있습니다.

스폰서

연구 기록 날짜

이 날짜는 ClinicalTrials.gov에 대한 연구 기록 및 요약 결과 제출의 진행 상황을 추적합니다. 연구 기록 및 보고된 결과는 공개 웹사이트에 게시되기 전에 특정 품질 관리 기준을 충족하는지 확인하기 위해 국립 의학 도서관(NLM)에서 검토합니다.

연구 주요 날짜

연구 시작 (추정된)

2027년 1월 1일

기본 완료 (추정된)

2028년 8월 1일

연구 완료 (추정된)

2028년 8월 1일

연구 등록 날짜

최초 제출

2026년 11월 9일

QC 기준을 충족하는 최초 제출

2026년 11월 16일

처음 게시됨 (실제)

2026년 11월 23일

연구 기록 업데이트

마지막 업데이트 게시됨 (실제)

2026년 11월 23일

QC 기준을 충족하는 마지막 업데이트 제출

2026년 11월 16일

마지막으로 확인됨

2026년 11월 1일

추가 정보

이 연구와 관련된 용어

개별 참가자 데이터(IPD) 계획

개별 참가자 데이터(IPD)를 공유할 계획입니까?

아니요

약물 및 장치 정보, 연구 문서

미국 FDA 규제 의약품 연구

아니

미국 FDA 규제 기기 제품 연구

예

미국에서 제조되어 미국에서 수출되는 제품

아니

이 정보는 변경 없이 clinicaltrials.gov 웹사이트에서 직접 가져온 것입니다. 귀하의 연구 세부 정보를 변경, 제거 또는 업데이트하도록 요청하는 경우 register@clinicaltrials.gov. 문의하십시오. 변경 사항이 clinicaltrials.gov에 구현되는 즉시 저희 웹사이트에도 자동으로 업데이트됩니다. .

구독하다