Eversense and Dexcom G5: Efficacy and Accuracy in Type 1 Diabetic Patients

March 26, 2020 updated by: Daniela Bruttomesso, University of Padova

Eversense and Dexcom G5 Comparison in Real Life: a Randomized Crossover Trial in Type 1 Diabetic Patients to Evaluate Differences in Accuracy, Efficacy and Quality of Life

Continuous Glucose Monitoring (CGM) systems improve glycaemic control in type 1 diabetic patients but they have different characteristics that could influence patients' quality of life and glycaemic control.

In this randomized cross over study investigators will compare 2 different CGM systems: Eversense implantable sensor (Senseonics, Germantown, MD, USA) and the standard transcutaneous sensor Dexcom G5 (Dexcom, San Diego, CA, USA). Investigators will evaluate sensors' accuracy, impact on quality of life and efficacy in optimizing glycaemic control.

The investigator's study's results might help clinicians choose the sensor and evaluate how sensor differences could impact glycaemic control.

Study Overview

Detailed Description

Good glycaemic control in type 1 diabetic patients prevents the onset and progression of chronic complications. Continuous Glucose Monitoring (CGM) systems help patients improve glycaemic control by providing real-time glucose levels, glycaemic tendency, glycaemic swing rate and by alerting the patient when the glucose value read by the sensor reaches a predefined threshold of hyper or hypoglycaemia.

Several CGM systems are available and they have different characteristics that could influence patients' quality of life and glycaemic control.

In this randomized cross over study investigators will compare 2 different CGM systems: Eversense implantable sensor (Senseonics, Germantown, MD, USA) and the standard transcutaneous sensor Dexcom G5 (Dexcom, San Diego, CA, USA). Investigators will evaluate sensors' accuracy, impact on quality of life and efficacy in optimizing glycaemic control.

Patients will use Dexcom G5 or Eversense for three months, respectively, in a randomized order. Accuracy will be evaluated comparing sensors values with capillary blood glucose at home.

Quality of life will be assessed at the beginning and at the end of each three-month period through validated questionnaires to underline differences in different sensors use.

Time spent in target (70-180 mg/dl), in hypoglycamiea and hyperglycemia will be evaluated with both sensors to assess differences in glycaemic control induced by different alarm system and by the presence of predictive alarms

Study Type

Interventional

Enrollment (Actual)

16

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

      • Padova, Italy
        • University of Padova

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years and older (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  • Male and female participants of at least 18 years of age
  • Diagnosis of type 1 diabetes mellitus (diagnosed according to World Health Organization criteria) for at least 1 year
  • Body Mass Index (BMI) <35 kg / m²
  • Availability to wear study equipment and to comply with the study protocol for its entire duration
  • HbA1c <10%
  • Signature of informed consent before any procedure related to the study.

Exclusion Criteria:

  • Pregnancy, breastfeeding, intention to undergo pregnancy or refusal to use contraceptive methods during the study period (for female subjects).
  • Known allergies to skin patches or disinfectants used during the study.
  • Skin lesions, irritation, redness, edema in sites where sensors can be applied, as this might interfere with sensor's placement or with the accuracy of the glycaemic value detection.
  • Use of drugs that may interfere with glucose metabolism (such as steroids) unless they are chronic therapies whose dosage has remained stable over the past 3 months and is expected to remain stable during the study period.
  • Use of acetaminophen or other drugs that could influence sensor accuracy
  • Severe medical or psychological conditions, which, in the opinion of the medical team, may compromise patients' safety while participating in the study.
  • Patients enrolled in other clinical trials.
  • Known disorders of adrenal glands, pancreatic tumors or insulinoma
  • Patient's inability to comply with the procedures of the study

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Other
  • Allocation: Randomized
  • Interventional Model: Crossover Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Active Comparator: Dexcom G5 - Eversense
Patients will wear first Dexcom G5 (Dexcom San Diego, CA, USA) transcutaneous sensor for 3 months, then Eversnese(Senseonics Inc, MD, USA) implantable sensor for 3 months Accuracy and efficacy will be evaluated
Patients will wear sensor for 3 months, monitor blood capillary values 4 times/day. At the beginning and at the end of the period HbA1c will be measured and questionnaires will be administered
Experimental: Eversense- Dexcom G5
Patients will wear first Eversense (Senseonics Inc, MD, USA) implantable sensor for 3 months, then Dexcom G5 (Dexcom San Diego, CA, USA) transcutaneous sensor for 3 months accuracy and efficacy will be evaluated

Patients will wear sensor for 3 months, monitor blood capillary values 4 times/day. At the beginning and at the end of the period HbA1c will be measured and questionnaires will be administered.

After 30-50 days of sensor implantation patients will wear also Dexcom G5 for a week to compare accuracy simultaneously

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
device accuracy
Time Frame: after 3 months, at the end of the study for each arm of the study
sensors' accuracy expressed in terms of MARD (mean absolute relative difference) versus capillary blood glucose in different glycaemic ranges
after 3 months, at the end of the study for each arm of the study

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Time spent in target
Time Frame: after 3 months, at the end of the study for each arm of the study
% Time spent in target (70-180 mg/dl) using each sensor to evaluate sensor efficacy
after 3 months, at the end of the study for each arm of the study
HbA1c
Time Frame: after 3 months, at the end of the study for each arm of the study
HbA1c changes using different sensors, to evaluate sensor efficacy
after 3 months, at the end of the study for each arm of the study
failure
Time Frame: after 3 months, at the end of the study for each arm of the study
% Sensors' failure rate to evaluate sensor duration
after 3 months, at the end of the study for each arm of the study
Adverse events
Time Frame: after 3 months, at the end of the study for each arm of the study
Adverse events (skin reactions, haematomas)
after 3 months, at the end of the study for each arm of the study
changes in quality of life
Time Frame: after 3 months, at the end of the study for each arm of the study
DTSQ questionnaire
after 3 months, at the end of the study for each arm of the study
changes in fear of hypoglycaemia
Time Frame: after 3 months, at the end of the study for each arm of the study
HFSII questionnaire
after 3 months, at the end of the study for each arm of the study

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

March 14, 2018

Primary Completion (Actual)

December 31, 2018

Study Completion (Actual)

September 1, 2019

Study Registration Dates

First Submitted

April 26, 2018

First Submitted That Met QC Criteria

July 27, 2018

First Posted (Actual)

August 3, 2018

Study Record Updates

Last Update Posted (Actual)

March 27, 2020

Last Update Submitted That Met QC Criteria

March 26, 2020

Last Verified

March 1, 2020

More Information

Terms related to this study

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

product manufactured in and exported from the U.S.

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.