Eversense and Dexcom G5: Efficacy and Accuracy in Type 1 Diabetic Patients
Eversense and Dexcom G5 Comparison in Real Life: a Randomized Crossover Trial in Type 1 Diabetic Patients to Evaluate Differences in Accuracy, Efficacy and Quality of Life
Continuous Glucose Monitoring (CGM) systems improve glycaemic control in type 1 diabetic patients but they have different characteristics that could influence patients' quality of life and glycaemic control.
In this randomized cross over study investigators will compare 2 different CGM systems: Eversense implantable sensor (Senseonics, Germantown, MD, USA) and the standard transcutaneous sensor Dexcom G5 (Dexcom, San Diego, CA, USA). Investigators will evaluate sensors' accuracy, impact on quality of life and efficacy in optimizing glycaemic control.
The investigator's study's results might help clinicians choose the sensor and evaluate how sensor differences could impact glycaemic control.
Tutkimuksen yleiskatsaus
Tila
Interventio / Hoito
Yksityiskohtainen kuvaus
Good glycaemic control in type 1 diabetic patients prevents the onset and progression of chronic complications. Continuous Glucose Monitoring (CGM) systems help patients improve glycaemic control by providing real-time glucose levels, glycaemic tendency, glycaemic swing rate and by alerting the patient when the glucose value read by the sensor reaches a predefined threshold of hyper or hypoglycaemia.
Several CGM systems are available and they have different characteristics that could influence patients' quality of life and glycaemic control.
In this randomized cross over study investigators will compare 2 different CGM systems: Eversense implantable sensor (Senseonics, Germantown, MD, USA) and the standard transcutaneous sensor Dexcom G5 (Dexcom, San Diego, CA, USA). Investigators will evaluate sensors' accuracy, impact on quality of life and efficacy in optimizing glycaemic control.
Patients will use Dexcom G5 or Eversense for three months, respectively, in a randomized order. Accuracy will be evaluated comparing sensors values with capillary blood glucose at home.
Quality of life will be assessed at the beginning and at the end of each three-month period through validated questionnaires to underline differences in different sensors use.
Time spent in target (70-180 mg/dl), in hypoglycamiea and hyperglycemia will be evaluated with both sensors to assess differences in glycaemic control induced by different alarm system and by the presence of predictive alarms
Opintotyyppi
Ilmoittautuminen (Todellinen)
Vaihe
- Ei sovellettavissa
Yhteystiedot ja paikat
Opiskelupaikat
-
-
-
Padova, Italia
- University of Padova
-
-
Osallistumiskriteerit
Kelpoisuusvaatimukset
Opintokelpoiset iät
Hyväksyy terveitä vapaaehtoisia
Sukupuolet, jotka voivat opiskella
Kuvaus
Inclusion Criteria:
- Male and female participants of at least 18 years of age
- Diagnosis of type 1 diabetes mellitus (diagnosed according to World Health Organization criteria) for at least 1 year
- Body Mass Index (BMI) <35 kg / m²
- Availability to wear study equipment and to comply with the study protocol for its entire duration
- HbA1c <10%
- Signature of informed consent before any procedure related to the study.
Exclusion Criteria:
- Pregnancy, breastfeeding, intention to undergo pregnancy or refusal to use contraceptive methods during the study period (for female subjects).
- Known allergies to skin patches or disinfectants used during the study.
- Skin lesions, irritation, redness, edema in sites where sensors can be applied, as this might interfere with sensor's placement or with the accuracy of the glycaemic value detection.
- Use of drugs that may interfere with glucose metabolism (such as steroids) unless they are chronic therapies whose dosage has remained stable over the past 3 months and is expected to remain stable during the study period.
- Use of acetaminophen or other drugs that could influence sensor accuracy
- Severe medical or psychological conditions, which, in the opinion of the medical team, may compromise patients' safety while participating in the study.
- Patients enrolled in other clinical trials.
- Known disorders of adrenal glands, pancreatic tumors or insulinoma
- Patient's inability to comply with the procedures of the study
Opintosuunnitelma
Miten tutkimus on suunniteltu?
Suunnittelun yksityiskohdat
- Ensisijainen käyttötarkoitus: Muut
- Jako: Satunnaistettu
- Inventiomalli: Crossover-tehtävä
- Naamiointi: Ei mitään (avoin tarra)
Aseet ja interventiot
Osallistujaryhmä / Arm |
Interventio / Hoito |
---|---|
Active Comparator: Dexcom G5 - Eversense
Patients will wear first Dexcom G5 (Dexcom San Diego, CA, USA) transcutaneous sensor for 3 months, then Eversnese(Senseonics Inc, MD, USA) implantable sensor for 3 months Accuracy and efficacy will be evaluated
|
Patients will wear sensor for 3 months, monitor blood capillary values 4 times/day.
At the beginning and at the end of the period HbA1c will be measured and questionnaires will be administered
|
Kokeellinen: Eversense- Dexcom G5
Patients will wear first Eversense (Senseonics Inc, MD, USA) implantable sensor for 3 months, then Dexcom G5 (Dexcom San Diego, CA, USA) transcutaneous sensor for 3 months accuracy and efficacy will be evaluated
|
Patients will wear sensor for 3 months, monitor blood capillary values 4 times/day. At the beginning and at the end of the period HbA1c will be measured and questionnaires will be administered. After 30-50 days of sensor implantation patients will wear also Dexcom G5 for a week to compare accuracy simultaneously |
Mitä tutkimuksessa mitataan?
Ensisijaiset tulostoimenpiteet
Tulosmittaus |
Toimenpiteen kuvaus |
Aikaikkuna |
---|---|---|
device accuracy
Aikaikkuna: after 3 months, at the end of the study for each arm of the study
|
sensors' accuracy expressed in terms of MARD (mean absolute relative difference) versus capillary blood glucose in different glycaemic ranges
|
after 3 months, at the end of the study for each arm of the study
|
Toissijaiset tulostoimenpiteet
Tulosmittaus |
Toimenpiteen kuvaus |
Aikaikkuna |
---|---|---|
Time spent in target
Aikaikkuna: after 3 months, at the end of the study for each arm of the study
|
% Time spent in target (70-180 mg/dl) using each sensor to evaluate sensor efficacy
|
after 3 months, at the end of the study for each arm of the study
|
HbA1c
Aikaikkuna: after 3 months, at the end of the study for each arm of the study
|
HbA1c changes using different sensors, to evaluate sensor efficacy
|
after 3 months, at the end of the study for each arm of the study
|
failure
Aikaikkuna: after 3 months, at the end of the study for each arm of the study
|
% Sensors' failure rate to evaluate sensor duration
|
after 3 months, at the end of the study for each arm of the study
|
Adverse events
Aikaikkuna: after 3 months, at the end of the study for each arm of the study
|
Adverse events (skin reactions, haematomas)
|
after 3 months, at the end of the study for each arm of the study
|
changes in quality of life
Aikaikkuna: after 3 months, at the end of the study for each arm of the study
|
DTSQ questionnaire
|
after 3 months, at the end of the study for each arm of the study
|
changes in fear of hypoglycaemia
Aikaikkuna: after 3 months, at the end of the study for each arm of the study
|
HFSII questionnaire
|
after 3 months, at the end of the study for each arm of the study
|
Yhteistyökumppanit ja tutkijat
Sponsori
Opintojen ennätyspäivät
Opi tärkeimmät päivämäärät
Opiskelun aloitus (Todellinen)
Ensisijainen valmistuminen (Todellinen)
Opintojen valmistuminen (Todellinen)
Opintoihin ilmoittautumispäivät
Ensimmäinen lähetetty
Ensimmäinen toimitettu, joka täytti QC-kriteerit
Ensimmäinen Lähetetty (Todellinen)
Tutkimustietojen päivitykset
Viimeisin päivitys julkaistu (Todellinen)
Viimeisin lähetetty päivitys, joka täytti QC-kriteerit
Viimeksi vahvistettu
Lisää tietoa
Tähän tutkimukseen liittyvät termit
Muita asiaankuuluvia MeSH-ehtoja
Muut tutkimustunnusnumerot
- 4270/AO/17
Lääke- ja laitetiedot, tutkimusasiakirjat
Tutkii yhdysvaltalaista FDA sääntelemää lääkevalmistetta
Tutkii yhdysvaltalaista FDA sääntelemää laitetuotetta
Yhdysvalloissa valmistettu ja sieltä viety tuote
Nämä tiedot haettiin suoraan verkkosivustolta clinicaltrials.gov ilman muutoksia. Jos sinulla on pyyntöjä muuttaa, poistaa tai päivittää tutkimustietojasi, ota yhteyttä register@clinicaltrials.gov. Heti kun muutos on otettu käyttöön osoitteessa clinicaltrials.gov, se päivitetään automaattisesti myös verkkosivustollemme .
Kliiniset tutkimukset Diabetes mellitus, tyyppi 1
-
University of Colorado, DenverMassachusetts General Hospital; Beta Bionics, Inc.ValmisDiabetes mellitus, tyyppi 1 | Tyypin 1 diabetes | Diabetes tyyppi 1 | Tyypin 1 diabetes mellitus | Autoimmuuni Diabetes | Diabetes mellitus, insuliinista riippuvainen | Nuorten diabetes | Diabetes, Autoimmuuni | Insuliiniriippuvainen diabetes mellitus 1 | Diabetes mellitus, insuliinista riippuvainen, 1 | Diabetes... ja muut ehdotYhdysvallat
-
University of California, San FranciscoJuvenile Diabetes Research FoundationValmisTyypin 1 diabetes mellitus | Diabetes mellitus, tyyppi I | Insuliiniriippuvainen diabetes mellitus 1 | Diabetes mellitus, insuliinista riippuvainen, 1 | IDDMYhdysvallat, Australia
-
Capillary Biomedical, Inc.ValmisDiabetes mellitus, tyyppi 1 | Tyypin 1 diabetes | Tyypin 1 diabetes mellitus | Diabetes mellitus, insuliinista riippuvainen, 1Australia
-
Spiden AGDCB Research AGRekrytointiTyypin 1 diabetes mellitus | Tyypin 1 diabetes mellitus hypoglykemialla | Tyypin 1 diabetes mellitus ja hyperglykemiaSveitsi
-
Capillary Biomedical, Inc.LopetettuTyypin 1 diabetes | Tyypin 1 diabetes mellitus | Diabetes mellitus, tyyppi I | Diabetes mellitus, insuliinista riippuvainen, 1 | IDDMItävalta
-
National Institute of Allergy and Infectious Diseases...PPD; Rho Federal Systems Division, Inc.; Immune Tolerance Network (ITN)ValmisTyypin 1 diabetes mellitus | T1DM | T1D | Uusi tyypin 1 diabetes mellitusYhdysvallat, Australia
-
Instytut Diabetologii Sp. z o.o.National Center for Research and Development, Poland; Nalecz Institute of...TuntematonTyypin 1 diabetes mellitus ja hyperglykemia | Tyypin 1 diabetes mellitus ja hypoglykemiaPuola
-
Ain Shams UniversityTuntematonTyypin 1 diabetes mellitus hypoglykemialla | tyypin 1 diabetesEgypti
-
University Hospital, CaenValmisTyypin 1 diabetes mellitus kypsyys alkaaRanska
-
SanofiValmisTyypin 1 diabetes - tyypin 2 diabetes mellitusUnkari, Venäjän federaatio, Saksa, Puola, Japani, Yhdysvallat, Suomi
Kliiniset tutkimukset Dexcom G5-Eversense
-
University of Colorado, DenverDexCom, Inc.ValmisDiabetes mellitus, tyyppi IYhdysvallat
-
Imperial College LondonValmisTyypin 1 diabetes | Tyypin 1 diabetes mellitus hypoglykemialla | Heikentynyt tietoisuus hypoglykemiasta
-
Science Consulting in DiabetesDexCom, Inc. a Delaware corporation, USA; Institut für Diabetes-Technologie...Valmis
-
Jessa HospitalValmisTyypin 1 diabetes mellitusBelgia
-
Charles University, Czech RepublicTuntematonDiabetes mellitus, tyyppi 1Tšekki
-
Spinal Simplicity LLCIlmoittautuminen kutsusta
-
Laval UniversityValmisSydänkirurgia | Mekaanisesti ventiloidut potilaat
-
University of VirginiaDexCom, Inc.; Tandem Diabetes Care, Inc.ValmisTyypin 1 diabetes mellitusYhdysvallat
-
Stanford UniversityValmis
-
Federal University of São PauloVanessa Marques FerreiraTuntematonSepelvaltimotauti | Akuutti keuhkovaurio | Sepelvaltimon ohitussiirto