Effect of Double-Tap Optimization on Valve Expansion and Hemodynamics in Balloon-expandable Transcatheter Aortic Valve Implantation: A Randomized Clinical Trial (DO-OPT TAVI Trial)
This study will compare two balloon inflation strategies during transcatheter aortic valve implantation (TAVI), a procedure used to replace a severely narrowed aortic valve through a catheter. A balloon-expandable replacement valve may not fully expand during its initial placement. An additional inflation using the same balloon and the same inflation volume, called "Double-Tap," may improve valve expansion and function. Its benefits and risks need to be evaluated in a randomized trial.
This prospective, multicenter study aims to include 250 adults with symptomatic severe aortic stenosis undergoing TAVI with an Edwards SAPIEN-family balloon-expandable valve. After successful valve implantation, participants for whom an additional inflation is considered technically feasible and clinically appropriate will be randomly assigned in equal numbers to Double-Tap or conventional single inflation. Patients requiring therapeutic post-dilatation for moderate or greater leakage around the implanted valve will not undergo randomization.
The main question is whether Double-Tap improves valve expansion immediately after TAVI, as measured using fluoroscopic X-ray images. The study will also compare blood flow through the valve, valve expansion and signs of valve thrombosis on cardiac computed tomography, procedure-related complications, and clinical outcomes. Participants will undergo echocardiography before discharge and at 3-6 and 12 months, and cardiac computed tomography at 3-6 months. Clinical follow-up will continue for at least 12 months, with annual follow-up thereafter until the last enrolled participant completes the 12-month visit.
A tanulmány áttekintése
Állapot
Körülmények
Beavatkozás / kezelés
Tanulmány típusa
Beiratkozás (Becsült)
Fázis
- Nem alkalmazható
Kapcsolatok és helyek
Tanulmányi kapcsolat
- Név: Young-Guk Ko, Professor
- Telefonszám: +82-22228-8460
- E-mail: ygko@yuhs.ac
Részvételi kritériumok
Jogosultsági kritériumok
Tanulmányozható életkorok
Egészséges önkénteseket fogad
Tanulmányozható nemek
Leírás
Inclusion Criteria:
- Patients ≥ 19 years old
- Patients undergoing TAVI using a Sapien family valve for severe AS
- Provision of informed consent
Exclusion criteria:
- Subannular calcification
- Bicuspid aortic valve with raphe calcification
- Valve-in-valve TAVI procedure
- Planned major surgery
- Cardiogenic shock or hemodynamic instability
- Chronic kidney disease stage 3b, 4, or 5 (eGFR <45mL/min)
- Thrombocytopenia (platelet count <100,000) or contraindication to antiplatelet therapy or anticoagulation despite clinical indications
Tanulási terv
Hogyan készül a tanulmány?
Tervezési részletek
- Elsődleges cél: Kezelés
- Kiosztás: Véletlenszerűsített
- Beavatkozó modell: Párhuzamos hozzárendelés
- Maszkolás: Nincs (Open Label)
Fegyverek és beavatkozások
Résztvevő csoport / kar |
Beavatkozás / kezelés |
|---|---|
|
Kísérleti: Double-Tap
Participants undergo TAVI with an Edwards SAPIEN-family balloon-expandable valve. After successful initial valve implantation and randomization, one additional balloon inflation is performed using the same delivery balloon and the same inflation volume used for initial deployment. The additional inflation is performed immediately after confirming appropriate valve position and procedural stability. |
Following successful implantation of an Edwards SAPIEN-family balloon-expandable transcatheter heart valve, one additional balloon inflation is performed using the same delivery balloon and the same inflation volume used for initial valve deployment. Balloon inflation volume is not increased for the Double-Tap inflation. The procedure is performed after randomization and confirmation of appropriate valve position and procedural stability. |
|
Aktív összehasonlító: No Double-Tap
Participants undergo standard TAVI with an Edwards SAPIEN-family balloon-expandable valve deployed using a single balloon inflation. After successful initial valve implantation and randomization, no planned additional balloon inflation for Double-Tap optimization is performed. |
An Edwards SAPIEN-family balloon-expandable transcatheter heart valve is implanted using a single balloon inflation according to standard practice and the manufacturer's instructions for use. No planned additional balloon inflation for Double-Tap optimization is performed after randomization. |
Mit mér a tanulmány?
Elsődleges eredményintézkedések
Eredménymérő |
Intézkedés leírása |
Időkeret |
|---|---|---|
|
Percentage valve expansion measured by fluoroscopy
Időkeret: immediately after TAVI procedure
|
Expansion of the implanted transcatheter heart valve, expressed as a percentage of the manufacturer's nominal cross-sectional area for the implanted valve size. Valve diameters are measured at the inflow, midframe, and outflow levels in three-cusp and cusp-overlap fluoroscopic views. At each level, the diameters from the two views are averaged, and the cross-sectional area is calculated using the circular area formula. Fluoroscopic images are evaluated by independent imaging core laboratory assessors blinded to treatment assignment and clinical outcomes. |
immediately after TAVI procedure
|
Együttműködők és nyomozók
Szponzor
Tanulmányi rekorddátumok
Tanulmány főbb dátumok
Tanulmány kezdete (Becsült)
Elsődleges befejezés (Becsült)
A tanulmány befejezése (Becsült)
Tanulmányi regisztráció dátumai
Először benyújtva
Először nyújtották be, amely megfelel a minőségbiztosítási kritériumoknak
Első közzététel (Tényleges)
Tanulmányi rekordok frissítései
Utolsó frissítés közzétéve (Tényleges)
Az utolsó frissítés elküldve, amely megfelel a minőségbiztosítási kritériumoknak
Utolsó ellenőrzés
Több információ
A tanulmányhoz kapcsolódó kifejezések
További vonatkozó MeSH feltételek
Egyéb vizsgálati azonosító számok
- 4-2026-1423
Terv az egyéni résztvevői adatokhoz (IPD)
Tervezi megosztani az egyéni résztvevői adatokat (IPD)?
Gyógyszer- és eszközinformációk, tanulmányi dokumentumok
Egy amerikai FDA által szabályozott gyógyszerkészítményt tanulmányoz
Egy amerikai FDA által szabályozott eszközterméket tanulmányoz
Ezt az információt közvetlenül a clinicaltrials.gov webhelyről szereztük be, változtatás nélkül. Ha bármilyen kérése van vizsgálati adatainak módosítására, eltávolítására vagy frissítésére, kérjük, írjon a következő címre: register@clinicaltrials.gov. Amint a változás bevezetésre kerül a clinicaltrials.gov oldalon, ez a webhelyünkön is automatikusan frissül. .