Investigating the Relationship Between Measures of the Hearing Nerve and Speech Perception Outcomes in Adult Nexa Cochlear Implant Recipients (NODE-SDN)
Prospective, Interventional, Multicentre Feasibility Investigation Evaluating Objective Measures and Clinical Outcomes in Adult Nexa® Cochlear Implant Recipients
This study aims to learn more about the information that can be obtained directly from measurements through a Nexa cochlear implant during and after surgery. Researchers will investigate whether these measurements can help understand how the implant is interacting with the inner ear and hearing nerve, where the electrodes are positioned inside the cochlea, and potentially how the implant should be programmed and how well a person detects sounds.
The study will also evaluate how well two methods agree in determining the position of the electrode contacts relative to the anatomy of the inner ear and the hearing nerve. One method uses measurements after surgery made through the implant itself. The other uses medical Computed Tomography(CT) imaging made after the surgery. The information collected will also contribute to a clinical research database that may be used to support future research and development. Participants will be told about this potential future use when they provide consent.
Studie Overzicht
Toestand
Conditie
Interventie / Behandeling
Studietype
Inschrijving (Geschat)
Fase
- Niet toepasbaar
Contacten en locaties
Studiecontact
- Naam: Geert De Ceulaer
- Telefoonnummer: +32486893006
- E-mail: gdeceulaer@cochlear.com
Studie Contact Back-up
- Naam: Loes Beckers
- E-mail: lbeckers@cochlear.com
Studie Locaties
-
-
Victoria
-
East Melbourne, Victoria, Australië, 3002
- The University of Melbourne, Royal Victorian Eye and Ear Hospital
-
Hoofdonderzoeker:
- Stephen O'Leary, MBBS, PhD, FRACS
-
-
-
-
Antwerp
-
Wilrijk, Antwerp, België, 2610
- Sint Augustinus ENT European Institute for ORL-HNS
-
Hoofdonderzoeker:
- Andrzej Zarowski, MD
-
-
-
-
Lower Saxony
-
Hanover, Lower Saxony, Duitsland, 30625
- Medizinische Hochschule Hannover
-
Hoofdonderzoeker:
- Andreas Büchner, PhD
-
-
-
-
Missouri
-
St Louis, Missouri, Verenigde Staten, 63141
- Washington University School of Medicine Ear, Nose & Throat Centre
-
Hoofdonderzoeker:
- Amit Walia, MD
-
-
Utah
-
Salt Lake City, Utah, Verenigde Staten, 84132
- University of Utah Hospitals and Clinics
-
Hoofdonderzoeker:
- Craig Buchman, MD
-
-
Deelname Criteria
Geschiktheidscriteria
Leeftijden die in aanmerking komen voor studie
Accepteert gezonde vrijwilligers
Geslachten die in aanmerking komen voor studie
Beschrijving
Inclusion Criteria:
- Aged 18 years or older (no upper age limit).
- Candidate for CI1032 cochlear implantation as per the local sensorineural hearing loss criteria.
- Sufficient proficiency in the language used to conduct the study procedures, as determined by the investigator.
- Availability of pre-op CT scan of the cochlear region meeting quality requirements as determined by the sponsor.
- Willing and able to provide written informed consent and to comply with all requirements of the protocol.
Exclusion Criteria:
- Previous cochlear implantation in the same ear.
- Cochlear implant in the contralateral ear.
- Have an unaided air conduction pure tone average (average of unaided thresholds at 0.5, 1, 2 and 4 kHz) ≤40 dB HL in the contralateral ear measured within the last 12 months
- Severe or greater sensorineural hearing loss prior to 2 years of age, as documented in medical history.
- Deafness due to lesions of the auditory nerve or central auditory pathway.
- Cochlear ossification or anatomical anomaly precluding full electrode array insertion.
- Diagnosis of auditory neuropathy.
- Additional conditions preventing or restricting participation in the audiological evaluations, as determined by the investigator.
- Pregnant or breastfeeding women.
- Unrealistic expectations on the part of the subject regarding the potential benefits, risks, and limitations of the surgical procedure and device.
- Unable or unwilling to comply with the requirements of the clinical investigation as determined by the Investigator.
- Investigator site personnel directly affiliated with this study and/or their immediate families; immediate family is defined as a spouse, parent, child, or sibling.
- Cochlear employees or employees of Contract Research Organisations or contractors engaged by Cochlear for the purposes of this investigation.
- Current participation, or participation in another interventional clinical study/trial in the past 30 days, involving an investigational drug or device (unless determined by the Sponsor to not impact this investigation).
Studie plan
Hoe is de studie opgezet?
Ontwerpdetails
- Primair doel: Diagnostisch
- Toewijzing: NVT
- Interventioneel model: Opdracht voor een enkele groep
- Masker: Geen (open label)
Wapens en interventies
Deelnemersgroep / Arm |
Interventie / Behandeling |
|---|---|
|
Experimenteel: Cochlear implant recipients
Adult participants receiving a commercial Cochlear Nucleus Nexa cochlear implant, who undergo additional investigational objective measure procedures performed intra- and post-operatively. |
Participants will be tested with intra- and post-operative objective measures, including Transimpedance Matrix (TIM) and Neural Response telemetry (NRT) recordings, using monopolar and focused multipolar stimulation. |
Wat meet het onderzoek?
Primaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
Correlation Between CT-Derived Electrode-to-Modiolus Distance and MAP T-Levels
Tijdsspanne: Postoperative CT assessment and cochlear implant programming assessment at 6 months post-activation
|
The correlation between electrode-to-modiolus distance (in millimeters, mm) derived from postoperative computed tomography (CT) imaging and cochlear implant MAP threshold (T) levels (in current level, CL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between electrode position and MAP T-levels. |
Postoperative CT assessment and cochlear implant programming assessment at 6 months post-activation
|
|
Correlation Between CT-Derived Electrode-to-Modiolus Distance and MAP C-Levels
Tijdsspanne: Postoperative CT assessment and cochlear implant programming assessment at 6 months post-activation
|
The correlation between electrode-to-modiolus distance (in millimeters, mm) derived from postoperative CT imaging and cochlear implant MAP maximum comfortable (C) levels (in current level, CL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between electrode position and MAP C-levels. |
Postoperative CT assessment and cochlear implant programming assessment at 6 months post-activation
|
|
Correlation Between CT-Derived Electrode-to-Modiolus Distance and Aided Pure-Tone Thresholds
Tijdsspanne: Postoperative CT assessment and aided hearing assessment at 6 months post-activation.
|
The correlation between electrode-to-modiolus distance (in millimeters, mm) derived from postoperative CT imaging and aided pure-tone thresholds (in dBHL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between electrode position and aided hearing thresholds. |
Postoperative CT assessment and aided hearing assessment at 6 months post-activation.
|
|
Correlation Between Monopolar T-NRT Thresholds and MAP T-Levels
Tijdsspanne: At 6 months post-activation.
|
The correlation between monopolar telemetry neural response threshold (T-NRT) measurements and cochlear implant MAP threshold (T) levels (in current level, CL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between monopolar T-NRT thresholds and MAP T-levels. |
At 6 months post-activation.
|
|
Correlation Between Monopolar T-NRT Thresholds and MAP C-Levels
Tijdsspanne: At 6 months post-activation.
|
The correlation between monopolar telemetry neural response threshold (T-NRT) measurements and cochlear implant MAP maximum comfortable (C) levels (in current level, CL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between monopolar T-NRT thresholds and MAP C-levels. |
At 6 months post-activation.
|
|
Correlation Between Monopolar T-NRT Thresholds and Aided Pure-Tone Thresholds
Tijdsspanne: At 6 months post-activation.
|
The correlation between monopolar telemetry neural response threshold (T-NRT) measurements (in current level, CL) and aided pure-tone thresholds (in dBHL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between monopolar T-NRT thresholds and aided hearing thresholds. |
At 6 months post-activation.
|
|
Correlation Between Focused T-NRT Thresholds and MAP T-Levels
Tijdsspanne: At 6 months post-activation.
|
The correlation between focused telemetry neural response threshold (T-NRT) measurements (in current level, CL) and cochlear implant MAP threshold (T) levels (in current level, CL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between focused T-NRT thresholds and MAP T-levels. |
At 6 months post-activation.
|
|
Correlation Between Focused T-NRT Thresholds and MAP C-Levels
Tijdsspanne: At 6 months post-activation.
|
The correlation between focused telemetry neural response threshold (T-NRT) measurements (in current level, CL) and cochlear implant MAP maximum comfortable (C) levels (in current level, CL) will be assessed. The correlation coefficient with 95% confidence intervals. will be reported to characterize the strength and direction of the association between focused T-NRT thresholds and MAP C-levels. |
At 6 months post-activation.
|
|
Correlation Between Focused T-NRT Thresholds and Aided Pure-Tone Thresholds
Tijdsspanne: At 6 months post-activation.
|
The correlation between focused telemetry neural response threshold (T-NRT) measurements (in current level, CL) and aided pure-tone thresholds (in dBHL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between focused T-NRT thresholds and aided hearing thresholds. |
At 6 months post-activation.
|
Secundaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
Mean Absolute Error Between TIM-Derived and CT-Derived Electrode-to-Modiolus Distance
Tijdsspanne: At 6 months post-activation
|
Postoperative electrode-to-modiolus distance for individual electrode contacts will be determined using postoperative computed tomography (CT) imaging and using Transimpedance Matrix (TIM) measurements analysed with an algorithm. The mean absolute error between paired TIM-derived and CT-derived electrode-to-modiolus distance measurements will be reported in millimetres (mm). |
At 6 months post-activation
|
Medewerkers en onderzoekers
Sponsor
Studie record data
Bestudeer belangrijke data
Studie start (Geschat)
Primaire voltooiing (Geschat)
Studie voltooiing (Geschat)
Studieregistratiedata
Eerst ingediend
Eerst ingediend dat voldeed aan de QC-criteria
Eerst geplaatst (Werkelijk)
Updates van studierecords
Laatste update geplaatst (Werkelijk)
Laatste update ingediend die voldeed aan QC-criteria
Laatst geverifieerd
Meer informatie
Termen gerelateerd aan deze studie
Trefwoorden
Aanvullende relevante MeSH-voorwaarden
Andere studie-ID-nummers
- AI02841
Plan Individuele Deelnemersgegevens (IPD)
Bent u van plan om gegevens van individuele deelnemers (IPD) te delen?
Informatie over medicijnen en apparaten, studiedocumenten
Bestudeert een door de Amerikaanse FDA gereguleerd geneesmiddel
Bestudeert een door de Amerikaanse FDA gereguleerd apparaatproduct
product vervaardigd in en geëxporteerd uit de V.S.
Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .