Early Feasibility Study of 3930M Defibrillation Lead Implantation Via the Middle Cardiac Vein in Patients With Tricuspid Valve Regurgitation (MCV3930 EFS)

3 september 2026 bijgewerkt door: Minneapolis Heart Institute Foundation

Early Feasibility Study of Medtronic OmniaSecure Model 3930M Defibrillation Lead Implantation Via the Middle Cardiac Vein in Patients With Tricuspid Valve Regurgitation: MCV3930 EFS

Prospective, single center, early feasibility study designed to evaluate the initial safety and device performance of the Medtronic 3930M Lead when implanted via the Middle Cardiac Vein in patients with Tricuspid Valve Regurgitation

Studie Overzicht

Toestand

Nog niet aan het werven

Interventie / Behandeling

Studietype

Ingrijpend

Inschrijving (Geschat)

5

Fase

  • Niet toepasbaar

Contacten en locaties

In dit gedeelte vindt u de contactgegevens van degenen die het onderzoek uitvoeren en informatie over waar dit onderzoek wordt uitgevoerd.

Studiecontact

Studie Contact Back-up

Studie Locaties

    • Minnesota
      • Minneapolis, Minnesota, Verenigde Staten, 55407
        • Minneapolis Heart Institute Foundation
        • Contact:
        • Contact:
        • Hoofdonderzoeker:
          • John S Zakaib, MD

Deelname Criteria

Onderzoekers zoeken naar mensen die aan een bepaalde beschrijving voldoen, de zogenaamde geschiktheidscriteria. Enkele voorbeelden van deze criteria zijn iemands algemene gezondheidstoestand of eerdere behandelingen.

Geschiktheidscriteria

Leeftijden die in aanmerking komen voor studie

  • Volwassen
  • Oudere volwassene

Accepteert gezonde vrijwilligers

Nee

Beschrijving

Inclusion Criteria:

  1. Subject meets current clinical practice guidelines for implantation of ICD or CRT-D and will undergo one of the following:

    • Medtronic CRT-D system implant
    • Medtronic ICD system implant (single or dual chamber)
  2. Anatomical suitability for middle cardiac vein lead placement based on pre procedural imaging.
  3. Subject meets both of the following:

    • At least Moderate Tricuspid Valve Regurgitation documented by cardiac imaging
    • A clinical goal to avoid worsening tricuspid valve regurgitation as determined by the treating physician and/or the patient
  4. Age 18 years or older.
  5. Subject is willing to undergo implant defibrillation testing
  6. Signed Patient Informed Consent Form (ICF).

Exclusion Criteria:

  1. Life expectancy < 6 months due to non-cardiac comorbidities or advanced frailty where the procedure is not aligned with goals of care
  2. Contraindication to transvenous lead implantation, including:

    • Severe or occlusive disease of the axillary/subclavian venous system
    • Coronary sinus or middle cardiac vein occlusion, severe stenosis, or anatomy prohibiting safe lead delivery
    • Known hypercoagulable state or bleeding disorder is not manageable with standard therapy.
  3. Subject is contraindicated for ≤1 mg dexamethasone acetate
  4. Patients with a medical condition that precludes them from undergoing defibrillation testing:

    • Pre-existing or suspected pneumothorax
    • Severe aortic stenosis
    • Current Intracardiac LA or LV thrombus
    • Severe proximal three-vessel or left main coronary artery disease without revascularization
    • Hemodynamic instability
    • Unstable angina
    • Recent stroke or transient ischemic attack (within the last 6 months)
    • Known inadequate external defibrillation
    • LVEDD >70 mm
    • Any other known medical condition not listed that precludes their participation in the opinion of the Investigator.
  5. Patient with evidence of active infection or undergoing treatment for an infection.
  6. Inability to provide informed consent, including:

    • Cognitive impairment without a legally authorized representative
    • Language or communication barriers that prevent adequate understanding of the study
  7. Unable to follow through study protocol for any reasons (social support, distance from research center, insurance denial, mental illness etc.) in the investigator's judgement
  8. Pregnant women or breastfeeding women, or women of childbearing potential and who are not on a reliable form of birth regulation method or abstinence
  9. End-Stage Renal Disease (ESRD) on dialysis
  10. NYHA IV
  11. Patient participation in another study that precludes concomitant study participation

Studie plan

Dit gedeelte bevat details van het studieplan, inclusief hoe de studie is opgezet en wat de studie meet.

Hoe is de studie opgezet?

Ontwerpdetails

  • Primair doel: Toestel Haalbaarheid
  • Toewijzing: NVT
  • Interventioneel model: Opdracht voor een enkele groep
  • Masker: Geen (open label)

Wapens en interventies

Deelnemersgroep / Arm
Interventie / Behandeling
Experimenteel: Interventional
Patients will undergo implant of the Medtronic 3930M Defibrillation lead via the Middle Cardiac Vein and undergo required electrical testing
Required electrical testing for pacing, sensing, impedance, and defibrillation will be completed

Wat meet het onderzoek?

Primaire uitkomstmaten

Uitkomstmaat
Maatregel Beschrijving
Tijdsspanne
Acute Procedural Success
Tijdsspanne: Index Procedure

Acute Procedural Success, defined as successful implantation of the investigational lead with all the following:

  1. Successful lead delivery and deployment at the intended MCV location within 4 deployment attempts
  2. Stable lead position at completion of the procedure
  3. Acceptable electrical performance:

    • Capture threshold: ≤ 2.0 V at 0.4 ms
    • Sensing amplitude: ≥ 4.0 mV
    • Impedance: 200-1500 Ω
  4. No device- or procedure-related Major Adverse Cardiovascular Events (MACE) through hospital discharge
  5. Absence of T-wave Oversensing (TWOS)
  6. Absence of P-wave Oversensing (PWOS)
  7. Absence of Myopotential Oversensing
  8. Successful Defibrillation Test (DFT) with 10J safety margin
  9. Absence of any serious adverse events related to the MCV implantation site.
Index Procedure

Secundaire uitkomstmaten

Uitkomstmaat
Maatregel Beschrijving
Tijdsspanne
Lead Electrical Performance and Stability through 90 Days
Tijdsspanne: Up to 90 days following implant

Evaluation of MCV lead electrical performance and stability through 90 days following implantation, including:

  • Stability of electrical performance (e.g., pacing threshold, sensing, impedance)
  • Lead stability (e.g., no dislodgement or migration)
  • Need for lead revision or system modification
Up to 90 days following implant
Incidence of Device- or Procedure-Related Complications
Tijdsspanne: Up to 90 days following implant

Number of participants experiencing device- or procedure-related complications through 90 days following implantation:

1. Incidence of device- or procedure-related major adverse events including, but not limited to:

  • Death (all-cause or procedure-related)
  • Cardiac perforation, pericardial effusion, or cardiac tamponade
  • Clinically significant arrhythmia requiring intervention
  • Device-related infection requiring intervention
  • Lead dislodgement requiring intervention
  • Coronary sinus or venous injury
  • Clinically significant phrenic nerve stimulation
  • System revision or extraction
Up to 90 days following implant
Freedom from Device-or Procedure-Related Serious Adverse Events
Tijdsspanne: Up to 90 days following implant
Proportion of participants free from device-or procedure-related serious adverse events through 90 days following implantation.
Up to 90 days following implant

Andere uitkomstmaten

Uitkomstmaat
Maatregel Beschrijving
Tijdsspanne
Tricuspid Valve Parameters
Tijdsspanne: 6 Month Post Implant
Tricuspid Valve Parameters: Tricuspid Valve function, including Tricuspid Valve Regurgitation post implant will be assessed at 6 months post implant or per TTVI protocol
6 Month Post Implant

Medewerkers en onderzoekers

Hier vindt u mensen en organisaties die betrokken zijn bij dit onderzoek.

Medewerkers

Onderzoekers

  • Hoofdonderzoeker: John S Zakaib, MD, Minneapolis Heart Institute Foundation

Studie record data

Deze datums volgen de voortgang van het onderzoeksdossier en de samenvatting van de ingediende resultaten bij ClinicalTrials.gov. Studieverslagen en gerapporteerde resultaten worden beoordeeld door de National Library of Medicine (NLM) om er zeker van te zijn dat ze voldoen aan specifieke kwaliteitscontrolenormen voordat ze op de openbare website worden geplaatst.

Bestudeer belangrijke data

Studie start (Geschat)

1 november 2026

Primaire voltooiing (Geschat)

1 januari 2028

Studie voltooiing (Geschat)

1 januari 2028

Studieregistratiedata

Eerst ingediend

27 augustus 2026

Eerst ingediend dat voldeed aan de QC-criteria

3 september 2026

Eerst geplaatst (Werkelijk)

10 september 2026

Updates van studierecords

Laatste update geplaatst (Werkelijk)

10 september 2026

Laatste update ingediend die voldeed aan QC-criteria

3 september 2026

Laatst geverifieerd

1 september 2026

Meer informatie

Termen gerelateerd aan deze studie

Informatie over medicijnen en apparaten, studiedocumenten

Bestudeert een door de Amerikaanse FDA gereguleerd geneesmiddel

Nee

Bestudeert een door de Amerikaanse FDA gereguleerd apparaatproduct

Ja

Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .