Fracture Internal X-fixation With Augmented Reality and Radiolucent Ring Retraction (FIXAR)
Fracture Internal X-fixation With Augmented Reality and Radiolucent Ring Retraction: A Prospective Randomised Controlled Trial
The surgical management of distal radius fractures via open reduction and internal fixation (ORIF) remains the gold standard for restoring wrist function. However, the procedure is traditionally characterized by repetitive workflow interruptions of the removal of metallic retractors to avoid radiological artifacts and the ergonomic strain of the "head-turn" required to view distant fluoroscopy monitors.
The Radius Radiolucent Ring Retractor (RRRR) is a single-use, 3D-printed nylon surgical device designed for hands-free retraction in arm and wrist surgeries. This device was developed at the University of Hong Kong and invention is held by HKU Versitech Limited.
This study aims to definitively evaluate the gadget-optimized distal radius fixation surgery with the RRRR and head-mounted Augmented Reality glasses display to optimize the surgical workflow. This trial will evaluate whether this system provides a more efficient alternative to conventional monitor-based fluoroscopy in modern orthopedic practice.
Studie Overzicht
Toestand
Conditie
Studietype
Inschrijving (Geschat)
Fase
- Niet toepasbaar
Contacten en locaties
Studiecontact
- Naam: Christian Prof. Fang
- Telefoonnummer: +852 22554581
- E-mail: cfang@hku.hk
Studie Locaties
-
-
-
Hong Kong, Hongkong
- Queen Mary Hospital, The University of Hong Kong
-
Contact:
- Christian Prof. Fang
- Telefoonnummer: +852 22554581
- E-mail: cfang@hku.hk
-
-
Deelname Criteria
Geschiktheidscriteria
Leeftijden die in aanmerking komen voor studie
Accepteert gezonde vrijwilligers
Geslachten die in aanmerking komen voor studie
Beschrijving
Inclusion Criteria:
- Adults (18-85 years)
- Distal radial fractures requiring Open Reduction and Internal Fixation.
- Patients that would be suitable for the volar approach.
Exclusion Criteria:
- Open fractures
- Active infection
- Polytrauma patients where the wrist is not the primary focus.
Studie plan
Hoe is de studie opgezet?
Ontwerpdetails
- Primair doel: Behandeling
- Toewijzing: Gerandomiseerd
- Interventioneel model: Parallelle opdracht
- Masker: Enkel
Wapens en interventies
Deelnemersgroep / Arm |
Interventie / Behandeling |
|---|---|
|
Experimenteel: Radius Radiolucent Ring Retractor and the Augmented Reality Glasses
The surgery will be completed with the combination of Radius Radiolucent Ring Retractor (RRRR) and Augmented Reality Glasses. The RRRR is a 3D-printed surgical tool designed for hands-free retraction in arm and wrist surgeries. The Augmented Reality Glasses for surgeons designed specifically for surgeons can display real-time X-ray images directly in the doctor's field of vision, allowing the doctor to focus on the wrist throughout the procedure without having to turn their head to look at the X-ray screen. |
The Radius Radiolucent Ring Retractor (RRRR) is a single-use, 3D-printed nylon surgical device designed for hands-free retraction in arm and wrist surgeries. Augmented Reality of C-Arm intraoperative fluoroscopic imaging integration with a head mount-display (HMD). |
|
Actieve vergelijker: Metal tools and standard X-ray monitor
Traditional method with metal retractors and standard X-ray monitor. |
Manual traditional metal retractor used by surgeons and assistants. Standard X-ray monitor that displays fluoroscopy images during the operation. |
Wat meet het onderzoek?
Primaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
Total Fluoroscopy Time
Tijdsspanne: Intra-operative
|
The total duration, measured in minutes or seconds, that the X-ray beam is actively turned on during a fluoroscopic procedure during the operation. |
Intra-operative
|
|
Functional Assessment measured by QuickDASH Score
Tijdsspanne: At post-op, 3 weeks, 6 weeks , 3 months and 6 months
|
The QuickDASH score is a validated, internationally recognized tool to assess functional outcomes of patients of the upper arm, especially in long term recovery assessments. The survey will assess the control group against intervention group in return to daily activities, assessment of overall upper limb function, symptoms, and quality of life. |
At post-op, 3 weeks, 6 weeks , 3 months and 6 months
|
Secundaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
Rating of placement of implant using 25 point Surgeon-Reported Outcome Measures
Tijdsspanne: Immediately after operation
|
The qualitative assessment of 1 to 5 of the placement of implant rated by the surgeon. |
Immediately after operation
|
|
Rating of ergonomics and visualisation of plate position using 25 point Surgeon-Reported Outcome Measures
Tijdsspanne: Immediately after operation
|
The qualitative assessment of 1 to 5 of the ergonomics and visualisation of plate position rated by the surgeon. |
Immediately after operation
|
|
Rating of retraction during fluoroscopy using 25 point Surgeon-Reported Outcome Measures
Tijdsspanne: Immediately after operation
|
The qualitative assessment of 1 to 5 of the retraction during fluoroscopy rated by the surgeon. |
Immediately after operation
|
|
Rating of fracture reduction using 25 point Surgeon-Reported Outcome Measures
Tijdsspanne: Immediately after operation
|
The qualitative assessment of 1 to 5 of the quality of fracture reduction rated by the surgeon. |
Immediately after operation
|
|
Rating of overall assistants' efforts using 25 point Surgeon-Reported Outcome Measures
Tijdsspanne: Immediately after operation
|
The qualitative assessment of 1 to 5 of the quality of overall assistants' efforts rated by the surgeon. |
Immediately after operation
|
|
Number of surgical assistants
Tijdsspanne: Intra-operative
|
The maximum number of surgical assistants required for retraction tasks at any single point during the surgery. |
Intra-operative
|
|
Number of times retractors were removed and replaced
Tijdsspanne: Intra-operative
|
The number of times the whole RRRR or the metal retractor is lifted out of the incision site, then putting it back count as 1. |
Intra-operative
|
|
Number of times the retractor was loosened from the surgical site
Tijdsspanne: Intra-operative
|
The number of times the traditional retractor or RRRR is not adequately retracting the surgical wound and providing a clear view of the fracture site, |
Intra-operative
|
|
Number of readjustments for fracture reduction
Tijdsspanne: Intra-operative
|
The number of times that a fracture reduction is changed during the operation. |
Intra-operative
|
|
Number of attempts for repositioning Kirschner wire
Tijdsspanne: Intra-operative
|
The numbers of times the Kirschner wire is removed from the incision site and then putting it back count as 1. |
Intra-operative
|
|
Maximum number of Kirschner wire used to maintain positioning and/or reduction
Tijdsspanne: Intra-operative
|
The maximum number of Kirschner wires at any time point that are used to maintain implant positioning and/or fracture reduction. |
Intra-operative
|
|
Number of repositioning of plates
Tijdsspanne: Intra-operative
|
The number of times the implant plate is moved and repositioned is recorded as 1. |
Intra-operative
|
|
Number of repositioning of screws
Tijdsspanne: Intra-operative
|
The number of times the screw is removed from the incision site and repositioned is recorded as 1. |
Intra-operative
|
|
Total number of screws used in the operation
Tijdsspanne: Intra-operative
|
The total number of screws that are used in the operation, including those that are implanted or wasted. |
Intra-operative
|
|
Number of head turns to assess x-ray images
Tijdsspanne: Intra-operative
|
The number of times the primary surgeon looks at the fluoroscopy monitor during fracture reduction. |
Intra-operative
|
|
Number of times the eyepiece focus is lost
Tijdsspanne: Intra-operative
|
The number of times the AR HMD eyepiece is out of focus is recorded as 1. |
Intra-operative
|
|
Duration of surgery
Tijdsspanne: Intra-operative
|
Defined from the time difference of the incision start time and the completion time of final suturing. |
Intra-operative
|
|
Incision Length in mm
Tijdsspanne: Intra-operative
|
Defined as the length in millimeters of the incision cut in the surgery. |
Intra-operative
|
|
Radial Inclination in degrees
Tijdsspanne: At post-op, 3 weeks, 6 weeks , 3 months and 6 months
|
Measurement of the coronal-plane slope of the distal radius articular surface, measured in pre-operative and post-operative wrist x-ray. |
At post-op, 3 weeks, 6 weeks , 3 months and 6 months
|
|
Radial Height in millimeters
Tijdsspanne: At post-op, 3 weeks, 6 weeks , 3 months and 6 months
|
Measurement of the axial distance between perpendicular lines at the radial styloid and ulnar border, quantifying fracture-induced radial shortening, measured in pre-operative and post-operative wrist x-ray. |
At post-op, 3 weeks, 6 weeks , 3 months and 6 months
|
|
Volar / Dorsal Angulation in degrees
Tijdsspanne: At post-op, 3 weeks, 6 weeks , 3 months and 6 months
|
Measurement of the tilt of the distal articular surface relative to the shaft, measured in pre-operative and post-operative wrist x-ray. |
At post-op, 3 weeks, 6 weeks , 3 months and 6 months
|
|
Articular Step-off in millimeters
Tijdsspanne: At post-op, 3 weeks, 6 weeks , 3 months and 6 months
|
Measurement of the vertical incongruity along the subchondral surface, measured in pre-operative and post-operative wrist x-ray. |
At post-op, 3 weeks, 6 weeks , 3 months and 6 months
|
Medewerkers en onderzoekers
Sponsor
Onderzoekers
- Hoofdonderzoeker: Christian Prof. Fang, Dept of Orthopaedics and Traumatology, Queen Mary Hospital
Publicaties en nuttige links
Algemene publicaties
- Fang C, Fang E, Yee DK, Kwan K, Leung G, Leung F. A comparison of six outcome measures across the recovery period after distal radius fixation-Which to use and when? J Orthop Surg (Hong Kong). 2021 Jan-Apr;29(1):2309499020971866. doi: 10.1177/2309499020971866.
- Al-Dahan T, Virani S, Asardag AN, Seoudi N, Izedonmwen I, Sherpa N, Murad MA, Ekungba-Adewole M, Seidahmed S, Gurung Y. Fluoroscopy Exposure During Distal Radius Fracture Fixation: An Audit of Current Practice. Cureus. 2025 Dec 20;17(12):e99719. doi: 10.7759/cureus.99719. eCollection 2025 Dec.
- von der Heide AM, Fallavollita P, Wang L, Sandner P, Navab N, Weidert S, Euler E. Camera-augmented mobile C-arm (CamC): A feasibility study of augmented reality imaging in the operating room. Int J Med Robot. 2018 Apr;14(2). doi: 10.1002/rcs.1885. Epub 2017 Dec 21.
- Chimenti PC, Mitten DJ. Google Glass as an Alternative to Standard Fluoroscopic Visualization for Percutaneous Fixation of Hand Fractures: A Pilot Study. Plast Reconstr Surg. 2015 Aug;136(2):328-330. doi: 10.1097/PRS.0000000000001453.
- Ortega G, Wolff A, Baumgaertner M, Kendoff D. Usefulness of a head mounted monitor device for viewing intraoperative fluoroscopy during orthopaedic procedures. Arch Orthop Trauma Surg. 2008 Oct;128(10):1123-6. doi: 10.1007/s00402-007-0500-y. Epub 2007 Nov 14.
- Laverdiere C, Corban J, Khoury J, Ge SM, Schupbach J, Harvey EJ, Reindl R, Martineau PA. Augmented reality in orthopaedics: a systematic review and a window on future possibilities. Bone Joint J. 2019 Dec;101-B(12):1479-1488. doi: 10.1302/0301-620X.101B12.BJJ-2019-0315.R1.
- Jud L, Fotouhi J, Andronic O, Aichmair A, Osgood G, Navab N, Farshad M. Applicability of augmented reality in orthopedic surgery - A systematic review. BMC Musculoskelet Disord. 2020 Feb 15;21(1):103. doi: 10.1186/s12891-020-3110-2.
- Shi Y, Wu X, Paydarfar JA, Halter RJ. An Imaging-Compatible Oral Retractor System for Transoral Robotic Surgery. Ann Biomed Eng. 2024 Sep;52(9):2473-2484. doi: 10.1007/s10439-024-03536-7. Epub 2024 May 25.
Studie record data
Bestudeer belangrijke data
Studie start (Geschat)
Primaire voltooiing (Geschat)
Studie voltooiing (Geschat)
Studieregistratiedata
Eerst ingediend
Eerst ingediend dat voldeed aan de QC-criteria
Eerst geplaatst (Werkelijk)
Updates van studierecords
Laatste update geplaatst (Werkelijk)
Laatste update ingediend die voldeed aan QC-criteria
Laatst geverifieerd
Meer informatie
Termen gerelateerd aan deze studie
Aanvullende relevante MeSH-voorwaarden
Andere studie-ID-nummers
- UW 26-371
Plan Individuele Deelnemersgegevens (IPD)
Bent u van plan om gegevens van individuele deelnemers (IPD) te delen?
Beschrijving IPD-plan
Anonymized dataset to be included as supplementary data in final publication
IPD-tijdsbestek voor delen
Within 1 year of study completion
IPD-toegangscriteria voor delen
Additional information available upon reasonable request of principal investigator
IPD delen Ondersteunend informatietype
- LEERPROTOCOOL
- SAP
- ICF
- MVO
Informatie over medicijnen en apparaten, studiedocumenten
Bestudeert een door de Amerikaanse FDA gereguleerd geneesmiddel
Bestudeert een door de Amerikaanse FDA gereguleerd apparaatproduct
Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .