Investigating the Relationship Between Measures of the Hearing Nerve and Speech Perception Outcomes in Adult Nexa Cochlear Implant Recipients (NODE-SDN)
Prospective, Interventional, Multicentre Feasibility Investigation Evaluating Objective Measures and Clinical Outcomes in Adult Nexa® Cochlear Implant Recipients
This study aims to learn more about the information that can be obtained directly from measurements through a Nexa cochlear implant during and after surgery. Researchers will investigate whether these measurements can help understand how the implant is interacting with the inner ear and hearing nerve, where the electrodes are positioned inside the cochlea, and potentially how the implant should be programmed and how well a person detects sounds.
The study will also evaluate how well two methods agree in determining the position of the electrode contacts relative to the anatomy of the inner ear and the hearing nerve. One method uses measurements after surgery made through the implant itself. The other uses medical Computed Tomography(CT) imaging made after the surgery. The information collected will also contribute to a clinical research database that may be used to support future research and development. Participants will be told about this potential future use when they provide consent.
Обзор исследования
Статус
Условия
Вмешательство/лечение
Тип исследования
Регистрация (Оцененный)
Фаза
- Непригодный
Контакты и местонахождение
Контакты исследования
- Имя: Geert De Ceulaer
- Номер телефона: +32486893006
- Электронная почта: gdeceulaer@cochlear.com
Учебное резервное копирование контактов
- Имя: Loes Beckers
- Электронная почта: lbeckers@cochlear.com
Места учебы
-
-
Victoria
-
East Melbourne, Victoria, Австралия, 3002
- The University of Melbourne, Royal Victorian Eye and Ear Hospital
-
Главный следователь:
- Stephen O'Leary, MBBS, PhD, FRACS
-
-
-
-
Antwerp
-
Wilrijk, Antwerp, Бельгия, 2610
- Sint Augustinus ENT European Institute for ORL-HNS
-
Главный следователь:
- Andrzej Zarowski, MD
-
-
-
-
Lower Saxony
-
Hanover, Lower Saxony, Германия, 30625
- Medizinische Hochschule Hannover
-
Главный следователь:
- Andreas Büchner, PhD
-
-
-
-
Missouri
-
St Louis, Missouri, Соединенные Штаты, 63141
- Washington University School of Medicine Ear, Nose & Throat Centre
-
Главный следователь:
- Amit Walia, MD
-
-
Utah
-
Salt Lake City, Utah, Соединенные Штаты, 84132
- University of Utah Hospitals and Clinics
-
Главный следователь:
- Craig Buchman, MD
-
-
Критерии участия
Критерии приемлемости
Возраст, подходящий для обучения
Принимает здоровых добровольцев
Полы, имеющие право на обучение
Описание
Inclusion Criteria:
- Aged 18 years or older (no upper age limit).
- Candidate for CI1032 cochlear implantation as per the local sensorineural hearing loss criteria.
- Sufficient proficiency in the language used to conduct the study procedures, as determined by the investigator.
- Availability of pre-op CT scan of the cochlear region meeting quality requirements as determined by the sponsor.
- Willing and able to provide written informed consent and to comply with all requirements of the protocol.
Exclusion Criteria:
- Previous cochlear implantation in the same ear.
- Cochlear implant in the contralateral ear.
- Have an unaided air conduction pure tone average (average of unaided thresholds at 0.5, 1, 2 and 4 kHz) ≤40 dB HL in the contralateral ear measured within the last 12 months
- Severe or greater sensorineural hearing loss prior to 2 years of age, as documented in medical history.
- Deafness due to lesions of the auditory nerve or central auditory pathway.
- Cochlear ossification or anatomical anomaly precluding full electrode array insertion.
- Diagnosis of auditory neuropathy.
- Additional conditions preventing or restricting participation in the audiological evaluations, as determined by the investigator.
- Pregnant or breastfeeding women.
- Unrealistic expectations on the part of the subject regarding the potential benefits, risks, and limitations of the surgical procedure and device.
- Unable or unwilling to comply with the requirements of the clinical investigation as determined by the Investigator.
- Investigator site personnel directly affiliated with this study and/or their immediate families; immediate family is defined as a spouse, parent, child, or sibling.
- Cochlear employees or employees of Contract Research Organisations or contractors engaged by Cochlear for the purposes of this investigation.
- Current participation, or participation in another interventional clinical study/trial in the past 30 days, involving an investigational drug or device (unless determined by the Sponsor to not impact this investigation).
Учебный план
Как устроено исследование?
Детали дизайна
- Основная цель: Диагностика
- Распределение: Н/Д
- Интервенционная модель: Одногрупповое задание
- Маскировка: Нет (открытая этикетка)
Оружие и интервенции
Группа участников / Армия |
Вмешательство/лечение |
|---|---|
|
Экспериментальный: Cochlear implant recipients
Adult participants receiving a commercial Cochlear Nucleus Nexa cochlear implant, who undergo additional investigational objective measure procedures performed intra- and post-operatively. |
Participants will be tested with intra- and post-operative objective measures, including Transimpedance Matrix (TIM) and Neural Response telemetry (NRT) recordings, using monopolar and focused multipolar stimulation. |
Что измеряет исследование?
Первичные показатели результатов
Мера результата |
Мера Описание |
Временное ограничение |
|---|---|---|
|
Correlation Between CT-Derived Electrode-to-Modiolus Distance and MAP T-Levels
Временное ограничение: Postoperative CT assessment and cochlear implant programming assessment at 6 months post-activation
|
The correlation between electrode-to-modiolus distance (in millimeters, mm) derived from postoperative computed tomography (CT) imaging and cochlear implant MAP threshold (T) levels (in current level, CL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between electrode position and MAP T-levels. |
Postoperative CT assessment and cochlear implant programming assessment at 6 months post-activation
|
|
Correlation Between CT-Derived Electrode-to-Modiolus Distance and MAP C-Levels
Временное ограничение: Postoperative CT assessment and cochlear implant programming assessment at 6 months post-activation
|
The correlation between electrode-to-modiolus distance (in millimeters, mm) derived from postoperative CT imaging and cochlear implant MAP maximum comfortable (C) levels (in current level, CL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between electrode position and MAP C-levels. |
Postoperative CT assessment and cochlear implant programming assessment at 6 months post-activation
|
|
Correlation Between CT-Derived Electrode-to-Modiolus Distance and Aided Pure-Tone Thresholds
Временное ограничение: Postoperative CT assessment and aided hearing assessment at 6 months post-activation.
|
The correlation between electrode-to-modiolus distance (in millimeters, mm) derived from postoperative CT imaging and aided pure-tone thresholds (in dBHL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between electrode position and aided hearing thresholds. |
Postoperative CT assessment and aided hearing assessment at 6 months post-activation.
|
|
Correlation Between Monopolar T-NRT Thresholds and MAP T-Levels
Временное ограничение: At 6 months post-activation.
|
The correlation between monopolar telemetry neural response threshold (T-NRT) measurements and cochlear implant MAP threshold (T) levels (in current level, CL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between monopolar T-NRT thresholds and MAP T-levels. |
At 6 months post-activation.
|
|
Correlation Between Monopolar T-NRT Thresholds and MAP C-Levels
Временное ограничение: At 6 months post-activation.
|
The correlation between monopolar telemetry neural response threshold (T-NRT) measurements and cochlear implant MAP maximum comfortable (C) levels (in current level, CL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between monopolar T-NRT thresholds and MAP C-levels. |
At 6 months post-activation.
|
|
Correlation Between Monopolar T-NRT Thresholds and Aided Pure-Tone Thresholds
Временное ограничение: At 6 months post-activation.
|
The correlation between monopolar telemetry neural response threshold (T-NRT) measurements (in current level, CL) and aided pure-tone thresholds (in dBHL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between monopolar T-NRT thresholds and aided hearing thresholds. |
At 6 months post-activation.
|
|
Correlation Between Focused T-NRT Thresholds and MAP T-Levels
Временное ограничение: At 6 months post-activation.
|
The correlation between focused telemetry neural response threshold (T-NRT) measurements (in current level, CL) and cochlear implant MAP threshold (T) levels (in current level, CL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between focused T-NRT thresholds and MAP T-levels. |
At 6 months post-activation.
|
|
Correlation Between Focused T-NRT Thresholds and MAP C-Levels
Временное ограничение: At 6 months post-activation.
|
The correlation between focused telemetry neural response threshold (T-NRT) measurements (in current level, CL) and cochlear implant MAP maximum comfortable (C) levels (in current level, CL) will be assessed. The correlation coefficient with 95% confidence intervals. will be reported to characterize the strength and direction of the association between focused T-NRT thresholds and MAP C-levels. |
At 6 months post-activation.
|
|
Correlation Between Focused T-NRT Thresholds and Aided Pure-Tone Thresholds
Временное ограничение: At 6 months post-activation.
|
The correlation between focused telemetry neural response threshold (T-NRT) measurements (in current level, CL) and aided pure-tone thresholds (in dBHL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between focused T-NRT thresholds and aided hearing thresholds. |
At 6 months post-activation.
|
Вторичные показатели результатов
Мера результата |
Мера Описание |
Временное ограничение |
|---|---|---|
|
Mean Absolute Error Between TIM-Derived and CT-Derived Electrode-to-Modiolus Distance
Временное ограничение: At 6 months post-activation
|
Postoperative electrode-to-modiolus distance for individual electrode contacts will be determined using postoperative computed tomography (CT) imaging and using Transimpedance Matrix (TIM) measurements analysed with an algorithm. The mean absolute error between paired TIM-derived and CT-derived electrode-to-modiolus distance measurements will be reported in millimetres (mm). |
At 6 months post-activation
|
Соавторы и исследователи
Спонсор
Даты записи исследования
Изучение основных дат
Начало исследования (Оцененный)
Первичное завершение (Оцененный)
Завершение исследования (Оцененный)
Даты регистрации исследования
Первый отправленный
Впервые представлено, что соответствует критериям контроля качества
Первый опубликованный (Действительный)
Обновления учебных записей
Последнее опубликованное обновление (Действительный)
Последнее отправленное обновление, отвечающее критериям контроля качества
Последняя проверка
Дополнительная информация
Термины, связанные с этим исследованием
Ключевые слова
Дополнительные соответствующие термины MeSH
Другие идентификационные номера исследования
- AI02841
Планирование данных отдельных участников (IPD)
Планируете делиться данными об отдельных участниках (IPD)?
Информация о лекарствах и устройствах, исследовательские документы
Изучает лекарственный продукт, регулируемый FDA США.
Изучает продукт устройства, регулируемый Управлением по санитарному надзору за качеством пищевых продуктов и медикаментов США.
продукт, произведенный в США и экспортированный из США.
Эта информация была получена непосредственно с веб-сайта clinicaltrials.gov без каких-либо изменений. Если у вас есть запросы на изменение, удаление или обновление сведений об исследовании, обращайтесь по адресу register@clinicaltrials.gov. Как только изменение будет реализовано на clinicaltrials.gov, оно будет автоматически обновлено и на нашем веб-сайте. .