Investigating the Relationship Between Measures of the Hearing Nerve and Speech Perception Outcomes in Adult Nexa Cochlear Implant Recipients (NODE-SDN)
Prospective, Interventional, Multicentre Feasibility Investigation Evaluating Objective Measures and Clinical Outcomes in Adult Nexa® Cochlear Implant Recipients
This study aims to learn more about the information that can be obtained directly from measurements through a Nexa cochlear implant during and after surgery. Researchers will investigate whether these measurements can help understand how the implant is interacting with the inner ear and hearing nerve, where the electrodes are positioned inside the cochlea, and potentially how the implant should be programmed and how well a person detects sounds.
The study will also evaluate how well two methods agree in determining the position of the electrode contacts relative to the anatomy of the inner ear and the hearing nerve. One method uses measurements after surgery made through the implant itself. The other uses medical Computed Tomography(CT) imaging made after the surgery. The information collected will also contribute to a clinical research database that may be used to support future research and development. Participants will be told about this potential future use when they provide consent.
Studieoversigt
Status
Betingelser
Intervention / Behandling
Undersøgelsestype
Tilmelding (Anslået)
Fase
- Ikke anvendelig
Kontakter og lokationer
Studiekontakt
- Navn: Geert De Ceulaer
- Telefonnummer: +32486893006
- E-mail: gdeceulaer@cochlear.com
Undersøgelse Kontakt Backup
- Navn: Loes Beckers
- E-mail: lbeckers@cochlear.com
Studiesteder
-
-
Victoria
-
East Melbourne, Victoria, Australien, 3002
- The University of Melbourne, Royal Victorian Eye and Ear Hospital
-
Ledende efterforsker:
- Stephen O'Leary, MBBS, PhD, FRACS
-
-
-
-
Antwerp
-
Wilrijk, Antwerp, Belgien, 2610
- Sint Augustinus ENT European Institute for ORL-HNS
-
Ledende efterforsker:
- Andrzej Zarowski, MD
-
-
-
-
Missouri
-
St Louis, Missouri, Forenede Stater, 63141
- Washington University School of Medicine Ear, Nose & Throat Centre
-
Ledende efterforsker:
- Amit Walia, MD
-
-
Utah
-
Salt Lake City, Utah, Forenede Stater, 84132
- University of Utah Hospitals and Clinics
-
Ledende efterforsker:
- Craig Buchman, MD
-
-
-
-
Lower Saxony
-
Hanover, Lower Saxony, Tyskland, 30625
- Medizinische Hochschule Hannover
-
Ledende efterforsker:
- Andreas Büchner, PhD
-
-
Deltagelseskriterier
Berettigelseskriterier
Aldre berettiget til at studere
Tager imod sunde frivillige
Køn, der er berettiget til at studere
Beskrivelse
Inclusion Criteria:
- Aged 18 years or older (no upper age limit).
- Candidate for CI1032 cochlear implantation as per the local sensorineural hearing loss criteria.
- Sufficient proficiency in the language used to conduct the study procedures, as determined by the investigator.
- Availability of pre-op CT scan of the cochlear region meeting quality requirements as determined by the sponsor.
- Willing and able to provide written informed consent and to comply with all requirements of the protocol.
Exclusion Criteria:
- Previous cochlear implantation in the same ear.
- Cochlear implant in the contralateral ear.
- Have an unaided air conduction pure tone average (average of unaided thresholds at 0.5, 1, 2 and 4 kHz) ≤40 dB HL in the contralateral ear measured within the last 12 months
- Severe or greater sensorineural hearing loss prior to 2 years of age, as documented in medical history.
- Deafness due to lesions of the auditory nerve or central auditory pathway.
- Cochlear ossification or anatomical anomaly precluding full electrode array insertion.
- Diagnosis of auditory neuropathy.
- Additional conditions preventing or restricting participation in the audiological evaluations, as determined by the investigator.
- Pregnant or breastfeeding women.
- Unrealistic expectations on the part of the subject regarding the potential benefits, risks, and limitations of the surgical procedure and device.
- Unable or unwilling to comply with the requirements of the clinical investigation as determined by the Investigator.
- Investigator site personnel directly affiliated with this study and/or their immediate families; immediate family is defined as a spouse, parent, child, or sibling.
- Cochlear employees or employees of Contract Research Organisations or contractors engaged by Cochlear for the purposes of this investigation.
- Current participation, or participation in another interventional clinical study/trial in the past 30 days, involving an investigational drug or device (unless determined by the Sponsor to not impact this investigation).
Studieplan
Hvordan er undersøgelsen tilrettelagt?
Design detaljer
- Primært formål: Diagnostisk
- Tildeling: N/A
- Interventionel model: Enkelt gruppeopgave
- Maskning: Ingen (Åben etiket)
Våben og indgreb
Deltagergruppe / Arm |
Intervention / Behandling |
|---|---|
|
Eksperimentel: Cochlear implant recipients
Adult participants receiving a commercial Cochlear Nucleus Nexa cochlear implant, who undergo additional investigational objective measure procedures performed intra- and post-operatively. |
Participants will be tested with intra- and post-operative objective measures, including Transimpedance Matrix (TIM) and Neural Response telemetry (NRT) recordings, using monopolar and focused multipolar stimulation. |
Hvad måler undersøgelsen?
Primære resultatmål
Resultatmål |
Foranstaltningsbeskrivelse |
Tidsramme |
|---|---|---|
|
Correlation Between CT-Derived Electrode-to-Modiolus Distance and MAP T-Levels
Tidsramme: Postoperative CT assessment and cochlear implant programming assessment at 6 months post-activation
|
The correlation between electrode-to-modiolus distance (in millimeters, mm) derived from postoperative computed tomography (CT) imaging and cochlear implant MAP threshold (T) levels (in current level, CL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between electrode position and MAP T-levels. |
Postoperative CT assessment and cochlear implant programming assessment at 6 months post-activation
|
|
Correlation Between CT-Derived Electrode-to-Modiolus Distance and MAP C-Levels
Tidsramme: Postoperative CT assessment and cochlear implant programming assessment at 6 months post-activation
|
The correlation between electrode-to-modiolus distance (in millimeters, mm) derived from postoperative CT imaging and cochlear implant MAP maximum comfortable (C) levels (in current level, CL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between electrode position and MAP C-levels. |
Postoperative CT assessment and cochlear implant programming assessment at 6 months post-activation
|
|
Correlation Between CT-Derived Electrode-to-Modiolus Distance and Aided Pure-Tone Thresholds
Tidsramme: Postoperative CT assessment and aided hearing assessment at 6 months post-activation.
|
The correlation between electrode-to-modiolus distance (in millimeters, mm) derived from postoperative CT imaging and aided pure-tone thresholds (in dBHL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between electrode position and aided hearing thresholds. |
Postoperative CT assessment and aided hearing assessment at 6 months post-activation.
|
|
Correlation Between Monopolar T-NRT Thresholds and MAP T-Levels
Tidsramme: At 6 months post-activation.
|
The correlation between monopolar telemetry neural response threshold (T-NRT) measurements and cochlear implant MAP threshold (T) levels (in current level, CL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between monopolar T-NRT thresholds and MAP T-levels. |
At 6 months post-activation.
|
|
Correlation Between Monopolar T-NRT Thresholds and MAP C-Levels
Tidsramme: At 6 months post-activation.
|
The correlation between monopolar telemetry neural response threshold (T-NRT) measurements and cochlear implant MAP maximum comfortable (C) levels (in current level, CL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between monopolar T-NRT thresholds and MAP C-levels. |
At 6 months post-activation.
|
|
Correlation Between Monopolar T-NRT Thresholds and Aided Pure-Tone Thresholds
Tidsramme: At 6 months post-activation.
|
The correlation between monopolar telemetry neural response threshold (T-NRT) measurements (in current level, CL) and aided pure-tone thresholds (in dBHL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between monopolar T-NRT thresholds and aided hearing thresholds. |
At 6 months post-activation.
|
|
Correlation Between Focused T-NRT Thresholds and MAP T-Levels
Tidsramme: At 6 months post-activation.
|
The correlation between focused telemetry neural response threshold (T-NRT) measurements (in current level, CL) and cochlear implant MAP threshold (T) levels (in current level, CL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between focused T-NRT thresholds and MAP T-levels. |
At 6 months post-activation.
|
|
Correlation Between Focused T-NRT Thresholds and MAP C-Levels
Tidsramme: At 6 months post-activation.
|
The correlation between focused telemetry neural response threshold (T-NRT) measurements (in current level, CL) and cochlear implant MAP maximum comfortable (C) levels (in current level, CL) will be assessed. The correlation coefficient with 95% confidence intervals. will be reported to characterize the strength and direction of the association between focused T-NRT thresholds and MAP C-levels. |
At 6 months post-activation.
|
|
Correlation Between Focused T-NRT Thresholds and Aided Pure-Tone Thresholds
Tidsramme: At 6 months post-activation.
|
The correlation between focused telemetry neural response threshold (T-NRT) measurements (in current level, CL) and aided pure-tone thresholds (in dBHL) will be assessed. The correlation coefficient with 95% confidence intervals will be reported to characterize the strength and direction of the association between focused T-NRT thresholds and aided hearing thresholds. |
At 6 months post-activation.
|
Sekundære resultatmål
Resultatmål |
Foranstaltningsbeskrivelse |
Tidsramme |
|---|---|---|
|
Mean Absolute Error Between TIM-Derived and CT-Derived Electrode-to-Modiolus Distance
Tidsramme: At 6 months post-activation
|
Postoperative electrode-to-modiolus distance for individual electrode contacts will be determined using postoperative computed tomography (CT) imaging and using Transimpedance Matrix (TIM) measurements analysed with an algorithm. The mean absolute error between paired TIM-derived and CT-derived electrode-to-modiolus distance measurements will be reported in millimetres (mm). |
At 6 months post-activation
|
Samarbejdspartnere og efterforskere
Sponsor
Datoer for undersøgelser
Studer store datoer
Studiestart (Anslået)
Primær færdiggørelse (Anslået)
Studieafslutning (Anslået)
Datoer for studieregistrering
Først indsendt
Først indsendt, der opfyldte QC-kriterier
Først opslået (Faktiske)
Opdateringer af undersøgelsesjournaler
Sidste opdatering sendt (Faktiske)
Sidste opdatering indsendt, der opfyldte kvalitetskontrolkriterier
Sidst verificeret
Mere information
Begreber relateret til denne undersøgelse
Nøgleord
Yderligere relevante MeSH-vilkår
Andre undersøgelses-id-numre
- AI02841
Plan for individuelle deltagerdata (IPD)
Planlægger du at dele individuelle deltagerdata (IPD)?
Lægemiddel- og udstyrsoplysninger, undersøgelsesdokumenter
Studerer et amerikansk FDA-reguleret lægemiddelprodukt
Studerer et amerikansk FDA-reguleret enhedsprodukt
produkt fremstillet i og eksporteret fra U.S.A.
Disse oplysninger blev hentet direkte fra webstedet clinicaltrials.gov uden ændringer. Hvis du har nogen anmodninger om at ændre, fjerne eller opdatere dine undersøgelsesoplysninger, bedes du kontakte register@clinicaltrials.gov. Så snart en ændring er implementeret på clinicaltrials.gov, vil denne også blive opdateret automatisk på vores hjemmeside .