Early Feasibility Study of 3930M Defibrillation Lead Implantation Via the Middle Cardiac Vein in Patients With Tricuspid Valve Regurgitation (MCV3930 EFS)

3 settembre 2026 aggiornato da: Minneapolis Heart Institute Foundation

Early Feasibility Study of Medtronic OmniaSecure Model 3930M Defibrillation Lead Implantation Via the Middle Cardiac Vein in Patients With Tricuspid Valve Regurgitation: MCV3930 EFS

Prospective, single center, early feasibility study designed to evaluate the initial safety and device performance of the Medtronic 3930M Lead when implanted via the Middle Cardiac Vein in patients with Tricuspid Valve Regurgitation

Panoramica dello studio

Stato

Non ancora reclutamento

Intervento / Trattamento

Tipo di studio

Interventistico

Iscrizione (Stimato)

5

Fase

  • Non applicabile

Contatti e Sedi

Questa sezione fornisce i recapiti di coloro che conducono lo studio e informazioni su dove viene condotto lo studio.

Contatto studio

Backup dei contatti dello studio

Luoghi di studio

    • Minnesota
      • Minneapolis, Minnesota, Stati Uniti, 55407
        • Minneapolis Heart Institute Foundation
        • Contatto:
        • Contatto:
        • Investigatore principale:
          • John S Zakaib, MD

Criteri di partecipazione

I ricercatori cercano persone che corrispondano a una certa descrizione, chiamata criteri di ammissibilità. Alcuni esempi di questi criteri sono le condizioni generali di salute di una persona o trattamenti precedenti.

Criteri di ammissibilità

Età idonea allo studio

  • Adulto
  • Adulto più anziano

Accetta volontari sani

No

Descrizione

Inclusion Criteria:

  1. Subject meets current clinical practice guidelines for implantation of ICD or CRT-D and will undergo one of the following:

    • Medtronic CRT-D system implant
    • Medtronic ICD system implant (single or dual chamber)
  2. Anatomical suitability for middle cardiac vein lead placement based on pre procedural imaging.
  3. Subject meets both of the following:

    • At least Moderate Tricuspid Valve Regurgitation documented by cardiac imaging
    • A clinical goal to avoid worsening tricuspid valve regurgitation as determined by the treating physician and/or the patient
  4. Age 18 years or older.
  5. Subject is willing to undergo implant defibrillation testing
  6. Signed Patient Informed Consent Form (ICF).

Exclusion Criteria:

  1. Life expectancy < 6 months due to non-cardiac comorbidities or advanced frailty where the procedure is not aligned with goals of care
  2. Contraindication to transvenous lead implantation, including:

    • Severe or occlusive disease of the axillary/subclavian venous system
    • Coronary sinus or middle cardiac vein occlusion, severe stenosis, or anatomy prohibiting safe lead delivery
    • Known hypercoagulable state or bleeding disorder is not manageable with standard therapy.
  3. Subject is contraindicated for ≤1 mg dexamethasone acetate
  4. Patients with a medical condition that precludes them from undergoing defibrillation testing:

    • Pre-existing or suspected pneumothorax
    • Severe aortic stenosis
    • Current Intracardiac LA or LV thrombus
    • Severe proximal three-vessel or left main coronary artery disease without revascularization
    • Hemodynamic instability
    • Unstable angina
    • Recent stroke or transient ischemic attack (within the last 6 months)
    • Known inadequate external defibrillation
    • LVEDD >70 mm
    • Any other known medical condition not listed that precludes their participation in the opinion of the Investigator.
  5. Patient with evidence of active infection or undergoing treatment for an infection.
  6. Inability to provide informed consent, including:

    • Cognitive impairment without a legally authorized representative
    • Language or communication barriers that prevent adequate understanding of the study
  7. Unable to follow through study protocol for any reasons (social support, distance from research center, insurance denial, mental illness etc.) in the investigator's judgement
  8. Pregnant women or breastfeeding women, or women of childbearing potential and who are not on a reliable form of birth regulation method or abstinence
  9. End-Stage Renal Disease (ESRD) on dialysis
  10. NYHA IV
  11. Patient participation in another study that precludes concomitant study participation

Piano di studio

Questa sezione fornisce i dettagli del piano di studio, compreso il modo in cui lo studio è progettato e ciò che lo studio sta misurando.

Come è strutturato lo studio?

Dettagli di progettazione

  • Scopo principale: Fattibilità del dispositivo
  • Assegnazione: N / A
  • Modello interventistico: Assegnazione di gruppo singolo
  • Mascheramento: Nessuno (etichetta aperta)

Armi e interventi

Gruppo di partecipanti / Arm
Intervento / Trattamento
Sperimentale: Interventional
Patients will undergo implant of the Medtronic 3930M Defibrillation lead via the Middle Cardiac Vein and undergo required electrical testing
Required electrical testing for pacing, sensing, impedance, and defibrillation will be completed

Cosa sta misurando lo studio?

Misure di risultato primarie

Misura del risultato
Misura Descrizione
Lasso di tempo
Acute Procedural Success
Lasso di tempo: Index Procedure

Acute Procedural Success, defined as successful implantation of the investigational lead with all the following:

  1. Successful lead delivery and deployment at the intended MCV location within 4 deployment attempts
  2. Stable lead position at completion of the procedure
  3. Acceptable electrical performance:

    • Capture threshold: ≤ 2.0 V at 0.4 ms
    • Sensing amplitude: ≥ 4.0 mV
    • Impedance: 200-1500 Ω
  4. No device- or procedure-related Major Adverse Cardiovascular Events (MACE) through hospital discharge
  5. Absence of T-wave Oversensing (TWOS)
  6. Absence of P-wave Oversensing (PWOS)
  7. Absence of Myopotential Oversensing
  8. Successful Defibrillation Test (DFT) with 10J safety margin
  9. Absence of any serious adverse events related to the MCV implantation site.
Index Procedure

Misure di risultato secondarie

Misura del risultato
Misura Descrizione
Lasso di tempo
Lead Electrical Performance and Stability through 90 Days
Lasso di tempo: Up to 90 days following implant

Evaluation of MCV lead electrical performance and stability through 90 days following implantation, including:

  • Stability of electrical performance (e.g., pacing threshold, sensing, impedance)
  • Lead stability (e.g., no dislodgement or migration)
  • Need for lead revision or system modification
Up to 90 days following implant
Incidence of Device- or Procedure-Related Complications
Lasso di tempo: Up to 90 days following implant

Number of participants experiencing device- or procedure-related complications through 90 days following implantation:

1. Incidence of device- or procedure-related major adverse events including, but not limited to:

  • Death (all-cause or procedure-related)
  • Cardiac perforation, pericardial effusion, or cardiac tamponade
  • Clinically significant arrhythmia requiring intervention
  • Device-related infection requiring intervention
  • Lead dislodgement requiring intervention
  • Coronary sinus or venous injury
  • Clinically significant phrenic nerve stimulation
  • System revision or extraction
Up to 90 days following implant
Freedom from Device-or Procedure-Related Serious Adverse Events
Lasso di tempo: Up to 90 days following implant
Proportion of participants free from device-or procedure-related serious adverse events through 90 days following implantation.
Up to 90 days following implant

Altre misure di risultato

Misura del risultato
Misura Descrizione
Lasso di tempo
Tricuspid Valve Parameters
Lasso di tempo: 6 Month Post Implant
Tricuspid Valve Parameters: Tricuspid Valve function, including Tricuspid Valve Regurgitation post implant will be assessed at 6 months post implant or per TTVI protocol
6 Month Post Implant

Collaboratori e investigatori

Qui è dove troverai le persone e le organizzazioni coinvolte in questo studio.

Collaboratori

Investigatori

  • Investigatore principale: John S Zakaib, MD, Minneapolis Heart Institute Foundation

Studiare le date dei record

Queste date tengono traccia dell'avanzamento della registrazione dello studio e dell'invio dei risultati di sintesi a ClinicalTrials.gov. I record degli studi e i risultati riportati vengono esaminati dalla National Library of Medicine (NLM) per assicurarsi che soddisfino specifici standard di controllo della qualità prima di essere pubblicati sul sito Web pubblico.

Studia le date principali

Inizio studio (Stimato)

1 novembre 2026

Completamento primario (Stimato)

1 gennaio 2028

Completamento dello studio (Stimato)

1 gennaio 2028

Date di iscrizione allo studio

Primo inviato

27 agosto 2026

Primo inviato che soddisfa i criteri di controllo qualità

3 settembre 2026

Primo Inserito (Effettivo)

10 settembre 2026

Aggiornamenti dei record di studio

Ultimo aggiornamento pubblicato (Effettivo)

10 settembre 2026

Ultimo aggiornamento inviato che soddisfa i criteri QC

3 settembre 2026

Ultimo verificato

1 settembre 2026

Maggiori informazioni

Termini relativi a questo studio

Informazioni su farmaci e dispositivi, documenti di studio

Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti

No

Studia un dispositivo regolamentato dalla FDA degli Stati Uniti

Sì

Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .