Effect of Double-Tap Optimization on Valve Expansion and Hemodynamics in Balloon-expandable Transcatheter Aortic Valve Implantation: A Randomized Clinical Trial (DO-OPT TAVI Trial)
This study will compare two balloon inflation strategies during transcatheter aortic valve implantation (TAVI), a procedure used to replace a severely narrowed aortic valve through a catheter. A balloon-expandable replacement valve may not fully expand during its initial placement. An additional inflation using the same balloon and the same inflation volume, called "Double-Tap," may improve valve expansion and function. Its benefits and risks need to be evaluated in a randomized trial.
This prospective, multicenter study aims to include 250 adults with symptomatic severe aortic stenosis undergoing TAVI with an Edwards SAPIEN-family balloon-expandable valve. After successful valve implantation, participants for whom an additional inflation is considered technically feasible and clinically appropriate will be randomly assigned in equal numbers to Double-Tap or conventional single inflation. Patients requiring therapeutic post-dilatation for moderate or greater leakage around the implanted valve will not undergo randomization.
The main question is whether Double-Tap improves valve expansion immediately after TAVI, as measured using fluoroscopic X-ray images. The study will also compare blood flow through the valve, valve expansion and signs of valve thrombosis on cardiac computed tomography, procedure-related complications, and clinical outcomes. Participants will undergo echocardiography before discharge and at 3-6 and 12 months, and cardiac computed tomography at 3-6 months. Clinical follow-up will continue for at least 12 months, with annual follow-up thereafter until the last enrolled participant completes the 12-month visit.
Studieöversikt
Status
Betingelser
Intervention / Behandling
Studietyp
Inskrivning (Beräknad)
Fas
- Inte tillämpbar
Kontakter och platser
Studiekontakt
- Namn: Young-Guk Ko, Professor
- Telefonnummer: +82-22228-8460
- E-post: ygko@yuhs.ac
Deltagandekriterier
Urvalskriterier
Åldrar som är berättigade till studier
Tar emot friska volontärer
Kön som är behöriga för studier
Beskrivning
Inclusion Criteria:
- Patients ≥ 19 years old
- Patients undergoing TAVI using a Sapien family valve for severe AS
- Provision of informed consent
Exclusion criteria:
- Subannular calcification
- Bicuspid aortic valve with raphe calcification
- Valve-in-valve TAVI procedure
- Planned major surgery
- Cardiogenic shock or hemodynamic instability
- Chronic kidney disease stage 3b, 4, or 5 (eGFR <45mL/min)
- Thrombocytopenia (platelet count <100,000) or contraindication to antiplatelet therapy or anticoagulation despite clinical indications
Studieplan
Hur är studien utformad?
Designdetaljer
- Primärt syfte: Behandling
- Tilldelning: Randomiserad
- Interventionsmodell: Parallellt uppdrag
- Maskning: Ingen (Open Label)
Vapen och interventioner
Deltagargrupp / Arm |
Intervention / Behandling |
|---|---|
|
Experimentell: Double-Tap
Participants undergo TAVI with an Edwards SAPIEN-family balloon-expandable valve. After successful initial valve implantation and randomization, one additional balloon inflation is performed using the same delivery balloon and the same inflation volume used for initial deployment. The additional inflation is performed immediately after confirming appropriate valve position and procedural stability. |
Following successful implantation of an Edwards SAPIEN-family balloon-expandable transcatheter heart valve, one additional balloon inflation is performed using the same delivery balloon and the same inflation volume used for initial valve deployment. Balloon inflation volume is not increased for the Double-Tap inflation. The procedure is performed after randomization and confirmation of appropriate valve position and procedural stability. |
|
Aktiv komparator: No Double-Tap
Participants undergo standard TAVI with an Edwards SAPIEN-family balloon-expandable valve deployed using a single balloon inflation. After successful initial valve implantation and randomization, no planned additional balloon inflation for Double-Tap optimization is performed. |
An Edwards SAPIEN-family balloon-expandable transcatheter heart valve is implanted using a single balloon inflation according to standard practice and the manufacturer's instructions for use. No planned additional balloon inflation for Double-Tap optimization is performed after randomization. |
Vad mäter studien?
Primära resultatmått
Resultatmått |
Åtgärdsbeskrivning |
Tidsram |
|---|---|---|
|
Percentage valve expansion measured by fluoroscopy
Tidsram: immediately after TAVI procedure
|
Expansion of the implanted transcatheter heart valve, expressed as a percentage of the manufacturer's nominal cross-sectional area for the implanted valve size. Valve diameters are measured at the inflow, midframe, and outflow levels in three-cusp and cusp-overlap fluoroscopic views. At each level, the diameters from the two views are averaged, and the cross-sectional area is calculated using the circular area formula. Fluoroscopic images are evaluated by independent imaging core laboratory assessors blinded to treatment assignment and clinical outcomes. |
immediately after TAVI procedure
|
Samarbetspartners och utredare
Sponsor
Studieavstämningsdatum
Studera stora datum
Studiestart (Beräknad)
Primärt slutförande (Beräknad)
Avslutad studie (Beräknad)
Studieregistreringsdatum
Först inskickad
Först inskickad som uppfyllde QC-kriterierna
Första postat (Faktisk)
Uppdateringar av studier
Senaste uppdatering publicerad (Faktisk)
Senaste inskickade uppdateringen som uppfyllde QC-kriterierna
Senast verifierad
Mer information
Termer relaterade till denna studie
Ytterligare relevanta MeSH-villkor
Andra studie-ID-nummer
- 4-2026-1423
Plan för individuella deltagardata (IPD)
Planerar du att dela individuella deltagardata (IPD)?
Läkemedels- och apparatinformation, studiedokument
Studerar en amerikansk FDA-reglerad läkemedelsprodukt
Studerar en amerikansk FDA-reglerad produktprodukt
Denna information hämtades direkt från webbplatsen clinicaltrials.gov utan några ändringar. Om du har några önskemål om att ändra, ta bort eller uppdatera dina studieuppgifter, vänligen kontakta register@clinicaltrials.gov. Så snart en ändring har implementerats på clinicaltrials.gov, kommer denna att uppdateras automatiskt även på vår webbplats .