Real-time Diagnosis of Diminutive Colorectal Polyps Using AI (COMET-OPTICAL)
Real Time Computer-aided Diagnosis (CADx) of Diminutive Colorectal Polyps Using Artificial Intelligence
Studie Overzicht
Toestand
Interventie / Behandeling
Gedetailleerde beschrijving
Correct endoscopic prediction of the histopathology and differentiation between benign, pre-malignant, and malignant colorectal polyps (optical diagnosis) remains difficult. Despite additional training, even experienced endoscopists continue to fail meeting international thresholds set for safe implementation of treatment strategies based on optical diagnosis.
Multiple machine learning techniques - computer-aided diagnosis (CADx) systems - have been developed for applications in medical imaging within colonoscopy and can improve endoscopic classification of colorectal polyps.
Aim of this study is to explore the feasibility of the workflow using AI4CRP (a CNN based CADx system) real-time in the endoscopy suite, and to investigate the real-time diagnostic performance of AI4CRP for the diagnosis of diminutive (<5mm) colorectal polyps. Secondary, the real-time performance of commercially available CADx systems will be investigated and compared with AI4CRP performance.
Studietype
Inschrijving (Verwacht)
Contacten en locaties
Studiecontact
- Naam: Quirine van der Zander, Drs MD
- Telefoonnummer: 031433882241
- E-mail: q.vanderzander@maastrichtuniversity.nl
Studie Contact Back-up
- Naam: Erik Schoon, Prof Dr MD
- Telefoonnummer: 031433882241
- E-mail: erik.schoon@catharinaziekenhuis.nl
Studie Locaties
-
-
Limburg
-
Maastricht, Limburg, Nederland, 6202AZ
- Werving
- Maastricht University Medical Center
-
Contact:
- Quirine van der Zander, Drs MD
- Telefoonnummer: 031433882241
- E-mail: q.vanderzander@maastrichtuniversity.nl
-
Contact:
- Erik Schoon, Prof Dr MD
- Telefoonnummer: 031433882241
- E-mail: erik.schoon@catharinaziekenhuis.nl
-
-
Noord-Brabant
-
Eindhoven, Noord-Brabant, Nederland, 5623 EJ
- Voltooid
- Catharina Ziekenhuis Eindhoven
-
-
Deelname Criteria
Geschiktheidscriteria
Leeftijden die in aanmerking komen voor studie
Accepteert gezonde vrijwilligers
Geslachten die in aanmerking komen voor studie
Bemonsteringsmethode
Studie Bevolking
Beschrijving
Inclusion Criteria:
- Age >18 years;
- Patients with at least one colorectal polyps encountered during colonoscopy;
- Patients referred for a colonoscopy by the Dutch bowel cancer screening program, patients undergoing a colonoscopy for endoscopic surveillance, or patients undergoing a colonoscopy because of complaints;
- Written informed consent.
Exclusion Criteria:
- Patients with prior history of inflammatory bowel diseases (IBD) or polyposis syndromes;
- Patients with inadequate bowel preparations after adequate washing, suctioning, and cleaning manoeuvres have been performed by the endoscopist;
- Patients undergoing an emergency colonoscopy;
- Written objection in the patient file for participation in scientific research.
Studie plan
Hoe is de studie opgezet?
Ontwerpdetails
Cohorten en interventies
Groep / Cohort |
Interventie / Behandeling |
|---|---|
|
Gastroenterology patients
Patient receiving a colonoscopy because of regular care will be considered eligible for inclusion if at least one diminutive colorectal polyp is encountered during the colonoscopy. Patients receive an endoscopic procedure in the context of the Dutch national screening program, because of gastrointestinal symptoms, or because of follow-up of previously diagnosed bowel diseases. Colonoscopies will be executed using Fujifilm endoscopy systems (Fujifilm® Corporation, Tokyo, Japan), using Pentax endoscopy systems (Pentax Medical®, Hamburg, Germany), and using Olympus endoscopy systems (Olympus®, Tokyo, Japan). |
Andere namen:
|
Wat meet het onderzoek?
Primaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
Technical feasibility of real-time use of AI4CRP.
Tijdsspanne: 6 months
|
The technical feasibility of real-time use of AI4CRP in the endoscopy suite regarding a proper reception of the video output from the local endoscopy processor towards AI4CRP (in high definition quality, without any delays in time).
|
6 months
|
|
User interface feasibility of real-time use of AI4CRP.
Tijdsspanne: 6 months
|
The user interface feasibility of real-time use of AI4CRP in the endoscopy suite regarding a correct alignment of the user interface of AI4CRP with the video output from the local endoscopy system (resizing image pixels and anonymization).
|
6 months
|
|
The diagnostic accuracy of AI4CRP per image modality (HDWL, BLI, LCI, i-scan).
Tijdsspanne: 1 year
|
The real-time diagnostic accuracy of AI4CRP per image modality (HDWL, BLI, LCI, i-scan).
Diagnostic accuracy defined as the percentage of correctly optically diagnosed colorectal polyps.
|
1 year
|
|
The sensitivity of AI4CRP per image modality (HDWL, BLI, LCI, i-scan).
Tijdsspanne: 1 year
|
The real-time sensitivity of AI4CRP per image modality (HDWL, BLI, LCI, i-scan).
|
1 year
|
|
The specificity of AI4CRP per image modality (HDWL, BLI, LCI, i-scan).
Tijdsspanne: 1 year
|
The real-time specificity of AI4CRP per image modality (HDWL, BLI, LCI, i-scan).
|
1 year
|
|
The negative predictive value of AI4CRP per image modality (HDWL, BLI, LCI, i-scan).
Tijdsspanne: 1 year
|
The real-time negative predictive value of AI4CRP per image modality (HDWL, BLI, LCI, i-scan).
|
1 year
|
|
The positive predictive value of AI4CRP per image modality (HDWL, BLI, LCI, i-scan).
Tijdsspanne: 1 year
|
The real-time positive predictive value of AI4CRP per image modality (HDWL, BLI, LCI, i-scan).
|
1 year
|
|
The Area Under ROC Curve (AUC) of AI4CRP per image modality (HDWL, BLI, LCI, i-scan).
Tijdsspanne: 1 year
|
The real-time Area Under ROC Curve (AUC) of AI4CRP per image modality (HDWL, BLI, LCI, i-scan).
|
1 year
|
Secundaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
The diagnostic accuracy of AI4CRP per polyp.
Tijdsspanne: 1 year
|
The real-time diagnostic accuracy of AI4CRP per polyp (comprising the combination of different imaging modalities).
|
1 year
|
|
The sensitivity of AI4CRP per polyp.
Tijdsspanne: 1 year
|
The real-time sensitivity of AI4CRP per polyp (comprising the combination of different imaging modalities).
|
1 year
|
|
The specificity of AI4CRP per polyp.
Tijdsspanne: 1 year
|
The real-time specificity of AI4CRP per polyp (comprising the combination of different imaging modalities).
|
1 year
|
|
The negative predictive value of AI4CRP per polyp.
Tijdsspanne: 1 year
|
The real-time negative predictive value of AI4CRP per polyp (comprising the combination of different imaging modalities).
|
1 year
|
|
The positive predictive value of AI4CRP per polyp.
Tijdsspanne: 1 year
|
The real-time positive predictive value of AI4CRP per polyp (comprising the combination of different imaging modalities).
|
1 year
|
|
The Area Under ROC Curve (AUC) of AI4CRP per polyp.
Tijdsspanne: 1 year
|
The real-time Area Under ROC Curve (AUC) of AI4CRP per polyp (comprising the combination of different imaging modalities).
|
1 year
|
|
The diagnostic accuracy of CAD EYE in BLI mode, per polyp.
Tijdsspanne: 1 year
|
The real-time diagnostic accuracy of CAD EYE in BLI mode, per polyp.
|
1 year
|
|
The sensitivity of CAD EYE in BLI mode, per polyp.
Tijdsspanne: 1 year
|
The real-time sensitivity of CAD EYE in BLI mode, per polyp.
|
1 year
|
|
The specificity of CAD EYE in BLI mode, per polyp.
Tijdsspanne: 1 year
|
The real-time specificity of CAD EYE in BLI mode, per polyp.
|
1 year
|
|
The negative predictive value of CAD EYE in BLI mode, per polyp.
Tijdsspanne: 1 year
|
The real-time negative predictive value of CAD EYE in BLI mode, per polyp.
|
1 year
|
|
The positive predictive value of CAD EYE in BLI mode, per polyp.
Tijdsspanne: 1 year
|
The real-time positive predictive value of CAD EYE in BLI mode, per polyp.
|
1 year
|
|
The Area Under ROC Curve (AUC) of CAD EYE in BLI mode, per polyp.
Tijdsspanne: 1 year
|
The real-time Area Under ROC Curve (AUC) of CAD EYE in BLI mode, per polyp.
|
1 year
|
|
The diagnostic accuracy of AI4CRP per patient.
Tijdsspanne: 1 year
|
The real-time diagnostic accuracy of AI4CRP per patient (in case of multiple polyps per patient).
|
1 year
|
|
The diagnostic accuracy of CAD EYE per patient.
Tijdsspanne: 1 year
|
The real-time diagnostic accuracy of CAD EYE per patient (in case of multiple polyps per patient).
|
1 year
|
|
The localization score of AI4CRP.
Tijdsspanne: 1 year
|
The localization score of AI4CRP regarding the number of images in which the heatmap produced by AI4CRP pointed out the area of interest (scale: correct, incorrect, or partly correct area of interest).
|
1 year
|
|
The difference in diagnostic accuracy of endoscopists per polyp before and after AI.
Tijdsspanne: 1 year
|
The difference in real-time diagnostic accuracy of endoscopists per polyp before and after AI.
|
1 year
|
|
The difference in sensitivity of endoscopists per polyp before and after AI.
Tijdsspanne: 1 year
|
The difference in real-time sensitivity of endoscopists per polyp before and after AI.
|
1 year
|
|
The difference in specificity of endoscopists per polyp before and after AI.
Tijdsspanne: 1 year
|
The difference in real-time specificity of endoscopists per polyp before and after AI.
|
1 year
|
|
The difference in negative predictive value of endoscopists per polyp before and after AI.
Tijdsspanne: 1 year
|
The difference in real-time negative predictive value of endoscopists per polyp before and after AI.
|
1 year
|
|
The difference in positive predictive value of endoscopists per polyp before and after AI.
Tijdsspanne: 1 year
|
The difference in real-time positive predictive value of endoscopists per polyp before and after AI.
|
1 year
|
|
The agreement in surveillance interval based on optical diagnosis and histopathology.
Tijdsspanne: 1 year
|
The agreement in surveillance interval based on optical diagnosis of diminutive colorectal polyps and histopathology of small and large colorectal polyps, compared to the surveillance interval based on histopathology of all colorectal polyps (diminutive, small, and large).
|
1 year
|
Medewerkers en onderzoekers
Onderzoekers
- Hoofdonderzoeker: Erik Schoon, Prof Dr MD, Maastricht Universitair Medisch Centrum
Studie record data
Bestudeer belangrijke data
Studie start (Werkelijk)
Primaire voltooiing (Verwacht)
Studie voltooiing (Verwacht)
Studieregistratiedata
Eerst ingediend
Eerst ingediend dat voldeed aan de QC-criteria
Eerst geplaatst (Werkelijk)
Updates van studierecords
Laatste update geplaatst (Werkelijk)
Laatste update ingediend die voldeed aan QC-criteria
Laatst geverifieerd
Meer informatie
Termen gerelateerd aan deze studie
Aanvullende relevante MeSH-voorwaarden
- Ziekten van het spijsverteringsstelsel
- Neoplasmata
- Neoplasmata per site
- Gastro-intestinale neoplasmata
- Neoplasmata van het spijsverteringsstelsel
- Gastro-intestinale aandoeningen
- Colon Ziekten
- Darmziekten
- Pathologische aandoeningen, anatomisch
- Intestinale neoplasmata
- Rectale ziekten
- Colorectale neoplasmata
- Poliepen
Andere studie-ID-nummers
- METC2021-3036
Plan Individuele Deelnemersgegevens (IPD)
Bent u van plan om gegevens van individuele deelnemers (IPD) te delen?
Beschrijving IPD-plan
Informatie over medicijnen en apparaten, studiedocumenten
Bestudeert een door de Amerikaanse FDA gereguleerd geneesmiddel
Bestudeert een door de Amerikaanse FDA gereguleerd apparaatproduct
Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .